From 18 February 2027, Article 11 of the EU Batteries Regulation (EU) 2023/1542 requires portable batteries built into products placed on the EU market to be removable and replaceable by the end-user; there is no blanket exemption for medical devices. Only professional medical imaging and radiotherapy devices, in vitro diagnostic medical devices (IVDs) and a few categories added by Delegated Regulation (EU) 2026/1770, such as certain wearables and on-body subcutaneous delivery systems, may instead be designed for replacement by independent professionals; the full derogation in Article 11(3) covers only cases where continuous power and a permanent connection are needed for safety or data integrity. Home-use blood pressure monitors, nebulisers and pulse oximeters therefore start from the default rule.
Key points
- There is no blanket exemption for medical devices. Article 11(2)(b) covers only professional medical imaging and radiotherapy devices, plus in vitro diagnostic medical devices (IVDs), and only relaxes the rule to "removable and replaceable by independent professionals". Home-use blood pressure monitors, nebulisers and pulse oximeters are not covered.
- The full derogation in Article 11(3) is narrow. Continuity of power supply must be necessary, and a permanent connection must be required for the safety of the user and the appliance or, for products whose main function is collecting and supplying data, for data integrity. The old Directive's "performance" and "medical" grounds were not carried over.
- The derogation list has been extended. Delegated Regulation (EU) 2026/1770, in force since 11 November 2026, adds wearables, on-body subcutaneous delivery systems and other categories to Article 11(2), each subject to a safety-necessity condition. The Commission guidelines cite implantable devices, hearing aids and test-strip blood glucose meters as relevant to Article 11(3), but are not legally binding.
- Related duties apply too: instructions and safety information supplied with the product and permanently online; spare batteries for at least five years after the last unit of the model is placed on the market; no software blocking of compatible batteries. From the same date all batteries must carry a QR code (Article 13(6)).
- Start with a battery inventory and a route decision, recorded in the technical documentation; assess any design change under MDR change management.
EU Batteries Regulation Article 11: background and scope
Regulation (EU) 2023/1542 was adopted on 12 July 2023, entered into force on 17 August 2023 and has applied generally since 18 February 2024. Article 96(2) separately provides that Article 11 applies from 18 February 2027. It replaces Directive 2006/66/EC, repealed with effect from 18 August 2025; under Article 95 the Directive's Article 11 (removal of batteries) continues to apply until 18 February 2027, so old and new rules run back to back.
Under Article 3, a portable battery is a battery that is sealed, weighs 5 kg or less, is not designed specifically for industrial use and is not an electric vehicle battery, an LMT battery or an SLI battery. Button cells, rechargeable lithium packs and AA/AAA cells in home-use devices generally fall here. LMT (light means of transport) batteries, up to 25 kg, are covered separately by Article 11(5) and are not discussed here.
The obligation falls on "any natural or legal person that places on the market products incorporating portable batteries". For products imported from outside the EU this is usually the EU importer, but removability is fixed at the design stage, so the manufacturer needs the rationale and documents ready.
| Topic | Old Directive Article 11 (as amended by 2013/56/EU) | New Regulation Article 11 |
|---|---|---|
| Default rule | Readily removable; if not by the end-user, then by qualified professionals independent of the manufacturer | End-user can remove and replace the whole battery with commercially available tools throughout the product's lifetime |
| Not applicable where | Safety, performance, medical or data integrity reasons require continuous power and a permanent connection | Continuous power and a permanent connection required for user and appliance safety, or for data integrity in data-collection products |
| Related duties | Instructions on safe removal | Instructions permanently online; five-year spare parts; no software blocking |
Removable and replaceable battery requirements under Article 11
| Requirement | Provision | Key point |
|---|---|---|
| Removable and replaceable by the end-user | Art. 11(1) | Applies to entire batteries, not individual cells; at any time during the product's lifetime |
| "Readily removable" | Art. 11(1) | Removable with commercially available tools; no specialised tools (unless supplied free with the product), proprietary tools, thermal energy or solvents |
| Instructions and safety information | Art. 11(1) | On use, removal and replacement; supplied with the product and permanently on a public website, easy for end-users to understand |
| "Readily replaceable" | Art. 11(6) | Can be substituted by another compatible battery without affecting functioning, performance or safety |
| Spare batteries | Art. 11(7) | At least five years after the last unit of the model is placed on the market, at a reasonable and non-discriminatory price |
| Software | Art. 11(8) | Software must not impede replacement with a compatible battery or key components |
Points from the Commission guidelines (C/2026/5531): the end-user is an adult without specific experience or qualifications in replacing batteries; "commercially available tools" follows EN 45554:2020 and covers both "basic tools or no tools" and "commercially available tools"; manuals should state the specifications a compatible battery must meet; and "parts-pairing" is given as an example of software that impedes replacement.
