When China reclassifies a medical device, existing registrations are handled under two documents. NMPA Announcement No. 52 of 2026 sets how initial registration, renewal, change registration and Class I filing carry over when a device moves up or down a class, or into or out of device regulation. Announcement No. 53 of 2026 issues the revised catalogue adjustment procedure, whose transition principle is generally 2 to 3 years for products brought into regulation as Class II or III or moved up a class, and in principle no more than 5; each reclassification notice sets the actual period. Both were published by China's National Medical Products Administration (NMPA) on 1 June 2026, together with an official interpretation. This article covers each scenario, uses Announcement No. 132 of 2025 as an example, and shows how to self-check in this database.

Key points

  • No. 52 covers the transition; No. 53 covers the procedure. No. 52 sets how registrations and filings carry over; No. 53 sets how catalogues are adjusted and how transition periods are set. Both took effect on publication, and No. 53 repealed Announcement No. 60 of 2021, the previous procedure.
  • The specific reclassification notice comes first. Where a notice sets its own rules on certificate validity or on registration and filing, those apply; otherwise No. 52 applies.
  • There is no fixed transition period. For products brought into regulation as Class II or III, or moved up a class, the principle is generally 2 to 3 years (usually 2 without a clinical trial, 3 with one); it may exceed 3 years but in principle not 5. Each notice sets the actual period.
  • The adjustment date is not the end of the transition period. From the adjustment date, new applications follow the new class. From the end of the transition period, products without the required registration or filing may not be manufactured, sold or imported.
  • Check scope first. Announcement No. 132 of 2025 adjusted catalogue entries for 31 types of medical devices. Check its annex item by item rather than guessing from product names.

Background and scope

Under China's Regulations on the Supervision and Administration of Medical Devices, devices are regulated in three risk classes. NMPA is responsible for the classification rules and the classification catalogue and adjusts them after analysing changes in risk, having heard from registrants, filers, manufacturers and distributors, users and industry associations.

According to the official interpretation, these reclassification requirements apply to adjustments of specific devices or whole product categories, which NMPA makes public through announcements. Changes to the class or regulatory status of an individual product for regulatory reasons follow the registration rules and NMPA's separate announcement on product classification determination instead. No. 52 expressly covers medical devices including in vitro diagnostic reagents.

DocumentDate signedPublished on NMPA websitePurpose
Announcement No. 132 of 202530 December 20254 January 2026Adjusts catalogue entries for 31 types of medical devices
Announcement No. 52 of 202629 May 20261 June 2026Registration and filing rules after a change in class or regulatory status
Announcement No. 53 of 202629 May 20261 June 2026Revised Working Procedure for Dynamic Adjustment of the Medical Device Classification Catalogue
Official interpretation—1 June 2026Explains both announcements, the transition principles and the meaning of the key dates

China device reclassification up or down: the four scenarios in Announcement No. 52

No. 52 has four parts: registration and filing after a change in class; products brought into device regulation; products taken out of device regulation; and other provisions. The Chinese text is at NMPA Announcement No. 52 of 2026.

Scenario 1: moved to a higher class

  • From the adjustment date, initial registration applications are accepted under the new class.
  • Initial or renewal applications accepted before the adjustment date but not yet decided may still be reviewed under the original class; if approved, the remarks field states the new class, and validity may not run past the end of any transition period.
  • For certificates approved before the adjustment date, the registrant must register under the new class, and the original certificate may not run past the end of the transition period. Exception: where technical progress has made clear a product should be Class III, an existing Class II certificate stays valid until expiry.
  • If a new-class application was filed while the original certificate was valid, and that certificate expires before the end of the transition period (if any), the registrant may renew under the old class with the original authority, provided the product is safe and effective with no serious adverse event or quality incident on the market. The renewal notes the new class and is likewise capped at the end of the transition period.
  • Once the new-class certificate is issued, the registrant should apply to cancel the old one; otherwise the original authority cancels it.
  • Where a Class I product moves up, the filer must register under the new class and withdraw the Class I filing; filings not withdrawn by the end of the transition period (if any) are publicly cancelled by the original filing authority. If the filed information or technical requirements change before then, the filer updates the filing with the original authority, enclosing an explanation and the notice.

