Whether a device can use the FDA 510(k) Third Party Review Program (3P510k) depends on its product code. As of 13 November 2026, FDA's device classification database shows that powered wheelchairs (ITI), three-wheeled motorized vehicles or scooters (INI), standup wheelchairs (IPL), mechanical wheelchairs (IOR) and portable oxygen generators (CAW, the code commonly used for oxygen concentrators) are eligible for third party review, while stair-climbing wheelchairs (IMK) are not. This article explains how the route works, how to check eligibility and review organizations on this site and on FDA's website, and whether test evidence differs between the two routes.
Key points
- Third party review is not an exemption. The submission is still a 510(k). An FDA-recognized review organization reviews it first and sends its review and a substantially equivalent (SE) or not substantially equivalent (NSE) recommendation to FDA, and FDA makes the final decision.
- Eligibility is set by product code. As of 13 November 2026, the FDA classification pages for ITI, INI, IPL, IOR and CAW read "Eligible for 510(k) Third Party Review Program", and the page for IMK reads "Not Third Party Eligible".
- Check in two steps: look for the third-party review flag on this site's product code page, then confirm on FDA's classification page and its list of review organizations. If the two differ, the FDA page at the time you check prevails.
- No 510(k) user fee is paid to FDA on the third party route; the review fee is agreed between the submitter and the review organization. Review organizations use the same review criteria as FDA, so the test evidence expected does not get lighter on this route.
- SUNGO is a testing laboratory, not an FDA-recognized third party review organization. It does not handle reviews and cannot speed them up.
What is FDA 510(k) third party review?
The legal basis is section 523 of the Federal Food, Drug, and Cosmetic Act (FD&C Act), codified at 21 U.S.C. 360m, "Accredited persons". FDA's 510(k) Third Party Review Program page says the program is formally known as the Accredited Persons Program, and that accredited 3P510k Review Organizations may review certain low-to-moderate risk devices. Use of the program is voluntary, and FDA states that approximately half of the 510(k)s it receives are eligible.
As FDA describes it, a 510(k) for an eligible device may first be submitted to a 3P510k Review Organization rather than directly to FDA. During its review, the organization may interact early with FDA to make sure it is using up-to-date standards and guidance, and it may request additional information from the submitter. When the review is complete, the organization sends FDA the original 510(k), its own review and an SE or NSE recommendation, and FDA makes the final determination based on that review and recommendation. FDA describes the program as intended to help yield more rapid 510(k) decisions and to let FDA focus its resources on higher risk devices. Actual timing depends on the quality of the submission and the organization's schedule, and this article does not estimate it.
On scope, FDA's page "How to Submit a 510(k) Through the Third Party Review Program" says most Class I and Class II devices are eligible, and the statute bars accredited persons from reviewing Class III devices. On 21 November 2024 FDA issued the final guidance "510(k) Third Party Review Program and Third Party Emergency Use Authorization (EUA) Review", which supersedes the 2020 third party review guidance. Refer to that guidance for eligibility, organization responsibilities and submission details.
Third party review vs. a 510(k) reviewed directly by FDA
| Item | 510(k) reviewed directly by FDA | 510(k) third party review |
|---|---|---|
| Scope | Devices that require a 510(k) | Only product codes on FDA's eligibility list; Class III devices are excluded |
| First submitted to | FDA | An FDA-recognized 3P510k Review Organization, with an authorization letter |
| Substantive review | FDA reviewers | The review organization, which recommends SE or NSE |
| Final decision | FDA | FDA, based on the organization's review and recommendation |
| Payment to FDA | 510(k) review fee under FDA's fee rules | No user fee to FDA |
| Review organization fee | None | Agreed with the organization and paid directly to it |
| Review criteria | FDA's 510(k) criteria | The same criteria FDA uses |
How to submit a 510(k) through third party review
FDA's "How to Submit a 510(k) Through the Third Party Review Program" page sets out four steps:
- Determine whether the device is eligible: check FDA's List of Devices for Third Party Review or the device classification database, or ask a review organization or FDA.
- Find a review organization that can review the product code: click through the eligibility list by device area to the product code, or check FDA's list of recognized review organizations, which shows the devices each organization is accredited to review.
- Obtain price quotes from one or more organizations and sign a review contract. The fee is set by agreement and paid directly to the organization; FDA does not collect a user fee for third party submissions.
