Blood glucose meter registration starts with two questions: who uses the product (a patient at home or a professional at the bedside), and whether the meter and strips count as separate devices. China regulates the meter as a Class II medical device and the strips as a Class II in vitro diagnostic (IVD) reagent, under separate catalogues. The United States places over-the-counter self-monitoring systems under 21 CFR 862.1345 with product code NBW, and a 510(k) is required. Under the EU IVDR, self-testing meters and strips are both Class C and need a notified body. This article compares the three markets, including GB/T 19634-2021, and shows how to look them up in this database.
Key points
- China: meter and strips sit in two catalogues. The meter is item 22-02-02 "blood glucose and related parameter analysers" in the Medical Device Classification Catalogue, Class II. Strips correspond to item 09001 "glucose test reagents" in the IVD Reagent Classification Catalogue, also Class II. Domestic manufacturers register with their provincial administration; imported products register with the NMPA.
- GB/T 19634-2021 is voluntary. It took effect on 1 May 2023, fully replaces the 2005 edition and is a non-equivalent adoption of ISO 15197:2013. The CMDE guideline for self-testing systems expects performance specifications to meet the current GB/T 19634.
- US: two product codes, two guidances. Over-the-counter (OTC) systems use NBW and prescription point-of-care meters use PZI. Both are Class II under 21 CFR 862.1345, both need a 510(k), and FDA's classification database marks both as not eligible for third-party review.
- FDA does not rely on ISO 15197. The OTC guidance finds its criteria insufficient to protect lay users and the point-of-care guidance does not recommend them. As of 13 November 2026, searching 15197 in the FDA Recognized Consensus Standards database returns no record.
- EU: self-testing meters and strips are Class C. The basis is IVDR Annex VIII Rule 4(a), and MDCG 2020-16 rev.5 gives meters and strips for capillary blood glucose self-testing as a Class C example. Self-testing and near-patient devices also need the technical documentation assessment in Annex IX Section 5.1.
Before registering a blood glucose meter: self-testing or point of care, meter or strips
Here a blood glucose monitoring system means a system that uses test strips to measure glucose in capillary whole blood or similar samples, usually a meter, strips and control solution. Glucose products that do not take a blood sample or do not use strips, and laboratory chemistry analysers, are out of scope.
The US and the EU split the pathway by user. FDA explains that hospital patients can be acutely ill and more likely to present interfering factors, while lay users vary widely in age, vision, dexterity and training, and their strips face less controlled storage; hence two separate guidances. IVDR Article 2 defines a "device for self-testing" (used by lay persons) and a "device for near-patient testing" (used outside a laboratory, generally near the patient, by a health professional). China's catalogues do not set different classes by user, but GB/T 19634 and the CMDE guideline both address systems "for self-testing".
FDA 510(k) for blood glucose meters: product codes NBW and PZI and two guidances
21 CFR 862.1345 identifies a glucose test system as a device intended to measure glucose quantitatively in blood and other body fluids, Class II (special controls). Only a product solely intended as a glucose tolerance drink is exempt from 510(k); a meter-and-strip system is not.
| Product code | FDA device name | Use setting | Pathway | Third-party review | Guidance |
|---|---|---|---|---|---|
| NBW | System, Test, Blood Glucose, Over The Counter | Home self-testing by lay users | 510(k) | Not eligible | SMBG OTC guidance |
| PZI | Prescription Use Blood Glucose Meter For Near-Patient Testing | Hospitals and other professional settings, multiple patients | 510(k) | Not eligible | BGMS point-of-care guidance |
| CGA, CFR, LFR | Named after the glucose oxidase, hexokinase and glucose dehydrogenase methods | The name states only the method | 510(k) | Eligible | Listed in the point-of-care guidance scope |
The 2020 point-of-care guidance lists CGA, CFR and LFR in its scope, while FDA's classification database also has PZI, defined as a near-patient meter for multiple-patient use across hospital and professional healthcare settings. Confirm the code for a given device against the classification database.
Both guidances were issued on 29 September 2020, replacing versions of 11 October 2016: Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use (OTC guidance) and Blood Glucose Monitoring Test Systems for Prescription Point-of-Care Use (point-of-care guidance). Their structure is largely the same:
- Bloodborne pathogen risk: cleaning and disinfection need separate validation, and the procedures must not degrade the device or its performance. OTC devices should be labelled for single-patient use; point-of-care devices are more likely to be used on multiple patients, which the guidance addresses specifically.
- Performance evaluation: precision, linearity, method comparison and user evaluation, and interference (endogenous and exogenous substances and hematocrit; the point-of-care guidance adds oxygen).
- Flex studies: strip stability, operating conditions, altitude, error codes outside the measuring range, short sample detection, sample perturbation, intermittent sampling and used strips. For mechanical vibration, shock, EMC and electrostatic discharge/electromagnetic interference, documentation of testing to an FDA-recognized standard is acceptable.
- Other sections: calibration and quality control materials, strip lot release criteria, third-party strips, software and labeling.
