What test report does a walker or crutch need for a Class I filing with China's NMPA? Under the Regulations on the Supervision and Administration of Medical Devices, a Class I filing must include a product test report, which can be the filer's own in-house report or a report from a qualified medical device testing institution. Three things go wrong most often: whether the product really fits the "auxiliary walking and standing devices" entry of the Class I catalogue, which edition of GB/T 14728 or GB/T 19545 the testing follows, and whether the testing body's CMA (China's mandatory approval for inspection bodies and laboratories) scope covers the standard used. This guide covers each point and shows how to look up China registration and filing records on this site.

Key points

  • The Catalogue of Class I Medical Devices (NMPA Announcement No. 158 of 2021, in force since 1 January 2022) lists medical crutches, walkers, framed walkers, wheeled walkers, walking tables and similar names under "19 Medical rehabilitation devices / 03 Walking aids / 02 Auxiliary walking and standing devices". The class is Class I, and the entry describes non-powered products.
  • Class I devices are managed by product filing. A domestic filer submits the dossier to the drug regulatory department of its local city-level government. The dossier must include a product test report from in-house testing or from a qualified medical device testing institution (Articles 13 to 15 of the Regulations).
  • The axillary crutch standard GB/T 19545.2-2026 has been in effect since 1 August 2026 and fully replaces GB/T 19545.2-2009. The framed and wheeled walker standards GB/T 14728.1-2024 and GB/T 14728.2-2024 have been in effect since 1 October 2025.
  • The CMA scope of Hefei SUNGO Medical Technology Co., Ltd. includes GB/T 14728.3-2008 (walking tables), GB/T 19545.2-2009 (axillary crutches) and GB/T 19545.4-2008 (tripod and multi-legged walking sticks). GB/T 14728.1-2024, GB/T 14728.2-2024 and GB/T 19545.1-2012 are within its CNAS scope only, with no CMA, so SUNGO cannot issue CMA reports for these three standards.
  • Biocompatibility content in these standards is outside the Hefei laboratory's testing scope. It can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report must name the subcontracted items and the subcontractor.

Why walkers and crutches are Class I filings: the catalogue entry

Article 13 of the Regulations (State Council Decree No. 739) states that Class I devices are managed by product filing, while Class II and Class III devices are managed by registration. Whether a walker or crutch can be filed depends on whether it matches an entry in the Catalogue of Class I Medical Devices, issued by NMPA Announcement No. 158 of 2021 and in force since 1 January 2022. The entry for walking aids reads, in summary:

ItemCatalogue content (summary)
Location19 Medical rehabilitation devices / 03 Walking aids / 02 Auxiliary walking and standing devices
Product descriptionLegs, handles, support pads, a support frame or arm cuffs; or a support platform, handles, wheels, height adjustment, brakes, a folding mechanism and a seat; non-powered product
Intended useAssisting people with mobility impairment to walk or stand, and for rehabilitation training
Example namesMedical crutch, walker, walking frame, framed walker, wheeled walker, walking table, transfer walker, standing frame and others
ClassClass I

The catalogue's notes for subcatalogue 19 matter when drafting the filing. "Typically consists of" in a product description only shows a representative structure; the filing should say "consists of" and list the actual components. The intended use can follow the catalogue or describe the actual product, but must not go beyond the intended use in the catalogue. A product that claims other uses or has a power source cannot simply be filed under this entry. This guide does not cover Class II or Class III registration.

Class I filing dossier and test report: in-house or external

Article 14 of the Regulations lists the filing documents: product risk analysis, product technical requirements, the product test report, clinical evaluation documents, instructions for use and draft labels, quality management system documents for development and production, and other documents needed to show safety and effectiveness. Clinical evaluation documents can be omitted where the product meets the exemption conditions in the Regulations. The test report must meet NMPA requirements and can come from two sources.

Report sourceLegal basisWhat to prepare
Filer's in-house test reportArticle 14: the filer's own in-house report is allowedThe filer tests the product against its technical requirements. The Provisions on In-House Testing for Medical Device Registration (NMPA Announcement No. 126 of 2021) address registration; for an in-house report in a Class I filing, follow the NMPA Class I filing document requirements and the local filing department.
External test reportArticle 14: a qualified medical device testing institution; Article 75: only recognised testing bodies may test medical devicesBefore placing the order, check the testing body's CMA scope schedule for the standard, its edition year and the required test items.