- Submit the 510(k) to the organization with a letter authorizing it to discuss the 510(k) with FDA and to forward it to FDA on the submitter's behalf. The letter should give the organization's name, the name and contact information of the person assigned to the review, and the device trade name.
On format, since 1 October 2023 FDA has required all 510(k) submissions, unless exempted, to be submitted electronically using eSTAR. The 2024 final guidance says submitters using third party review should also prepare the 510(k) in eSTAR, and the review organization then submits to FDA through the CDRH Portal. Confirm delivery arrangements and timing with the organization you choose.
Common mistakes
- Treating third party review as an exemption or a lighter review. The criteria are the same as FDA's, and FDA keeps the final decision. FDA's program page states that if the organization did not appropriately apply the 510(k) decision criteria, or if there are substantive quality issues with its documentation, FDA may need to re-review all or part of the submission. FDA also states it is updating the program to avoid routine re-review of submissions already reviewed by an organization.
- Judging eligibility by device name alone. Similar names can map to different FDA product codes, and eligibility, the classification regulation and the applicable guidance all follow the code. FDA's eligibility list tells manufacturers to determine classification at the product code level that most appropriately identifies their device.
- Sending an ineligible device to a review organization. The 2024 guidance says organizations should confirm eligibility before accepting a submission. If an organization submits an ineligible device to FDA, FDA places the submission on hold, and the same device cannot be submitted directly to FDA or to another organization until the file is withdrawn by the organization or deleted by FDA.
- Submitting the same 510(k) to several organizations in turn. The 2024 guidance asks organizations to guard against this "forum shopping" and to confirm the submission has not been presented to another organization. Getting quotes from several organizations is fine; the formal submission goes to one.
How to check eligibility in the database
The FDA product code index on this site uses classification data from openFDA (public domain, CC0). As of 13 November 2026 it holds 7091 product codes, grouped by FDA medical specialty. The steps are:
- Open the product code index on this site (path /en/regulatory/us/codes/) and pick a specialty. Powered wheelchairs, scooters and mechanical wheelchairs are under Physical Medicine (220 codes as of 13 November 2026), while CAW is under Anesthesiology (229 codes), so check both. Each specialty page lists the codes with device name, class, 21 CFR section and the number of 510(k) clearances in the last three years.
- Open a code page, for example the ITI powered wheelchair page or the CAW oxygen generator page. It shows the device class, 21 CFR section, medical specialty and marketing pathway; eligible codes carry a "Third-party review eligible" flag. It also shows the number and list of 510(k) clearances in the last three years (some summaries can be downloaded directly) and FDA's all-years total. Only the last three years are listed in detail on this site.
- Use the "Official FDA classification page" link at the bottom of the code page to check FDA's classification database (see FDA's ITI classification page as an example). Its Third Party Review field reads "Eligible for 510(k) Third Party Review Program" or "Not Third Party Eligible", and for eligible codes the review organizations recognized for that code are listed beneath it as Accredited Persons. The same page lists the recognized consensus standards and guidance for the code.
- To see an organization's overall scope and contact details, use FDA's list of recognized 510(k) third party review organizations. As of 13 November 2026, that page showed a database update date of 9 November 2026. This article does not compare or recommend any organization.
As of 13 November 2026, the code pages on this site and FDA's classification pages compare as follows (counts are the last-three-year 510(k) figures shown on this site's code pages):
| Code | FDA device name | Class | 21 CFR | This site's code page | FDA classification page | 510(k)s, last 3 years |
|---|---|---|---|---|---|---|
| ITI | Wheelchair, Powered | II | 890.3860 | Third-party review eligible | Eligible | 98 |
| INI | Vehicle, Motorized 3-Wheeled | II | 890.3800 | Third-party review eligible | Eligible | 40 |
| IPL | Wheelchair, Standup | II | 890.3900 | Third-party review eligible | Eligible | 1 |
| IOR | Wheelchair, Mechanical | I | 890.3850 | Third-party review eligible | Eligible | 26 |
| CAW | Generator, Oxygen, Portable | II | 868.5440 | Third-party review eligible | Eligible | 22 |
| IMK | Wheelchair, Stair Climbing | II | 890.3890 | No flag | Not Third Party Eligible | 1 |
Points to note when reading the table:
- IOR is Class I, but 21 CFR 890.3850 reads only "Class I (general controls)" with no 510(k) exemption, and FDA's classification page lists the submission type as 510(k). Class I does not by itself mean exempt.
