To check an FDA recognized consensus standard, search FDA's Recognized Consensus Standards database and open the entry's Supplementary Information Sheet (SIS): it gives the recognized version and recognition number, complete or partial recognition, the old version's transition date, the list number and the date of entry. Whether FDA recognizes the version a 510(k) cites, and for how long the old version is accepted, comes from this sheet. Below, Recognition List 066 of August 24, 2026 serves as the worked example.
Key takeaways
- FDA recognizes a specific version. A new edition, or even the same edition with an amendment folded in, gets a new recognition number. The cited version and the recognition number must match.
- Check the extent of recognition first. The SIS says "Complete standard" or "Partial recognition"; a partial recognition lists the clauses not recognized and why.
- Look up each transition period individually. In List 066, the outgoing versions of ISO 10993-1, ISO 10993-12, ISO 20417 and IEC 60601-2-2 are accepted until July 1, 2029, but for ISO 10079-1 only until July 2, 2028. The Federal Register notice does not list these dates.
- The database moves before the notice. The new List 066 numbers were entered on May 25, 2026; the notice appeared on August 24, 2026. Under FDA guidance, a standard with a recognition number in the database, including one whose recognition is pending publication, can already be the subject of a declaration of conformity.
- Two FDA recognition numbers on one of our standard pages usually signal a transition, but cut-off dates and the extent of recognition must be confirmed in FDA's database.
What FDA recognized consensus standards are
The legal basis is section 514(c) of the FD&C Act, as amended by the FDA Modernization Act of 1997 and the 21st Century Cures Act: FDA recognizes, by Federal Register publication, all or part of standards from nationally or internationally recognized standards organizations, and a person may submit a declaration of conformity (DoC) to them to meet premarket or other requirements.
FDA explains the mechanism in the guidance Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices; as of November 12, 2026, FDA's website still shows the final version of September 2018. The guidance states that using consensus standards is not mandatory unless a standard has been incorporated by reference into a regulation, and that a DoC may cover only standards with an FDA recognition number, which the DoC should state. DoCs and test reports are beyond this article; standards usually cover only part of a submission, and FDA reviews the whole.
Changes are announced in the Federal Register as numbered recognition lists. List 065 was published on February 19, 2026 and List 066 (FR Doc 2026-17229, 91 FR 54715, Docket FDA-2004-N-0451) on August 24, 2026, applicable the same day.
How to search the FDA recognized standards database: reading ISO 10079-1 (1-203)
The FDA Recognized Consensus Standards search page searches by organization, designation number, recognition number, keyword, ASCA inclusion, specialty task group, product code, regulation number and date of entry. Results show date of entry, group, recognition number, extent (Complete or Partial), designation and title, exportable to Excel; a title opens the SIS. Main fields of the List 066 entry for ISO 10079-1, electrically powered suction equipment (viewed November 12, 2026):
| SIS field | Entry 1-203 | Meaning |
|---|---|---|
| FR Recognition List Number | 066 | Recognition list the entry belongs to |
| Date of Entry | 05/25/2026 | Database entry date; can precede the notice |
| FR Recognition Number | 1-203 | "1" is the Anesthesiology group; then the number within the group |
| Standard | ISO 10079-1 Fourth edition 2022-03 [Including AMD1:2026] | Exact version, including amendments |
| Extent of Recognition | Complete standard | Recognized in full |
| Transition Period | Old entry 1-157 accepted until July 2, 2028 | Last date for the old version |
| Public Law, CFR Citation(s) and Procode(s) | 11 product codes under 21 CFR 868, 870, 874, 878 | Examples only; others may apply |
| Relevant FDA Guidance and/or Supportive Publications | Related guidances and companion standards | Read alongside |
| FDA Specialty Task Group (STG) | Anesthesiology | Responsible group |
The first part of a recognition number is the STG number shown in the search page's drop-down: Anesthesiology 1, Biocompatibility 2, General I (QS/RM) 5, General Hospital/General Plastic Surgery 6, Physical Medicine 16 and General II (ES/EMC) 19. ISO 10079-1 also shows that the edition need not change: both entries are the 2022-03 fourth edition, but adding Amendment 1 (2026) moved the number from 1-157 to 1-203. For horizontal standards the product code field may list regulations instead; the ISO 20417 entry (5-149) lists 21 CFR 660, 801, 809 and 820.
