The first hurdle for selling an oxygen concentrator in Canada is the Medical Device Licence (MDL). Under section 26 of the Medical Devices Regulations (SOR/98-282), a Class II, III or IV device may be imported or sold in Canada only if its manufacturer holds a licence for that device. Under Rule 11 of Schedule 1, an oxygen concentrator starts as Class II; as of 13 November 2026, all 36 active licences in Health Canada's licence data whose name contains "oxygen concentrator" are Class II. A Class II MDL application needs a standards list, an attestation by a senior official, the label and a quality management system certificate, and ISO 80601-2-69 is the particular safety standard for oxygen concentrators on that list.

Key points

  • Who applies: the MDL is applied for and held by the device manufacturer (subsection 32(1)). A Class II, III or IV device cannot be imported or sold in Canada without one (section 26).
  • Which class: Rule 11 of Schedule 1 places an active device intended to administer drugs, body fluids or other substances to the body in Class II. If the administration is potentially hazardous, the device is Class III, and a device that controls treatment through a closed loop system is Class IV. Existing oxygen concentrator licences are all Class II, but the manufacturer still has to justify the class against the rules and state it in the application (paragraph 32(1)(b)).
  • What a Class II application contains: besides the device name, class, identifier and manufacturer details, it needs a description of the intended uses, a list of the standards complied with in manufacture, an attestation by a senior official that the manufacturer holds objective evidence, a copy of the label, and a quality management system certificate to CAN/CSA-ISO 13485 (subsection 32(2)).
  • Test reports: subsection 32(2) does not list test reports among the Class II documents, but the manufacturer must keep objective evidence (subsection 9(2)), and Health Canada may ask for more information or for samples when the file is insufficient (section 35).
  • Standards list: ISO 80601-2-69 is the particular standard for oxygen concentrators. List it together with the general standard IEC 60601-1 and the applicable collateral standards, with editions. Which entries and editions Health Canada's list of recognized standards contains should be checked against the current list on Health Canada's website.

Scope

This article is for manufacturers, including Chinese manufacturers, preparing to sell oxygen concentrators in Canada, and for the regulatory and quality staff who prepare the licence file and the standards list. Medical device market authorization in Canada rests on SOR/98-282, made under the Food and Drugs Act and administered by Health Canada. As of 13 November 2026, the consolidated text of SOR/98-282 on the Justice Laws website is current to 19 October 2026, and the provisions cited here follow that text.

Two licences need to be kept apart. The MDL answers the question "may this device be sold in Canada?" and is held by the manufacturer. The Medical Device Establishment Licence (MDEL), which importers and distributors need for their import and sale activities, is a separate set of requirements and is not covered here. Nor does this article cover the arrangements for quality system audits. It covers three things: how to settle the class of an oxygen concentrator, what a Class II MDL application contains, and how to write the standards list.

What class is an oxygen concentrator in Canada?

Section 6 of SOR/98-282 classifies medical devices into Classes I to IV using the rules in Schedule 1, with Class I the lowest risk and Class IV the highest. Section 7 provides that where a device can be classified into more than one class, the class representing the higher risk applies. Schedule 1 has specific rules for active devices. The ones directly relevant to oxygen concentrators are these:

RuleWhat it saysWhat it means for an oxygen concentrator
Rule 11(1)An active device, including any dedicated software, intended to administer drugs, body fluids or other substances to the body or withdraw them from the body is Class IIAn oxygen concentrator delivers oxygen to the patient, so its starting class under this rule is Class II
Rule 11(2)If the administration or withdrawal is potentially hazardous, taking into account its nature, the substance involved and the part of the body concerned, the device is Class IIIAssess against the intended use and the risk analysis
Rule 11(3)A device described in 11(2) that is intended to control the treatment of a patient's condition through a closed loop system is Class IVAssess separately if the design includes closed-loop control
Rule 12Any other active device is Class IApplies only to active devices not covered by the earlier rules

Health Canada's Medical Devices Active Licence Listing (MDALL) is a useful cross-check. As of 13 November 2026, a count through Health Canada's public data interface found 36 active licences whose licence name contains "oxygen concentrator", and every one of them carries Class II as its risk class. This shows that Class II is the usual class for licensed oxygen concentrators. It does not replace an assessment of your own product: if the device has an unusual intended use, patient population or control method, the application should still give the reasoning under Rule 11 point by point.

