Taiwan TFDA powered wheelchair registration is a Class 2 route. Under Taiwan's Regulations Governing the Classification of Medical Devices, powered wheelchairs (item code O.3860) and motorized vehicles for medical purposes, the scooter item (O.3800), are both Class 2, so they need registration and market approval from the Taiwan Food and Drug Administration (TFDA) and a medical device license before import. For test data, TFDA's Pre-clinical Testing Guidance for Powered Wheelchair cites the ISO 7176 series for performance and ISO 7176-21:2009 and IEC 60601-1-2:2014 for electromagnetic compatibility (EMC). This article works from the regulatory texts: item check, dossier, test editions and timelines.
Key points
- Item and class: O.3860 powered wheelchair and O.3800 motorized vehicle for medical purposes are both Class 2. Both identifications state a maximum speed limit of 10 km/h. O.3860 also covers add-on power accessories for manual wheelchairs.
- Legal basis: Taiwan's Medical Devices Act has applied since 1 May 2021. Article 25 requires registration and market approval and a license before manufacture or import (announced items use listing instead). The old Pharmaceutical Affairs Act, Article 40, is no longer the basis.
- Testing basis: the Pre-clinical Testing Guidance for Powered Wheelchair (announced 4 May 2011, amended 13 December 2017) lists 19 specification items, an EMC test and 11 performance tests. Where a referenced standard has been revised, the guidance lets the manufacturer test to the new edition.
- Manufacturers in mainland China should plan two things early: proof of compliance with Taiwan's medical device quality management system regulations, and, for devices made in mainland China, an import permit from Taiwan's trade authority before filing.
- This database shows Taiwan TFDA licence totals, the class split and per-company counts, but not totals by item code.
What class is a powered wheelchair in Taiwan?
Taiwan's Medical Devices Act was promulgated on 15 January 2020 and has applied since 1 May 2021. The Regulations Governing the Classification of Medical Devices, made under Article 3 of the Act and amended on 22 August 2023, sort devices into Class 1 (low risk), Class 2 (medium) and Class 3 (high), with items listed in an annex. The relevant items, summarised from the annex:
| Item code | Name | Class | Identification |
|---|---|---|---|
| O.3860 | Powered wheelchair | Class 2 | Wheeled, battery-powered device for medical use by patients restricted to a sitting position; maximum speed limit 10 km/h; includes add-on power accessories for manual wheelchairs |
| O.3800 | Motorized vehicle for medical purposes | Class 2 | Gasoline or battery-powered device for medical use by patients with impaired mobility; maximum speed limit 10 km/h |
| O.3850 | Mechanical wheelchair | Class 1 | Manual wheelchair; must meet CNS 14964-8, ISO 7176-8 or an equivalent international standard |
| O.3890, O.3900 | Stair-climbing wheelchair, standup wheelchair | Class 2 | Separate items, not O.3860 |
Two points are easy to miss. First, the speed limit is part of the identification, so a product designed to exceed 10 km/h no longer matches these items. Under Article 6 of the Regulations, a device whose function, intended use or working principle falls outside the identification of the listed items is classified as Class 3. If unsure, apply to the authority for a classification under Article 5. Second, the current Chinese name of O.3800 differs from the wording in older documents such as the 2019 draft amendment; the item code is the same.
What documents does a Taiwan registration need?
The dossier list is Appendix 2 of the Regulations Governing Issuance of Medical Device License, Listing and Annual Declaration (amended 27 November 2023). The import column for Class 2 devices, read with the notes, is summarised below.
| Document | Class 2 import | Notes |
|---|---|---|
| Application form; draft labels, instructions or packaging | Required | Original labelling, a draft Chinese leaflet and colour product pictures |
| Copy of the medical device firm licence | Required | The importer must be a medical device dealer whose licensed business includes import |
| Original letter of authorization from the foreign manufacturer | Required | Names the authorized device firm and the product models; valid for one year from issue |
| Original manufacture and free sale certificate of the country of origin | Case by case | States the device name, models, manufacturer and sale status in its home market |
| Proof that the manufacturer complies with the medical device quality management system regulations | Required | A copy of the certificate issued by the authority |
| Pre-clinical test and factory quality-control specifications and methods, raw test records and test reports | Case by case | Where similar devices are already approved, Note 7 allows substitute documents |
| Structure, materials, specifications, performance, intended use and drawings | Required | For instruments, operation and service manuals can be used |
| Clinical evidence | Case by case | Where similar devices are already approved, it can generally be omitted |
Three notes in Appendix 2 matter directly for powered wheelchairs:
- Note 7(5): a contract laboratory performing biocompatibility, electrical safety, EMC or sterility testing must meet either ISO/IEC 17025 or Good Laboratory Practice (GLP).
- Note 7(3): O.3860 powered wheelchair is listed in Appendix 3. If the applicant already holds a valid license for a similar device of the same item from the same manufacturer, a Class 2 product comparison and pre-clinical test data conformity statement can replace the pre-clinical test data (devices containing drugs excepted). O.3800 is not in Appendix 3.
