The EU AI Act's high-risk obligations for AI medical devices apply from 2 August 2028, not 2 August 2027, after Regulation (EU) 2026/1744 amended the timetable. Only the date moved: the MDR and IVDR remain in Section A of Annex I, so AI-enabled devices that need a notified body are still high-risk AI systems, assessed within the MDR/IVDR procedure.
AI-enabled devices sold in the EU must meet the MDR (Regulation (EU) 2017/745) or the IVDR (Regulation (EU) 2017/746) and the Artificial Intelligence Act (Regulation (EU) 2024/1689, the AI Act). The change came from Regulation (EU) 2026/1744, signed on 8 July 2026 and known as the Digital Omnibus on AI.
Key points
- One more year. Regulation (EU) 2026/1744, published in the Official Journal on 24 July 2026 and in force since 27 July 2026, moves Chapter III, Sections 1 to 3 for high-risk AI under Article 6(1) and Annex I, which covers MDR and IVDR devices, from 2 August 2027 to 2 August 2028. The Annex III route moves to 2 December 2027.
- A delay, not an exemption. The MDR and IVDR remain in Section A of Annex I. AI-enabled devices that need a notified body are still high-risk AI systems.
- Assessment still runs through the MDR/IVDR. Under the amended Article 43(3), the MDR/IVDR conformity assessment procedure applies and the AI Act's Chapter III, Section 2 requirements form part of it. Notified bodies notified under Section A legislation must apply for AI Act designation by 28 January 2028.
- "Section B" is only a proposal. The Commission's MDR/IVDR revision proposal, COM(2025) 1023, would move both to Section B; Regulation (EU) 2026/1744 did not.
- Use the time to merge documentation. AI Act requirements on data governance, logging, transparency, human oversight and monitoring can be built into existing MDR/IVDR files.
How the EU AI Act applies to medical devices: two high-risk routes
The AI Act entered into force on 1 August 2024 and applies in stages. It has two high-risk routes. Article 6(2) and Annex III list areas of use. Article 6(1) has two conditions: the AI system is a safety component of a product covered by the legislation listed in Annex I, or is itself such a product; and that product must undergo third-party conformity assessment. In Section A of Annex I, point 11 is the MDR and point 12 is the IVDR.
MDCG 2025-6 (also AIB 2025-1), a joint document of the AI Board and the Medical Device Coordination Group published in June 2025, applies this to device classes. The AI Act does not move a device into a higher risk class; the MDR/IVDR class decides whether the AI system is high-risk. The table is based on Table 1 of that document, which is non-exhaustive; a row printed there as "IVDR Class A)" is read from context as sterile Class A.
| MDR/IVDR classification | Notified body involved | Article 6(1) met |
|---|---|---|
| MDR Class I (non-sterile, non-measuring, not reusable surgical) | No | No |
| MDR Class I (sterile, measuring or reusable surgical) | Yes | Yes |
| MDR Class IIa, IIb, III | Yes | Yes |
| MDR Annex XVI products (except non-invasive products classified as Class I under the guidance) | Yes | Yes |
| IVDR Class A (non-sterile) | No | No |
| IVDR Class A (sterile) | Yes | Yes |
| IVDR Class B, C, D | Yes | Yes |
| In-house devices under Article 5(5) | No | No |
MDCG 2025-6 adds that most medical device AI (MDAI) is MDR Class IIa, IVDR Class B or higher. Regulation (EU) 2026/1744 also rewrote the "safety component" definition and added Article 6(1a) to (1c): AI used solely for non-safety aspects such as user assistance, performance optimisation, service efficiency, automation, convenience or quality control is not a safety component unless its failure would endanger health and safety; and a product needing third-party assessment only for non-health-and-safety risks, such as electromagnetic interference, does not meet condition (b) of Article 6(1). This mainly concerns hardware with embedded AI.
