The IVDR uses seven classification rules in Annex VIII to place in vitro diagnostic (IVD) devices in Classes A to D. Only non-sterile Class A devices are fully self-declared; sterile Class A and Classes B, C and D all need a notified body. Legacy devices (with an IVDD certificate, or with an IVDD declaration of conformity drawn up before 26 May 2022 and needing a notified body under the IVDR) can use the Article 110 transitional period only if six conditions are all met. Class B and sterile Class A devices self-declared under the IVDD must lodge an application by 26 May 2027 and sign a written agreement by 26 September 2027, and may be placed on the market until 31 December 2029 at the latest.
Key points
- Classification follows intended purpose. Annex VIII has seven rules placing devices in Classes A to D in order of increasing risk. Where several rules or intended purposes apply, the higher class applies.
- Only non-sterile Class A devices are fully self-declared. Sterile Class A needs a notified body for sterility aspects only; Classes B, C and D all need one.
- The transitional period covers legacy devices only. New devices (no IVDD certificate and no IVDD declaration of conformity) and non-sterile Class A devices have none; the IVDR has applied to them since 26 May 2022.
- The conditions are cumulative. Continued IVDD compliance, no significant changes, no unacceptable risk, an IVDR quality management system by 26 May 2025, an application on time and a written agreement on time: all six must be met.
- Class B and sterile Class A still face two deadlines in 2027. Those self-declared under the IVDD need an application by 26 May 2027 and a written agreement by 26 September 2027. Both Class C deadlines expired in 2026.
Background and scope
Regulation (EU) 2017/746 (IVDR) has applied since 26 May 2022, replacing Directive 98/79/EC (IVDD). Under Article 9 of the IVDD, only devices in Annex II List A and List B and self-testing devices required a notified body; most other IVDs were self-declared under Annex III. Under the IVDR, many previously self-declared reagents and instruments need a notified body for the first time.
Article 110 has been amended three times. Regulation (EU) 2022/112 introduced class-based transitional periods. Regulation (EU) 2023/607 rewrote Article 110(4) so that devices lawfully placed on the market may continue to be made available, with no sell-off date. Regulation (EU) 2024/1860, in force since 9 July 2024, turned the former end dates into application deadlines, extended placing on the market to the end of 2027, 2028 or 2029, and added quality management system, application and written agreement conditions to the existing ones (continued IVDD compliance, no significant changes). Provisions cited here follow the consolidated IVDR text of 10 January 2025 on EUR-Lex.
On 16 December 2025 the European Commission proposed a regulation to simplify the MDR and IVDR, COM(2025) 1023. As of 12 November 2026, the European Parliament's Legislative Observatory showed it still awaiting a committee decision, so it is not law. For Article 110 the proposal only adds new paragraphs, such as one on orphan devices, and does not change the existing deadlines.
IVDR classification rules in Annex VIII
Implementing rules: classification follows intended purpose; devices used in combination and accessories are classified separately; software driving a device, calibrators and controls with assigned values take the class of their device; the higher class applies where several rules apply; and the rules cover first-line, confirmatory and supplemental assays alike. The examples below come from MDCG 2020-16 rev.5 (September 2026). MDCG guidance is not legally binding.
| Rule | Class | What it covers | Examples in rev.5 |
|---|---|---|---|
| Rule 1 | D | Transmissible agents in blood, tissues or organs, tested for suitability for transfusion, transplantation or cell administration; agents causing a life-threatening disease with a high risk of propagation; infectious load of a life-threatening disease where monitoring is critical to patient management | HBsAg, anti-HCV and anti-HIV 1/2 donor screening tests; HIV and hepatitis C viral load tests |
| Rule 2 | C or D | Blood grouping, foeto-maternal blood group incompatibility and tissue typing for transfusion or transplantation are Class C; the listed ABO, Rhesus, Kell, Kidd and Duffy markers are Class D | Class D: foetal RhD typing kit; Class C: HLA typing reagents for transplantation |
| Rule 3 | C | Thirteen indents, (a) to (m), including sexually transmitted agents, companion diagnostics, cancer screening, diagnosis or staging, human genetic testing and congenital disorder screening | Chlamydia trachomatis and HPV tests; companion diagnostics |
| Rule 4 | B or C | Self-tests are Class C, except pregnancy, fertility and cholesterol tests and urine tests for glucose, erythrocytes, leucocytes and bacteria (Class B); near-patient tests are classified in their own right | Self-testing glucose meters and strips: Class C; fingerprick HIV antibody self-test: Class D under Rule 1 |
| Rule 5 | A | General laboratory products, accessories without critical characteristics, buffers, washing solutions, general culture media and stains intended for a specific examination; IVD instruments; specimen receptacles | Nucleic acid purification kits; immunoassay analysers, PCR thermocyclers, clinical chemistry analysers |
| Rule 6 | B | Devices not covered by the rules above | C-reactive protein tests; Helicobacter pylori tests |
| Rule 7 | B | Controls without a quantitative or qualitative assigned value | Unassigned control sera |
Note the limit of Rule 5: per rev.5, an instrument that measures independently without additional reagents (for example cell counters, or blood gas or glucose sensors) is classified by the intended purpose of the analysis, not automatically as Class A.