The OTC guidance notes that many manufacturers base their studies on ISO 15197, but FDA considers its criteria insufficient to protect lay users, for example in the hypoglycemic range and across strip lots, and recommends studies along the lines of the guidance. The point-of-care guidance states that FDA does not recommend the ISO 15197 criteria. For the statistical details, refer to the FDA documents.
Three further points are easy to miss. First, CLIA: per the point-of-care guidance, 510(k) clearance of an OTC system allows automatic CLIA-waived categorization, while FDA expects a prescription point-of-care system to be categorized as moderate complexity on clearance and recommends a dual 510(k)/CLIA waiver submission. Second, a company making compatible third-party strips should explain in its 510(k) how it manages the risk of meter design changes; FDA suggests an agreement with the meter maker. Third, the preface of both PDFs notes that under FDA's 2023 device software guidance, these devices should generally follow the Basic Documentation Level.
FDA's classification database links NBW to these recognized standards: CLSI EP37 (1st and 2nd editions), CLSI GP42 7th edition and IEEE 11073-10417 (2015 and 2023).
IVDR classification of blood glucose meters: why Class C
The IVDR (Regulation (EU) 2017/746) has applied since 26 May 2022. Annex VIII Rule 4(a) makes devices for self-testing Class C, except devices for pregnancy, fertility and cholesterol testing and for detecting glucose, erythrocytes, leucocytes and bacteria in urine, which are Class B. That "glucose" exception is about urine, not blood glucose.
Implementing rule 1.2 applies the rules separately to each device used in combination. MDCG 2020-16 rev.5 (September 2026) gives this Rule 4(a) example: meters and strips (containing integrated testing reagent) for self-testing of capillary blood glucose are Class C. Near-patient devices are "classified in their own right" under Rule 4(b); the MDCG examples put blood glucose reagents and strips for patient monitoring in Class C under Rule 3, and the Rule 3(k) examples include blood glucose tests for managing patients with diabetes. So self-testing meters and strips are Class C; point-of-care strips are Class C in the MDCG example, while a point-of-care meter has to be classified under Rule 4(b) and Rule 5.
Under Article 48(7), Class C devices follow Annex IX Chapters I (quality management system) and III, plus a technical documentation assessment under Section 4 for at least one representative device per generic device group. Self-testing and near-patient devices also go through Annex IX Section 5.1: the application to the notified body includes test reports (including studies with intended users), an example of the device where practicable, and data showing suitability for self-testing or near-patient testing.
Article 10(10) requires information supplied with self-testing and near-patient devices to be easily understandable and in the official Union language(s) determined by the Member State where the device is made available. Whether legacy devices under the old IVD Directive can stay on the market depends on the transitional provisions in IVDR Article 110, not covered here. ISO 15197 is the international standard commonly used for self-testing systems; check the ISO website for its current edition.
China blood glucose meter registration: two catalogues and GB/T 19634-2021
| Product | Catalogue | Entry | Class |
|---|---|---|---|
| Blood glucose meter | Medical Device Classification Catalogue | 22-02-02 blood glucose and related parameter analysers | Class II |
| Test strips | IVD Reagent Classification Catalogue (NMPA Announcement No. 58 of 2024) | 09 reagents for carbohydrate testing; 09001 glucose test reagents | Class II |
| Control solution | Calibrators and controls are not listed | Used with Class II or III reagents: same class as the reagent | Follows the reagent |
The registration review guideline for self-testing blood glucose monitoring systems (2023 revision) from the NMPA Center for Medical Device Evaluation (CMDE) states that the meter is Class II (code 22-02-02) and blood glucose test reagents are Class II (code 6840), with control materials following the reagent. It also expects performance specifications to meet the current GB/T 19634, and the strips and the meter to be filed separately under the applicable rules. It covers strip-based systems for self-testing by lay persons; systems for professional use may follow its applicable parts.
The IVD Reagent Classification Catalogue was issued with NMPA Announcement No. 58 of 2024 (dated 10 May 2024) as a revision of the 2013 "6840" IVD reagent sub-catalogue. The accompanying Notice No. 17 of 2024 keeps classification code 6840; from 1 January 2025, first registrations are accepted under the new catalogue and the old sub-catalogue is repealed; certificates approved before then stay valid for their term.
Under Article 16 of the Regulations on the Supervision and Administration of Medical Devices, a domestic applicant files a Class II registration with its provincial administration, and an overseas applicant exporting Class II or III devices to China files with the NMPA through a designated legal entity in China.
The self-testing guideline above was issued with CMDE Notice No. 1 of 2024; for the meter there is also the registration review guideline for blood glucose meters (2024 revision), issued with CMDE Notice No. 21 of 2024. Refer to the Chinese originals for requirements.
GB/T 19634-2021 (In vitro diagnostic test systems: general technical requirements for blood-glucose monitoring systems for self-testing) was published on 11 October 2021 and took effect on 1 May 2023, under technical committee TC136. As of 13 November 2026, the national standards information platform shows the plan for its Amendment No. 1 at the "under review" stage, to take effect on publication.