- Note 13(3): for devices whose place of origin is mainland China, an import permit from Taiwan's trade authority must be obtained before the registration application is filed.
Full text: Regulations Governing Issuance of Medical Device License, Listing and Annual Declaration.
TFDA pre-clinical testing guidance for powered wheelchairs: test items and ISO 7176 editions
On 13 December 2017 TFDA amended 25 pre-clinical testing guidances by announcement FDA-Device No. 1061609014, including the Pre-clinical Testing Guidance for Powered Wheelchair. It applies to battery-powered wheeled wheelchairs under item O.3860. As of 13 November 2026, the administrative rules list in TFDA's medical device regulations area shows no later version. Four of its explanatory notes affect the dossier directly:
- The guidance is a recommendation and does not cover clinical trials; reviewers may ask for more, depending on claims and design.
- Pre-clinical data should include the test specification (acceptance range and its basis), the method, raw test records and the test report.
- If a listed test is not done, literature or a scientific evaluation must be submitted. If a test method differs from the reference method, its equivalence or basis must be explained.
- If a referenced standard has been revised, withdrawn or replaced, the manufacturer may test to the new edition.
The product description section asks for 19 specification items: battery specification, battery power indicator, charger, charge time, user mass, maximum drive distance, maximum speed, motor, brake, climbing angle, seat height, seat depth, seat width, front wheel size, rear wheel size, brake wheel size, drive and braking method, minimum turning space, and weight including battery.
The safety and performance section has two groups: EMC testing and performance testing. The table matches each item to its reference standard by name and compares the accreditation of the SUNGO Hefei laboratory (CNAS L24279 and IAS TL-1119 only; IAS accredits whole standards and the certificate does not state editions).
| Test item | Reference standard in the guidance | Hefei laboratory accreditation |
|---|---|---|
| EMC | ISO 7176-21:2009, IEC 60601-1-2:2014, CNS 14964-21:2007 | ISO 7176-21 and IEC 60601-1-2 in IAS scope only; CNS not in scope |
| (1) Static stability | ISO 7176-1:2014 | CNAS 2014 edition; IAS |
| (2) Dynamic stability | ISO 7176-2:2001 | CNAS 2017 edition; IAS |
| (3) Efficiency of brakes | ISO 7176-3:2012 | CNAS 2012 edition; IAS |
| (4) Energy consumption | ISO 7176-4:2008 | CNAS 2008 edition; IAS |
| (5) Dimensions, mass and manoeuvring space | ISO 7176-5:2008 | CNAS 2008 edition; IAS |
| Also listed: seat and wheel dimensions | ISO 7176-7:1998 | CNAS 1998 edition; IAS |
| (6) Maximum speed, acceleration and deceleration | ISO 7176-6:2001 | CNAS 2018 edition (maximum speed on a horizontal surface); IAS |
| (7) Static, impact and fatigue strengths | ISO 7176-8:2014 | CNAS 2014 edition; IAS |
| (8) Climatic tests | ISO 7176-9:2009 | CNAS 2009 edition; IAS |
| (9) Obstacle-climbing ability | ISO 7176-10:2008 | CNAS 2008 edition; IAS |
| (10) Power and control systems | ISO 7176-14:2008 | CNAS 2022 edition; IAS |
| (11) Resistance to ignition of postural support devices | ISO 7176-16:2012 | IAS only |
| (12) Where transport of the user in a vehicle is claimed | Taiwan vehicle rules, including point 67 of Taiwan's vehicle safety testing standard; ISO 7176-19:2008 is listed in the references | Not in certificate scope |
| Taiwan national standards for each item | CNS 14964 series (editions vary by part) | Not in certificate scope |
Edition gaps do not mean going back to the old edition. ISO 7176-2:2001, ISO 7176-6:2001 and ISO 7176-14:2008 have all been revised, and note 7 of the guidance allows testing to the new edition; the report should state which edition was used. Watch ISO 7176-6 in particular: the 2018 edition covers maximum speed only, while item (6) of the guidance also names acceleration and deceleration, so the dossier should explain how those two are addressed when the new edition is used. If reviewers ask for a specific edition, their request prevails.
For scooters, as of 13 November 2026 there is no pre-clinical testing guidance for O.3800 in TFDA's regulations area. The titles of ISO 7176-4:2008 and ISO 7176-21:2009 cited in the guidance both cover scooters, as does the title of ISO 7176-14:2022, so the applicable items can be used as a reference, subject to the reviewers' requirements.
Power kits for manual wheelchairs have their own guidance
O.3860 covers add-on power accessories for manual wheelchairs. On 29 April 2020 TFDA announced the Pre-clinical Testing Guidance for Power Kits on Mechanical Wheelchair (announcement FDA-Device No. 1091602214). It asks for the compatible manual wheelchairs to be specified and adds biocompatibility testing of parts in contact with the body (cytotoxicity, sensitization, irritation), software validation and quick-release clamping force (JIS D9301:2013). It cites the 2018 edition of ISO 7176-6.