Scenario 2: moved to a lower class

  • When a certificate approved before the adjustment date needs renewal, the registrant applies under the new class to the corresponding authority for renewal or a Class I filing; the new certificate or filing record notes the original registration number in its remarks field.
  • Where a product moves to Class I, it may be filed as Class I from the adjustment date. For an initial registration accepted before that date, the applicant may still obtain the certificate or ask for the review to be terminated, with the file archived.
  • The competent authority may change. Under the Regulations, domestic Class II registrations go to the provincial regulator, and Class III and imported Class II and III registrations to NMPA; domestic Class I filings go to the municipal regulator, and imported Class I filings to NMPA.

Scenarios 3 and 4: brought into or taken out of device regulation

Products newly regulated as devices need registration or a Class I filing; from the end of the transition period (if any), products without either may not be manufactured, sold or imported. Products taken out of device regulation stop being regulated as devices from the status change date; registrations and filings are no longer accepted, and pending applications are terminated.

ItemMoved upMoved down
New applications from the adjustment dateAccepted under the new classAccepted under the new class; Class I filing available if moved to Class I
Accepted before the adjustment date, not yet decidedMay be approved under the old class with the new class noted; capped at the end of any transition periodMay be approved under the old class with the new class noted; review may be terminated if moved to Class I
Existing certificate expiringRegister under the new class; renewal under the old class possible, if conditions are met, up to the end of the transition periodRenew or file under the new class, noting the original number
Change registrationApply to the original authority; new class notedApply to the original authority; new class noted

Changes and responsibilities

Changes to certificates issued before the adjustment date, or renewed under the old class after the adjustment under Article 1(3) of No. 52, go to the original registration authority, and the change documents note the new class. During the transition period, registrants, filers and related companies should actively pursue registration or filing and remain responsible for quality and safety. Provincial regulators keep a register of affected products and supervise implementation.

How long is the reclassification transition period? Announcement No. 53

The revised procedure has 13 articles. According to the official interpretation, it changes four things. First, it adds the IVD classification rules and the IVD registration and filing measures as legal bases, and brings the Class I Product Catalogue and the IVD Classification Catalogue into scope. Second, it sets transition-period principles. Third, it details each step; for example, public comment may be shortened for products urgently needed for regulation. Fourth, it assigns duties: NMPA's medical device standards management centre compiles classification determinations into initial proposals, the executive committee of the classification technical committee reviews whole sub-catalogue proposals, and its specialist groups review proposals for specific product types.

ScenarioTransition principle (Article 10 of the procedure, as quoted in the interpretation)
Brought into regulation as Class II or III, or moved to a higher classGenerally 2 to 3 years
Of which no clinical trial neededGenerally 2 years
Of which clinical trial neededGenerally 3 years
Special casesMay exceed 3 years; in principle no more than 5 years

According to the interpretation, whether a period is set, and for how long, is decided after weighing product risk, industry development, clinical need and stakeholder views, so check each notice. See the official interpretation (in Chinese).

Class adjustment date vs end of the transition period

According to the interpretation, the "class adjustment date" and "status adjustment date" are the implementation dates set in the notice, not the end of the transition period. New applications follow the new class from the adjustment date. From the end of the transition period, products without the required registration or filing may not be manufactured, sold or imported, and certificates issued or renewed under the old class under No. 52 cannot remain valid beyond that date.

The interpretation also covers linking old and new numbers: a registrant or filer who needs the remarks field of the certificate or filing record to show the original registration or filing number, for example for tendering and procurement, should say so when submitting the initial registration, renewal or filing.