China, US and EU compared
| Item | China | United States | European Union |
|---|---|---|---|
| Legal basis | Medical device regulations and the two catalogues | 21 CFR 862.1345 | Regulation (EU) 2017/746, Annex VIII |
| Home self-testing | Meter Class II, strips Class II | NBW, Class II, 510(k) | Rule 4(a), Class C |
| Professional point of care | Entries do not distinguish self-testing from professional use; same classes; the self-testing guideline may be followed where applicable | PZI (or method codes CGA, CFR, LFR), 510(k) | Rule 4(b); strips Class C in the MDCG example |
| Meter and strips | Two catalogues, registered separately | One test system; third-party strips file their own 510(k) | Classified separately; both Class C for self-testing |
| Who reviews | Provincial administration (domestic Class II) or NMPA (imported) | FDA; NBW and PZI not eligible for third-party review | Notified body, including Annex IX Section 5.1 |
| Main references | GB/T 19634-2021, CMDE guidelines | The two 2020 FDA guidances | IVDR Annex I requirements, MDCG guidance |
Blood glucose meter EMC and electrical safety: IEC 61326-2-6 and IEC 61010-1
The EMC standard specific to IVD medical equipment is IEC 61326-2-6; the Chinese standard of the same kind is GB/T 18268.26. As of 13 November 2026, the FDA Recognized Consensus Standards database lists Edition 4.0 of June 2025 (19-57, complete recognition) and Edition 3.0 of October 2020 (19-43, partial recognition; clauses 3, 5.3.101, 6 and 9 not recognized). FDA accepts declarations of conformity to 19-43 until 2 July 2028. For electrical safety, the database lists IEC 61010-1 Edition 3.1 (19-34); which parts apply depends on the standard text and FDA's review.
How to look up blood glucose meter codes, standards and data in this database
- Open the NBW product code page for the device class, 21 CFR section, medical specialty and pathway. As of 13 November 2026 it shows 13 510(k) clearances in the last three years (10 summaries archived for download) and 510 across all years, with a link to FDA's classification page.
- For point-of-care meters, enter PZI in the Regulation search. The PZI page carries FDA's definition and, as of the same date, 4 clearances in the last three years and 14 in all. CGA, CFR and LFR work the same way; their pages carry a "Third-party review eligible" flag that NBW and PZI lack.
- The FDA product code overview groups 7,091 codes by specialty as of the same date; NBW, PZI, CGA, CFR and LFR all sit in Clinical Chemistry (527 codes).
- Open the GB/T 19634 standard page for the 2005 and 2021 editions and their effective dates in China. Entering 61326-2-6 in the Testing search shows the 2020 and 2025 editions with FDA recognition numbers 19-43 and 19-57. Standard texts are copyrighted; the site gives bibliographic data and official sources only.
- The category data article Home & self-testing: registration footprint of 42 companies summarises where companies in this category are based and which markets they are registered in.
Test evidence and the SUNGO laboratory
Of the testing discussed here, the part that maps to SUNGO's accreditation is IVD equipment EMC: the IAS accreditation of the Hefei laboratory (TL-1119) covers IEC 61326-2-6, EN IEC 61326-2-6 and GB/T 18268.26. Electrical safety to IEC 61010-1 and IEC 61010-2-101, accuracy testing to ISO 15197 or GB/T 19634, user evaluation and clinical evaluation are outside SUNGO's accreditation. Whether a regulator or notified body accepts a report depends on its own requirements. See the SUNGO laboratory page.
FAQ
Can a blood glucose meter 510(k) use third-party review?
No. FDA's classification database marks both NBW and PZI as "Not Third Party Eligible", so the 510(k) goes directly to FDA.
Does meeting ISO 15197 satisfy FDA 510(k) expectations?
Not by itself. The OTC guidance finds the ISO 15197 criteria insufficient for lay users and the point-of-care guidance does not recommend them; design studies to the FDA guidances.
What class is a blood glucose meter under the EU IVDR?
Self-testing meters and strips are Class C under Rule 4(a) and need a notified body. The Class B "glucose" exception in Rule 4(a) covers urine testing, not blood glucose.
Is GB/T 19634-2021 mandatory, and when did it take effect?
It is a voluntary national standard, in effect since 1 May 2023, fully replacing GB/T 19634-2005. However, the CMDE guideline for self-testing blood glucose monitoring systems expects the performance specifications in a submission to meet the current GB/T 19634.
Do the meter and test strips need separate registrations?
In China, yes: the meter is under 22-02-02 and the strips under 09001, and the CMDE guideline expects them to be filed separately under the applicable rules. In the EU they are classified separately. In the US, FDA evaluates them as one test system, and a maker of compatible strips files its own 510(k).
The key is fixing the intended use first. In the US, home self-testing and professional point of care map to different codes and guidances; in the EU, self-testing meters and strips are Class C; in China, the meter and strips each follow their own catalogue. This article is based on official public information as of 13 November 2026; the latest documents from the regulators prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.