Do wheelchair accessories such as cushions, armrests, footrests and anti-tip devices need an FDA 510(k) when exported to the United States on their own? In most cases, no. These accessories mostly map to product codes under 21 CFR 890.3920 (wheelchair component) or 890.3910 (wheelchair accessory), and both sections are Class I and 510(k) exempt. The exemption is limited by 21 CFR 890.9, however, and the two sections differ on quality system requirements. Below: a code map, the limits of the exemption, the remaining duties, and how to check codes in this database.
Key Points
- Stand-alone wheelchair accessories usually map to codes under 21 CFR 890.3910 (wheelchair accessory) or 890.3920 (wheelchair component). Both are Class I, both state a 510(k) exemption, and both are subject to 890.9.
- The 890.3910 exemption applies only if the device is not intended for use as a protective restraint as defined in 21 CFR 880.6760. The section also exempts the device from the Part 820 quality management system requirements, except for those on records and complaint files under 820.35.
- Section 890.3920 has no quality system exemption. In the openFDA classification data, all 13 codes under it carry a GMP exemption flag of "N", including cushion (IMP), armrest (IML), footrest (IMM) and anti-tip device (IMR).
- Exempt does not mean no obligations. Establishment registration, device listing, labeling and other general controls still apply, and a 510(k) is still required when the intended use or fundamental technology goes beyond legally marketed devices of the same type.
- Test reports requested by buyers (flammability, strength, anti-tip) must match the accessory's function and test object, and cannot support conclusions about a complete wheelchair.
How FDA Classifies Wheelchair Accessories: 890.3910 and 890.3920
FDA places wheelchairs and their accessories in 21 CFR Part 890 (physical medicine devices), under the Physical Medicine panel. Two sections matter most for stand-alone accessories:
- 21 CFR 890.3910 Wheelchair accessory: a device sold separately from a wheelchair to meet the specific needs of a patient who uses one. Examples in the text include armboards, lapboards, pusher cuffs, crutch and cane holders, overhead suspension slings, head and trunk supports, and blanket and leg rest straps.
- 21 CFR 890.3920 Wheelchair component: a device generally sold as an integral part of a wheelchair that may also be sold separately as a replacement part. Examples include armrests, narrowing attachments, belts, extension brakes, curb climbers, cushions, antitip devices, footrests, handrims, hill holders, leg rests, and heel and toe loops.
Roughly, a replacement part of the chair points to 890.3920, and an add-on meeting a specific user need points to 890.3910; the final answer depends on the code's device name and section in the FDA database and on the accessory's intended use. Complete chairs, such as mechanical (890.3850) and powered (890.3860) wheelchairs, are outside this article's scope; an accessory sold together with a complete chair must be assessed as part of that device.
Wheelchair Accessory FDA Product Codes at a Glance
As of 13 November 2026, the openFDA classification data lists 8 product codes under 890.3910 and 13 under 890.3920, all Class I and all marked 510(k) exempt. The table picks out codes for common accessories.
| Accessory | Product code (FDA device name) | Section | 510(k) | GMP exempt flag |
|---|---|---|---|---|
| Wheelchair cushion | IMP (Cushion, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Armrest | IML (Armrest, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Footrest | IMM (Footrest, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Anti-tip device | IMR (Device, Anti-Tip, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Handrim | IMO (Handrim, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Extension brake, curb climber | IMW (Brake, Extension, Wheelchair); IMN (Climber, Curb, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Other components | KNN (Components, Wheelchair) | 890.3920 | Exempt; 890.9 applies | N |
| Armboard, lapboard | IMY (Armboard, Wheelchair); IMX (Board, Lap, Wheelchair) | 890.3910 | Exempt unless a protective restraint; 890.9 applies | Y |
| Head and trunk support | IMS (Support, Head And Trunk, Wheelchair) | 890.3910 | Exempt unless a protective restraint; 890.9 applies | Y |
| Non-protective restraint | KID (Restraint, Wheelchair, Non-Protective) | 890.3910 | Exempt unless a protective restraint; 890.9 applies | Y |
| Other accessories | KNO (Accessories, Wheelchair) | 890.3910 | Exempt unless a protective restraint; 890.9 applies | Y |
Note two things. Similar names can sit under different sections: the armrest (IML) is under 890.3920 and the armboard (IMY) under 890.3910, with GMP flags N and Y. And KID is a non-protective restraint. The 890.3910 exemption depends on the device not being used as a protective restraint, which FDA classifies separately in 21 CFR 880.6760 as a device that limits the patient's movements to the extent necessary for treatment, examination, or protection of the patient or others. If a restraint-type accessory's claims touch on that use, reassess the pathway against both sections and the classification database.
