To bill Medicare under a power wheelchair or scooter HCPCS code, an FDA 510(k) is not enough. The product also needs a written coding verification review (CVR) from the Pricing, Data Analysis, and Coding (PDAC) contractor, and its RESNA test results must meet the minimum figures that CMS Policy Article A52498 sets for that group. This article lists the thresholds for power wheelchair (PWC) Groups 1 to 5 and power operated vehicle (POV) Groups 1 and 2, maps them to the ANSI/RESNA WC-1, WC-2 and ISO 7176 test methods, and shows how to look up the related FDA product codes and RESNA records on this site.

Key points

  • 510(k) and PDAC coding are separate tracks. A 510(k) decides whether the device can be marketed in the United States; Medicare billing depends on PDAC coding verification and the coverage conditions in LCD L33789. Under A52498, a power mobility device with no written CVR, or one PDAC has found meets no code, must be billed with code K0899.
  • Each group has written thresholds. A52498 divides PWCs into five groups and POVs into two, each with requirements for top speed, range, obstacle climb, dynamic stability incline and dimensions. All PWCs and POVs must also meet a 200,000-cycle fatigue test and a 6,666-cycle drop test, and Group 5 must pass crash testing (WC-19).
  • A code does not guarantee coverage. LCD L33789 states that Group 4 PWCs and Group 2 POVs have added capabilities that are not needed for use in the home, so they are denied as not reasonable and necessary if provided. A PWC or POV used only outside the home is not covered.
  • RESNA defines the test methods. A52498 bases performance testing on RESNA test parameters and ANSI/RESNA acceptance criteria. Speed, range, obstacle climb and dynamic stability correspond to WC-2 Sections 6, 4, 10 and 2; dimensions, fatigue and drop methods sit in WC-1 Sections 5 and 8.
  • Watch the revision date. A52498 revision R20 and L33789 revision R10 take effect on 20 December 2026. Their notes cover Group 2 seat elevator power wheelchairs, replacement rules and similar items, with no listed change to the group performance thresholds.

What is Medicare PDAC coding, and how does it differ from FDA 510(k)?

On the FDA side, 21 CFR 890.3860 classifies the powered wheelchair as Class II (performance standards), product code ITI. Mobility scooters fall under 21 CFR 890.3800, motorized three-wheeled vehicle (product code INI), also Class II. Both product codes go through 510(k). A 510(k) review looks at substantial equivalence to a legally marketed predicate device. The CMS group thresholds are not part of that review.

On the CMS side there are two documents. LCD L33789, Power Mobility Devices, sets the coverage conditions. Policy Article A52498, Power Mobility Devices - Policy Article, sets the coverage and payment rules, documentation requirements and coding guidelines. The contractors listed on both pages are the DME MACs (Durable Medical Equipment Medicare Administrative Contractors) for Jurisdictions A, B, C and D.

The coding rule in A52498 is plain. The POV and PWC base codes it lists may be used only for products with a written CVR from PDAC, and those products appear on the Product Classification List (PCL) on the PDAC website. A claim line billed under such a code for a product not on that code's PCL is denied as incorrect coding. For CVR documents and testing requirements, A52498 directs manufacturers and suppliers to the PDAC website or to PDAC directly.

This article covers only PWC and POV base thresholds and test methods. Manual wheelchairs, power assist systems (E0986), custom power wheelchair bases (K0013) and accessories are out of scope.

PWC Group 1–5 and POV Group 1–2 performance thresholds

A52498 states that there are five PWC groups and two POV groups, divided by performance. Each group is subdivided by beneficiary weight capacity, seat type, portability and power seating capability. The figures below are copied from the A52498 page as of 13 November 2026. Speed, range and dimensions stay in the original US customary units and are not converted.

GroupCodes listed in A52498Minimum top speed (mph)Minimum range (miles)Minimum obstacle climb (mm)Dynamic stability inclineDimensions and other (selected)
POV Group 1K0800–K080235206°Length ≤48 in, width ≤28 in, radius pivot turn ≤54 in
POV Group 2K0806–K0808410507.5°Length ≤48 in, width ≤28 in, radius pivot turn ≤54 in
PWC Group 1K0813–K081635206°Length ≤40 in, width ≤24 in
PWC Group 2K0820–K0831, K0835–K084337406°Length ≤48 in, width ≤34 in
PWC Group 3K0848–K08644.512607.5°Length ≤48 in, width ≤34 in
PWC Group 4K0868–K0871, K0877–K0880, K0884–K0886616759°Length ≤48 in, width ≤34 in
PWC Group 5K0890, K0891412609°Length ≤48 in, width ≤34 in; crash testing passed

Outside the table, every POV and PWC must meet a fatigue test of 200,000 cycles and a drop test of 6,666 cycles. For the portable PWCs in Groups 1 and 2 (K0813, K0814, K0820, K0821), the largest single component may not exceed 55 pounds. Group 5 is the pediatric group (L33789 calls it a Group 5 (Pediatric) PWC); it also requires at least five one-inch seat width options, at least three one-inch seat depth options and at least 3 inches of growth adjustment for width, depth and back height, among other items.

