A mobility scooter sold in the United States normally needs an FDA 510(k). FDA classifies it under product code INI, regulation 21 CFR 890.3800 "Motorized three-wheeled vehicle," Class II, with the option of 510(k) third-party review. As for which test reports the 510(k) needs, the 30 INI 510(k) summaries archived in this database give a direct answer: the ISO 7176 parts on stability, brakes, range, dimensions, speed, strength, climatic conditions and obstacle climbing appear in all 30, and electromagnetic compatibility (EMC), ignition resistance of seating parts and biocompatibility appear in most of them. This guide covers the code, the test list, version changes, the FDA guidance and how to search this database.
Key points
- The FDA product code for mobility scooters is INI, regulation 21 CFR 890.3800, Class II, 510(k) pathway. The FDA classification page (last updated November 9, 2026, according to the page) shows the code is not GMP exempt and is eligible for 510(k) third-party review.
- The code name says "3-wheeled," yet at least 19 of the 30 archived INI summaries describe devices that include a four-wheel configuration with two front and two rear wheels. Decide the code from the device identification and from comparable cleared 510(k)s, not from the wheel count.
- All 30 summaries mention ISO 7176-1, -2, -3, -4, -5, -6, -7, -8, -9, -10, -13 and -15. IEC 60601-1-2 and ISO 16840-10 each appear in 29, ISO 7176-25 and the ISO 10993 series each in 26, and ISO 7176-21 in 25. None mentions ISO 7176-19.
- On May 25, 2026, FDA added ISO 7176-21:2025 to its recognized consensus standards (recognition number 16-240). Declarations of conformity to the 2009 edition (16-166) will be accepted only until July 2, 2028. Where the archived summaries state an edition of ISO 7176-21, it is the 2009 edition.
- FDA's July 26, 1995 510(k) guidance for wheelchairs and motorized three-wheeled vehicles is still listed on the INI classification page. Its items on tip-over, braking, charger safety, range, speed, EMC and software are a useful framework, but the guidance is not binding.
Which FDA product code covers mobility scooters: INI and 21 CFR 890.3800
The identification paragraph of 21 CFR 890.3800 defines a motorized three-wheeled vehicle as a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons. The classification paragraph reads "Class II (performance standards)." This is the first test for INI: the intended use in the submission should match "intended for medical purposes" and "transportation by disabled persons."
Within 21 CFR Part 890, "Physical Medicine Devices," each wheelchair-related section has its own product code. The FDA recognized standards entry for ISO 7176-21:2025 lists, among others, INI (890.3800), mechanical wheelchair IOR (890.3850), powered wheelchair ITI (890.3860), special grade wheelchair IQC (890.3880), stair-climbing wheelchair IMK (890.3890) and standup wheelchair IPL (890.3900). For a product between a scooter and a powered wheelchair, compare both identification texts and the codes used by comparable cleared 510(k)s.
The INI device name is "Vehicle, Motorized 3-Wheeled," but cleared submissions show that four-wheel scooters are also filed under INI. Of the 30 archived INI summaries, at least 19 describe devices with two front wheels and two rear wheels, and some submissions cover both three-wheel and four-wheel models.
INI code at a glance
| Item | INI (FDA classification database) |
|---|---|
| Device name | Vehicle, Motorized 3-Wheeled |
| Regulation | 21 CFR 890.3800 Motorized three-wheeled vehicle |
| Device class | Class II |
| Submission type | 510(k) |
| Review panel | Physical Medicine |
| Premarket review | Office of Neurological and Physical Medicine Devices (OHT5), Neuromodulation and Physical Medicine Devices (DHT5B) |
| GMP exempt | No |
| Implanted / life-sustaining device | No / No |
| Third-party review | Eligible for the 510(k) Third Party Review Program |
| Guidance document | 1995 510(k) guidance for wheelchairs and motorized three-wheeled vehicles |
This information comes from the FDA classification page for INI, which shows a last-updated date of November 9, 2026. The live FDA page governs.
