Whether a rollator or mobility scooter is a medical device in the EU depends on the intended purpose the manufacturer states on the label, in the instructions for use and in promotional and sales materials, not on the product name. A product claimed to compensate for a disability or serve another medical purpose is a medical device under the MDR (Regulation (EU) 2017/745). A product sold to consumers without medical claims falls under the General Product Safety Regulation, GPSR (Regulation (EU) 2023/988), which has applied since 13 December 2024. This article explains the boundary, the GPSR obligations and the test standards that can serve as safety evidence.

Key points

  • Status follows the intended purpose. Under MDR Article 2(12), the intended purpose is the use stated by the manufacturer on the label, in the instructions for use or in promotional or sales materials or statements. The medical purposes in Article 2(1) include "compensation for, an injury or disability". Claims on an online product page count too.
  • Rollators and scooters sold to consumers without medical claims are covered by the GPSR, which has applied since 13 December 2024 and repealed Directives 2001/95/EC and 87/357/EEC on that date.
  • Core GPSR obligations for manufacturers: an internal risk analysis and technical documentation (kept for 10 years), a type, batch or serial number, the manufacturer's name and addresses, and instructions and safety information in a language consumers understand. A responsible economic operator must be established in the EU, and online offers must show specified information.
  • As medical devices, rollators and scooters usually fall into Class I (Rule 1 for non-active devices, Rule 13 for active ones). The manufacturer draws up technical documentation, issues the EU declaration of conformity and affixes the CE marking; a manufacturer outside the EU must designate a sole authorised representative.
  • Selling a product with medical claims as a consumer item does not stop it being treated as a medical device. Under MDR Article 97, competent authorities can require the non-compliance to be ended within a set period and, failing that, restrict or prohibit the product or require its recall or withdrawal.

What the EU GPSR covers: scope and relationship with the MDR

Article 3 of the GPSR defines a "product" as an item intended for consumers or likely, under reasonably foreseeable conditions, to be used by consumers even if not intended for them. Article 2(2) excludes medicinal products, food, feed, living plants and animals and genetically modified organisms, animal by-products, plant protection products, consumer transport equipment operated by a service provider, aircraft and antiques. Medical devices are not on that list.

The relationship between the two regulations is set by GPSR Article 2(1). Where Union law already has specific safety provisions with the same objective, the GPSR covers only the aspects and risks they do not cover. For products subject to Union harmonisation legislation (the MDR is listed in Annex I to Regulation (EU) 2019/1020), Chapter II on safety requirements does not apply to the risks that legislation covers, and Chapter III Section 1 (obligations of manufacturers, importers, distributors and others), Chapters V and VII and Chapters IX to XI do not apply. A rollator placed on the market under the MDR is therefore mainly governed by the MDR, but whether and how the remaining GPSR chapters apply (for example on distance sales information, online marketplaces and recalls) depends on the text and the case. The official text prevails.

Is a rollator a medical device? Intended purpose and claims decide

MDR Article 2(1) defines a medical device as any instrument, apparatus, appliance, software, implant, reagent, material or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more specific medical purposes. For rollators and scooters, the relevant purpose is diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability.

Regulators look at everything the manufacturer says, because Article 2(12) takes the intended purpose from the label, the instructions for use and promotional or sales materials or statements. A practical self-check:

  • Do the label, instructions, packaging, listing title or product page say the product helps people with reduced mobility, people with disabilities or post-operative patients to walk or transfer?
  • Do marketing images and copy place the product in a medical, rehabilitation or care setting?
  • Are statements consistent across channels and languages, or does one page on its own describe a medical use?

The EU-level reference is the MDCG Borderline and Classification Working Group's Manual on borderline and classification. Version 5 (April 2026), the public version as of 13 November 2026, has a section on the borderline between medical devices and general consumer products, but it contains no product entries, and the manual has no entry on rollators or scooters. Entry 1.1.7.1, a rescue bag for patient transport, shows how the phrase is applied: the product was qualified as a medical device because its purpose was alleviation of, or compensation for, an injury or disability, and placed in Class I under Rule 1. The manual is not legally binding; national competent authorities and courts decide individual cases. Under MDR Article 4, the Commission can also decide by implementing act whether a product falls within the medical device definition.