Weight capacity is a separate axis. In A52498, Standard Duty, Heavy Duty and similar terms describe weight capacity, not performance: a Group 3 Heavy Duty power wheelchair has Group 3 performance and a beneficiary weight capacity of 301 to 450 pounds. A device need not carry the full weight of its class, but its capacity must fall within the range; a 400-pound capacity is coded as Heavy Duty.

Which groups Medicare does not cover: Group 4 PWCs and Group 2 POVs

The two groups with the highest thresholds fall outside home-use coverage. LCD L33789 states that Group 2 POVs (K0806–K0808) and Group 4 PWCs have added capabilities that are not needed for use in the home, so if they are provided they are denied as not reasonable and necessary. A52498 adds that a POV or PWC used only outside the home is denied as noncovered.

For the other groups, payment depends on the beneficiary meeting the L33789 coverage criteria. That is work for the practitioner and the supplier, but it shapes which group a product is supplied under:

  • The treating practitioner must complete the face-to-face encounter within six months before the order date on the written order prior to delivery (WOPD) for the base item, and the base item needs that WOPD before delivery. The practitioner who completes the face-to-face encounter must be the one who writes the order for the base item.
  • Group 2 single power option and multiple power option PWCs, Group 3 PWCs and Group 5 PWCs also require a specialty evaluation by a professional with no financial relationship with the supplier, and direct, in-person involvement in wheelchair selection by a RESNA-certified Assistive Technology Professional (ATP) employed by the supplier.
  • Revision R10 of L33789, which takes effect on 20 December 2026, adds coverage criteria for non-complex Group 2 seat elevator PWCs (K0830, K0831), again including the specialty evaluation and ATP involvement.

Which RESNA and ISO 7176 tests cover each threshold

A52498 defines performance testing as the RESNA-based test parameters used to test power mobility devices. The device should meet or exceed the listed performance and durability figures for its category, and need not be tested with all possible accessories. Acceptance criteria are defined by ANSI/RESNA testing protocols.

CMS figureA52498 definition (summarised)RESNA methodISO 7176 part with the same title
Top end speedDetermined by the RESNA test for maximum speed on a flat hard surfaceWC-2 Section 6ISO 7176-6
RangeMinimum distance on a single battery charge, determined by the appropriate RESNA test for rangeWC-2 Section 4 (energy consumption for theoretical distance range)ISO 7176-4
Obstacle climbHeight of a solid obstruction climbed in the standing and/or 0.5 m run-up RESNA testWC-2 Section 10ISO 7176-10
Dynamic stability inclineMinimum slope at which the device stays stable in the most common seating configuration at the required weight capacityWC-2 Section 2ISO 7176-2
Dimensions, radius pivot turnPivot turn equivalent to the "minimum turning radius" in the ANSI/RESNA bulletinsWC-1 Section 5ISO 7176-5
Fatigue 200,000 cycles, drop 6,666 cyclesCounts only; no method or clause namedWC-1 Section 8 (static, impact and fatigue strengths)ISO 7176-8
Crash testing (Group 5 only)Successful completion of WC-19 testingWC-19Not mapped here

RESNA WC-1 and WC-2 section numbers match the ISO 7176 part numbers, with the same or similar titles; WC-2 Section 2 and ISO 7176-2, for example, both cover dynamic stability of electrically powered wheelchairs. Whether the texts are fully identical, and whether PDAC accepts reports issued to ISO 7176, should be checked against the standards and PDAC requirements. CMS gives the fatigue and drop counts without naming clauses. ISO 7176-8 fatigue testing includes a double-roller test and a drop test; check the standard text for how the counts map to clauses.

On editions, the RESNA wheelchair standards committee page states that the next version of WC-1 and WC-2 has been approved for publication by ANSI (page content as of 13 November 2026). Confirm the edition cited in the test plan before testing and check it against PDAC requirements.

What to line up before testing

  • Pick the target code first. A52498 codes are split by group, weight capacity, seat type, portability and power options. The tested configuration should match the descriptor of the code being sought.
  • Every item must pass. A52498 lists what each group "must meet". If any item falls short, the product cannot be coded in that group.
  • Test dynamic stability at the right weight. The incline is defined at the required weight capacity in the most common seating configuration, so a product aimed at Heavy Duty or another class needs testing prepared for that capacity.
  • Do not convert the thresholds yourself. The thresholds are in mph, miles, millimetres and degrees. If a report gives results in metric units, compare against the CMS figures as written.
  • A 510(k) report is not coding evidence by itself. Even where the methods are the same, check each result against the target group's thresholds.
  • Check coverage first. For the Medicare home-use channel, read the L33789 denial rules for Group 4 PWCs and Group 2 POVs.