FDA partial recognition: what it means
A partial entry lists the parts not recognized and, under Rationale, the reason, for example a conflict with an FDA guidance or another recognized standard. The guidance notes that submitters often miss this, which leads to requests for more information. List 066 shows two patterns:
- The new version is still partial. ISO 10993-1 Sixth edition 2025-11 (2-313) does not recognize the phrase "consumer products" in clause 6.5.11.3 or clause 6.9; the rationale points respectively to FDA's guidance on the use of ISO 10993-1 and to ISO 14971:2019. Two notes encourage contacting the review offices about implementing related clauses or requirements before starting a biological evaluation. The old entry 2-258 was also partial, but excluded Table A.1 in Annex A, so the new SIS must be read again.
- The extent changes with the version. The old IEC 60601-2-2 entry 6-389 (Edition 6.0) excluded subclauses 202.7 and 202.8 (electromagnetic emissions and immunity); the new entry 6-521 (Edition 6.1 consolidated) is complete. The notice flags this row with "Extent of recognition", as it does for ISO 23500-1 and ISO 23500-4.
For aspects addressed by unrecognized parts, follow the guidance or standard the SIS points to, and if in doubt consult the review team in advance, for example through a Pre-Submission.
FDA recognized standards transition period: how long the old version works
The transition period appears in the SIS of both the outgoing and the new entry: FDA will accept declarations of conformity to the old version until a stated date, and not after (shortened below to "accepted until"). For submissions after that date, FDA expects testing to a version with a current recognition number. The List 066 notice gives only old and new numbers; the dates below were checked entry by entry:
| Standard (new version) | Old → new | Extent (new) | Old version accepted until |
|---|---|---|---|
| ISO 10079-1 Fourth edition 2022-03 [incl. AMD1:2026] | 1-157 → 1-203 | Complete | July 2, 2028 |
| ISO 10993-1 Sixth edition 2025-11 | 2-258 → 2-313 | Partial | July 1, 2029 |
| ISO 10993-12 Fifth edition 2021-01 [incl. AMD1:2025] | 2-289 → 2-314 | Complete | July 1, 2029 |
| ISO 20417 Second edition 2026-03 | 5-135 → 5-149 | Complete | July 1, 2029 |
| IEC 60601-2-2 Edition 6.1 2023-02 consolidated | 6-389 → 6-521 | Complete (old: partial) | July 1, 2029 |
| FIRST CVSS 3.1 (to be superseded by CVSS 4.0, 13-140) | 13-142 | Complete (13-142) | July 2, 2028 (extended by List 066) |
Dates differ within one list, so do not derive them from the publication date. The guidance adds that the old version may be used during the transition, but if the new version raises significant questions the old one did not address, its use should be justified.
Date of entry and list number: the database updates before the Federal Register
Per the guidance, once FDA decides to recognize a standard it adds a recognition number and SIS to the database before the Federal Register notice, and such pending standards can already be the subject of a declaration of conformity. List 066 followed this pattern.
As of November 12, 2026, the last list published in the Federal Register is still List 066, yet the database already shows an SIS marked List 067: a date-of-entry search from May 26, 2026 returns ISO 10993-18, 2-298 (entered September 7, 2026, partial recognition). That number was already announced in List 059 on August 2, 2023, so database records change ahead of the next notice; rely on the SIS as it stands when you cite it.
