Canada's MDEL amendments (SOR/2026-110), Health Canada's second phase of changes to the establishment licensing provisions of the Medical Devices Regulations (SOR/98-282), come into force on 1 December 2026, together with a new version of guidance document GUI-0016. The core change: a foreign distributor that sells solely to MDEL-holding Canadian importers no longer needs its own MDEL, while the importer must report each of its suppliers at application and annual review. Class II to IV devices still need a manufacturer-held MDL.
For importers sourcing from China, this article sets out who needs a medical device establishment licence (MDEL) and who needs a medical device licence (MDL), what changes for Chinese trading companies, and what importers must file and maintain, based on the regulatory text and guidance. It ends with how to check licensing in this database.
Key points
- Two licences, two subjects. An MDL is held by the manufacturer for a Class II, III or IV device and covers the product. An MDEL is issued to importers, distributors, and Class I manufacturers that import or distribute their own devices; it covers the activity. There is no MDL for Class I devices.
- Foreign distributor licensing is relaxed. From 1 December 2026, an MDEL holder importing from someone other than the manufacturer no longer has to ensure that supplier holds an MDEL. Under the new GUI-0016, a distributor outside Canada selling solely to MDEL-holding importers needs no MDEL; one that also sells to hospitals, retailers or similar buyers still does.
- Supplier information becomes mandatory. MDEL applications must list the name, address and device classes of every supplier other than the manufacturer, updated at the annual licence review due before 1 April each year.
- Attestations become explicit obligations. Several attestations in section 45 are removed; sections 52, 58, 58.1, 59 and 61.2 now state which documented procedures each party must establish, implement and maintain.
- The MDL pathway does not change. Class II to IV devices still need a manufacturer-held MDL, with a CAN/CSA-ISO 13485 quality management system certificate.
Background and scope
Canada classifies devices into Classes I to IV by risk. GUI-0016 gives bandages, crutches and canes, and manual toothbrushes as Class I; powered toothbrushes and hearing aids as Class II; hip implants as Class III; and artificial hearts as Class IV.
Establishment licensing sits in sections 44 to 51.1 of the Regulations. Phase I amendments came into force on 14 December 2024, covering recalls and a new power to impose terms and conditions on an MDEL (section 48.1). SOR/2026-110 is Phase II. Its Regulatory Impact Analysis Statement (RIAS) names three problems: duplicate licensing of foreign distributors and their Canadian importers; incomplete supplier data, since Health Canada had collected it only voluntarily since February 2020; and inspection findings that documented procedures were often missing or out of date despite attestations. The RIAS also states that importation of devices for investigational testing under Part 3 is not affected.
The roles that matter (the last three as defined in GUI-0016):
- Manufacturer: a person who sells a device under their own name or trademark and is responsible for its design, manufacture, labelling, packaging or intended purpose, whether or not those tasks are done on their behalf (section 1 of the Regulations).
- Importer: a person in Canada, other than the manufacturer, responsible for the device being brought into Canada for sale.
- Distributor: a person, other than a manufacturer, importer or retailer, who sells a device in Canada for resale or non-personal use; a person outside Canada selling devices into Canada is also a distributor.
- Supplier: any person, other than the manufacturer, who sells a device to an MDEL holder for import or sale in Canada.
MDEL vs MDL: what is the difference?
| Item | MDL | MDEL |
|---|---|---|
| Holder | The legal manufacturer named on the label | Importers and distributors; Class I manufacturers importing or distributing their own devices |
| Subject | Specific Class II, III or IV devices | Import and distribution activity; not an approval of any device |
| Legal basis | Sections 26 and 32 | Sections 44, 45 and 46.1 |
| Application | Class-specific content plus a quality management system certificate | Form FRM-0292: establishment details, activities, manufacturer and supplier lists, device classes, addresses where procedures are in place |
| Public listing | MDALL | MDEL Listing |
Section 32 requires a quality management system certificate, which section 1 defines as one issued by a registrar recognized under section 32.1; Health Canada's recognition list names Auditing Organizations, and the SOR/2026-110 RIAS states that Class II to IV manufacturers hold their certification under the Medical Device Single Audit Program (MDSAP). Separately, the SOR/2024-238 provisions have applied since 1 January 2026, allowing the Minister to impose or amend terms and conditions on a Class II to IV MDL at any time (section 36(2)). That is a different change from the MDEL amendments.
