A manufacturer based outside the UK must appoint a single UK Responsible Person (UKRP) to sell medical devices in Great Britain (England, Wales and Scotland), and the UKRP registers the devices with the MHRA. Under current law, CE marked devices are accepted until 30 June 2028 or 30 June 2030 at the latest, depending on the EU legislation they meet; indefinite recognition of EU MDR and EU IVDR CE marking, decided by the government on 21 October 2026, was not yet law as of 12 November 2026. Below: UKRP and importer duties, MHRA registration, CE deadlines, Northern Ireland, the 2026 reforms and how to check UK data in our database. Test reports and detailed post-market surveillance (PMS) rules are out of scope.
Key points
- Start with geography: Great Britain applies the Medical Devices Regulations 2002 (UK MDR 2002, SI 2002/618); Northern Ireland applies the EU MDR (Regulation (EU) 2017/745) and EU IVDR (Regulation (EU) 2017/746).
- The UKRP is mandatory: a non-UK manufacturer appoints one UKRP for all its devices, who registers them in the MHRA Device Online Registration System (DORS); unregistered devices cannot lawfully be placed on the Great Britain market.
- CE is accepted, with deadlines in current law: MDD and AIMDD devices until certificate expiry or 30 June 2028, whichever is sooner; IVDD devices until certificate expiry or 30 June 2030, whichever is sooner; EU MDR and EU IVDR devices until 30 June 2030.
- Indefinite CE recognition is not yet law: the 21 October 2026 decision needs legislation approved by Parliament and had not taken effect as of 12 November 2026.
- CE devices still need the UK steps: a UKRP, MHRA registration and PMS compliance, as the government response confirms.
UK medical device regulations: Great Britain vs Northern Ireland
Under the Windsor Framework, Northern Ireland applies the EU device regulations; Great Britain applies the amended UK MDR 2002, which originally implemented the MDD, AIMDD and IVDD, so UKCA requirements are still based on those directives.
| Item | Great Britain | Northern Ireland |
|---|---|---|
| Applicable law | UK MDR 2002 (as amended) | EU MDR since 26 May 2021; EU IVDR since 26 May 2022 |
| Marking | UKCA, or CE within the transitional deadlines | CE, plus the UK(NI) indication where a UK notified body performs mandatory assessment; UKCA not recognised |
| Representative of a non-UK manufacturer | A single UKRP | An authorised representative based in the EU or Northern Ireland |
| Device registration | MHRA registration before placing on the market | Since 28 May 2026, non-custom-made devices are registered in EUDAMED and MHRA registration is not required; custom-made devices remain registered with the MHRA |
Qualifying Northern Ireland goods with a valid CE marking already have indefinite access to Great Britain under the UK Internal Market Act 2020, separate from the 2026 CE decision. The rest of this article refers to Great Britain. See the MHRA guidance Regulating medical devices in the UK.
What is a UK Responsible Person (UKRP) and what must it do?
Regulation 7A of the UK MDR 2002 (regulations 21A and 33A for active implantables and IVDs) requires a manufacturer based outside the UK to appoint a sole UKRP, which must provide written evidence of the manufacturer's authority. The letter of designation uploaded at registration is a legal contract stating that the UKRP is the manufacturer's exclusive UKRP and listing its mandatory tasks.
The duties in regulation 7A(3), in summary (the legal text prevails):
- ensure the declaration of conformity and technical documentation have been drawn up and, where applicable, the conformity assessment procedure carried out;
- keep copies of the technical documentation, declaration of conformity and any certificate, including amendments, available for inspection;
- provide the MHRA on request with the information and documents needed to demonstrate conformity;
- provide samples or access where it has them; otherwise pass the request to the manufacturer and tell the MHRA whether the manufacturer will comply;
- cooperate with the MHRA on preventive or corrective action against device risks;
- immediately inform the manufacturer of complaints and reports of suspected incidents;
- if the manufacturer breaches its obligations, terminate the relationship and inform the MHRA and, if applicable, the UK Approved Body.
- Labelling: where UKCA is affixed (including dual CE and UKCA marking), the UKRP's name and address must appear on the labelling, outer packaging or instructions for use. CE-only devices do not need it.
- Importers and distributors: do not need to appoint a UKRP, but may act as one.
- Legacy accounts: DORS accounts of former Great Britain-based authorised representatives not updated to UKRP by 30 March 2026 have been closed.
UK importer vs UKRP: who does what
| Task | UKRP | Great Britain importer (not acting as UKRP) |
|---|---|---|
| MHRA device registration | Submits and maintains it for the non-UK manufacturer | Not responsible |
| Importer details | The manufacturer or UKRP gives the MHRA the importer's details, including its Great Britain place of business | Must tell the manufacturer or UKRP it intends to import |
| Technical documentation and MHRA requests | Keeps copies; answers requests for information and samples | No UKRP duties under regulation 7A |
| Storage, transport and marking checks | Not set out in regulation 7A | Storage and transport obligations; checks for CE or UKCA marking |
| Name and address on the label | Required where UKCA is affixed | Not required unless acting as UKRP for a UKCA device |
Distributors that supply only within Great Britain and do not import do not need to register with the MHRA.