Cushions need one more check. Besides IMP, the database has two other cushion codes. KIC (Cushion, Flotation) corresponds to the flotation cushion in 21 CFR 890.3175, a cushion filled with water, air, gel or another flotation medium and used on a seat to lessen the likelihood of skin ulcers; it is Class I, exempt subject to 890.9, with a GMP flag of N. MOC (Cushion, Flotation, Therapeutic) is listed under 21 CFR 890.5170 (powered flotation therapy bed) as Class II, also marked 510(k) exempt. A cushion with pressure-ulcer claims or a moving medium should not default to IMP; compare its actual claims with each of these sections.
890.3910 vs 890.3920: Is the Quality System Exempt?
Both sections are 510(k) exempt; the real difference lies in 21 CFR Part 820. Since 2 February 2026, Part 820 has been the Quality Management System Regulation (QMSR), which incorporates ISO 13485:2016 by reference and adds FDA's own requirements. Under the current text:
| Item | 890.3910 codes (e.g. KNO, IMY) | 890.3920 codes (e.g. IMP, IML) |
|---|---|---|
| 510(k) | Exempt (not a protective restraint; 890.9 applies) | Exempt (890.9 applies) |
| Part 820 quality management system | Exempt, except records and complaint files under 820.35 | Not exempted; a QMS must be established and maintained under the QMSR |
| Establishment registration and device listing (Part 807) | Applies | Applies |
| US agent for foreign establishments | Applies | Applies |
| Other general controls (adulteration, misbranding and labeling, records and reports, etc.) | Applies | Applies |
FDA's web page on Class I and Class II device exemptions states that a 510(k)-exempt device must still comply with other regulatory controls unless its classification regulation expressly exempts it, and that all devices are subject to the QMSR unless Part 820 notes an exception or exemption. Under 21 CFR 807.40, a foreign establishment must register and list its devices and name one United States agent in its registration. So two "exempt" products can carry different duties: a maker of cushions, armrests or footrests (890.3920) must run a QMS under the QMSR, while a maker of armboards or head and trunk supports (890.3910) must still meet the 820.35 requirements on records and complaint files.
When the 510(k) Exemption No Longer Applies: 21 CFR 890.9
21 CFR 890.9 limits the exemption for a Class I or II generic type to devices that have existing or reasonably foreseeable characteristics of commercially distributed devices in that type. Even if a code is marked exempt, a 510(k) is still required before marketing when:
- the device is intended for a use different from that of a legally marketed device in the generic type, for example a different medical purpose, or lay use where the former intended use was by health care professionals only; or
- the modified device operates using a different fundamental scientific technology from a legally marketed device in the generic type.
Paragraph (c) of 890.9 covers in vitro diagnostic devices and does not apply here. For accessories, the usual trouble spot is the claims. An ordinary cushion that adds therapeutic claims in its instructions, labeling or promotion may fall into a "different intended use", and an accessory using a technical principle not found in marketed devices of its type needs checking against paragraph (b). A regulatory professional should compare the actual claim wording with 890.9. If a 510(k) turns out to be necessary, see Choosing a predicate device for a wheelchair 510(k).
Exporting a Wheelchair Accessory on Its Own: A Five-Step Check
- Define the intended use and claims: what each accessory does, who uses it, and what the instructions, labeling and packaging say.
- Find the code by function in the FDA classification database or this site's code library, and check the regulation number and device name.
- Check the conditions: for 890.3910 codes, rule out protective restraint use; for every code, compare claims and technology with marketed devices under 890.9.
- Confirm quality system duties: QMS under the QMSR for 890.3920 codes; at least the 820.35 records and complaint file requirements for 890.3910 codes.
- Complete registration and listing, designate a US agent if based outside the United States, then confirm which test reports the buyer requires.
If a model comes in several sizes or materials with different claims, assess each variant separately with its own test object; one variant's report should not stand in for another.
Common Mistakes
- Reading "510(k) exempt" as "nothing to do". Registration, listing, labeling, and records and reports still apply, and 890.3920 codes are also subject to the QMSR.
- Assuming similar accessories share the same quality system duties. The armrest (IML) and armboard (IMY) have different GMP flags.
- Filing every cushion under IMP. Cushions with pressure-ulcer claims or a moving medium need checking against 890.3175, 890.5170 and similar sections.
- Using a complete-chair report for a stand-alone accessory, or the reverse. The test objects differ, so the conclusions are not interchangeable.
How to Look Up Wheelchair Accessory FDA Product Codes in the Database
This site's FDA product code library uses data from the openFDA device classification database. As of 13 November 2026, it holds 7,091 codes grouped by medical specialty, 220 of them in Physical Medicine. To check an accessory:
- Open the Physical Medicine product code list, which shows each code's device name, class and 21 CFR section. Most wheelchair accessory codes have "Wheelchair" in the name, and the 21 CFR column shows whether a code sits under 890.3910, 890.3920 or another section.
