FDA annual registration for fiscal year 2027 (FY2027) must be completed between 1 October and 31 December 2026: pay the annual fee on the DFUF website to obtain a PIN and PCN, then submit annual registration in FURLS and review all listings, even if nothing has changed. Foreign establishments should also check that the details of their one designated U.S. agent are current. This article covers who must register, the steps, the U.S. agent's role, listing review, missed deadlines, and how U.S. buyers can check whether a supplier's registration is current. Fee amounts are not covered.

Key points

  • Required even if nothing has changed. 21 CFR 807.22 requires annual registration between 1 October and 31 December of each fiscal year, with all listings reviewed in the same period.
  • Pay first, then register. Pay on the Device Facility User Fee (DFUF) website to obtain a PIN and PCN, then submit in the Device Registration and Listing Module (DRLM) of FURLS. Until the PIN and PCN are entered, registration is incomplete.
  • The U.S. agent's role is limited. Under 21 CFR 807.40, the agent helps FDA communicate, answers questions about imported devices and helps schedule inspections. FDA states the agent has no responsibility for adverse event reporting (MDR) or for submitting 510(k)s.
  • FDA does not issue registration certificates, and registration is not approval. A database entry does not denote FDA approval, clearance or authorization of the establishment or its devices.
  • 31 December is a hard deadline. Establishments that miss it go into "failed to register" status and are not considered active. Under 807.40(c), a device may not be imported or offered for import unless it is listed and made at a registered foreign establishment.

Who must complete FDA annual registration

Registration and listing fall under 21 CFR Part 807. The FDA fiscal year runs from 1 October to 30 September (807.3(u)), so FY2027 is 1 October 2026 to 30 September 2027. For overseas firms the core provision is 807.40: a foreign establishment that manufactures, prepares, propagates, compounds or processes a device imported or offered for import into the United States must register, list and designate a U.S. agent.

FDA's Who Must Register, List and Pay the Fee page sets out which activities trigger each obligation. The categories most relevant to Chinese suppliers are below (the FDA page prevails):

Foreign establishment typeRegisterListPay fee
Manufacturer (including kit assemblers)YesYesYes
Contract manufacturer (including contract packagers)YesYesYes
Contract sterilizerYesYesYes
Foreign exporter (located outside the U.S., exporting to the U.S.)YesYesYes
Specification developer, relabeler or repackagerYesYesYes
Maker of accessories or components packaged or labeled for end usersYesYesYes
Maker of components, not finished devices, supplied only to a finished device manufacturerNoNoNo
Device under investigation under an IDENoNoNo

Two points are often missed. First, contract manufacturers register and list in their own right: FDA's FAQ says contract manufacturers and contract sterilizers of finished devices must do so whether they distribute the device themselves or return it to the manufacturer or specification developer, and that the device must first be listed by the manufacturer, specification developer or another listing party. Second, a trading company exporting to the United States as a foreign exporter must also register, list and pay. On the U.S. side, an initial importer registers and pays but does not list; it must identify the manufacturers of the devices it imports.

FDA published the FY2027 medical device user fee rates in the Federal Register on 30 July 2026 (document 2026-15335); they apply from 1 October 2026 to 30 September 2027. For the amounts, see FDA's MDUFA fees page.

FDA annual registration deadline: FY2027 timeline

DateWhat to doBasis
1 October 2026FY2027 begins; the annual registration window opens and FY2027 fee rates apply21 CFR 807.22(b)(1); FY2027 user fee notice
Several days before registeringPay on the DFUF website: the PIN comes with the order; the PCN is emailed once payment is processed, typically after several daysFDA payment process page
1 October to 31 December 2026Complete "Annual Registration" in FURLS/DRLM: review registration information, all listings and, for foreign establishments, known importers; enter the PIN and PCN807.22(b)(1), (b)(3); 807.41
31 December 2026Window closes; FDA treats the previous fiscal year's registration as active through the end of the calendar yearFDA "When to Register and List" page
From 1 January 2027Late establishments are in "failed to register" or "failed to list" status, not considered active, and may not appear on the FDA website until they submit and FDA processes the information807.22(c)
Any timeUpdate registration information within 30 days of a change; report changes to the U.S. agent's name, address or phone within 10 business days; update listings when devices are added, changed or discontinued807.22(b)(2), (b)(4); 807.40(b)(3)