Also, the list number on an SIS may not reflect every change: List 066 extended the CVSS 3.1 transition (13-142), yet that SIS still shows List 064. And standards withdrawn without replacement disappear from search: List 066 withdrew ASTM F1582-98 (Reapproved 2016), 11-175, and ISO 8828, 11-290; searching either number returns "No records were found". If a standard cannot be found, check the lists for a withdrawal.
What FDA Recognition List 066 changed, by specialty group
Table 1 has 32 rows: 29 replacements, 2 withdrawals without replacement and 1 transition extension. Table 2 adds 35 entries, 31 of them dental (4-369 to 4-399).
| Group | Main changes (old → new) |
|---|---|
| Anesthesiology | ISO 10079-1 [incl. AMD1:2026], 1-157 → 1-203 |
| Biocompatibility | ISO 10993-1:2025, 2-258 → 2-313; ISO 10993-12 [incl. AMD1:2025], 2-289 → 2-314 |
| Cardiovascular | ISO 5840-1/-2/-3 [incl. AMD1:2025], 3-174/3-175/3-173 → 3-205/3-206/3-204; ASTM F2942-25, 3-189 → 3-207 |
| Dental/ENT | ISO 4823 Sixth edition, 4-278 → 4-368; 31 new ANSI/ADA and ISO entries (4-369 to 4-399) |
| General I (QS/RM) | ISO 20417:2026, 5-135 → 5-149 |
| General Hospital/General Plastic Surgery | IEC 60601-2-2, 6-389 → 6-521 (extent changed) |
| Materials | ASTM F3260-25, F799-25, F2503-26 (8-647 to 8-649); new ASTM F3766-25 (8-650) |
| Nanotechnology | ISO 21362 First edition 2026-02, 18-16 → 18-25 |
| OB-Gyn/Gastroenterology/Urology | ISO 23500-1 to -5 Second editions, 9-133 to 9-137 → 9-152 to 9-156 (extent changed for -1, -4) |
| Orthopedic | ASTM F1582-98 (11-175) and ISO 8828 (11-290) withdrawn; 6 replacements (11-441 to 11-446) |
| Physical Medicine | ISO 7176-21 Third edition, 16-166 → 16-240; new ISO 7176-31:2023 (16-241) |
| Radiology | IEC 63465 Edition 1.0 2026-01 replaces 12-51 and 12-303 as 12-386 |
| Neurology | New IEC 80601-2-26 (17-23) |
| Software/Informatics | IEEE 11073-10207 [incl. Cor 1-2025], 13-106 → 13-157; CVSS 3.1 (13-142) transition extended; new IEEE 1752.1-2021 (13-158) |
| Tissue Engineering | ASTM F2064-25, 15-52 → 15-70 |
General II (ES/EMC), In Vitro Diagnostics, Ophthalmic and Sterility have no changes in this list.
Three checks for a 510(k) in preparation or under review
- Version and number match. The version cited in the submission must match the recognition number: ISO 10993-1:2018 is 2-258, ISO 10993-1:2025 is 2-313. Do not mix them.
- Cut-off dates. Compare the planned submission date with each outgoing entry's date. Under the guidance, if a standard changes during active review, FDA generally continues against the previously recognized version, but may ask for the relevant portions of the new version, or alternative data with a rationale, if it raises a new safety or effectiveness issue relevant to the decision.
- Affected documents. The ISO 20417 change affects the version cited in labeling, instructions for use and other information supplied by the manufacturer; the ISO 10993-1 change affects the version cited in the biological evaluation report and brings in the partial-recognition notes of the new SIS. Old versions remain acceptable during the transition, so the switch can follow the submission plan.
How ASCA relates to FDA recognized standards
The Accreditation Scheme for Conformity Assessment (ASCA) is a voluntary FDA program. FDA grants ASCA Recognition to accreditation bodies, which assess testing laboratories against ISO/IEC 17025:2017 and ASCA specifications; FDA then grants qualified laboratories ASCA Accreditation. When a submission includes a DoC with an ASCA Summary Test Report, FDA says it will have confidence in the test methods and results and does not intend to request more information on testing methodologies. ASCA covers only selected biocompatibility and basic safety and essential performance standards, marked "Included in ASCA" in the SIS (for example 2-314 and 6-521).
