Mexico COFEPRIS medical device registration means obtaining a sanitary registration (registro sanitario) from COFEPRIS, the Federal Commission for the Protection against Sanitary Risks. Devices approved by the US FDA, Health Canada or Japan can be filed under equivalence agreements, and the abbreviated pathway agreement of 18 July 2025 (Vía Regulatoria Abreviada, or Reliance) widened the reference authorities to IMDRF Management Committee members and full MDSAP members. Based on the June 2026 COFEPRIS guide, this article covers who files, which route applies, what to submit and where to look up data.

Key points

  • Making, selling and distributing a device in Mexico requires a sanitary registration. Article 83 of the Reglamento de Insumos para la Salud (RIS) sorts devices into Class I, II and III by risk, applying the RIS and FEUM classification criteria.
  • The file needs a notarial power of attorney in favor of a legal representative established in Mexico, or a RUPA; if the device is not made by the applicant's parent, subsidiary or affiliate, the foreign legal manufacturer also issues a letter of representation to the applicant company established in Mexico.
  • The June 2026 guide covers three agreements: the 2010 FDA and Health Canada agreement, the 2012 Japan agreement and the 18 July 2025 abbreviated pathway agreement, matched to options A to F on the application form.
  • A US 510(k) or PMA does not convert automatically into a Mexican registration: authenticated FDA documents, Spanish labeling and technical files are still required, plus adverse event reports for Class II and III, and COFEPRIS reviews the file.
  • The ssy database holds no Mexican registration data. Use the COFEPRIS registration viewer launched on 20 August 2026 for Mexico, and the ssy FDA product code library for US classification and clearance status.

Mexico medical device registration: legal basis and Class I, II, III

COFEPRIS is an autonomous body under Mexico's Ministry of Health; domestic and imported devices may be sold in Mexico only after it grants a sanitary registration. Its December 2023 device registration guide cites RIS Article 82: medical equipment, prostheses, orthoses, functional aids, diagnostic agents, dental supplies, wound-care materials, hygiene products and other medical-use devices need a registration for manufacture, sale and distribution.

The June 2026 guide also cites the General Health Law, labeling standard NOM-137-SSA1, GMP standard NOM-241-SSA1, technovigilance standard NOM-240-SSA1 and the Medical Devices Supplement of the Mexican Pharmacopoeia (FEUM), in their current versions.

For classification, the 2023 guide restates the three classes of RIS Article 83 and points to Appendix II (normative) of FEUM Medical Devices Supplement 5.0, Volume II, for the classification criteria:

Mexican classRIS Article 83 (summary)
Class IKnown in medical practice, with proven safety and efficacy; generally not introduced into the body
Class IIKnown in medical practice, may vary in material or concentration; generally introduced into the body for less than 30 days
Class IIINew or newly accepted in medical practice, or introduced into the body and remaining there for more than 30 days

The 2023 guide adds that some low-risk products, and products that are not health supplies, appear in a dedicated listing agreement worth checking first, and that registrations granted from 2005 onwards are valid for five years and must then be extended (prórroga).

Who applies: the applicant in Mexico and the letter of representation

Under the June 2026 guide, the FDA, Health Canada, Japan and abbreviated pathway sections all require:

  • proof of legal standing: the original or a certified copy of a notarial power of attorney in favor of a legal representative established in Mexico, or a simple, valid copy of the RUPA (single registry of accredited persons);
  • the current COFEPRIS application form for authorizations, certificates and visits, FF-COFEPRIS-01, signed by the owner or legal representative;
  • proof of payment of the fee under the Federal Duties Law (check the amount in current COFEPRIS notices).

The FDA, Health Canada and Japan sections also require a letter of representation (carta de representación) when the device is not made by the applicant's parent, subsidiary or affiliate. Issued by the foreign legal manufacturer with an apostille or Mexican consular legalization, it grants marketing and distribution rights to the company established in Mexico that files the application. In other words, a foreign manufacturer applies through a company established in Mexico. Under the abbreviated pathway, if the certificate of free sale does not cover the legal manufacturer, a document notarized in the country of origin must show the responsibilities among the manufacturers.

The 2023 standard-route guide also asks for the applicant's operating notice (aviso de funcionamiento) and sanitary responsible person notice (responsable sanitario), under RIS Article 181 and General Health Law Article 259. The guides and COFEPRIS's review govern applicant requirements.