How to complete FDA annual registration: PIN and PCN first, then FURLS

Since 1 October 2012, the fee must be paid before registration can be completed. The FY2027 user fee notice adds that an establishment cannot register without a PIN and PCN, and is not legally registered for FY2027 until the registration and payment steps are done. Based on FDA's pages:

  1. Place a DFUF order. In the FY2027 establishment registration fee store on the DFUF website, order for the number of establishments requiring payment. The PIN is in the order's upper right-hand corner.
  2. Pay. Use a credit card, ACH electronic check or wire transfer. A wire must quote the PIN or the payment may not be applied to the facility; add any sending-bank charges so FDA receives the full amount.
  3. Wait for the PCN. FDA emails it once payment is processed. This usually takes several days, so do not leave payment to the last week of December.
  4. Log in to FURLS. Use the existing account, open DRLM and choose "Annual Registration". FDA warns that creating a new account blocks access to the existing registration and delays completion.
  5. Review every item. Check establishment name and address, owner or operator, official correspondent, U.S. agent and each listing. A foreign establishment must also review its known U.S. importers per exported device and any agents or brokers used to import, entering anything missing during the annual period, or the registration cannot be completed.
  6. Certify, submit and enter the PIN and PCN. If no prompt appears, FDA asks firms to email the registration and listing mailbox.
  7. Keep the confirmation. FDA considers the establishment registered once the confirmation screen appears, and also states registration is complete when the fee is paid, information submitted electronically and FDA's email confirms all requirements are met. Keep both.

Also: FURLS passwords expire every 90 days (FDA FAQ), so check the login early; and FDA contacts registrants by email, so allow @fda.gov, @cdrh.fda.gov and @fda.hhs.gov through spam filters.

FDA U.S. agent requirements: what the agent does and does not do

Under 807.40(b), each foreign establishment submits its U.S. agent's details at initial registration and in updates, and may designate only one. FDA's U.S. Agents page describes the role as limited:

ItemRequirementBasis
NumberOne U.S. agent per foreign establishment; it may, but need not, also be the official correspondent807.40(b)
EligibilityResides or keeps a place of business in the U.S.; no post office box and not just an answering service; someone must answer the phone during normal business hours807.3(s); 807.40(b)(1); U.S. Agents page
Designation and confirmationThe establishment enters the agent in FURLS and the system emails the agent to confirm; if the agent declines or does not respond within 10 business days, the establishment is notified and must designate a new oneU.S. Agents page
ResponsibilitiesOn FDA's request, assist with communications, answer questions about devices imported or offered for import, and help schedule inspections; if FDA cannot reach the establishment, information given to the agent counts as given to the establishment807.40(b)(2)
Not responsible forAdverse event reporting under 21 CFR Part 803 (MDR); submitting 510(k)sU.S. Agents page
ChangesChanges to the agent's name, address or phone are reported within 10 business days by the establishment or the agent807.40(b)(3)

In practice, agree a forwarding deadline with the agent, since documents given to it count as delivered, and arrange MDR reporting and 510(k) submissions separately. Use the annual window to confirm the agent's contact details and that it has accepted the designation in the system.

FDA device listing: annual review, including 510(k)-exempt devices

807.22(b)(3) requires all listings to be reviewed and updated between 1 October and 31 December each fiscal year, with accuracy confirmed even if nothing changed. Key rules:

  • 510(k)-exempt devices must be listed too. 807.25(g)(2) requires the product code for exempt devices; FDA's instructions say the premarket submission number is left blank and the device is selected by product code.
  • Devices needing a 510(k) are listed only after clearance. FDA says such devices can be listed only after the 510(k), PMA or other submission is cleared or approved; if it is the only device, do not register the establishment before then.
  • Listing numbers. One per product code for exempt devices; one per premarket submission number for non-exempt devices (807.28(f)). A device under a product code or submission number not yet listed needs a new listing (807.28(b)).
  • All proprietary names. FDA's FAQ requires every name used to market the device in the U.S., including private-label names; "multiple" and "various" are not accepted, and a model or catalog number may be used.
  • Discontinue listings for discontinued devices. When all devices under a product code or submission number leave commercial distribution, discontinue the listing, and reactivate it if distribution resumes (807.28(d), (e)).

FDA registration certificates: FDA does not issue them

FDA's "Important Reminders about Registration and Listing" page says, in substance:

  • FDA does not "certify" devices or issue registration certificates to device facilities; a database entry after registration and listing does not denote approval, clearance or authorization of the facility or its devices.
  • Firms offering registration services are not affiliated with FDA, and FDA has not contracted any firm to register facilities, does not use outside contractors to notify or bill industry about registration, and does not require firms to hire anyone to register online.
  • The FDA logo is for official FDA use, not for private-sector materials.