- Open individual code pages, such as the IMP (Cushion, Wheelchair) code page and the KNO (Accessories, Wheelchair) code page. Each shows the class, 21 CFR section, medical specialty, marketing pathway and a link to the official FDA classification page. The KNO page carries a "GMP exempt" flag and the IMP page does not, reflecting the quality system difference between the two sections.
- In the regulation search, a product code (for example IMP) gives the most precise result, and a word from the English device name (for example Cushion) lists several cushion codes at once, including IMP, KIC and MOC. As of 13 November 2026, a regulation number such as 890.3920 returns no codes, so use the specialty list to find codes by section.
- The FDA 510(k) count on each code page covers all years. As of 13 November 2026, the KNO page shows 35 and the IMP page shows 2; in the openFDA 510(k) data, the latest decisions for the two codes date from 1987 and 1989. A historical count does not mean products under the code must file a 510(k).
Code data here is for reference only, not a registration strategy; the FDA database and current regulation text prevail.
Test Evidence and SUNGO Lab
Buyers or importers may ask for flammability, strength or stability reports depending on the accessory. Below are the accessory-related entries in the accreditation scope of Hefei SUNGO Medical Technology Co., Ltd. (the Hefei laboratory); the certificates govern the exact scope. The Shanghai laboratory's scope does not include these items.
- Flammability (resistance to ignition) of cushions and postural support devices: the Hefei CNAS (L24279) scope lists clauses 7.2 (horizontal test), 7.3 (vertical test) and 7.4 (small postural support devices) of ISO 16840-10:2021/Amd 1:2024, and the IAS (TL-1119) scope lists ISO 16840-10 as a whole standard without a year. Check the edition first: according to the ISO website, ISO 16840-10:2026 (edition 3) was published in November 2026, and the 2021 edition and its 2024 amendment were withdrawn on 9 November 2026. The older ISO 7176-16:2012 was withdrawn on 8 June 2021 and replaced by ISO 16840-10:2021. Confirm in writing which edition the buyer requires before placing the order.
- Strength of armrests, footrests and similar parts: the Hefei CNAS scope lists ISO 7176-8:2014 clause 8 (static strength), clause 9 (impact strength), 10.3 (double-drum fatigue), 10.4 (drop fatigue) and 10.5 (fatigue of manually operated parking brakes). The CMA (260015342945) and IAS scopes also list ISO 7176-8.
- Anti-tip devices: the Hefei CNAS scope lists ISO 7176-1:2014 clause 11, static stability with forward or rearward anti-tip devices. The CMA and IAS scopes also list ISO 7176-1.
On the certificates, the test object for these items is wheelchairs and mobility scooters. For a stand-alone accessory, confirm with the laboratory beforehand whether the sample is within the accredited scope, which wheelchair it is mounted on, and what the report covers; the results cannot be extended to a complete wheelchair.
Outside SUNGO's accreditation scope: cushion performance under ISO 16840-2, -3 and -6, and the motor-vehicle and restraint-related ISO 7176-19 and ISO 10542. If a buyer asks for biocompatibility data on cushions or other skin-contacting accessories, SUNGO can only subcontract the tests to a laboratory with the appropriate qualifications, with the client's prior consent and with the subcontracted items and subcontractor stated in the report; see Biocompatibility of Wheelchair Cushions, Armrests and Handgrips. An introduction to SUNGO's laboratories is on the SUNGO Lab page; the accreditation scope is defined by each certificate's schedule.
Frequently Asked Questions
Does a wheelchair cushion need a 510(k) for the US market?
An ordinary cushion maps to IMP under 21 CFR 890.3920: Class I and 510(k) exempt, subject to 890.9. Claims or technology beyond marketed cushions still need a 510(k), and pressure-ulcer claims also call for a check against 890.3175 and related sections.
Do 510(k)-exempt wheelchair accessories still need FDA registration?
Yes. The exemption covers premarket notification only. Registration, listing, labeling and other general controls still apply, and a foreign establishment must designate a US agent.
What is the difference between 21 CFR 890.3910 and 890.3920?
Both are 510(k) exempt. Section 890.3910 also exempts devices from the Part 820 quality management system requirements, except the 820.35 requirements on records and complaint files; 890.3920 does not, so the QMSR applies.
How do I find the FDA product code for a wheelchair accessory?
Use the 21 CFR column in this site's Physical Medicine list, or search by code or English device name, then check the code page and treat the official FDA classification page as final.
Can a stand-alone accessory report replace a complete-wheelchair report?
No. It only shows that accessory's results under those test conditions; check the sample description, standard edition and accreditation scope item by item.
In short: find the product code by function, clear the 890.9 and quality system hurdles, then prepare the reports the buyer asks for. This article is based on official public information as of 13 November 2026; the latest documents from the regulator prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.