Class II MDL application: what to submit

Section 32 sets out the content of an MDL application. Every class supplies the basic information in subsection 32(1), and a Class II application adds the items in subsection 32(2). By provision:

ProvisionRequirementWhat to prepare for an oxygen concentrator
32(1)(a)–(e)Device name; class; device identifier, including the identifier of any device that is part of a system, family or group; manufacturer name and address as on the label; name and address of the manufacturing site, if differentList the identifiers of the main unit, the model range and the supplied accessories, consistent with the label
32(2)(a)A description of the medical conditions, purposes and uses for which the device is manufactured, sold or representedKeep it consistent with the intended use on the label and in the instructions
32(2)(b)A list of the standards complied with in the manufacture of the device to satisfy sections 10 to 20Give each standard number and edition (see the next section)
32(2)(c)An attestation by a senior official that the manufacturer has objective evidence that the device meets sections 10 to 20The test reports, risk management file and other evidence behind the attestation must be retrievable
32(2)(d)A copy of the device labelThe label must also meet the labelling requirements starting at section 21
32(2)(f)A quality management system certificate showing that the system under which the device is manufactured meets the National Standard of Canada CAN/CSA-ISO 13485The certificate must come from a registrar recognized by Health Canada under section 32.1 and is valid for no more than three years (section 32.2)

Class III asks for considerably more. Subsection 32(3) requires a list of the standards complied with in the design and manufacture of the device, a summary of all studies the manufacturer relies on and the conclusions drawn from them, and the number of units sold outside Canada with a summary of reported problems and recalls, among other items. If an oxygen concentrator were classified as Class III, the depth of the file would be quite different, which is why the class has to be settled first.

Not having to attach test reports to a Class II application does not mean testing can be skipped. Subsection 9(2) requires the manufacturer to keep objective evidence that the device meets the safety and effectiveness requirements of sections 10 to 20. Section 35 allows Health Canada, where the submitted information is insufficient to make a determination, to ask the manufacturer for additional information by a specified date, and to require samples of the device during the review. Every standard on the list needs evidence that can be produced on request.

How to write the standards list for an oxygen concentrator MDL

The standards list is the part of a Class II application most directly tied to testing. For an oxygen concentrator, the list is usually built around the general standard plus the particular standard, with collateral standards added according to the product's functions:

StandardTitle (summary)When it goes on the list
IEC 60601-1Medical electrical equipment: general requirements for basic safety and essential performanceThe base standard for medical electrical equipment
ISO 80601-2-69Medical electrical equipment, Part 2-69: particular requirements for the basic safety and essential performance of oxygen concentrator equipmentThe particular standard for oxygen concentrators, used together with IEC 60601-1
IEC 60601-1-2Collateral standard: electromagnetic disturbances, requirements and testsNormally needs to be considered for medical electrical equipment
IEC 60601-1-8Collateral standard: general requirements, tests and guidance for alarm systemsWhen the device has an alarm system
IEC 60601-1-11Collateral standard: medical electrical equipment and systems used in the home healthcare environmentWhen the device is intended for home use

Three points matter when writing the list. First, give the edition year. Particular, general and collateral standards are tied to each other by edition, and the edition on the list should match the edition actually used in the test report; the correspondence follows the normative references in the standards themselves. Second, Health Canada publishes a separate List of Recognized Standards for Medical Devices. Which entries it contains, which editions it recognizes and any notes attached should be taken from the current list on Health Canada's website, not from a draft or a third-party copy. Third, Canada adopts some international standards as National Standards of Canada, for example CAN/CSA-ISO 13485, which section 32 cites directly. If the applicant chooses to list a Canadian national edition, check that edition's text for differences from the ISO or IEC edition rather than assuming they are identical.