EU AI Act timeline for medical devices: what Regulation (EU) 2026/1744 changed
The change is in the third paragraph of Article 113. Chapter III, Sections 1, 2 and 3 (except Article 6(5)) now apply as follows:
| High-risk route | Original date | After the amendment |
|---|---|---|
| Article 6(1) and Annex I (including MDR/IVDR devices) | 2 August 2027 | 2 August 2028 |
| Article 6(2) and Annex III | 2 August 2026 | 2 December 2027 |
Section 1 holds the classification rules, Section 2 the requirements of Articles 8 to 15, and Section 3 the obligations of providers, deployers and other parties. Key dates:
| Date | What happens | Legal basis |
|---|---|---|
| 2 February 2025 | Chapters I and II apply, including Article 4 (AI literacy) and Article 5 (prohibited practices) | Art. 113(a) |
| 2 August 2025 | Chapter III, Section 4 (notifying authorities and notified bodies) and other provisions apply | Art. 113(b) |
| 27 July 2026 | Regulation (EU) 2026/1744 enters into force | Its Article 4 |
| 2 August 2026 | General application date; Article 50 transparency obligations apply (check the legal text) | Art. 113, second paragraph |
| 2 December 2026 | New Article 5 prohibitions added by Regulation (EU) 2026/1744 (Article 5(1), first subparagraph, points (ba) and (bb), and Article 5(1a) and (1b)) apply | Art. 113(a) |
| 1 August 2027 | Commission guidelines on Articles 8(2), 9(10) and 17(3) and Section A legislation due | Art. 96(1)(g) |
| 2 August 2027 | Commission delegated acts due on limiting requirements where sector law gives equivalent protection | Art. 2(13) |
| 2 September 2027 | Commission guidance and template on the post-market monitoring plan due | Art. 72(3) |
| 2 December 2027 | High-risk obligations apply to Annex III systems | Art. 113(c)(i) |
| 28 January 2028 | Notified bodies under Section A legislation must have applied for AI Act designation | Art. 43(3) |
| 2 August 2028 | High-risk obligations apply to Annex I systems, including MDR/IVDR devices | Art. 113(c)(ii) |
The start date moved; the scope did not shrink
The only Annex I change concerns machinery: point 1 of Section A (Directive 2006/42/EC) was deleted and Regulation (EU) 2023/1230 added to Section B. The MDR and IVDR stay in Section A. Under the amended Article 43(3), the provider follows the conformity assessment procedure of the relevant Section A legislation; the Chapter III, Section 2 requirements form part of that assessment, the Article 17 quality management system is also assessed, and points 3, 4.3, 4.4, 4.5, the fifth paragraph of 4.6 and point 5 of Annex VII apply. Classification as high-risk under Article 6(1) does not affect the choice of conformity assessment procedure the MDR/IVDR give manufacturers, who need not switch to third-party assessment only because the product includes a high-risk AI safety component; a system covered by both Section A legislation and Annex III follows the Section A procedure.
AI Act notified body designation: 28 January 2028 deadline
Notified bodies under the MDR/IVDR may assess AI systems against Chapter III, Section 2 if their compliance with Article 31(4), (5), (10) and (11) was assessed in their existing notification. Article 43(3) continues: "such notified bodies which have been notified under the Union harmonisation legislation in Section A of Annex I, shall apply for designation in accordance with Section 4 of this Chapter by 28 January 2028". The new Article 28(8) offers a single application and unified assessment, and the new Annex XIV codes AIP 0111 and AIP 0112 cover points 11 and 12 of Section A; ask your notified body early about its plans.
Simplifications and clarifications
- Article 2(13): where Section A legislation gives equivalent or higher protection, specific requirements in Articles 9 to 15 and 17 to 25 may be limited without reducing overall protection; the delegated acts will set the details.
- Article 11(1): SMEs, including start-ups, and small mid-cap enterprises (SMCs) may use a simplified Commission form for the Annex IV elements, which notified bodies must accept.
- Former Article 10(5), on special categories of personal data for bias correction, is replaced by Article 4a; one condition is that other data, including synthetic or anonymised data, cannot do the job effectively.