Article 15 sets out where to file. A domestic filer submits to the drug regulatory department of its local city-level government; an overseas filer acts through a designated legal entity in China, which submits to the NMPA. The filing is complete once a compliant dossier is submitted, and the information is published on the NMPA online government service platform within 5 working days. The full text is on the Chinese government portal (Regulations on the Supervision and Administration of Medical Devices).

GB/T 14728 and GB/T 19545: current editions for walkers and crutches

Walkers and crutches are covered by two groups of recommended national standards: GB/T 14728 (walking aids manipulated by both arms: framed walkers, wheeled walkers, walking tables) and GB/T 19545 (walking aids manipulated by one arm: elbow crutches, axillary crutches, tripod and multi-legged walking sticks and others). They are the main reference for product technical requirements, and the manufacturer decides which clauses to adopt. The status column comes from the national standards information public service platform as of 13 November 2026; the right-hand columns show coverage in the Hefei laboratory's certificate schedules. The Shanghai laboratory's scope does not include these standards.

StandardProduct (identical adoption)StatusHefei CMAHefei CNASHefei IAS (TL-1119)
GB/T 14728.1-2024Framed walkers (ISO 11199-1:2021)Current; in effect since 1 October 2025; fully replaces the 2006 editionNoneYes, excluding clause 7.3 (biocompatibility)Yes, except clause 7.3
GB/T 14728.2-2024Wheeled walkers (ISO 11199-2:2021)Current; in effect since 1 October 2025; fully replaces the 2008 editionNoneYes, excluding clause 7.3 (biocompatibility)Yes, except clause 7.3
GB/T 14728.3-2008Walking tables (ISO 11199-3:2005)Current; in effect since 1 September 2009; confirmed at review on 5 September 2025Yes; clause 4.9 (materials and finish) excludes biocompatibilityYes, same limitationYes, except clause 4.9
GB/T 19545.1-2012Elbow crutches (ISO 11334-1:2007)Current; in effect since 1 May 2013; confirmed at review on 5 September 2025NoneYesYes, whole standard
GB/T 19545.2-2026Axillary crutchesCurrent; in effect since 1 August 2026; fully replaces the 2009 editionSchedule lists the 2009 editionSchedule lists the 2009 editionLists GB/T 19545.2 with no edition year
GB/T 19545.4-2008Tripod and multi-legged walking sticks (ISO 11334-4:1999)Current; the review on 22 May 2026 concluded it should be revised, and revision plan 20260828-T-314 is being draftedYes; the materials item excludes biocompatibilityYes; clause 4.4 (materials) excludes biocompatibilityYes, except clause 4.4

Three points follow. First, the standards for framed walkers, wheeled walkers and elbow crutches are covered at Hefei by CNAS only, with no CMA. If a filing needs an external report bearing the CMA mark, SUNGO cannot issue one for these three standards, and the filer needs a testing body whose CMA scope covers them. Second, the axillary crutch standard has a new edition, but the Hefei CMA and CNAS schedules list the abolished 2009 edition. Whether a report to the 2026 edition can carry an accreditation mark depends on the certificate schedule in force at the time, so confirm this in writing before ordering. Third, biocompatibility content can only be subcontracted. Dates for the new axillary crutch standard can be checked on the GB/T 19545.2-2026 page of the national standards platform.

Common mistakes in walker and crutch filings

  • Treating a Class I filing as a registration. The filing is complete once a compliant dossier is submitted; there is no registration review and approval step. The dossier still has to be complete and accurate, including the test report.
  • Checking the standard number but not the edition year. GB/T 19545.2 has moved to the 2026 edition, and GB/T 14728.1 and .2 to 2024 editions. The edition years in the technical requirements, the test report and the testing body's schedule must match.
  • Using this entry when the use or design goes beyond it. The entry covers non-powered products for assisted walking, standing and rehabilitation training. Products with power or other claimed uses need their class confirmed first.
  • Reading the number of companies in search results as a filing count. Device rows are a sample, as explained below.

How to search the database for China walker and crutch filing records

This site's registration database includes China registration certificates and filing records taken from the NMPA's public data. It shows which companies hold records for similar products, but it cannot be used to count them.

