To look up powered wheelchair adverse events and recalls in the United States, query FDA's openFDA endpoints for MAUDE reports and recall records by product code ITI (Wheelchair, Powered) and INI (Vehicle, Motorized 3-Wheeled). In data retrieved on 13 November 2026, the top ITI problem codes for 2021 to 2025 were electrical/electronic problems, failure to shut off and device remains activated, and most recalls since 2021 involved software or firmware, speed controls or electrical connections, which map mainly to ISO 7176-14 test items. Only aggregate figures are given; no company, brand or model is named.
Key points
- By FDA date received, ITI has 1,881 MAUDE records for 2021 to 2026 (2026 up to the 4 November database update) and INI has 205. ITI had 651 records in 2025 against 146 in 2024; almost all of the increase is Malfunction reports carrying the same three problem codes.
- For 2021 to 2025, the most frequent ITI problem codes were Electrical /Electronic Property Problem (573), Failure to Shut Off (560) and Device Remains Activated (558). 558 reports carry all three, so read them as one group of reports, not three separate problems.
- openFDA holds 71 ITI recall product records in 41 recall events. Of the 10 events initiated since 2021, five involve software or firmware; the rest involve speed controls, a battery harness, a seat part and Bluetooth security.
- FDA states that MAUDE data is not intended to evaluate adverse event rates, to evaluate a change in event rates over time, or to compare event rates across devices. The figures show what was reported, not how common a failure is.
- These failures map mainly to test items in ISO 7176-14 (switches, power loss, controller faults, software), ISO 7176-8 (strength and fatigue), ISO 7176-3 (brakes) and ISO 7176-25 and ISO 7176-31 (batteries and chargers).
Background and scope: what ITI and INI cover
ITI is the powered wheelchair (21 CFR 890.3860) and INI the motorized three-wheeled vehicle (21 CFR 890.3800). Both are Class II (performance standards) and require a 510(k). FDA defines a powered wheelchair as a battery-operated wheeled device that provides mobility to persons restricted to a sitting position (stair-climbing devices use code IMK), and a motorized three-wheeled vehicle as a gasoline-fueled or battery-powered device used for outside transportation by disabled persons.
The FDA classification page for ITI lists recognized consensus standards, including ISO 7176-3 (third edition, 2012), ISO 7176-8 (second edition, 2014), ISO 7176-14 (third edition, 2022), ISO 7176-25 (second edition, 2022), ISO 7176-31 (first edition, 2023-05) and ISO 7176-19 (third edition, 2022). ISO 7176-21 is listed in both the second edition (2009) and the third edition (2025-04). The INI page lists the same ISO 7176 parts, except that it does not include ISO 7176-19.
Only ITI and INI are covered here. ITI recall records include power-assist devices added to manual wheelchairs as well as complete wheelchairs.
What MAUDE data can and cannot tell you
MAUDE holds reports from mandatory reporters (manufacturers, importers, device user facilities) and voluntary reporters (such as health care professionals, patients and consumers). The openFDA documentation calls it a passive surveillance system whose data may be incomplete, inaccurate, untimely, unverified or biased. Because of under-reporting and unknown frequency of device use, incidence cannot be determined from it alone, and a report cannot establish a causal relationship between product and reaction. It also notes that widely used products may have more reports, and that increases in report totals are likely caused by improved reporting.
FDA's MAUDE page adds that MDR data is not intended to evaluate rates of adverse events, to evaluate a change in event rates over time, or to compare adverse event occurrence rates across devices. A report is classified as a death only if the reporter believes the device was or may have been a cause or factor, but submitting a report does not by itself show that the device caused the event.
Powered wheelchair MAUDE adverse events by year
The table counts records by date_received. Data was retrieved from openFDA on 13 November 2026 (database last updated 4 November 2026). Event type is chosen by the reporter; a few ITI records with a blank event type or Other are not shown separately.
| Year received | ITI total | ITI Malfunction | ITI Injury | ITI Death | INI total |
|---|---|---|---|---|---|
| 2021 | 166 | 42 | 111 | 12 | 41 |
| 2022 | 135 | 33 | 100 | 2 | 41 |
| 2023 | 223 | 88 | 129 | 6 | 32 |
| 2024 | 146 | 19 | 123 | 4 | 31 |
| 2025 | 651 | 518 | 129 | 3 | 27 |
| 2026 (to the 4 November update) | 560 | 412 | 139 | 5 | 33 |
ITI injury reports stayed between 100 and 139 a year; the 2025 and 2026 increase came mainly from malfunction reports. In 2025, 506 reports carried all three of Failure to Shut Off, Device Remains Activated and Electrical /Electronic Property Problem, 501 of them malfunctions. Without them, 145 records remain for 2025, close to the 146 of 2024. In 2026, up to the database update, 363 reports carried the same three codes. Under FDA's guidance, this change cannot be read as a rise in event rates or a change in product quality.