Key dates

DateEvent
1 October 2015Original effective date of A52498 and L33789
5 November to 19 December 2026Notice period for revision R10 of L33789
20 December 2026A52498 R20 and L33789 R10 take effect: coverage criteria and documentation for non-complex Group 2 seat elevator PWCs, adjusted replacement rules; no listed change to group thresholds

How to check the database

For the FDA side, the product code pages on this site show the classification and clearance records (figures as displayed on 13 November 2026):

  • ITI product code page: Wheelchair, Powered; 21 CFR 890.3860; Class 2; 510(k); flagged as eligible for third-party review. It shows 98 clearances in the last three years (70 summaries archived for download) and 586 in all years according to FDA, plus the FDA classification definition and the device names of recent clearances by date.
  • INI product code page: Vehicle, Motorized 3-Wheeled; 21 CFR 890.3800; Class 2; 510(k). It shows 40 clearances in the last three years (30 summaries archived) and 351 in all years. Most recent device names are Mobility Scooter, Electric Scooter or similar.
  • RESNA standards list: 23 records, covering 12 sections of WC-1, 8 sections of WC-2 and 3 sections of WC-4. Each record page shows the edition year and the FDA recognition number; WC-2 Section 2, for example, is listed as the 2019 edition with FDA recognition number 16-208. The list also includes WC-4 Section 19 (Wheelchairs used as seats in motor vehicles). For how it relates to the WC-19 named by CMS, check RESNA and PDAC guidance.

You can also search by number: the regulatory search covers regulations and guidance, FDA product codes and 510(k) records from the last three years, and the testing search covers standard records and testing bodies. Standard text is copyrighted, so the site provides records and official links only. The CMS thresholds, code lists and the PDAC PCL are not held here; check them in A52498, L33789 and on the PDAC website. For why a tested range figure differs from what users experience, see Why Tested Range Beats Real-World Range on Powered Wheelchairs.

Test evidence and SUNGO (Hefei laboratory)

The relevant laboratory is Hefei SUNGO Medical Technology Co., Ltd.; the Shanghai laboratory's accreditation does not cover wheelchairs. According to the certificates:

  • IAS TL-1119: ANSI/RESNA WC-1 and WC-2 as full standards, and ISO 7176-2, -4, -5, -6, -8 and -10 as full standards. The RESNA standards are accredited by IAS only, not by CNAS, so accredited reports to RESNA can only be IAS reports.
  • CNAS L24279: ISO 7176-2:2017 chapters 8, 9 and 10 (rearward, forward and lateral dynamic stability); ISO 7176-4:2008 clause 7.1 continuous driving test and clause 7.2 manoeuvrability test; ISO 7176-5:2008 clauses 8.2 to 8.17 (overall length, overall width, turning diameter and others); ISO 7176-6:2018 maximum speed on the level; ISO 7176-8:2014 chapters 8 and 9 plus clause 10.3 double-roller fatigue, clause 10.4 drop fatigue and clause 10.5 parking brake fatigue; ISO 7176-10:2008 clauses 7.1 to 7.6 (six obstacle-climbing items).
  • CMA 260015342945: covers the corresponding items of the ISO 7176 parts above; the certificate states it is valid within the People's Republic of China.

CNAS and IAS are both signatories to the ILAC MRA. ISO 7176-19 and WC-19 crash testing are outside SUNGO's accreditation, so the Group 5 crash test must be placed with another body that has that capability. SUNGO does not handle PDAC applications and does not guarantee coding outcomes; whether PDAC accepts a given report is for PDAC to decide.

Frequently asked questions

Does an FDA 510(k) mean a power wheelchair can be billed to Medicare?

No. A 510(k) allows marketing. Under A52498 the base codes need a written PDAC CVR, and the beneficiary must also meet the L33789 coverage criteria.

What testing does PDAC coding verification require?

A52498 lists top speed, range, obstacle climb, dynamic stability incline, dimensions, 200,000-cycle fatigue and 6,666-cycle drop tests, plus crash testing for Group 5, with ANSI/RESNA methods. For submission details, A52498 points to the PDAC website or PDAC.

How is a power wheelchair without PDAC coding billed?

Under A52498, a product without a written CVR, or one that PDAC has found does not meet the requirements of any code, must be billed with code K0899.

Does Medicare cover Group 4 power wheelchairs?

L33789 states that Group 4 PWCs have added capabilities not needed for use in the home, so they are denied as not reasonable and necessary if provided. The same applies to Group 2 POVs.

Can SUNGO perform the WC-19 crash test required for Group 5?

No. ISO 7176-19 and WC-19 crash testing are outside SUNGO's accreditation, so another body with that capability is needed.

The Medicare thresholds are set in the CMS documents; testing turns them into evidence. Settle the target code and coverage position against A52498 and L33789, plan tests to the RESNA methods, then confirm submission requirements with PDAC. This article is based on official public information as of 13 November 2026. Refer to the latest documents from the regulatory authorities.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.