21 CFR 807.39 is explicit: registration or a registration number does not denote approval of the establishment or its products, and any representation implying official approval because of registration is misleading and constitutes misbranding. So marketing materials should not say "FDA certified" or "FDA approved factory", and a service provider's "FDA registration certificate" is not an FDA document. To show status, give customers the registration and listing numbers to check in FDA's database.

How to verify a supplier's FDA registration

A U.S. importer checking a Chinese supplier would typically take these steps; suppliers can use the same steps as a self-check:

  1. Ask for the establishment registration number (or owner/operator number) and the listing numbers of the devices being bought; FDA states these are needed to import a listed device.
  2. Search FDA's Establishment Registration & Device Listing database by registration or FEI number, establishment name, product code, country and similar fields. FDA says it is updated weekly, usually on Mondays.
  3. Match the listings: FDA's public registration and listing data records listings by product code, and non-exempt devices also carry a 510(k) or other premarket submission number; confirm the code or number for the devices you buy appears under that establishment.
  4. Mind the timing: from October to December, the previous year's registration counts as active through 31 December. From January 2027, an establishment without FY2027 registration is in "failed to register" status and its information may no longer appear. With weekly updates, late-December registrations may show late, so re-check year-end orders in January.

How to look it up in the database

This database's U.S. data comes from FDA openFDA (public domain, CC0). Figures below are as displayed on 12 November 2026:

  • Overview: the United States section of Browse by registration market shows 26,108 companies with records and 282,111 listed device types (Class I 159,301; Class II 112,161; Class III 5,441; unclassified 5,208). Units differ between markets, so do not compare totals across markets.
  • One company: search the name in Search. The company page's United States (FDA openFDA) panel shows listed device types with the Class I/II/III split, registration record and 510(k) counts, 510(k) clearances by year and listed device names (first 8 only). For a competitor, this shows how many device types it lists in the U.S. and which kinds.
  • Product codes: FDA device product codes holds 7,091 codes by medical specialty. Each code page shows device name, class, 21 CFR section, marketing pathway (such as "510(K) Exempt"), any GMP-exempt flag, FDA's all-years 510(k) count and a link to FDA's classification page. Confirm code and pathway here before listing an exempt device.
  • Look up by number or name: device names on company pages are English classification names; search them in Regulation search, which covers regulations and guidance, FDA product codes and recent 510(k) clearances. Searching by number is most precise.
  • The regulation record: the 21 CFR 807 record gives the regulation's basic details and a link to the official eCFR text.

For aggregate figures, see Chinese device manufacturers in the United States and United States device registration in numbers (data as of 7 September 2026).

Three limitations: device rows are a sample of up to 8 per company per market, so a missing record does not mean the company lacks it; company pages do not show whether FY2027 registration is complete; and for current status, purchasing or compliance decisions, FDA's official database prevails.

FAQ

Is FDA annual registration required if nothing has changed?

Yes. 807.22 requires annual registration between 1 October and 31 December every fiscal year, and confirmation that listings are accurate, whether or not anything changed.

Can an FDA U.S. agent file MDR reports or submit 510(k)s?

Neither is part of the agent's regulatory role; FDA states the agent has no responsibility for either. The establishment must arrange both separately.

Is an FDA registration certificate from a service provider valid?

It is not an FDA document. FDA issues no registration certificates to device facilities, and registration is not approval; status is shown by the records in FDA's database.

Can a small business get the FDA annual registration fee waived?

FDA's FY2027 fees page and fee notice say FDA may, but need not, waive the annual registration fee for a business that qualifies through the Small Business Determination (SBD) program and shows the fee is a financial hardship. Initial registration is excluded, and FDA decides eligibility.

What happens if FDA annual registration is missed by 31 December?

Under 807.22(c), the establishment stays in "failed to register" status until it submits and FDA processes the information. Devices from an unregistered foreign establishment may not be imported, so pay and register promptly and confirm status with FDA's registration and listing help desk.

FY2027 annual registration comes down to three things: pay and enter the PIN and PCN before 31 December; review registration information, listings and importers; and confirm the U.S. agent's details are valid and the designation accepted. Leaving a few days for payment to clear avoids a last-week scramble. This article is based on official public information available as of 12 November 2026; the latest documents from the regulator prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.