  • On the search page, enter the Chinese product name, because China records carry Chinese names: for example 助行器 (walker), 腋拐 (axillary crutch), 拐杖 (crutch) or 手杖 (walking stick). English terms such as "walker" mostly match company names in other markets. The search can cover everything or be limited to company name, registration number, product or country/region; matching is fuzzy by default, and Exact or quotes give literal matching. Device rows show the number, product name and device class.
  • Device rows are a sample of up to 8 per company per market, not the full set, and a missing number does not mean the company does not hold it. Results therefore cannot be read as a filing total, and this guide gives no filing total or provincial breakdown for walkers or crutches.
  • The registration markets page gives China totals. As of 13 November 2026 it shows China NMPA as "partial", with 38,717 companies matched and at least 315,469 records (207 companies were capped at the first 300 records, so this is a lower bound): 176,169 Class I, 116,981 Class II and 20,759 Class III. These figures cover all product types, not walking aids alone.
  • A free account adds member search filters by country, registration market and device class, sorting, and CSV export (100 rows a month, with source attribution). For China, only basic fields such as number, company, product and date can be exported.
  • For standard editions, use the testing search page: enter GB/T 14728 or GB/T 19545 to open record pages showing edition years and the implementation date. As of 13 November 2026 there are no record pages for GB/T 14728.3 or GB/T 19545.4. The site does not provide standard texts, and the national standards platform governs current status.

As of 13 November 2026, the China NMPA section of this site's regulatory database is marked as still being compiled. For the catalogue, filing requirements and classification decisions, refer to the NMPA's original documents. For an overview of China data in this database, see China device registration in numbers (data as of 7 September 2026).

Test evidence and SUNGO

Hefei SUNGO Medical Technology Co., Ltd. holds CMA 260015342945, CNAS L24279 and IAS TL-1119, and can issue GB/T 14728 and GB/T 19545 test reports within the scopes shown in the table. Reports bearing the CMA mark are limited to the items in the schedules for GB/T 14728.3-2008, GB/T 19545.2-2009 and GB/T 19545.4-2008. CNAS and IAS are both signatories to the ILAC MRA. Biocompatibility content is outside the Hefei laboratory's testing scope and can only be subcontracted to a suitably qualified laboratory, with the client's prior consent and with the subcontracted items and subcontractor named in the report. See the SUNGO Lab page for an introduction; the accredited scope is the certificate schedule published by the accreditation and approval bodies.

Can a powered walker be filed as Class I?

The catalogue entry describes non-powered products. Walking aids with powered height adjustment, powered drive or similar functions fall outside its description and cannot simply be filed under it; check the Medical Device Classification Catalogue and the NMPA's published classification decisions. Under Article 23 of the Regulations, for a newly developed device not yet in the classification catalogue, the applicant can determine the class under the classification rules and apply to the NMPA for class confirmation. This guide gives no class conclusion for such products.

Frequently asked questions

Where is a Class I walker filing submitted in China?

A domestic filer submits to the drug regulatory department of its local city-level government; an overseas filer acts through a designated legal entity in China that submits to the NMPA (Article 15). The filing is complete once a compliant dossier is submitted.

Can the test report for a Class I filing come from in-house testing?

Yes. Article 14 allows the filer's own in-house report or a report from a qualified medical device testing institution. Detailed requirements are set by the NMPA Class I filing document requirements.

Does an external test report have to come from a CMA-approved body?

Article 75 states that only testing bodies recognised by the State Council's certification and accreditation authority together with the national drug regulator may test medical devices, so choose a body whose CMA scope covers the standard used. SUNGO cannot issue CMA reports for GB/T 14728.1-2024, GB/T 14728.2-2024 or GB/T 19545.1-2012.

Can an axillary crutch report to GB/T 19545.2-2009 still be used?

GB/T 19545.2-2009 was abolished on 1 August 2026 and replaced by GB/T 19545.2-2026. Testing a new filing to the current edition is safer; whether an old-edition report is still accepted should be confirmed with the local filing department.

Can this site show the total number of walker filings?

Not directly. Device rows are a sample of up to 8 per company per market; they show which companies hold similar records, but not a total.

For a Class I walker or crutch filing in China, first confirm the product fits the catalogue entry and its intended use, then test to the current edition of the national standard, and check an external testing body's CMA schedule and edition years item by item. This guide is based on official public information as of 13 November 2026. The latest documents issued by the regulators govern.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.