NMPA registration self-testing means the applicant tests its own device against the registered product technical requirements and issues the test report itself, under the Administrative Provisions on Self-Testing for Medical Device Registration published with NMPA Announcement No. 126 of 2021. Items the applicant cannot test can be entrusted to a qualified medical device testing body, and the entrusted report should carry the CMA mark. This guide follows the provisions, the NMPA interpretation and the Guideline for Inspection of Registration Self-Testing issued in 2025. It explains the requirements for each path, the qualification an entrusted body needs, how multi-model products show coverage, and what to file.

Key points

  • Article 14 of the Regulations on the Supervision and Administration of Medical Devices allows the registration test report to be the applicant's self-test report or a report from a qualified medical device testing body. When self-testing, items the applicant cannot test may be entrusted, and the applicant then compiles one complete self-test report.
  • An entrusted body must meet Article 75 of the Regulations. The NMPA interpretation states that a medical device testing body should hold a CMA (inspection body and laboratory qualification) certificate and that an entrusted report should carry the CMA mark. Where the mark cannot be applied, NMPA circular Yao Jian Zong Ke Wai Han [2020] No. 746 applies.
  • The self-test report must cover every item in the product technical requirements. Where items were entrusted, the remarks name the entrusted items and the testing body, and the original entrusted reports and the body's qualification documents are attached.
  • Multi-model products need a difference analysis and a model coverage statement.
  • The Guideline for Inspection of Registration Self-Testing took effect on publication on 16 September 2025 and details how regulators inspect self-testing capability. It does not apply where every test item is entrusted.

Legal basis for NMPA registration self-testing

The Regulations on the Supervision and Administration of Medical Devices (State Council Order No. 739) have applied since 1 June 2021. Article 14 allows the applicant or filer to submit its own self-test report. According to the NMPA interpretation, the 2014 Regulations required a registration test report from a testing body qualified for medical device testing; the new Regulations opened the door to self-testing and strengthened the applicant's own responsibility.

NMPA published the Administrative Provisions on Self-Testing for Medical Device Registration (Announcement No. 126 of 2021) on 21 October 2021, effective on publication. They apply to the registration of medical devices, including in vitro diagnostic reagents, and have six parts: self-testing capability, the self-test report, entrusted testing, filing documents, on-site inspection, and responsibilities. Change registrations and renewals submit self-test reports as the relevant rules require.

On 12 September 2025 the NMPA General Department issued the Guideline for Inspection of Registration Self-Testing (Yao Jian Zong Xie Zhu Han [2025] No. 499). It was posted on the NMPA website on 16 September 2025 and took effect on publication. It guides regulators inspecting self-testing for Class II and Class III devices, can serve as a reference for applicants, and is not enforced as a mandatory regulation.

What NMPA self-testing requires

An applicant that self-tests must have the capability to do so, bring self-testing into its medical device quality management system, and take primary responsibility for the self-test report. The main requirements are:

  • Personnel: testing staff must be formally employed full-time staff who work only for the applicant. Testing, review and approval staff must be authorised by the applicant.
  • Equipment and environment: equipment must suit the test methods and be metrologically traceable. Specialist laboratories, such as biological evaluation, electromagnetic compatibility (EMC), biosafety and IVD laboratories, must meet the environmental conditions of their specialty.
  • Samples, quality control and records: a sample management procedure, evaluation of measurement uncertainty where applicable, and archived quality and raw test records.
  • Test basis: testing follows the product technical requirements of the device to be registered, and test methods must be verified or validated.

Entrusted testing: what qualification the testing body needs and how the CMA mark works

The provisions state that an applicant submitting a self-test report may entrust items it cannot test to a qualified medical device testing body, and that the body must meet Article 75 of the Regulations. Article 75 provides that only testing bodies recognised by the State Council certification and accreditation authority together with the State Council drug regulator may test medical devices. The NMPA interpretation adds that a medical device testing body should hold a CMA certificate and that an entrusted report should carry the CMA mark.