Per its revision history, rev.5 changed only Rule 7 from rev.4, clarifying the rationale and revising the examples. Under rev.5, analyte presence or ranges indicated for standalone quality control materials are not "assigned values", so such controls are Class B unless other rules apply. Controls with manufacturer-assigned values, including assay kit validity controls, take the class of their device.
Which IVDR classes need a notified body
| Class | Conformity assessment route (Article 48) | Notified body |
|---|---|---|
| A, non-sterile | Technical documentation under Annexes II and III, then an EU declaration of conformity | Not required |
| A, sterile | Annex IX or Annex XI | Required, limited to establishing, securing and maintaining sterile conditions |
| B | Annex IX Chapters I and III, with technical documentation assessment of at least one representative device per category | Required |
| C | Annex IX Chapters I and III, with at least one representative device per generic device group assessed, or Annex X with Annex XI (except Section 5); companion diagnostics require consulting a medicines authority or the EMA | Required |
| D | Annex IX Chapters I, II (except Section 5) and III, or Annex X with Annex XI; where an EU reference laboratory is designated, it verifies performance by laboratory testing | Required |
Self-testing and near-patient devices in Classes B, C and D also undergo assessment under Annex IX Section 5.1.
Which devices get the IVDR transitional period
- Devices with an IVDD notified body certificate (Article 110(2) and (3a)). IVDD certificates issued from 25 May 2017, valid on 26 May 2022 and not withdrawn remain valid until 31 December 2027. Certificates that expired before 9 July 2024 count only if a written agreement was signed before expiry, or a competent authority granted a derogation under Article 54(1) or required the conformity assessment procedure under Article 92(1). These devices, whatever their class, may be placed on the market until 31 December 2027.
- Devices self-declared under the IVDD that need a notified body under the IVDR (Article 110(3b)). The IVDD declaration of conformity must predate 26 May 2022. Latest placing on the market: Class D, 31 December 2027; Class C, 31 December 2028; Class B and sterile Class A, 31 December 2029.
Both groups must meet every condition in Article 110(3c):
| Condition | Requirement | Timing |
|---|---|---|
| (a) | The device continues to comply with the IVDD | Throughout |
| (b) | No significant changes in design or intended purpose | Throughout |
| (c) | No unacceptable risk to the health or safety of patients, users or other persons, or to public health | Throughout |
| (d) | Quality management system in place under Article 10(8) | By 26 May 2025 |
| (e) | Formal application lodged with a notified body by the manufacturer or authorised representative, for the device or a substitute device | By class, see below |
| (f) | Written agreement signed by the notified body and the manufacturer | By class, see below |
IVDR transition deadlines by class
| Legacy group | Application | Written agreement | Latest placing on the market | Status |
|---|---|---|---|---|
| IVDD certificate (any class) | 26 May 2025 | 26 September 2025 | 31 December 2027 | Both expired |
| Self-declared, IVDR Class D | 26 May 2025 | 26 September 2025 | 31 December 2027 | Both expired |
| Self-declared, IVDR Class C | 26 May 2026 | 26 September 2026 | 31 December 2028 | Both expired in 2026 |
| Self-declared, IVDR Class B or sterile Class A | 26 May 2027 | 26 September 2027 | 31 December 2029 | Both fall in 2027 |
The quality management system deadline of 26 May 2025 has passed for every group. Under Article 110(4), devices lawfully placed on the market within these periods may continue to be made available afterwards, with no sell-off date.
The Commission's Q&A on the extension of the IVDR transitional periods (July 2024, not legally binding) clarifies some common misreadings:
- For self-declared devices, if no IVDR quality management system was in place by 26 May 2025, the transitional period ended on that date.
- Devices left out of an application lose the transition when the application deadline passes. Withdrawing the application or ending the agreement after the deadline also ends it, unless the application moves at the same time to another notified body (not where the switch follows a refusal).
- Administrative changes such as a new name or address are generally not design changes; transfer to a different manufacturer is not covered unless both belong to the same larger organisation.