Are medical devices exempt from battery removability?
Article 11(2): removable by independent professionals only
Delegated Regulation (EU) 2026/1770, adopted by the Commission on 14 July 2026 under Article 11(4), replaces Article 11(2). The amended list reads as follows:
| Point | Products | Relevance to medical devices |
|---|---|---|
| (a) | Appliances, including wearable devices, specifically designed to operate primarily in an environment regularly subject to splashing water, water streams or water immersion, and intended to be washable or rinseable | May apply to individual devices |
| (b) | Professional medical imaging and radiotherapy devices (medical devices within the meaning of MDR Article 2(1)), and in vitro diagnostic medical devices as defined in IVDR Article 2(2) | Directly relevant |
| (c) | Wearables whose safety, durability or water resistance may be compromised by user access to the battery, and that are too small for safe end-user replacement or rely on a compact sealed enclosure for functional integrity | May apply to wearable devices |
| (g) | On-body delivery systems specifically designed for the subcutaneous administration of medicinal products | Directly relevant |
| (d)(e)(f)(h) | Electric toys with rechargeable batteries (until 31 July 2030), food-contact wireless thermometer probes, Directive 2014/34/EU explosive-atmosphere equipment, roof-mounted telematics for agricultural and construction machinery | Not relevant |
First, every point except (b) applies only where required to ensure the safety of the user and the appliance; point (b) has no such condition, and on its wording its IVD limb is not limited to professional use. Second, this is a partial derogation: per the guidelines, the battery must still be removable and replaceable by independent professionals with commercially available tools. Independent professionals are described as independent operators competent and qualified to repair the product, working on a commercial basis and/or in commercial premises.
For wet appliances under point (a), the guidelines list five indicators: designed specifically for that environment; that environment is the primary one; intended to be washable or rinseable; documentary evidence that end-user removal would compromise safety; and documentary evidence that redesign with current technology is not possible. Splashing water, water streams and immersion are equated with IPX4, IPX5/6 and IPX7, but an IP rating alone is not enough.
Article 11(3): conditions for the full derogation
The Article 11(1) obligations do not apply at all where continuity of power supply is necessary and a permanent connection between product and battery is required for the safety of the user and the appliance or, for products whose main function is to collect and supply data, for data integrity. The medical points in the Commission guidelines C/2026/5531 are:
- Life-saving, life-sustaining and safety-critical devices are typical cases; a risk-averse approach is proposed for medical devices, using the classification in MDR Article 51 and Annex VIII and IVDR Article 47 and Annex VIII.
- Examples include implantable devices (such as pacemakers and implantable cardioverter defibrillators) and certain IVDs (such as instruments detecting transmissible agents in blood screening for transfusion, and blood glucose meters used with test strips); surgical utensils and endoscope washer-disinfectors are also mentioned.
- Hearing aids are considered relevant to Article 11(3). Charging cases whose main function is powering other products, such as hearing aid charging cases, can be considered portable batteries themselves, so Article 11 does not apply to them.
- For data integrity, the main function must be collecting and supplying data and data integrity must be at risk. Backup batteries powering volatile memory or an internal clock (such as CMOS batteries) qualify regardless of device class; on-body wearable glucose biosensor systems are also cited. It does not apply where data collection is only an additional feature, or where non-volatile memory removes the risk of data loss.
Where blood pressure monitors and other home-use devices fall
| Product type | Possible route | Key point |
|---|---|---|
| Electronic blood pressure monitors, nebulisers, pulse oximeters, electronic thermometers, home physiotherapy devices | Article 11(1) by default | Not covered by point (b) or cited as Article 11(3) examples; a full-derogation claim must show continuous supply and a permanent connection are essential for safety; check wearable forms against point (c) |
| Blood glucose meters used with test strips | Article 11(2)(b); also cited by the guidelines for Article 11(3) | Depends on the product's status under the IVDR |
| Hearing aids | Considered relevant to Article 11(3) by the guidelines | Charging cases can be treated as portable batteries |
| On-body subcutaneous delivery systems | Article 11(2)(g) | Subject to the safety-necessity condition |
| Devices with CMOS or real-time clock backup batteries | The backup battery can rely on Article 11(3) | Other batteries in the product must be assessed separately |
| Devices using AA, AAA or similar general-use cells in a user-accessible compartment | Article 11(1) | Check that instructions and safety information are permanently available online |
Five common misreadings about medical device exemptions
- "Medical devices are exempt." Article 11(2) names only professional imaging and radiotherapy devices, IVDs and on-body subcutaneous delivery systems; everything else must be checked against the wet-appliance, wearable and Article 11(3) conditions.
- "The old medical-reasons exemption still works." Article 11 of the old Directive applies only until 18 February 2027, and the new Regulation contains no "performance" or "medical" ground.