Announcement No. 132: how the reclassification of 31 device types is handled

The implementation section of No. 132 says:

  • From publication, applications are accepted under the new class.
  • Initial and renewal applications already accepted are reviewed under the class in which they were accepted, with the new class noted on approval.
  • Certificates moved from Class III to Class II stay valid until expiry; renewal under the new class must be filed with the corresponding authority at least 6 months before expiry.
  • Certificates moved from Class II to Class I stay valid until expiry, and the product must be filed before the certificate expires.
  • Change registrations during validity go to the original authority, and the change documents note the new class.

No. 132 predates No. 52 and has its own implementation rules; in line with No. 52's principle that a notice's specific provisions prevail, products within its scope follow No. 132 first. See NMPA Announcement No. 132 of 2025; the annex is the authoritative list of affected products.

A five-step self-check

  1. List every certificate and Class I filing with product name, class, catalogue entry, expiry date and issuing authority.
  2. Compare each item with the reclassification annex, not by product name alone.
  3. Identify the scenario (up, down, in or out) and note the adjustment date and end of any transition period.
  4. For upward moves, plan the dossier, registration testing and any clinical evaluation back from the transition deadline; for downward moves, plan renewal or filing before expiry and confirm the new authority.
  5. Raise any need to link old and new numbers at submission, and update labels, IFUs and quality documents that cite class or registration number.

How to check in the database

Figures below are as displayed on the site as of 12 November 2026.

  • Registrations: On Browse by registration market, China (NMPA) is marked "partial": 38,717 companies matched and at least 315,469 China registrations in total. The total is a lower bound, because 207 companies were capped at the first 300 records. By class: Class I 176,169, Class II 116,981, Class III 20,759. For a breakdown, see China device registration in numbers.
  • Search: Search covers company name, registration number, product and country. A company page shows its China registration count, Class I/II/III split, data date and sample records. Device rows are a sample of up to 8 per company per market, so a missing number does not mean the company lacks it. Certificates issued before an adjustment still carry the class at issue, so do not infer the new class from the database; the notice's annex is the reference.
  • Regulations: The Regulations database lists 282 regulations and guidance documents and 589 update records. China (NMPA) is marked "in progress", so use the NMPA links above for Nos. 52, 53 and 132. For comparison abroad, the search box covers FDA product codes (classification, class and market pathway) and documents such as 21 CFR 860 and MDCG 2021-24 rev.1.
  • Standards: The Standards index holds 4,792 entries, including 1,606 YY/T, 628 GB/T, 503 YY and 224 GB standards. Each entry gives bibliographic data, version status, applicable categories and laboratories, but not the text. Use it to check current versions before registration testing after an upward move.

FAQ

How long is the transition period after a medical device reclassification in China?

Under Article 10 of the procedure, generally 2 to 3 years for products brought into regulation as Class II or III or moved up a class (usually 2 without a clinical trial, 3 with one), and in principle no more than 5. It is a principle, not a fixed term; each notice sets its own.

My Class III certificate was moved to Class II. Is it still valid?

Under No. 132, it stays valid until expiry; renewal under the new class is due at least 6 months before expiry, with the provincial regulator for domestic Class II products. Other notices may set different rules.

Is the class adjustment date the same as the end of the transition period?

No. The adjustment date is the implementation date set in the notice, from which applications follow the new class. From the end of the transition period (if any), products without a registration or filing may not be manufactured, sold or imported.

What happens to an initial application accepted before the adjustment date?

Under No. 52, the regulator may continue the review under the old class and note the new class; with a transition period, validity is capped at its end. For moves to Class I, the applicant may instead ask to terminate the review and file as Class I.

What if a product is taken out of device regulation while an application is pending?

From the status change date, the review is terminated and the file archived; no new registrations or filings are accepted.

Summary

To handle a China medical device reclassification, check the notice's annex, separate the adjustment date from the end of the transition period, and plan registration, renewal, change and filing steps for your scenario. Information on the Sungo laboratory's accreditation scope is available on the Sungo laboratory page. This article is based on official public information as of 12 November 2026; the latest documents from the regulator prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.