Where the CMA mark cannot be applied, circular No. 746 of 2020 applies. It is the NMPA General Department notice clarifying how qualification marks are applied under the Medical Device Testing Work Specification. A Q&A published on the Fujian Provincial Medical Products Administration website on 29 July 2025 summarises it as follows. Where the technical requirements fully adopt national or industry standards, the testing body must hold CMA approval for those standards and the report cover carries the CMA mark. Where they do not, or only partly, rely on national or industry standards, the report remarks carry a self-declaration of testing capability. Refer to the circular itself for the exact wording.

The applicant must also assess the entrusted body's qualification and capability, keep a list of approved testing bodies with assessment records, and make sure the self-tested and entrusted samples are the same.

PathWho issues the reportKey requirements
Full self-testingThe applicant, including testing by an authorised group laboratory or by the contract manufacturerFull-item report in the Annex 1 template; self-test capability statement, equipment table (Annex 2) and testing staff table (Annex 3)
Self-testing plus partial entrusted testingThe applicant compiles the outside reports into one complete self-test reportRemarks name the entrusted items and testing body; original entrusted reports and the body's qualification documents attached
Fully entrustedA qualified medical device testing bodyThe body meets Article 75; CMA mark or capability self-declaration under circular No. 746; the 2025 inspection guideline does not apply

Can a group laboratory or contract manufacturer self-test for the applicant?

The provisions state that where a domestic applicant's group company in China, or one of its subsidiaries, has a laboratory accredited by CNAS (China National Accreditation Service for Conformity Assessment), that laboratory may self-test for the applicant with the group's authorisation, and the applicant issues the self-test report. The 2025 guideline describes this case as a group laboratory that holds inspection body qualification (CMA) or laboratory accreditation, and inspectors check the group authorisation and whether the laboratory's qualification and scope cover the items.

The NMPA interpretation explains that group authorisation is required to avoid unclear responsibility between subsidiaries testing for each other. A domestic applicant may use only a group laboratory located in China, and a laboratory may not accept items outside its own scope. If the applicant's own testing laboratory is CNAS-accredited, the applicant may skip the capability statement and quality system documents but must submit the accreditation evidence and supporting information on its testing scope.

An applicant that uses a contract manufacturer may have the manufacturer perform the self-testing, with the applicant issuing the report. The 2025 guideline adds that the contract manufacturer should not pass the self-test items on to a third party.

How to show coverage across several models

The products tested within one registration unit must represent the safety and effectiveness of the other products in that unit. The model coverage statement includes an explanation of typicality and a difference analysis between the covered models and the main tested model. According to the interpretation, testing the items where models differ is one way to show that the performance figures in the self-test report cover all models.

The 2025 self-testing inspection guideline: how regulators check

Under the guideline, regulators organise a self-testing inspection when needed, using the registration file and focusing on personnel, equipment and environmental facilities, sample management, quality control of testing and record control. Each requirement is checked against the product technical requirements and recorded in the on-site inspection report.

Where the registration quality system inspection is done by document review, the self-testing inspection may also rely on document review, with on-site checks only when needed, in the cases the guideline lists. Examples are a self-testing laboratory whose relevant scope is CNAS-accredited with the items inside that scope; the same items for a similar product having already passed an on-site capability check with no major change to the quality system; and simple test items, such as visual checks of appearance or measuring dimensions or weight with simple gauges. The guideline also says the applicant should run at least one full internal audit and one management review each year.

Filing documents and timeline

Date or stageItem
2020Circular No. 746 clarifies how test reports carry the qualification (CMA) mark
1 June 2021The Regulations (State Council Order No. 739) take effect
21 October 2021Announcement No. 126 publishes the self-testing provisions, effective on publication
16 September 2025The Guideline for Inspection of Registration Self-Testing (signed 12 September 2025) is published and takes effect
At filingSelf-test report; self-test capability statement; quality system documents (equipment table, testing staff table and others); model coverage statement; self-declaration of report authenticity; qualification documents of any entrusted body
Quality system inspectionRegulators check each self-testing capability requirement and record the result in the on-site inspection report

How active devices handle electrical safety and EMC

Announcement 126 does not say which items must be self-tested and which must be entrusted. Medical electrical safety (for example GB 9706.1) and EMC (for example YY 9706.102) need demanding equipment and environments. The provisions list EMC as specialist testing, and on-site inspection looks at the facilities, environment and monitoring records. A company without suitable in-house facilities may choose to self-test performance items and entrust electrical safety and EMC to a qualified testing body. This is one possible arrangement, not a requirement of the rules.