MHRA device registration: how it works
All devices, including IVDs, custom-made devices and systems or procedure packs, must be registered before being placed on the Great Britain market, meaning the first supply of each individual device, paid or free. The MHRA only accepts registrations from UK-based manufacturers; for a non-UK manufacturer, the UKRP submits it in DORS. According to the MHRA device registration guidance, the information required includes (non-exhaustively):
- Manufacturer: legal name and address as on the labelling, company type, administrative contacts, UKRP letter of designation;
- Device: applicable legislation, class, GMDN code and term, Basic UDI-DI and UDI-DI where applicable, device name, model, catalogue number, UK Approved Body or EU notified body where applicable, and attributes such as sterility, latex and MRI compatibility;
- Documents: conformity assessment certificates or self-declarations.
- GMDN terms can be picked in DORS without GMDN Agency membership; not every GMDN term is a UK medical device.
- Each application covers up to 100 devices (GMDN) and 20,000 products in total.
- A device is registered only once the MHRA confirms it; the MHRA may first request technical documentation.
- Registration is not MHRA accreditation, certification or approval; such claims and MHRA logos are not permitted.
- Since 1 April 2026, fees are charged annually by GMDN Level 2 category, and the former "renew registration" function was removed on that date. See the MHRA fees page for amounts.
Updating a registration
Changes must be notified to the MHRA. A new UKRP, a change of legislation (for example MDD to EU MDR), or a change of GMDN assignment, class, sterility or brand name requires re-registration; addresses, contacts, the letter of designation, added products and new certificates under the same legislation are updated directly in DORS. Discontinued devices can be unregistered, but post-market obligations continue.
Once registered, the manufacturer's name, address, MHRA reference number, GMDN terms and codes and brand names are published on the MHRA Public Access Registration Database (PARD, pard.mhra.gov.uk), with the UKRP's name and address shown in the manufacturer record. IVDs under performance evaluation are not published.
How long is CE marking accepted in Great Britain?
SI 2023/627, made on 9 June 2023, substituted regulation 1ZA of the UK MDR 2002: regulations 19B (MDD) and 30A (AIMDD) cease at 23:59 on 30 June 2028; regulations 19C (EU MDR), 44ZA (IVDD) and 44ZB (EU IVDR) cease at 23:59 on 30 June 2030. The MHRA guidance gives these deadlines:
| Basis of CE marking | GB placing allowed until | Notes |
|---|---|---|
| MDD or AIMDD, valid certificate and declaration | Certificate expiry or 30 June 2028, whichever is sooner | Issued before 26 May 2021; certificates still valid under EU MDR Article 120 can be relied on |
| IVDD | Certificate expiry or 30 June 2030, whichever is sooner | Certificates still valid under EU IVDR Article 110 can be relied on |
| EU MDR or EU IVDR (including custom-made devices under the EU MDR) | 30 June 2030 | Notified body certificate must not have expired or been withdrawn |
| Class I devices self-declared under the MDD before 26 May 2021 that need a notified body under the EU MDR (for example up-classified devices, reusable surgical instruments) | 30 June 2028 | Class I devices self-declared under the EU MDR: until 30 June 2030 |
| Class I sterile or measuring devices with a valid MDD certificate | 30 June 2028 | Expired certificates cannot be relied on unless deemed valid by the EU |
Class I devices and general IVDs needing no notified body under either the directives or the EU MDR or IVDR cannot rely on directive-based CE; they must be self-declared under the EU MDR or IVDR, or go through UKCA. MDD or AIMDD custom-made devices can no longer be placed on the Great Britain market but may be registered under the UK MDR 2002.
Where UKCA requires third-party assessment, it must be done by an MHRA-designated UK Approved Body, which cannot issue CE certificates. Non-sterile, non-measuring Class I devices and general IVDs can be self-certified for UKCA.