What was not postponed
Articles 4 and 5 have applied since 2 February 2025 (the prohibitions newly added by Regulation (EU) 2026/1744 apply from 2 December 2026); Article 4 now requires measures "to support" AI literacy. Article 50 applies with the general date of 2 August 2026. MDCG 2025-6 notes that Articles 5 and 50 apply regardless of MDR/IVDR class. The MDR and IVDR continue to apply and were not amended by Regulation (EU) 2026/1744.
Common misreadings
- "Delayed" read as "medical AI is outside the AI Act". Only the date changed.
- Treating the proposal as law. Under the amended Article 2(2), products under Section B legislation are directly subject only to Article 6(1), Article 60a and Articles 102 to 112. Whether devices end up there depends on the proposal's fate.
- Copying dates from MDCG 2025-6. Question 31 uses the original 2 August 2027 date; the template that Question 34 expects as an implementing act by 2 February 2026 is now guidance due by 2 September 2027.
MDCG 2025-6: how the MDR/IVDR and the AI Act apply together
The Q&A is not a Commission document and is not legally binding; the AI Board and the MDCG both consist of representatives of all Member States. Key points:
- One set of technical documentation (Article 11(2)); MDR/IVDR sampling rules for technical documentation review still apply (Questions 12, 13).
- AI Act quality and risk management requirements target the AI system and can be integrated into existing MDR/IVDR systems (Questions 6, 7).
- Training, validation and testing data should be relevant, sufficiently representative, as error-free and complete as possible, and examined for bias (Questions 8 to 11).
- High-risk MDAI must log events automatically over its lifetime, complementing UDI traceability (Questions 9, 21).
- Pre-determined changes assessed at the initial conformity assessment and described in the technical documentation are not a substantial modification (Question 30).
- In-house devices are not high-risk under Article 6(1), though other provisions such as the prohibitions still apply (Question 35).
As of 12 November 2026, the Commission's MDCG guidance list still shows the June 2025 version.
CE marking AI medical devices: documents that must meet both the MDR/IVDR and the AI Act
| Area | MDR/IVDR basis | AI Act additions | Integration basis |
|---|---|---|---|
| Technical documentation | Annexes II and III | Annex IV: development methods (including third-party pre-trained systems or tools), algorithm logic and design choices, training data, human oversight assessment, pre-determined changes, validation metrics, cybersecurity measures | Art. 11(2) |
| Risk management | Annex I requirements | Risks to health, safety and fundamental rights; impact on persons under 18 and other vulnerable groups | Art. 9(10) |
| Data governance | Clinical or performance evaluation | Data quality, representativeness and bias (Art. 10); special categories of data under Art. 4a | MDCG 2025-6, Q8 to Q11 |
| Quality management system | MDR/IVDR QMS | AI-related aspects in Article 17 | Art. 17(3) |
| Post-market monitoring | Post-market surveillance and vigilance | Article 72 plan; AI Act serious incident reporting limited to infringements of fundamental-rights obligations (Art. 73(10)) | Art. 72(4) |
A practical order of work:
- List products containing an AI system as defined in Article 3(1), with their MDR/IVDR class and whether a notified body is involved.
- Ask the notified body how it will apply for AI Act designation before 28 January 2028.
- Check Articles 9 to 15, 17 and 72 against existing files and record the gaps.
- Plan data access: under Annex VII point 4.3 the notified body gets full access to training, validation and testing data where relevant, via API where appropriate; under point 4.5, if other means are insufficient, it may on reasoned request access the trained models and parameters, subject to IP and trade secret protection.
- For third-party models or tools, agree the information, technical access and assistance in writing, as the amended Article 25(4) requires.
- Under Article 22, providers outside the EU must appoint an EU authorised representative by written mandate before making a high-risk AI system available.