Medical device or consumer product: comparing the two routes

ItemMedical device (MDR)Consumer product (GPSR)
When it appliesIntended purpose includes a medical purpose such as compensation for a disability (MDR Article 2(1))No medical purpose; intended for or likely to be used by consumers (GPSR Article 3)
Classification and assessmentUsually Class I (Rule 1 or Rule 13); technical documentation under Annexes II and III and a self-issued declaration of conformity; a notified body is involved only for sterile devices, devices with a measuring function or reusable surgical instrumentsNo classification or notified body procedure; the manufacturer meets the general safety requirement (Article 5), carries out an internal risk analysis and draws up technical documentation (Article 9(2))
MarkingCE marking (MDR Article 20)The GPSR itself does not deal with CE marking; other applicable EU legislation governs
Responsible party in the EUA non-EU manufacturer designates a sole authorised representative (MDR Article 11)An economic operator established in the EU (Article 16): an EU manufacturer, importer, authorised representative or fulfilment service provider
Document retentionAt least 10 years after the last device is placed on the market (15 for implantables)10 years after placing on the market (Article 9(3))
Accidents and safety issuesSerious incidents and field safety corrective actions reported under MDR Article 87Accidents and dangerous products notified through the Safety Business Gateway (Articles 9(8) and 20)

GPSR obligations: what selling as a consumer product requires

GPSR Article 5 allows economic operators to place or make available only safe products. Article 6(1) requires the safety assessment to consider the categories of consumers, in particular vulnerable consumers such as children, older people and persons with disabilities. Rollator and scooter users are largely older people, so this applies even without medical claims. The main manufacturer obligations:

ProvisionRequirement
Article 9(2)Before placing on the market, an internal risk analysis and technical documentation with at least a general description and the essential safety characteristics; where appropriate, the risk analysis and solutions adopted (including test report outcomes) and the standards or other elements applied
Article 9(3)Keep the documentation up to date for 10 years and provide it to market surveillance authorities on request
Article 9(5)A type, batch or serial number or other identifier visible to consumers, or on the packaging or an accompanying document
Article 9(6)Manufacturer's name, registered trade name or trade mark, postal and electronic address
Article 9(7)Clear instructions and safety information in a language consumers easily understand, as determined by the Member State of sale
Article 9(8) and 9(11)For a dangerous product, immediate corrective measures (including withdrawal or recall), informing consumers and notifying through the Safety Business Gateway; public channels for complaints
Article 16A responsible economic operator established in the EU, with its name and contact details on the product, packaging, parcel or an accompanying document
Article 19Online offers must show the manufacturer's name and contact addresses, the EU responsible person for non-EU manufacturers, product identification including a picture, and warnings and safety information
Article 20Accidents caused by the product notified through the Safety Business Gateway to the authorities where they occurred

Importers, distributors and online marketplaces have their own obligations (Articles 11, 12, 22 and others). A mobility scooter sold as a consumer product may also be subject to other EU product legislation; the boundary with vehicle legislation is outside this article. The official texts prevail.

GPSR key dates

DateEvent
10 May 2023Regulation (EU) 2023/988 adopted
23 May 2023Published in the Official Journal (OJ L 135); in force from the twentieth day after publication
13 December 2024Applies; Directives 2001/95/EC and 87/357/EEC repealed (Articles 50 and 52)
Products placed on the market before 13 December 2024Member States may not impede products that conform to Directive 2001/95/EC and were placed on the market before that date (Article 51)

Common mistakes

  • Renaming the product a "shopping trolley" or "senior scooter", or deleting medical words from the manual. MDR Article 2(12) counts promotional and sales materials, so medical wording on product pages, videos and images is taken into account.
  • Assuming a consumer product needs no responsible person in the EU. GPSR Article 16 requires one, with its details on the product, packaging or similar.
  • Assuming the GPSR is irrelevant once a product is sold as a medical device. GPSR Article 2(1) excludes only certain chapters.
  • Treating one test report as compliance. The GPSR requires a risk analysis and technical documentation; a test report is one piece of evidence in that file, does not decide the product's status and is not an MDR conformity assessment.