COFEPRIS equivalence agreements and the reliance pathway in the June 2026 guide

The legal basis for equivalence is RIS Article 161 Bis, which lets the Ministry of Health issue general provisions recognizing foreign authorities' requirements, tests and evaluation procedures as equivalent to Mexican rules. The June 2026 COFEPRIS guide on registration through equivalence agreements and the abbreviated pathway (file name dated 12 June 2026) sets out these routes:

RouteBasis (publication date in the DOF, Mexico's official gazette)Foreign approval relied onForm option
FDA equivalenceAgreement of 26 October 2010FDA 510(k), PMA, or registration and listing of a 510(k)-exempt deviceA, B, C (2010 agreement)
Health Canada equivalenceAgreement of 26 October 2010 (same as FDA)Health Canada medical device licence; the Canada slide of the COFEPRIS deck reads "Requisitos Clase 2, 3, 4"A, B, C (2010 agreement)
Japan equivalenceAgreement of 25 January 2012Certification by a registered certification body for designated controlled Class II devices (option D), or an MHLW approval letter for Class II, III and IV devices (option E)D, E
Abbreviated pathway (Reliance)Agreement of 18 July 2025Approval by a reference regulatory authority (ARR): IMDRF Management Committee member countries and full MDSAP membersF

The guide notes that option F applies only to filings under the 2025 agreement, with the whole dossier aligned to it, and that this pathway can also rely on FDA, Health Canada and MHLW approvals. Documents that do not match the chosen agreement may affect the technical review.

Which agreements are in force or repealed should be read from the timeline in the COFEPRIS slide deck released with the guide (RELIANCE_12062026.pdf), labeled as follows:

Date on the timelineName on the timelineLabel in the file
26 October 2010FDA and Health Canada agreementNo separate label; included in the scope of the June 2026 guide
25 January 2012Japan agreementNo separate label; included in the scope of the June 2026 guide
22 June 2021Equivalence agreement (Acuerdo Equivalencias)No separate label; not listed in the scope of the June 2026 guide
11-06-2025 (as written)Shown under the abbreviated pathway agreement segmentDEROGADO (repealed)
18 July 2025Abbreviated regulatory pathway agreementVIGENTE (in force)

The deck does not say which instrument the 11-06-2025 label refers to, and this article does not guess; the June 2026 guide rests on the 2010, 2012 and 18 July 2025 agreements.

On filing codes, the deck maps the former COFEPRIS-04-001-D to 04-001-H to the new code COFEPRIS-04-050, registration by equivalence with reference regulatory authorities. The guide states that 04-050 was not yet enabled for digital filing when it was issued; a USB drive does not replace the paper dossier, which must be filed physically.

What the FDA equivalence route asks for: 510(k), PMA and exempt devices

Under the 2010 agreement, the FDA document depends on the device's US class and must carry an apostille issued by a competent US authority:

US classFDA document required
Class I (510(k)-exempt, general controls)FDA establishment registration and device listing showing the device is listed, its risk class and its "510(k) exempt" status
Class II510(k) substantial equivalence letter
Class IIIPMA approval letter

Other legal documents include an FDA Certificate to Foreign Government (CFG) covering the device, models, legal manufacturer or site and codes, and the Establishment Inspection Report (EIR) from FDA's last inspection. Without an EIR, the guide accepts an inspection report from an MDSAP-recognized organization's last visit, or a valid ISO 13485:2016 (or later) certificate from an MDSAP-recognized organization. If FDA has not inspected the site or has not provided the report, a valid ISO 13485:2016 (or later) certificate from an authorized body plus an explicit GMP compliance statement in the CFG may be filed. Class II and III devices also need a postmarket adverse event report, which may include MAUDE reports, translated into Spanish and signed by the manufacturer's quality lead.

The technical file centers on a manufacturer-signed monograph: description, materials, specifications, manufacturing summary, sterilization, biocompatibility summaries where applicable, stability, packaging, models and clinical summaries. Equivalents to the certificate of analysis include batch production records, an electrical test conformity certificate with test summaries (such as a CB report) and a declaration of conformity with test summaries; with an electrical test certificate, medical equipment should cover applicable IEC 60601 electrical safety, EMC and functional tests. Trade name, manufacturer name and address must match across documents.

The COFEPRIS deck lists frequent problems on the FDA route: the 21 CFR regulation cited for an exempt device does not match the product; the 510(k) or PMA filed does not correspond to the device, so sameness is not shown; the device already holds a registration through the standard route and the new filing does not mention it; and stability is stated as "complies" without a study summary.

Reliance pathway checklist under the 2025 agreement

Per the COFEPRIS deck, the 18 July 2025 agreement entered into force 30 business days after DOF publication and recognizes decisions of full MDSAP members and of IMDRF Management Committee member countries, listed as Australia, Brazil, Canada, China, the European Union, Japan, Russia, Singapore, South Korea, Switzerland, the United Kingdom and the United States. Whether a specific approval can be relied on depends on the agreement text and on COFEPRIS's review. The June 2026 guide asks for the following under option F:

No.Administrative and legalNo.Technical
1.1Form FF-COFEPRIS-012.1Introductory description: generic and trade names, technical specifications, presentations, intended use, category and risk-based classification, accessory list
1.2Notarial power of attorney or RUPA2.2Description and specifications of main components, accessories and formulation, with drawings or images and a list of patient-contacting materials
1.3Proof of fee payment2.3Manufacturing process summary or flow chart
1.4Draft Spanish label per NOM-137-SSA12.4Conformity assessment submitted to the reference authority (summaries of processes and tests)
1.5Instructions for use or operating manual in Spanish2.5Primary and secondary packaging materials
1.6GMP certificate or equivalent; if no validity is stated, issued no more than 30 months earlier2.6Clinical evaluation summary (clinical studies or literature)
1.7Certificate of free sale or equivalent issued by the authority relied on; if no validity is stated, issued no more than 30 months earlier2.7Risk analysis report under harmonized international standards
2.8Certificate of analysis or finished-device certificate
2.9Technovigilance report in Spanish, signed by the manufacturer's quality lead and the sanitary responsible person

For GMP evidence, the deck cites the GMP document guidelines published in the DOF on 20 March 2025 (Lineamientos BPF): for foreign-made devices, an MDSAP audit certificate, an ISO 13485:2016 certificate from an accredited body, a CE certificate issued in the EU, or a statement in the certificate of free sale by the national regulatory authority or ministry.

The notes at the end of the guide state that documents may be in Spanish or English; other languages need a translation, by a certified expert translator for legal documents (RIS Article 153). Legal documents need an apostille, or Mexican consular legalization outside the Hague Convention. Electronically issued legal documents whose authenticity can be verified on the foreign authority's official website may be filed without apostille or consular legalization, in print or scanned, with the full access path and login details.

How to check the data in our database

  1. Confirm the US classification and pathway. The FDA device product code library listed 7,091 product codes as of 12 November 2026, grouped by FDA medical specialty. Each code page shows the class, 21 CFR section, specialty, marketing pathway (510(k), PMA or exempt), FDA's all-time 510(k) count and a link to FDA's classification page. Before using the FDA route, check that the 21 CFR section and pathway match the device; a mismatch is one of the problems COFEPRIS lists.
  2. Check the company's US listings. On the browse-by-market page, the US FDA column showed 26,108 companies and 282,111 listed devices as of 12 November 2026. Site search covers company names, registration numbers and products; device details are a sample of up to eight per company per market, so a missing record proves nothing. See also Chinese device manufacturers in United States: 5,817 companies, 49,334 listed device types and, for the Health Canada route, Chinese device manufacturers in Canada: 238 companies, 2,778 active device licences.
  3. Read the Mexico country page correctly. As of 12 November 2026, the Mexico country page listed 350 Mexico-based companies, 349 with registration records found here, showing their record counts in China, the EU, the US, the UK, Brazil, Saudi Arabia and Canada. It reflects Mexico-based companies' registrations in other markets, not COFEPRIS sanitary registrations.
  4. Note that ssy does not cover Mexico. The 14 registration-data markets on the data sources and licensing page do not include Mexico.
  5. Look up Mexican registrations at COFEPRIS. On 20 August 2026 COFEPRIS launched the Visor de Registros Sanitarios de Dispositivos Médicos, at registros.cofepris.gob.mx/BRSDM/default2.aspx. Per the press release, it shows current registrations that meet legal requirements and is updated continuously; the COFEPRIS home page describes search by generic name (denominación genérica), trade name (denominación distintiva) and official key (clave oficial). The page itself governs fields and results.

FAQ

Do you need a local company to register a medical device in Mexico?

The June 2026 COFEPRIS guide asks for a notarial power of attorney in favor of a legal representative established in Mexico, or a RUPA; a foreign manufacturer that is not an affiliate of the applicant also issues a letter of representation to the applicant company established in Mexico.

Can an FDA 510(k) be used directly to register in Mexico?

Not directly. It can support a filing under the 2010 FDA agreement or the 2025 abbreviated pathway, but authenticated FDA documents, Spanish labeling and a technical file are still required, and COFEPRIS reviews the file.

Can a Chinese NMPA registration support the Mexican reliance pathway?

The COFEPRIS deck lists China among IMDRF Management Committee members, but whether a specific NMPA registration qualifies depends on the 18 July 2025 agreement text and COFEPRIS's review; a certificate of free sale or equivalent from that authority is also required.

How long is a COFEPRIS sanitary registration valid?

The 2023 COFEPRIS guide states that device registrations granted from 2005 onwards are valid for five years and must be extended when they expire.

Where can I search registered medical devices in Mexico?

Use the COFEPRIS device registration viewer launched on 20 August 2026, searching by generic name, trade name or official key.

The equivalence and reliance routes reduce repeat review, but applicant, Spanish-language and authentication requirements remain, and the dossier must match the chosen agreement item by item. Full requirements are on the COFEPRIS abbreviated regulatory pathway (Reliance) page and in the June 2026 guide. This article is based on official public information as of 12 November 2026; the latest regulator documents prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.