Mobility scooter 510(k) testing requirements: what 30 summaries cite
As of November 13, 2026, this database's INI code page lists 40 510(k)s, and summaries for 30 of them are archived on the site. We read all 30 (one is a scan, checked page by page) and counted the summaries in which each standard number appears:
| Standard | Subject | Summaries mentioning it (of 30) |
|---|---|---|
| ISO 7176-1, -2 | Static and dynamic stability | 30 |
| ISO 7176-3 | Effectiveness of brakes | 30 |
| ISO 7176-4 | Energy consumption and theoretical range | 30 |
| ISO 7176-5, -7 | Dimensions, mass and maneuvering space; seating and wheel dimensions | 30 |
| ISO 7176-6 | Maximum speed | 30 |
| ISO 7176-8 | Static, impact and fatigue strength | 30 |
| ISO 7176-9 | Climatic tests | 30 |
| ISO 7176-10 | Obstacle-climbing ability | 30 |
| ISO 7176-13 | Coefficient of friction of test surfaces | 30 |
| ISO 7176-15 | Information disclosure, documentation and labeling | 30 |
| ISO 7176-11 | Test dummies | 29 |
| ISO 7176-14 | Power and control systems | 29 |
| ISO 16840-10 | Ignition resistance of postural support devices | 29 |
| IEC 60601-1-2 | EMC of medical electrical equipment | 29 |
| ISO 7176-25 | Lead-acid batteries and chargers | 26 |
| ISO 10993 series | Biological evaluation | 26 |
| ISO 7176-21 | EMC of powered wheelchairs, scooters and battery chargers | 25 |
| ISO 7176-22 | Set-up procedures | 20 |
| ISO 7176-16 | Ignition resistance of postural support devices (not listed on the INI page) | 14 |
| IEC 62133 | Safety of secondary cells and batteries with alkaline or other non-acid electrolytes | 8 |
| ISO 7176-31 | Lithium-ion battery systems and chargers | 5 |
| IEC 60601-1 (incl. ANSI/AAMI ES 60601-1) | General requirements for medical electrical equipment safety | 4 |
| ANSI/RESNA WC series | U.S. wheelchair standards | 3 |
| ISO 7176-19 | Wheelchairs used as seats in motor vehicles (crash testing) | 0 |
The count includes predicate comparison tables, so "mentioned" does not always mean "tested on the new device," and a summary does not show every test. Even so, the numbers show:
- The basic ISO 7176 parts are common to all: they appear in all 30 summaries, for three-wheel and four-wheel designs alike.
- EMC is usually covered by two standards side by side: most summaries cite both ISO 7176-21 and IEC 60601-1-2. Only one of the 30 lists neither separately, and that one cites EN 12184:2022.
- Battery standards follow the battery type: ISO 7176-25 covers lead-acid batteries and chargers. For lithium batteries, 8 summaries mention IEC 62133 and 5 mention ISO 7176-31, with 3 mentioning both.
- Biocompatibility is easy to miss: 26 summaries mention the ISO 10993 series, usually stating that materials in direct contact with the user meet ISO 10993-5, ISO 10993-10 and similar requirements.
- Clinical studies are usually not part of the submission: 24 of the 30 state explicitly that no clinical study was conducted, and the rest also rely on non-clinical testing to support substantial equivalence.
FDA recognized consensus standards and the ISO 7176-21 transition
The recognized consensus standards listed on the INI classification page include ISO 7176 parts -1, -2, -3, -4, -5, -6, -7, -8, -9, -10, -11, -13, -14 (2022 edition), -15, -21 (two entries, 2009 and 2025 editions), -22, -25 (2022 edition), -30 and -31 (2023 edition); several sections of ANSI/RESNA WC-1:2019 and WC-2:2019 and some sections of WC-4:2017; and ISO 16840-10. Two version changes need attention:
- ISO 7176-21: On May 25, 2026, FDA recognized the third edition (April 2025) as recognition number 16-240. According to the transition statement in that entry, declarations of conformity to the 2009 edition (16-166) will be accepted until July 2, 2028, and not after that date.