A recognized standard and an ASCA laboratory are independent facts: the first is checked in the standards database, the second in the ASCA-Accredited Testing Laboratories list on FDA's ASCA page, including status and covered standards. In an update of September 18, 2025, FDA gave certain laboratories an "FDA Initiated Withdrawal" status, and their ASCA Summary Test Reports are no longer accepted.
How to look it up in our database
This site is a bibliographic database and does not replace FDA's database. Page contents are as viewed on November 12, 2026.
- Find the standard. Enter a number or title on the testing search page (for example "ISO 10993-1" or "20417"); it covers standards records and testing bodies. Or browse the standards index by publishing body: 4,792 standards with pages as of November 12, 2026.
- Read the "FDA recognition no." field. The ISO 10993-1 page shows editions 2018 and 2025 with 2-258 and 2-313; the ISO 20417 page shows 2021 and 2026 with 5-135 and 5-149; the IEC 61326-2-6 page shows 19-43 and 19-57 (19-57 came with List 065; FDA's database shows 19-43 accepted until July 2, 2028). Two numbers usually mean a transition. One number does not rule one out: the ISO 10079-1 page shows only 1-203, yet FDA's database shows 1-157 accepted until July 2, 2028. Always confirm dates and extent in FDA's SIS. The "Laboratories covering this standard" section lists bodies recorded here, not the ASCA list; the site does not provide standard texts.
- Trace the notice. On the United States regulations page, the update row dated 2026-08-24 records List 066 and links to the Federal Register original (FR Doc 2026-17229). The page shows an update date of 2026-09-18; for later notices, rely on the Federal Register.
- Work from product codes. Look up SIS product codes on the FDA device product codes page (7,091 codes as of November 12, 2026). Each code page shows device class, 21 CFR section, marketing pathway and 510(k) clearances within a three-year data window, with summary PDFs for some records. Where a summary lists cited standards, compare those versions with the SIS.
FAQ
Is the recognized version always the newest edition of a standard?
Not necessarily. The guidance notes that FDA recognizes specific versions, which may not be the most current edition or cover every clause; go by the version and extent stated in the SIS.
What does it mean if a standard is not in FDA's recognized standards database?
It may never have been recognized, may have been withdrawn, or may have been assessed and not recognized. The no-results page suggests a keyword search (the standard may be referenced in another SIS) or the Table of Non-Recognized Standards, and the List 066 notice states that any person may recommend standards to FDA for recognition.
Does a cleared device need a new 510(k) when a recognized standard changes?
The guidance says changes in recognized standards do not retroactively affect clearance or approval, and a cleared device remains eligible as a 510(k) predicate. A new 510(k) citing that device as a predicate cannot likewise rely on a declaration of conformity to the withdrawn standard, and FDA will likely recommend the version with a current recognition number.
Can I use a standard that has an FDA recognition number but no Federal Register notice yet?
Under the guidance, FDA adds a recognition number and SIS once it decides to recognize a standard; such pending standards can already be the subject of a declaration of conformity. Check their extent and transition period first.
Is ISO 13485 a recognized consensus standard in the same sense?
No. ISO 13485:2016 is incorporated by reference into the Quality Management System Regulation (QMSR) through 21 CFR 820.7, so it is a regulatory requirement; recognized consensus standards are in principle voluntary. See FDA replaced QSR with QMSR.
In short, open the SIS for every standard you rely on: does the version match the number, is recognition partial, until when is the old version accepted, and has the entry been published? List 066 touches widely used standards for biological evaluation, labeling information and high-frequency surgical equipment, so 510(k) teams may want to check their standards list entry by entry. This article is based on official public information available as of November 12, 2026; the regulators' latest documents prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.