How to look it up in the database

Start with the EU medical device regulations page, which lists regulations and guidance by type, number, title and status. As of 13 November 2026 it shows 172 regulations and guidance entries, updated 18 September 2026. The list includes the borderline and classification manual, the Helsinki Procedure (under which Member States exchange views on borderline and classification questions; the manual records the agreements reached) and the classification guidance MDCG 2021-24 rev.1. The page covers medical device legislation and does not include the GPSR; search EUR-Lex for Regulation (EU) 2023/988 to read it.

Next, open the borderline and classification manual entry, which shows the document type, issuer, status and a link to the official Commission text, so the version can be checked before downloading.

Then, in testing search, enter "ISO 11199-2" or "EN 12184" to find the standard record, version years and official sources; the site does not provide standard texts. As of 13 November 2026 the standards overview lists 4,792 standards with their own pages.

If it is unclear whether a product is a wheelchair or a walking aid, see Wheelchair or Walking Aid? Settling the Classification Boundary, which covers the line between two kinds of assistive product rather than between medical devices and consumer products.

Test evidence: ISO 11199-2 and EN 12184

Either route needs safety evidence. GPSR Article 9(2) states that, where appropriate, technical documentation includes the risk analysis, the solutions adopted and the outcome of reports on tests conducted by the manufacturer or another party on its behalf. Two product standards are commonly used: ISO 11199-2 (Assistive products for walking manipulated by both arms, Requirements and test methods, Part 2: Rollators) and EN 12184 (Electrically powered wheelchairs, scooters and their chargers, Requirements and test methods). Under GPSR Article 7, conformity with European standards referenced in the Official Journal gives a presumption of conformity with the general safety requirement; otherwise, Article 8 lists other European and international standards among the elements to consider. Whether a standard has been referenced should be checked in the Official Journal.

In the accredited scope of Hefei SUNGO Medical Technology Co., Ltd. (the Hefei laboratory), CNAS accreditation L24279 lists 22 clauses of ISO 11199-2:2021/Amd 1:2024 and 47 clauses of EN 12184:2022, and IAS accreditation TL-1119 lists ISO 11199-2 (except 7.3 Biocompatibility and toxicity) and EN 12184 (except 6.6 Biocompatibility and toxicity, and 12.3.1). CNAS and IAS are both signatories to the ILAC MRA. These standards are not within the Hefei CMA scope, and the Shanghai laboratory's accreditation and CMA scopes do not include them. Biocompatibility and toxicity clauses are outside the accredited scope; where such testing is needed, it is subcontracted to a laboratory with appropriate qualifications, with the client's prior consent, and the subcontracted items and subcontractor are stated in the report. Test items are described on the walking aid and crutch testing page, the mobility scooter testing page and in Rollator Testing to ISO 11199-2: Scope, Items and Sample Prep. A test report is one piece of evidence; it does not determine the product's status or guarantee acceptance by any regulator.

Frequently asked questions

Is a rollator a medical device in the EU?

Not necessarily. If the label, instructions or sales materials claim a medical purpose such as compensating for a disability, it is a medical device under the MDR; sold only as a shopping or travel aid without medical claims, it falls under the GPSR. The competent authority of the Member State of sale decides individual cases.

Does a mobility scooter sold in the EU need to comply with the GPSR?

A scooter sold to consumers without medical claims is covered by the GPSR. Where other EU product legislation also applies, the GPSR covers only the aspects and risks that legislation does not; see GPSR Article 2.

What is the GPSR responsible person?

GPSR Article 16 requires an economic operator established in the EU to be responsible for the product: an EU manufacturer, importer, authorised representative or fulfilment service provider. Its name and contact details must appear on the product, packaging, parcel or an accompanying document.

How long must GPSR technical documentation be kept?

Ten years after the product is placed on the market (Article 9(3)), available to market surveillance authorities on request. Importers keep a copy for the same period (Article 11(6)).

What if a consumer product has medical claims on its sales page?

Align claims across all channels and check that labels, instructions and sales pages match the actual intended purpose. If the medical use is to stay, the product has to comply with the MDR. This article is not legal advice.

Which regulation a rollator or scooter is placed on the EU market under is decided first by the manufacturer's own claims, then supported by technical documentation and test evidence. This article is based on official public information available as of 13 November 2026; the latest documents from the regulators prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.