- ISO 7176-31: The first edition (2023) of the lithium-ion battery system and charger standard was recognized on May 25, 2026, as recognition number 16-241.
The summaries also cite older editions such as ISO 7176-14:2008 and ISO 7176-25:2013, while the classification page lists the 2022 editions of -14 and -25. Check each edition against the FDA recognized standards database instead of copying an older summary.
Content listed in the 1995 FDA guidance
The guidance linked from the INI classification page is "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles," dated July 26, 1995 (reformatted December 18, 1997). Its opening note says it was written before FDA implemented Good Guidance Practices on February 27, 1997, does not bind FDA or the public, and allows an alternative approach that satisfies the applicable statute and regulations. The main items relevant to scooters are:
- Device description: all components and accessories; engineering diagrams with length, width and height, or front, rear and side views; dimensions and weight; operating surfaces; and maximum weight capacity.
- Tip-over tests: results of tilt-over tests performed laterally, posteriorly and anteriorly with a user of maximum allowable weight and maximum allowable seat height.
- Materials: the materials of each component, material strength tests on the frame and wheels, and flame-retardance test results for the upholstery.
- Labeling: draft labeling and operator instructions; electromagnetic interference labeling per FDA's March 26, 1994 letter to wheelchair manufacturers; the prescription statement under 21 CFR 801.109; and cleaning instructions.
- Additional items for powered devices: the type of motor; a braking system drawing with electrical schematics, how the device stops if the electrical brake fails, and braking time and distance; the electrical safety test results for the battery charger and the ground leakage values (the guidance sets a maximum); the number and type of batteries; and the maximum distance on a full charge.
- Speed and incline: speed limits in slow, normal and high control (the guidance states a maximum of 6 mph); the maximum safe incline anteriorly, posteriorly and sideways with the center of gravity at its highest; the highest curb clearance; and the maximum safe speed and minimum turning radius to prevent tip-over.
- EMC: documentation following the April 3, 1995 draft EMC addition to ISO 7176-14.
- Software: for control software, a level-of-concern determination under FDA's August 29, 1991 reviewer guidance for computer-controlled devices, with design, verification and validation documentation and a software hazard analysis.
- Substantial equivalence: identification of the predicate device (its 510(k) number may be used) and a comparison table covering intended use, functions, and physical and safety characteristics.
These items reflect 1995 practice; the EMC item, for example, refers to a 1995 draft, while the classification page now lists ISO 7176-21. Use the guidance as a framework and take test editions from the recognized standards.
How to look up mobility scooter 510(k)s in this database
- Open the INI product code page. As of November 13, 2026, it shows Class 2, 21 CFR 890.3800, the Physical Medicine specialty, a 510(k) pathway and the third-party review flag. In its three-year window it lists 40 510(k)s (8 in 2024, 15 in 2025 and 17 in 2026), with 30 summaries archived for direct download, and 351 510(k)s for the code across all years.
- Open any entry to reach its record page, with the date received, decision date, decision, review pathway, advisory committee, the archived summary PDF and a link to FDA's official record. Where no summary is archived, the page states why, for example that the applicant filed a 510(k) Statement, so FDA does not publish a summary.
- In each summary, read the "Non-Clinical Tests" section, note the standards and editions, and compare them with the recognized standards above.
- If you are unsure whether a product belongs under INI or another code, open the Physical Medicine code list and compare the device names and 21 CFR sections of neighboring codes such as ITI (powered wheelchair) and IOR (mechanical wheelchair). As of November 13, 2026, the FDA product code library on this site holds 7,091 codes, 220 of them in the Physical Medicine group.