UK 2026 reforms: CE recognition, pre-market rules and the Health Bill
| Strand | Official documents and dates | Status as of 12 November 2026 |
|---|---|---|
| Indefinite recognition of CE marked devices | Consultation 16 February to 10 April 2026; government response 21 October 2026 | Government will legislate, subject to Parliament; not in force, current deadlines apply |
| Pre-market regulations, including international reliance | Routes-to-market response 22 July 2025; MHRA implementation page updated 12 March 2026 | The MHRA said it would follow in 2026; no draft or made text on legislation.gov.uk |
| Three Health Bill amendments | Announced by the MHRA on 1 September 2026 | Bill at committee stage in the House of Lords; the MHRA says no immediate regulatory change |
1. CE recognition: decision published, legislation pending
The consultation drew 1,056 responses on three proposals: aligning MDD timelines with the EU; indefinite recognition of EU MDR and EU IVDR devices; and a reliance route for devices classified higher in Great Britain than in the EU. The government response takes forward the first two: MDD devices accepted until the end of the EU transition period (31 December 2028 at the time of the response), and devices CE marked under the EU MDR and EU IVDR recognised indefinitely. The classification-based reliance route is dropped. CE marked devices will still need a UKRP, MHRA registration and PMS compliance, with technical documentation and the certificate of conformity available in English.
The government says that, subject to Parliament, the changes will be made before the earliest transition deadline, 30 June 2028. Commencement will be set by the eventual regulations; until then the deadlines above apply.
2. Pre-market regulations and international reliance: still in drafting
The 22 July 2025 response named Australia, Canada, the EU and the USA as comparable regulator jurisdictions, and said the government intends to implement Routes 1, 3 and 4, extending Route 4 to software as a medical device and to implantable Class IIb and III devices compliant with US 510(k) requirements, subject to entire equivalence. Reliance routes for CE marked devices were not implemented at that time and were left for further consultation. The MHRA says a pre-market instrument will follow in 2026, introducing international reliance, UDI and some reclassification. As of 12 November 2026 it was not law; the final text will set the conditions.
3. Health Bill amendments: powers first, details later
The amendments announced on 1 September 2026 cover clearer MHRA powers to share information with UK bodies and international regulators; letting legislation track updates to technical standards and international guidelines, with simpler consultation for minor changes; and amending the Medicines and Medical Devices Act 2021 to enable a future licensing regime, under which devices entering Great Britain by a domestic route could need an MHRA licence instead of third-party conformity assessment. The MHRA says there is no immediate change, consultation and parliamentary scrutiny would follow, and UKCA remains meanwhile.
How to check UK registrations in our database
Our database carries MHRA PARD data, listed as a Tier A source (Open Government Licence v3.0) on the data sources and licences page.
- UK market overview: the "UK United Kingdom" section of the markets page shows, as of 12 November 2026, 1,214 companies matched and 19,367 registered device records, broken down by class, for example 6,444 Class I, 4,424 Class IIa, 1,458 Class IIb, 521 Class III and 4,632 IVD general. A table lists each company's record counts in China, the EU, the US, the UK, Brazil, Saudi Arabia and Canada (first 60 shown).
- Similar products and companies: the search page takes product or company names, with a scope filter and fuzzy matching by default. Company pages show the MHRA PARD record count under "Registrations by market", the UK responsible person (where recorded) under "Regulatory identity", and UK records in "Registration details (sample)". Device rows are a sample of up to 8 per company per market; a missing row does not mean no registration.
- Legislation and updates: as of 12 November 2026 the United Kingdom regulations page lists 36 regulations and guidance entries and 271 update records (page dated 18 September 2026), including SI 2002/618 and SI 2023/627, each linked to the official text. For the 21 October 2026 consultation outcome, use the MHRA page above.
- Confirm in PARD: PARD, whose published fields are described above, is the reference; as of 12 November 2026 the MHRA guidance says it cannot be searched by representative.
For an overall picture, see our earlier articles United Kingdom device registration in numbers and Chinese device manufacturers in the United Kingdom; their figures reflect the data when they were written.
FAQ
Do I need UKCA if I already have a CE certificate?
Under current law, not within the deadlines above. The government has decided to legislate for indefinite recognition of EU MDR and EU IVDR CE marking, but the 2028 and 2030 deadlines apply until the new regulations take effect.
Can a UK importer act as the UKRP?
Yes; the MHRA says nothing prevents an importer or distributor from also acting as UKRP. It still needs a letter of designation and takes on all regulation 7A duties.
Does a CE-only device need the UKRP on the label?
No. The UKRP's name and address must appear on the labelling, outer packaging or instructions for use only where UKCA is affixed, including dual marking.
Is a UKRP needed to supply Northern Ireland?
No. Northern Ireland applies the EU MDR and EU IVDR, so manufacturers outside the EU and Northern Ireland need an EU or Northern Ireland authorised representative; since 28 May 2026 non-custom-made devices are registered in EUDAMED.
Can a registered device be called "MHRA approved"?
No. The MHRA states that registration is not accreditation, certification or approval, and its logos may not be used in marketing, packaging or instructions for use.
In short: confirm the CE or UKCA basis and its deadline, appoint a single UKRP, and have it register and maintain the devices in DORS; the 2026 reforms change nothing until the law changes. To see the accreditation scope of the SUNGO laboratories, visit the SUNGO Lab page. This article is based on official information publicly available as of 12 November 2026; the latest regulator documents prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.