Taiwan registration process and timelines
An application goes through formal review, then substantive review; the license is collected after approval. Deadlines:
| Stage | Rule | Basis |
|---|---|---|
| Formal review | Missing documents that can be supplemented must be supplied within 1 month of notice, or the application is rejected | Regulations, Art. 7 |
| Substantive review | Incomplete documents that can be supplemented must be supplied within 3 months of notice, or the application is rejected | Regulations, Art. 8 |
| License collection | Within 3 months of the approval notice, pay the certificate fee, prepare labels, instructions or packaging as approved, and collect the license; otherwise the license may be revoked | Regulations, Art. 11 |
| Labelling on the market | Chinese labelling on the smallest sales package and a Chinese instruction leaflet; the license number and other items must appear on the label, instructions or packaging | Act, Arts. 32 and 33 |
| Extension | A license is valid for up to 5 years; to keep importing, apply for extension within the 6 months before expiry, up to 5 years each time | Act, Art. 27; Regulations, Art. 17 |
Common mistakes: following process notes written under the old Pharmaceutical Affairs Act; overlooking the 10 km/h limit; and preparing only test reports while missing the quality management system proof, the letter of authorization or the import permit.
How to check Taiwan licences in this database
Taiwan data here comes from TFDA's open data platform (data.fda.gov.tw). The data sources page lists it as tier A under the Open Government Data License, Taiwan, version 1.0 (page updated 18 September 2026). Three ways to use it:
- Market totals: on the markets page, the "TW Taiwan" section shows, as of 13 November 2026, 4,475 companies matched and 22,350 valid device licences in total: Class I 8,402, Class II 12,680, Class III 1,258 and 10 unclassified. For how to read this market, see Taiwan device registration in numbers (data as of 7 September 2026).
- One company: use the search page by company name, registration number or product (exact match is available). On the company page, the TW Taiwan TFDA box in the registration overview shows the number of valid device licences and the class split.
- Records and export: device rows shown publicly on a company page are a sample of up to 8 per market. Taiwan sample rows are visible in the member area with a free account, which can filter by country, registration market and management class and export CSV (Taiwan with full fields, 100 rows a month, source attribution added automatically).
The public pages do not total licences by item code, so this article gives no count for O.3860 or O.3800. A missing licence here does not mean the company lacks it; check status with TFDA's official search.
Test evidence and the SUNGO Hefei laboratory
SUNGO's wheelchair testing is done by Hefei SUNGO Medical Technology Co., Ltd. (the Hefei laboratory); the Shanghai laboratory has no wheelchair accreditation. The Hefei laboratory holds CNAS certificate L24279, CMA certificate 260015342945 and IAS certificate TL-1119. CNAS and IAS are both signatories to the ILAC MRA. Coverage relevant to this guidance:
- CNAS: 17 parts of ISO 7176, including the parts used here: -1, -2:2017, -3, -4, -5, -6:2018, -7, -8, -9, -10 and -14:2022. Of the 17, 14 are also in CMA scope; -14, -15 and -25 are not.
- IAS (whole standards): 20 parts of ISO 7176, adding -16, -21 and -30 to the CNAS list, plus IEC 60601-1-2.
- EMC (ISO 7176-21 and IEC 60601-1-2) is accredited only under the Hefei IAS certificate TL-1119, not under CNAS or CMA.
- Not in certificate scope: ISO 7176-19, the CNS 14964 series, and the vehicle rules referred to in item (12) of the guidance.
- For the biocompatibility tests in the power kit guidance, SUNGO does not test in-house. Where needed, the work is subcontracted to a laboratory with the relevant qualifications, with the client's prior agreement, and the report names the subcontracted items and the subcontractor.
Accreditation scope does not mean TFDA will accept any particular report; whether a dossier meets Note 7(5) of Appendix 2 is decided in TFDA's review.
Frequently asked questions
What approval does a powered wheelchair need for export to Taiwan?
A TFDA medical device license through Class 2 registration and market approval, with quality management system proof; devices made in mainland China first need the import permit.
What class is a mobility scooter in Taiwan?
O.3800 motorized vehicle for medical purposes is Class 2, with a maximum speed limit of 10 km/h in its identification. As of 13 November 2026, TFDA has no pre-clinical testing guidance specific to O.3800.
Can newer ISO 7176 editions be used for the TFDA powered wheelchair guidance?
Note 7 of the guidance allows testing to the new edition when a referenced standard is revised, withdrawn or replaced. State the edition in the report and explain the scope change in ISO 7176-6.
What does TFDA require of the laboratory for EMC reports?
Note 7(5) of Appendix 2 requires a contract laboratory doing EMC and similar tests to meet ISO/IEC 17025 or GLP. Whether a specific report is accepted is decided in TFDA's review.
How long is a Taiwan medical device license valid?
Up to 5 years from the date of issue. To keep importing, apply for extension within the 6 months before expiry; each extension is up to 5 years.
Confirm the item and speed limit against the annex first, then assemble the documents required by Appendix 2. This article is based on official public information available as of 13 November 2026. Refer to the latest documents from the regulator.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.