- "IP67 qualifies for the wet-appliance derogation." The guidelines say an IP rating alone is not enough.
- "Once derogated, nothing else applies." Article 11(2) still requires removability by independent professionals. Article 11(3) disapplies only the paragraph 1 obligations and is silent on spare parts (paragraph 7) and software (paragraph 8); check the text and with the authorities before assuming those are lifted.
- "The guidelines are the law." They state that they reflect the Commission's views and are not legally binding; final interpretation rests with the Court of Justice of the EU.
Key dates: 18 February 2027 and the battery QR code
| Date | Event |
|---|---|
| 17 August 2023 | Regulation (EU) 2023/1542 enters into force |
| 18 February 2024 | Regulation applies generally |
| 10 January 2025 | Commission guidelines on Article 11 (C/2025/214) published |
| 18 August 2025 | Directive 2006/66/EC repealed; its Article 11 continues to apply |
| 22 October 2026 | Delegated Regulation (EU) 2026/1770 and updated guidelines (C/2026/5531) published |
| 11 November 2026 | Delegated Regulation (EU) 2026/1770 enters into force |
| 18 February 2027 | Article 11 applies; old Directive Article 11 ceases to apply; all batteries must carry a QR code |
The QR code follows Annex VI Part C. For portable batteries it links to label information, the EU declaration of conformity, waste battery information and other items listed in Article 13(6). Where the nature and size of the battery do not allow marking, the code goes on the packaging and accompanying documents (Article 13(7)).
What medical device exporters should prepare
- Battery inventory: for each model sold in the EU, list battery type, chemistry, weight and mounting, including backup batteries and button cells.
- Route decision: assign each battery to Article 11(1), a point of Article 11(2), or Article 11(3), with legal basis, guidance and risk analysis in the technical documentation.
- Design and documents: confirm the compartment opens with commercially available tools; prepare instructions, safety information and compatible battery specifications, kept permanently on a public website.
- Spare parts and software: arrange spare battery supply for at least five years and check firmware for pairing checks that would block compatible batteries.
- Battery supplier: agree who is responsible for the QR code and label information, and where they will appear.
- MDR change control: changing the battery compartment or enclosure is a design change under MDR change management; whether to notify the notified body depends on the MDR text and the notified body's requirements.
How to check this in the database
The database covers the legal basis and EU registration scale, not battery design:
- Legal basis: as of 12 November 2026, the EU regulations page lists 172 regulations and guidance entries with their type, number or title and status; each entry page gives the issuing body and a link to the official text. The MDR and IVDR definitions cited in Article 11(2)(b) can be reached from the MDR (32017R0745) and IVDR (32017R0746) entries in that list. As of the same date the library does not include Regulation (EU) 2023/1542; use the EUR-Lex links above.
- Scale of EU registrations: as of 12 November 2026, the EU row on the browse-by-market page shows EUDAMED as live, with 1,838 companies matched and at least 150,958 UDI-DI device records in total (447 companies were capped at the first 300 records, so the total is a lower bound), broken down by Class I, IIa, IIb, III and the IVD classes.
- Companies and products: use Search by company name, registration number or product name; company pages show record counts per market. Device rows are a sample of up to eight per company per market, so a missing record does not mean it does not exist.
- Category and country data: the articles on the home and self-testing category and on Chinese device manufacturers in the EU show the EU registration footprint of comparable products and of Chinese manufacturers.
FAQ
Can a blood pressure monitor still use a built-in, non-removable lithium battery after 18 February 2027?
It is not a professional imaging or radiotherapy device or an IVD under Article 11(2)(b). Unless it is shown and documented to meet another point of Article 11(2) (such as wearables) or Article 11(3), units newly placed on the market must let the end-user remove and replace the battery.
Do products placed on the market before 18 February 2027 need to change?
Article 11 applies from 18 February 2027 and attaches to placing on the market. The guidelines state that the Article 11(7) spare-parts requirement does not apply to products placed on the market earlier; treatment of specific batches depends on the legal text and Member State authorities.
What if the product is not on the derogation list but end-user battery replacement is unsafe?
Assess whether Article 11(3) can be met and document it. Also watch for Commission calls for application: the guidelines say these will be published regularly, applicants have three months to submit technical evidence, and new derogations may follow by delegated act.
Who is responsible for the EU battery QR code: device maker or battery supplier?
Article 13(6) is a marking requirement on the battery itself. Agree with your battery supplier who is responsible, where the information sits, and whether a very small battery will carry the code on packaging and accompanying documents instead.
Exporters of home-use active devices to the EU should complete their battery inventory, route decisions and documentation before 18 February 2027. This article is based on official public information as of 12 November 2026; the latest documents from the regulators prevail. For SUNGO Lab's accreditation scope, see the SUNGO Lab page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.