- CAW corresponds to 21 CFR 868.5440, portable oxygen generator, which the regulation defines as a device intended to release oxygen for respiratory therapy by either a chemical reaction or physical means (for example, a molecular sieve). The last-three-year entries on this site's CAW page include both "Portable Oxygen Concentrator" and "Oxygen Concentrator" without the word portable. Confirm the code for a specific product against FDA's classification database and the chosen predicate.
- INI's FDA device name is Vehicle, Motorized 3-Wheeled, and 21 CFR 890.3800 defines it as a gasoline-fueled or battery-powered device used for outside transportation by disabled persons. For four-wheeled or other scooter designs, confirm the code against the definition and the predicate first.
- FDA's definition for ITI notes that the device type is not intended to climb stairs and points to IMK. IMK is "Not Third Party Eligible" on FDA's classification page, so a 510(k) for it goes directly to FDA.
- IPL and IMK each have only one 510(k) in the last three years, so there are few comparable submissions; check eligibility and predicates against FDA's own pages.
Does test evidence differ between the two routes?
No. FDA states that review organizations use the same 510(k) criteria as FDA, so the performance, electrical safety, EMC and biocompatibility evidence you need depends on the device, the chosen predicate and the applicable FDA guidance and recognized consensus standards, not on who reviews it. The recognized standards on FDA's classification pages are a good starting point. The classification pages for ITI, INI, IOR, IPL and IMK all list the "Guidance Document for the Preparation of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles". The recognized standards on the ITI page include several parts of ISO 7176 and several sections of ANSI/RESNA WC-1 and WC-2:2019. The CAW page lists ISO 80601-2-69 (second edition, 2020) and ISO 18562-1 to -4 (2024, biocompatibility evaluation of breathing gas pathways). For how wheelchair test reports are used in FDA and CE filings, see Can a Wheelchair Test Report Be Used Directly for FDA and CE Filings?
The parts of Hefei SUNGO Medical Technology Co., Ltd.'s accreditation that relate to these products are as follows; the certificates and their scope schedules prevail:
- ISO 7176 series (wheelchairs): CNAS (L24279) covers 17 parts, 14 of which are also within the CMA scope (260015342945); IAS (TL-1119) covers 20 parts.
- ANSI/RESNA WC-1 and WC-2: within the IAS (TL-1119) scope only.
- ISO 80601-2-69, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8 and IEC 60601-1-11 for oxygen concentrators: within the IAS (TL-1119) scope only. Reports for these are issued under IAS accreditation and cannot be issued as CNAS or CMA reports.
- Shanghai SUNGO Medical Technology Co., Ltd. does not test these product types.
Outside SUNGO's accredited scope are ISO 7176-19, ISO 7176-20, ISO 10542 and the ISO 18562 series listed on the CAW page. For biocompatibility-related tests, SUNGO can only subcontract to a laboratory with the appropriate qualifications, with the client's prior agreement, and the subcontracted items and subcontractor are stated in the report. Whether FDA or a review organization accepts a given report depends on its review, and SUNGO makes no commitment on that.
Frequently asked questions
How do I check if my product code is eligible for 510(k) third party review?
Look for the "Third-party review eligible" flag on this site's product code page, then open FDA's classification page and read the Third Party Review field. If they differ, the FDA page prevails.
Do I pay the FDA 510(k) user fee with third party review?
No. FDA states it does not collect a user fee for third party submissions. The review fee is agreed with the review organization and paid directly to it.
Does FDA re-review a 510(k) after third party review?
FDA makes the final decision. If the organization did not appropriately apply the 510(k) decision criteria or its documentation has substantive quality issues, FDA may re-review all or part of the submission; FDA states it is updating the program to avoid routine re-review.
If my product code is eligible, can any review organization review it?
No. Choose an organization recognized for that product code; FDA's classification page lists them under the Third Party Review field.
Can a stair-climbing wheelchair use third party review?
No. IMK is "Not Third Party Eligible" on FDA's classification page, so its 510(k) goes directly to FDA.
This article is based on official public information available as of 13 November 2026; the latest documents from the regulatory authority prevail. Product code eligibility and the list of review organizations are updated, so check FDA's pages again before you submit. For the scope of SUNGO's laboratory accreditation, see the testing laboratory page.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.