Which licence for devices sourced from China: three common models
| Model | Chinese party | Canadian importer | From 1 December 2026 |
|---|---|---|---|
| Class I device, manufacturer sells through an MDEL-holding importer | No MDL for Class I. Exempt from an MDEL if it imports or distributes solely through an MDEL holder (section 44(2)(d)); under GUI-0016, a manufacturer using this exemption cannot import or sell other companies' devices | Holds an MDEL; need not ensure the Class I manufacturer holds one (section 44(4)(a)) | Licensing unchanged; manufacturer and importer each need documented incident reporting procedures (section 59(3)) |
| Class II device, manufacturer holds an MDL | Holds the MDL; exempt from an MDEL when selling its own licensed devices (section 44(2)(c)), again without selling other companies' devices | Holds an MDEL; need not ensure the Class II to IV manufacturer holds one (section 44(4)(b)) | Licensing unchanged; importer procedures now spelled out |
| Trading company (not the manufacturer) | Before: importers had to ensure suppliers held an MDEL (section 44(3)), so every foreign distributor needed one. After: selling solely to MDEL-holding importers, no MDEL; also selling to hospitals, retailers and similar buyers, MDEL still required | Class II to IV devices still need a manufacturer-held MDL; importer lists the trading company as a supplier | Biggest change: duplicate licensing replaced by importer supplier reporting |
The RIAS is direct: foreign distributors that only sell through Canadian importers will no longer need an MDEL; those also selling to Canadian health care facilities or retailers still will. Health Canada was already applying a risk-based approach to this duplicative requirement for MDEL-holding importers; the amendments put that into the Regulations.
What SOR/2026-110 changes, provision by provision
1. Section 44(4): exceptions to the verification duty
Section 44(3), requiring an importer to ensure its supplier holds an MDEL, is unchanged. The replaced section 44(4) lists three exceptions: an MDEL holder importing a Class I device from the manufacturer; anyone importing a Class II, III or IV device from the manufacturer; and an MDEL holder importing any device from a person other than the manufacturer. The third is new. The section 44(2) exemptions for retailers, health care facilities and manufacturers are unchanged, so an importer without an MDEL, such as a hospital or retailer buying directly from abroad, must still ensure the foreign distributor holds one.
2. Section 45: application content
- Paragraph 45(e), medical specialities per manufacturer, is repealed.
- New paragraphs 45(f.1) and (f.2): the name and address of each person, other than the manufacturer, who sells devices to the establishment for importation or distribution, and the classes of devices each supplies.
- The attestations in paragraphs 45(g) to (i) are repealed; paragraph 45(j) now asks for the address of each building where the procedures below are in place.
Section 46.1(1) now requires the annual review to include the section 45 information and any updates. The RIAS estimates about 2,300 importers and distributors will submit supplier information, about 800 of whom already did so voluntarily.
3. Documented procedures
| Provision | Applies to | Procedures to establish, implement and maintain |
|---|---|---|
| 52(3) (new) | Manufacturers, importers, distributors; all classes | Maintaining distribution records |
| 58 (reworded) | Manufacturers, importers, distributors; all classes | Effective and timely investigation of reported problems and recall |
| 58.1 (new) | Manufacturers, importers, distributors; Class II to IV | As applicable, handling, storage, delivery, installation, servicing and corrective action |
| 59(3) (new) | Manufacturers and importers; Class I | Preliminary and final incident reports under 59(1) and (1.1) |
| 59(4) (new) | Manufacturers and importers; Class II to IV | Preliminary and final incident reports under 59(1) |
| 61.2(4) (new) | MDL holders; importers of Class II to IV | Timely provision of serious-risk information under 61.2(2), such as risks communicated by foreign regulators, labelling changes and recalls |
The RIAS clarifies that these obligations are not new, only explicit; that procedures are not submitted, only the addresses where they are in place; and that Class I parties need no section 58.1-type procedures. Recall responsibility does not shift: Health Canada said the importer has always been the entity responsible for conducting a recall under the Regulations.
4. Transition and coming into force
Section 10 sets coming into force on the 180th day after registration; registration on 4 June 2026 gives 1 December 2026. Under section 9, an MDEL application filed under the former section 45 and undecided before that day will not be issued unless, by that day and in the form established by the Minister, the applicant provides the paragraph 45(f.1) and (f.2) information.
What the new GUI-0016 adds
The new GUI-0016 was issued on 30 October 2026, takes effect on 1 December 2026, and replaces version 10 dated 14 December 2024; until then it is marked for reference only. Relevant points:
- Licensing by activity. Table 1: a non-manufacturer outside Canada selling solely to MDEL-holding importers needs no MDEL; one selling to health care facilities, retailers or other persons who do not need an MDEL (other than the ultimate consumer) needs one; a manufacturer selling only its own Class II to IV devices needs only an MDL.
- The label decides who the manufacturer is. Where manufacturing is contracted out, the establishment named on the label is the legal manufacturer: it holds the MDL for Class II to IV devices, and for Class I devices an MDEL unless they are solely distributed through another MDEL holder.
- Applying and amending. Form FRM-0292 is sent through the form's Submit function. Changes to manufacturer or supplier lists go in the annual review, or in an amendment application between 1 April and mid-November.