Top 10 ITI problem codes (received 2021 to 2025)
- Electrical /Electronic Property Problem: 573
- Failure to Shut Off: 560
- Device Remains Activated: 558
- Appropriate Device Problem Term/Code Not Available: 345
- Mechanical Problem: 92
- Use of Device Problem: 73
- Adverse Event Without Identified Device or Use Problem: 49
- Unintended Movement: 43
- Insufficient Device Problem Information: 36
- Break: 20
Next come Detachment of Device or Device Component (19), Unintended System Motion (18) and Device Tipped Over (14). One report can carry several codes, so code counts exceed report counts. The 172 INI records for 2021 to 2025 are mostly Injury reports (145), and 138 carry Appropriate Device Problem Term/Code Not Available; the few informative codes are Fire (6), Detachment (4) and Battery Problem (1).
Powered wheelchair FDA recalls: 71 records, 41 recall events
The openFDA recall endpoint lists one record per product, and one recall event (res_event_number) can cover several records. By product_code, ITI has 71 records in 41 recall events, initiated (event_date_initiated) between 2003 and 2026. At retrieval, 42 records were Terminated and 29 were Open, Classified (classified by FDA and not yet terminated). INI has 22 records in 14 recall events, and the latest initiation date is in 2021. Recall records and MAUDE reports are counted differently and cannot be converted into each other.
ITI recalls initiated since 2021 comprise 29 records in 10 events. Reasons by year of initiation:
- 2021: a seat baseplate pin broke in the field (the same event also covers INI); firmware followed an incorrect deceleration profile, decelerating harder than programmed.
- 2022: the battery harness could disconnect in shipment, over large obstacles or after a jolt, cutting power.
- 2023: a watch app for a power-assist device could fail, leaving the motor running without the user being able to stop it by tap gestures.
- 2024: firmware lacked a speed restriction with the seat raised more than 50 mm, risking a tip-over or fall; a faulty circuit board in a power-assist speed control remote could cause loss of control.
- 2025: firmware missing drive inhibit logic let the chair drive under restricted conditions; a faulty connection between a speed control dial and the motor could cause loss of control; a cybersecurity vulnerability was found in Bluetooth Low Energy communication.
- 2026: with "Remind Me Later" selected, reminder pop-ups could appear during use, causing an unexpected park brake stop while driving or a control system crash at rest. This one event has 13 product records.
Earlier ITI recalls mostly concern structural parts (seat-back welds, wheel hub bolts, seat mounting bolts, backrest bars), pinched or shorted wiring, charger harness short circuits, batteries overheating during charging and controller settings, plus one longer than expected stopping distance. INI recalls include over-inflated plastic tyres that could shatter, a fractured throttle spring that kept the scooter from stopping, overheating battery cables and control units, and weak seat-base welds. Apart from the 2021 seat baseplate pin event, all INI records date from 2014 or earlier.
Which ISO 7176 test items match these failures
The mapping below is this article's own grouping by failure mechanism. Clause numbers in brackets are those listed in the CNAS scope of SUNGO's Hefei laboratory and have not been checked one by one against the published standards; confirm them against the standard text before testing. For more on the ISO 7176-14 test items, see ISO 7176-14: power and control systems on powered wheelchairs.
| Failure category | MAUDE problem codes (ITI, 2021 to 2025) | Recall examples since 2021 | Related test items |
|---|---|---|---|
| Will not switch off, keeps running, unintended motion | Failure to Shut Off 560; Device Remains Activated 558; Unintended Movement 43; Unintended System Motion 18 | Power-assist motor running on after a watch app failure; speed control board or dial faults risking loss of control | ISO 7176-14: on/off switch (8.1), switching off while driving (8.7), controller command signal processing fault (7.2), controller output device fault (7.3) |
| Software and firmware behaviour | No software code in the top 15 | Wrong deceleration profile; no seat-raised speed limit; missing drive inhibit logic; reminder pop-up causing a park brake stop or crash; Bluetooth vulnerability | ISO 7176-14: software fault (8.16), display device (8.8); for seat-raised stability, also ISO 7176-2 rearward, forward and lateral dynamic stability (8, 9, 10). Cybersecurity is outside these test items and needs separate handling in design and risk management |
| Electrical faults, wiring and loss of power | Electrical /Electronic Property Problem 573 (558 of these also carry Failure to Shut Off and Device Remains Activated) | Battery harness disconnecting after a jolt, causing loss of power | ISO 7176-14: stopping ability on power loss (7.4), wiring routing (12.2), connectors (10.5, 11.3), circuit protection (9.3); for electromagnetic compatibility, ISO 7176-21 |
| Structural breakage and detached parts | Mechanical Problem 92; Break 20; Detachment of Device or Device Component 19 | Seat baseplate pin breaking | ISO 7176-8: static strength (8), impact strength (9), double-drum fatigue (10.3), drop fatigue (10.4) |
| Batteries and charging | Not in the ITI top 15; INI has Fire 6 and Battery Problem 1 | None since 2021; earlier, charger harness short circuits and batteries overheating during charging | ISO 7176-14: drive inhibit during charging (8.9); ISO 7176-25: charging a mismatched battery pack (5.2.6), charging capability (5.3); ISO 7176-31: on-board battery charger (5.2.1.3), charger performance (5.2.1.5) |
| Braking and stopping | No braking code in the top 15 | Reminder pop-up causing an unexpected park brake stop while driving; earlier, a longer than expected stopping distance | ISO 7176-3: parking brake (7.2), running brake in normal operation (7.3), emergency stop (7.5); ISO 7176-14: brakes (8.12) |
One failure category usually needs several tests. Loose harnesses or bolts also depend on assembly, incoming inspection and final inspection; a single type test report cannot show they will not occur in production.