Class B and sterile Class A: working back from 2027
| Step | What to do | Basis |
|---|---|---|
| Check eligibility | Confirm an IVDD declaration of conformity dated before 26 May 2022, Class B or sterile Class A under Annex VIII, a quality management system in place by 26 May 2025, and no significant changes | Article 110(3b), (3c) |
| By 26 May 2027 | Lodge a formal application. Per the Commission Q&A, full technical documentation is not needed yet, but the application must identify the manufacturer and devices covered (and the legacy device any substitute device replaces), with quality management system documentation and a timeline for technical documentation | Article 110(3c)(e) |
| By 26 September 2027 | Sign the written agreement, including the schedule for further documentation | Article 110(3c)(f) |
| After 31 December 2029 | The transitional period ends; further placing on the market requires an IVDR certificate. Conditions (a) to (c) apply throughout | Article 110(3b) |
The Regulation sets deadlines only; it does not guarantee notified body reviews finish by them. For sterile Class A, proposal COM(2025) 1023 would delete the second subparagraph of Article 48(10), removing notified body involvement. It had not been adopted as of 12 November 2026, and its final text and timing are uncertain, so preparation should still follow the two 2027 deadlines.
IVDR vs MDR transition
The MDR transitional period was extended by Regulation (EU) 2023/607 and ends on 31 December 2027 or 31 December 2028 depending on risk class. The IVDR timeline comes from Regulation 2024/1860, with application deadlines in three tiers, the last in 2027. The two sets of dates and conditions cannot be applied to each other.
How to check a supplier's IVDR legacy status
The Commission Q&A lists ways to show a legacy IVD is covered: a manufacturer's self-declaration confirming the conditions are met, stating the transition end date and listing the devices and certificates; an optional notified body confirmation letter that the application was received and a written agreement signed; or copies of the application and agreement. For self-declared Class B and sterile Class A devices, no application before 26 May 2027 is not in itself a problem, but the manufacturer should be able to show an IVDD declaration of conformity dated before 26 May 2022 and its quality management system documentation; after 26 May 2027, proof of the application; and after 26 September 2027, the written agreement or a notified body confirmation letter.
How to look it up in the database
All figures are as shown on the pages as of 12 November 2026.
- Find regulations and guidance. Regulation search accepts reference numbers. "MDCG 2020-16" finds the classification guidance, "MDCG 2022-6" the significant-changes guidance, "MDCG 2022-8" the legacy-device guidance and "32017R0746" the IVDR entry. Entries show the number, type, issuer, status and official text link; the MDCG 2020-16 entry is labelled rev.4, but its link opens rev.5.
- Browse by jurisdiction. The EU regulations page lists 172 regulations and guidance documents (marked as updated 18 September 2026), including Regulations 2017/746, 2022/112 and 2023/607 and MDCG 2020-16, 2022-6 and 2022-8.
- See EU record volumes. The EU section of Browse by market shows 1,838 companies in the database with EUDAMED matches, totalling at least 150,958 UDI-DI records (447 companies were capped at 300 records, so this is a lower bound). These include IVD Class A 6,011, Class B 1,321, Class C 1,229 and Class D 21 records, plus four labels whose names correspond to IVDD categories: IVD general 9,741, Annex II List A 109, List B 289 and self-testing 493. These totals cover companies in the database, not all of EUDAMED.
- Find companies in the same field. Under "By category", open "Clinical Diagnostics & Therapeutic Technologies", then "Laboratory, IVD & Diagnostics" to see 417 companies, with subcategories such as Point-of-care & near-patient testing (82) and Home & self-testing (42). Each company row shows record counts by market, including the EU. See also Laboratory, IVD & Diagnostics: registration footprint of 417 companies.
Registration records cannot show whether a device meets the Article 110 conditions; that needs the manufacturer's declaration and notified body documents.
FAQ
When does an IVDR Class B device need a notified body?
A new Class B device needs one before being placed on the market. A legacy Class B device self-declared under the IVDD needs an application by 26 May 2027 and a written agreement by 26 September 2027 to stay in the transition, and an IVDR certificate after 31 December 2029.
Can an IVD first launched after 26 May 2022 use the transitional period?
No. A new device must complete IVDR conformity assessment before being placed on the market.
What if there was no IVDR quality management system by 26 May 2025?
No. Per the Commission Q&A, the transition for a self-declared device ended on 26 May 2025; it now needs an IVDR certificate.
Can a legacy IVD be changed during the IVDR transition?
Significant changes in design or intended purpose end eligibility. MDCG 2022-6 gives guidance on what counts as significant; a revision to align it with Regulation 2024/1860 is planned, so use the latest official version.
If the simplification proposal is adopted, will sterile Class A devices no longer need a notified body?
The proposal includes this change, but as of 12 November 2026 it had not been adopted and the final text may differ; Article 48(10) as it stands applies.
Classify, choose the conformity assessment route, then check each transitional condition. Legacy Class B and sterile Class A devices still face 26 May 2027 and 26 September 2027. This article is based on official public information available as of 12 November 2026; the latest documents from the regulators prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.