The new Article 42(3) gives a cybersecurity presumption for products within the Cyber Resilience Act, but MDCG 2025-6 (Question 22) notes that medical devices and IVDs are outside that Act, so cybersecurity must still be shown under the MDR/IVDR and Article 15. Regulation (EU) 2026/1744 does not touch the MDR transitional rules.
How to check this in the database
Pages and figures as of 12 November 2026:
- EU regulations /en/regulatory/eu/: 172 regulations and guidance entries (marked updated 2026-09-18), including the MDR (32017R0745), the IVDR (32017R0746), MDCG 2019-11 rev.1 (software classification), MDCG 2020-1 (software evaluation) and MDCG 2019-16 rev.1 (cybersecurity). The AI Act and Regulation (EU) 2026/1744 are not listed; use the EUR-Lex link below.
- MDCG 2025-6 entry /en/regulatory/eu/mdcg-2025-6/: number, type, issuer, status and official link, last verified 2026-09-08.
- Regulation search /en/regulatory/search/: regulations and guidance, FDA product codes and recent 510(k) clearances. "MDCG 2025-6" finds this guidance directly; "2017/745" lists the MDR, its corrigenda and guidance entries that cite it in their titles.
- EU registration data in the "EU European Union" block of /en/markets/: EUDAMED live, 1,838 database companies matched in EUDAMED, at least 150,958 UDI-DI records (a lower bound, as some companies were capped at 300 records). Class I 69,396, IIa 45,342, IIb 12,846, III 4,141; IVD A 6,011, B 1,321, C 1,229, D 21. Article 6(1) follows the MDR/IVDR class, but EUDAMED records do not flag AI; check your own product list.
- Company pages: search a name at /en/search/; the "EU European Union EUDAMED" block shows the UDI-DI count and class breakdown. Device rows are a sample of up to 8 per company per market.
See also the Digital Medicine, Data & Artificial Intelligence category footprint and Chinese device manufacturers in the European Union, with figures as of their stated cut-off dates.
FAQ
When does the EU AI Act apply to medical devices? Can it wait until 2 August 2028?
The high-risk obligations (Chapter III, Sections 1 to 3) apply to MDR/IVDR devices from 2 August 2028, but devices are not outside the AI Act before then: Articles 4 and 5 already apply, Article 50 applies from 2 August 2026, and the MDR/IVDR still apply.
Is MDR Class I AI software high-risk?
It depends on notified body involvement. Per MDCG 2025-6 Table 1, Class I products that are non-sterile, non-measuring and not reusable surgical instruments do not meet Article 6(1); sterile, measuring or reusable surgical ones do.
Do AI medical devices need a separate AI Act notified body?
Under Article 43(3), the assessment happens within the MDR/IVDR procedure, by notified bodies under the MDR/IVDR that meet the relevant Article 31 requirements. They must apply for AI Act designation by 28 January 2028.
Do devices already on the market need reassessment?
The amended Article 111(2) applies the Regulation to high-risk systems placed on the market before the Chapter III date only if their design changes significantly after that date. MDCG 2025-6 (Question 31) treats "placing on the market" per individual product, not per type, but was written for the original 2027 date; follow later guidance.
Will the MDR/IVDR move to Section B of AI Act Annex I?
That is in the Commission proposal COM(2025) 1023, which has not been adopted. As of 12 November 2026, the European Parliament's Legislative Observatory (OEIL) shows procedure 2025/0404(COD) at the "Awaiting committee decision" stage; under current law the MDR and IVDR remain in Section A, so preparation should not wait on the proposal.
Regulation (EU) 2026/1744 adds one year, but an AI device needing a notified body is still a high-risk AI system. Use it to confirm notified body plans, integrate the AI Act into MDR/IVDR documentation, and track the delegated acts and guidance due in 2027 and progress on harmonised standards. The legal text is on EUR-Lex: Regulation (EU) 2026/1744; MDCG 2025-6 is a PDF from the European Commission. This article is based on official public information available as of 12 November 2026; the latest documents from the regulators prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.