Using the data in risk management and test planning
First, check the controller, switches, software functions (seat-raised speed limits, reminders, Bluetooth), batteries, charger and structural parts against the table and mark the failure modes that apply. Second, record for each the consequence, the protective measures, and the test item and acceptance basis that verify them. Third, record hardware, firmware and software versions in the submission documents, and after any firmware or software change, assess which tests to repeat. Several recall descriptions name the affected firmware or software version, so version records are part of the evidence.
How to look it up in the database
The ssy FDA product code pages show classification information and 510(k) records by code. The ITI code page shows the device name, Class 2, 21 CFR 890.3860, medical specialty, the 510(k) pathway, a third-party review flag and the FDA definition, then lists 510(k) clearances in a three-year window (number, decision date, applicant, device name, and summary PDFs for some). On 13 November 2026 the ITI page showed 98 clearances in that window, 70 archived summaries, 586 clearances in all years per FDA, and a latest decision date of 21 August 2026. The INI code page showed 40, 30, 351 and 10 August 2026 respectively. Both link to the official FDA classification page. The product code index showed 7,091 codes on the same date.
The ssy site does not offer MAUDE or recall search. To reproduce these figures, query openFDA directly. For adverse events, use api.fda.gov/device/event.json with search=device.device_report_product_code:ITI AND date_received:[20250101 TO 20251231], plus count=event_type.exact or count=product_problems.exact. For recalls, use api.fda.gov/device/recall.json with search=product_code:ITI and group records by res_event_number. FDA's MAUDE web search can also filter by Product Code and date received. openFDA adverse event data is updated weekly, so record the query string and retrieval date each time. The site's 510(k) lists help identify comparable devices under the same code but contain no adverse event information.
Test evidence and SUNGO
At SUNGO, the ISO 7176 tests in the table are performed by the Hefei laboratory (Hefei SUNGO Medical Technology Co., Ltd.); the Shanghai laboratory's scope does not include ISO 7176. Per the certificate scopes, CNAS (L24279) covers parts including ISO 7176-3:2012, ISO 7176-8:2014, ISO 7176-14:2022 (49 items), ISO 7176-25:2022 and ISO 7176-31:2023, for wheelchairs and mobility scooters. CMA (260015342945) covers ISO 7176-3, -8 and -31, but not -14 or -25. IAS (TL-1119) covers 20 parts of ISO 7176; ISO 7176-21 (electromagnetic compatibility) is within the IAS scope only and is not accredited by CNAS or CMA. CNAS and IAS are both signatories to the ILAC MRA.
ISO 7176-19 and IEC 62133 for battery cells are outside SUNGO's accreditation. The failure figures in this article come from public FDA data and do not mean that SUNGO has reproduced any reported event. For an introduction to the laboratory, see the SUNGO laboratory page; the official CNAS, CMA and IAS directories govern the accreditation scope.
Frequently asked questions
How do I search FDA adverse events for powered wheelchairs?
Use the openFDA device/event endpoint and filter by device.device_report_product_code:ITI and date_received, or enter product code ITI and a date range in FDA's MAUDE web search. For motorized three-wheeled vehicles, use product code INI.
Can MAUDE report counts be used to compare brand quality?
No. FDA states that MDR data is not intended to evaluate event rates, changes in rates over time or differences between devices, and report counts are also affected by under-reporting and the number of users.
What is the difference between FDA recall records and MAUDE reports?
MAUDE holds adverse event reports that are not extensively validated. A recall record reflects a correction or removal action a firm has initiated, with a reason and a status. Recall records are listed by product, so group them by event first; the 13 ITI records initiated in 2026 are one event.
Which tests should a powered wheelchair 510(k) cover first?
Start from the recognized consensus standards FDA lists for ITI, then set the scope from the product's design and risk analysis. On the public data, control systems and software (ISO 7176-14), strength (ISO 7176-8), brakes (ISO 7176-3) and batteries and chargers (ISO 7176-25, ISO 7176-31) deserve priority.
FDA adverse event and recall data helps find blind spots in design verification but does not support conclusions on its own. State the product code, date field and retrieval date, count by report and recall event, and tie each failure category to test items. This article is based on official public information available as of 13 November 2026. The latest documents from the regulatory authorities prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.