Mind the data lag: the latest decision on this site's INI page is dated August 10, 2026, while openFDA (data updated November 2, 2026) shows 21 INI 510(k)s in 2026, the latest decided on September 29, 2026. Verify specific submissions in FDA's 510(k) database. Data on this site are a search aid, not regulatory strategy advice.
Test evidence and SUNGO
Most scooter 510(k) tests are ISO 7176 bench tests. SUNGO's wheelchair and scooter testing is done by the Hefei laboratory (Hefei SUNGO Medical Technology Co., Ltd.); the Shanghai laboratory does not perform these tests. In the Hefei CNAS scope schedule, the test object listed for the ISO 7176 parts is "wheelchairs and electric mobility scooters." CNAS and IAS are both signatories to the ILAC MRA. The table maps the standards counted above to the Hefei scope:
| Standard | Hefei CNAS (L24279) | Hefei IAS (TL-1119) |
|---|---|---|
| ISO 7176-1, -2, -3, -4, -5, -7, -8, -9, -10, -11, -13, -22 | In scope, for the editions and clauses listed in the certificate | In scope |
| ISO 7176-6 | Maximum speed on a level surface only | In scope |
| ISO 7176-14, -15, -25, -31 | In scope (2022, 1996, 2022 and 2023 editions respectively, for the clauses listed in the certificate) | In scope |
| ISO 16840-10 | Clauses 7.2, 7.3 and 7.4 | In scope |
| ISO 7176-16, ISO 7176-21 | Not in scope | In scope |
| IEC 60601-1, IEC 60601-1-2 | Not in scope | In scope |
| ANSI/RESNA WC-1, WC-2 | Not in scope | In scope |
| ISO 7176-19, ISO 10542, IEC 62133, UN 38.3 | Not in scope | Not in scope |
| ISO 10993 series biological tests | Not in scope | Not in scope |
The IAS scope lists whole standards without edition years; the certificate governs the edition. Items only in the IAS scope, such as ISO 7176-21 EMC and IEC 60601-1 safety, cannot be issued as CNAS-accredited reports. Crash testing and battery-cell safety testing need another body with the appropriate accreditation. Biological tests for biocompatibility can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report states the subcontracted items and the subcontractor. The certificates govern the scope, which can be viewed on the SUNGO laboratory page of this site.
Frequently asked questions
Are four-wheel mobility scooters also filed under INI?
Cleared submissions say yes: at least 19 of the 30 archived INI summaries describe devices with two front and two rear wheels. FDA's decision on each submission sets the final code.
Is a mobility scooter exempt from 510(k)?
No exemption applies. As of November 13, 2026, the FDA classification page shows the INI submission type as 510(k), and the text of 21 CFR 890.3800 contains no exemption.
Can a mobility scooter 510(k) use third-party review?
Yes. FDA describes the program as voluntary: an FDA-accredited third-party review organization reviews the submission and sends FDA a recommendation, and FDA makes the final decision. Payment is between the submitter and the organization, with no separate FDA user fee. In openFDA data, all 44 INI 510(k)s decided from April 18, 2024 to September 29, 2026 carry a third-party flag of "N."
Is the 1995 FDA guidance still useful?
It is not binding, and documents it cites, such as the 1995 EMC draft, predate current standards. FDA still lists it for INI, so it works as a completeness checklist; test editions should follow the FDA recognized standards database.
Can a SUNGO test report be used directly in an FDA 510(k)?
A test report is one part of a 510(k). Whether FDA accepts it depends on the complete submission and FDA's review, and this guide does not promise that any report will be accepted. The accreditation mark on a report depends on which certificate covers the item.
To prepare a mobility scooter 510(k), confirm the code and comparable submissions on this site's INI page, check editions against the FDA recognized standards database, and complete the description, labeling and comparison material with the 1995 guidance as a framework. This guide is based on official public information as of November 13, 2026. Refer to the latest documents from the regulatory authority.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.