- Terms and conditions, shortages. Health Canada may impose terms and conditions on an MDEL under section 48.1. Manufacturers and importers of specified medical devices must report shortages and discontinuations under sections 62.21 to 62.25.
Canada MDEL amendments timeline
| Date | Event |
|---|---|
| 9 November 2024 | Phase II notice of intent published for consultation |
| 14 December 2024 | Phase I amendments in force |
| 8 November 2025 | Proposal prepublished in Canada Gazette, Part I (70-day comment period) |
| 1 January 2026 | SOR/2024-238 terms and conditions provisions for device licences in force |
| 4 June 2026 | SOR/2026-110 registered (P.C. 2026-550) |
| 17 June 2026 | Published in Canada Gazette, Part II |
| 30 October 2026 | New GUI-0016 issued |
| 1 December 2026 | SOR/2026-110 and new GUI-0016 in force |
| Before 1 April 2027 | First annual review under the new rules, including supplier information |
Common misreadings
- "An MDEL lets us sell the product." It approves no device; Class II to IV devices depend on the manufacturer's MDL.
- "Foreign distributors no longer need licences." Only those selling solely to MDEL-holding importers are exempt.
- "A trading company using its own brand is just a distributor." Selling under its own name makes it the manufacturer; for Class II to IV it must hold the MDL.
- "Class I manufacturers never need an MDEL." One selling directly to Canadian retailers or health care facilities does.
How to check this in the database
Canadian data in this database comes from Health Canada's Medical Devices Active Licence Listing (MDALL) and covers device licences; MDEL holdings are not included.
- Open Browse by registration market and find "CA Canada". As of 12 November 2026 it shows 4,711 companies matched and 57,726 active device licences in total, including 42,525 Class II, 11,681 Class III and 3,490 Class IV. The page notes that record units differ between markets.
- Open Browse by country and select China. The company list has a "Canada" column; a number there means this database has found Canadian licence records for that company in MDALL, so you can check prospective suppliers one by one.
- Use Search by company name, product name or registration number. A company page's "CA Canada MDALL" entry shows active device licences and the count including historic records; the verification section shows Canadian records confirmed and excluded. Device rows are a sample of up to 8 per company per market, so a missing licence number does not mean the company lacks it.
- Confirm in the official MDALL (Active Licence Search; Archived Licence Search for licences no longer in force). To check a partner's MDEL, use Health Canada's MDEL Listing, searchable by licence number, company ID, or company name, activity and country.
- For regulatory text, see Canadian regulations. As of 12 November 2026 it lists 6 entries; the Part 1 entry summarizes establishment licensing, distribution records, incident reporting and recalls, and links to the official text. SOR/2026-110 is not listed separately; until it takes effect, its text appears under "Amendments Not in Force" at the end of the consolidated Regulations on the Justice Laws website.
- See Data sources: Canadian data is used under the Open Government Licence – Canada, and members can export all fields.
Two earlier articles, Canada device registration in numbers and Chinese device manufacturers in Canada, give background; each states its data date.
FAQ
Does a foreign distributor selling only to MDEL-holding Canadian importers still need an MDEL?
From 1 December 2026, the new GUI-0016 lists a distributor outside Canada selling solely to MDEL holders, such as a Chinese trading partner supplying only you, as not requiring an MDEL, and the RIAS says Health Canada will notify foreign distributors that they may no longer need one. If the partner also supplies Canadian hospitals or retailers directly, it still does; whether to cancel depends on its actual customers and Health Canada's notice.
Will our supplier list be made public?
The RIAS states that supplier information will be treated as confidential business information under the Food and Drugs Act and as Protected B within Health Canada's internal database.
We filed an MDEL application before 1 December 2026 with no decision yet. What now?
Under section 9 of SOR/2026-110, the licence will not be issued unless the applicant provides the supplier information by the coming-into-force day (1 December 2026), in the form established by the Minister.
Does an MDL application need an MDSAP certificate?
Section 32 requires a CAN/CSA-ISO 13485 certificate from a registrar recognized under section 32.1. Health Canada's MDSAP page states that the transition to MDSAP was completed in 2019 and that all manufacturers holding medical device licences in Canada participate in it, and the SOR/2026-110 RIAS says Class II to IV manufacturers hold their certification under MDSAP. Follow Health Canada's MDL application guidance for specifics.
In practice, the change sits between foreign trading companies and Canadian importers: a trading company selling only to licensed importers no longer needs an MDEL, while the importer reports each supplier and keeps documented procedures for distribution records, investigation and recall, and incident reporting. The MDL pathway is unchanged, so device class and the manufacturer named on the label remain the first things to confirm. See SOR/2026-110 and the new GUI-0016. This article is based on official public information available as of 12 November 2026; the regulators' latest documents prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.