Common mistakes

  • Choosing standards before settling the class. The class decides which subsection of section 32 applies, and the depth of a Class II file differs greatly from a Class III file, so the class comes first.
  • Assuming Class II needs no testing. Reports are not required with the application, but the objective evidence must be kept, and Health Canada can ask for it.
  • Listing IEC 60601-1 without ISO 80601-2-69. The general standard does not cover the particular requirements for oxygen concentrators; one does not replace the other.
  • Listing standard numbers without editions, or editions that do not match the reports. The attestation and the evidence then no longer line up.
  • Carrying a classification from another country straight into Canada. Canada classifies under its own Schedule 1 rules, and the application must state the class and reasoning under those rules.

How to check this in the database

Before preparing the application, you can use this site to check three things: how Canadian licences are distributed, which oxygen concentrator makers already have Canadian records, and the bibliographic record for ISO 80601-2-69. All figures below are as shown on the pages on 13 November 2026.

  1. The Canada block on the browse by registration market page is connected to Health Canada's MDALL. It shows 4,711 companies in this database with Canadian records, holding 57,726 active device licence records in total, broken down by Classes I to IV. The page notes that company counts and record totals are different units and should not be compared directly.
  2. In site search, set the scope to "Product". A search for "oxygen concentrator" returns 100 companies, and the Chinese term for oxygen concentrator returns 197. Each result lists the markets where the company has records; on the company page, the "CA Canada MDALL" box shows the number of active Canadian device licences. Device rows are a sample of up to 8 per company per market, not the full set, so a missing licence does not mean the company does not hold it.
  3. The ISO 80601-2-69 record shows the edition year 2020 and FDA recognition number 1-148, and points to the official channel for the standard's text. This site does not provide the full text.

To check a specific licence, use Health Canada's MDALL active licence search, which can search by company name, licence name, device name, company ID, licence number or device identifier. The regulatory text is in the full text of SOR/98-282. For market-level figures, see Canada device registration in numbers and Chinese device manufacturers in Canada. Data on this site comes from public regulatory databases, is for reference only, and the official record prevails.

Test evidence and SUNGO

The standards on the list are usually backed by test reports as objective evidence. Hefei SUNGO Medical Technology Co., Ltd. holds IAS accreditation (certificate TL-1119) whose scope includes ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269, together with IEC/EN 60601-1, IEC/EN 60601-1-2, IEC/EN 60601-1-8 and IEC/EN 60601-1-11 and the corresponding Chinese standards; the scope listed on the certificate governs. IAS is a signatory to the ILAC MRA.

The limits need to be stated as well. These medical electrical safety and EMC items fall only within the Hefei laboratory's IAS accreditation, not within CNAS or CMA accreditation, and Canadian national editions (CSA editions) are outside SUNGO's certificates. Whether a test report meets Health Canada's review requirements is for the regulator to decide. The accreditation scope is on the SUNGO laboratory page.

Frequently asked questions

Does an oxygen concentrator need an MDL to be sold in Canada?

Yes. Under Rule 11 of Schedule 1 its starting class is Class II, and a Class II, III or IV device may be imported or sold in Canada only if the manufacturer holds a licence for it (section 26 of SOR/98-282).

What class is an oxygen concentrator in Canada?

Class II by default (Rule 11(1)); Class III if the administration is potentially hazardous, and Class IV if it controls treatment through a closed loop system. As of 13 November 2026, all 36 active MDALL licences whose name contains "oxygen concentrator" are Class II.

Do I need to submit test reports with a Class II MDL application?

Subsection 32(2) does not list test reports as required documents; what is submitted is the standards list and an attestation by a senior official. The manufacturer must still keep objective evidence, and Health Canada may ask for more information or samples under section 35.

How does ISO 80601-2-69 relate to IEC 60601-1?

IEC 60601-1 sets the general requirements for medical electrical equipment, and ISO 80601-2-69 sets the particular requirements for oxygen concentrators. They are used together and do not replace each other. The edition correspondence follows the standards' own text.

How can I check whether an oxygen concentrator already has a Canadian licence?

MDALL is the authoritative source; search it by device name, licence name or licence number. This site's search can show whether a company has Canadian records, but device rows are sampled and do not replace the official record.

The order for an oxygen concentrator MDL is: settle the class under Rule 11, assemble the documents required by section 32, then make sure the standards list and the objective evidence match. This article is based on official public information available as of 13 November 2026. Refer to the latest documents issued by the regulator.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.