Before entrusting, check three things item by item: the standard number and edition cited in the technical requirements; whether the body's CMA certificate schedule covers that standard, edition and clause; and whether the report will carry the CMA mark or a capability self-declaration. For a worked example on wheelchairs, see Defining the NMPA Registration Test List for Wheelchairs.

How to check this in our database

As of 13 November 2026, the Regulations database lists 282 regulations and guidance documents and 589 update records. The China NMPA entry is marked as in progress, so Chinese regulations are not yet indexed, and searching for "126" or "self-test" will not return the announcement, the interpretation or the 2025 guideline. Search the NMPA website by document title or number for the official text.

The site is useful for checking standards. Open Testing search and enter a standard number cited in the technical requirements, such as "9706.102" or "12996". The standard page shows the title, edition years and China implementation dates, which helps you confirm the edition an entrusted body must cover. Standard texts are copyrighted, so the site gives bibliographic records and official links only.

Do not rely on any website's description of a testing body, including the bodies listed on our standard pages. According to the Fujian Provincial Medical Products Administration Q&A, certificate and capability information can be looked up on the national certification and accreditation public information platform, reached from the website of the Certification and Accreditation Administration of China (CNCA). Then check the body's CMA certificate schedule item by item.

SUNGO laboratory scope

SUNGO has two laboratories, in Shanghai and Hefei, run by two separate legal entities: Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd. The scope relevant to this topic is below; confirm each item against the certificate schedules:

  • The schedule of the Hefei laboratory's CMA certificate 260015342945 covers some items of GB/T 12996-2024 (electric wheelchairs), some parts of the GB/T 18029 and ISO 7176 series, and some items of Chinese standards for walking tables, axillary crutches and multi-legged walking sticks.
  • The schedule of the Shanghai laboratory's CMA certificate 240015340087 covers some items of Chinese standards for medical face masks, medical gloves, dressings, gauze and absorbent cotton, and medical device packaging materials.
  • Medical electrical safety (GB 9706.1, YY 9706.108, YY 9706.111) and EMC (YY 9706.102, GB/T 18029.21) are covered only by the Hefei laboratory's IAS accreditation (TL-1119), with no CMA and no CNAS accreditation. These reports cannot be used as entrusted test reports for China registration and are for export and research and development verification only.

Whether an entrusted body meets Article 75 and whether a report is accepted is for the applicant to assess under the rules and for the regulator to decide in review. An introduction to the laboratory is on the SUNGO laboratory page.

Frequently asked questions

Does an NMPA registration test report need the CMA mark?

Under the NMPA interpretation, an entrusted report should carry the CMA mark. Where it cannot, circular No. 746 applies, and if the technical requirements do not, or only partly, rely on national or industry standards, the report remarks carry a capability self-declaration. The applicant's own self-test report follows the Annex 1 template and signature requirements.

Can I self-test some items and entrust the rest?

Yes. Items you cannot test may be entrusted, but the self-test report you file must cover every item in the technical requirements, name the entrusted items in the remarks, and attach the original entrusted reports.

Can a group laboratory self-test for a subsidiary?

Yes, with the group company's authorisation and a laboratory holding the relevant accreditation or qualification. A domestic applicant may use only a group laboratory in China, and the applicant issues the report.

Can a report from a laboratory with CNAS or IAS accreditation but no CMA be used for registration?

Not as an entrusted test report, because the interpretation requires the entrusted medical device testing body to hold a CMA certificate. An authorised CNAS-accredited group laboratory testing on the applicant's behalf is a different case, and the applicant issues that report. SUNGO's medical electrical safety and EMC testing is covered only by the Hefei IAS accreditation, so those reports cannot serve as entrusted reports for China registration.

Does every model have to be tested?

Not necessarily. The rules require a difference analysis and a model coverage statement, and testing the differing items is one way to show coverage.

Whether you self-test, entrust or combine the two, the deciding questions are whether your self-testing capability will stand up to inspection and whether the entrusted body's qualified scope covers the entrusted items. This guide is based on official public information as of 13 November 2026. The latest regulatory documents from the authorities prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.