Does a shower chair or commode chair need an FDA 510(k) for the US, and does it need a notified body for CE marking in the EU? In short: in the US these products most often use product code ILS, "Adaptor, Hygiene", under 21 CFR 890.5050, which is Class I and 510(k) exempt, but establishment registration and device listing still apply. In the EU, a non-active product intended to compensate for a disability is generally Class I under Rule 1 of MDR Annex VIII, and the manufacturer issues its own declaration of conformity without a notified body certificate. This article sets out the classification basis, quality system requirements and the role of ISO 17966 test reports in both markets, and shows how to look up FDA product codes in this database.

Key takeaways

  • US: ILS falls under 21 CFR 890.5050 (daily activity assist device). It is Class I and exempt from 510(k), subject to the limitations in 890.9. If not labeled as sterile, it is also exempt from most of part 820, except the records and complaint file requirements under § 820.35.
  • US: Wheelchair commode attachments (INB, 21 CFR 890.3920) and chairs with casters (INM, 21 CFR 890.3100) are also 510(k) exempt, but their regulations grant no part 820 exemption, so a quality management system under the QMSR is required.
  • A 510(k) exemption is not a registration exemption. A foreign establishment still registers and lists under 21 CFR 807, pays the annual registration fee and designates one US agent.
  • EU: A non-active shower or commode chair intended to compensate for a disability is generally Class I under Rule 1. The manufacturer draws up technical documentation and issues an EU declaration of conformity. A notified body is involved only for Class I devices placed on the market sterile, with a measuring function, or that are reusable surgical instruments. Non-EU manufacturers must designate an EU authorised representative.
  • ISO 17966, GB/T 43386, MZ/T 119 and GB/T 24434 reports are not documents that FDA or the MDR names as mandatory. They are test evidence for the technical file and for buyer acceptance.

Are shower chairs and commode chairs medical devices?

This article covers assistive products for personal hygiene: shower chairs and shower stools, commode chairs (including mobile commodes with a pail), raised toilet seats, and wheeled shower commode chairs. In both the US and the EU, whether such a product is a medical device turns on its intended use, not its appearance.

21 CFR 890.5050 defines a daily activity assist device as a modified adaptor or utensil (for example a dressing, grooming, recreational activity, transfer, eating or homemaking aid) that is intended for medical purposes to assist a patient to perform a specific function. Article 2(1) of the EU MDR includes, among medical purposes, the diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability. How the instructions, labels and marketing describe the intended use therefore decides whether the product enters device regulation in either market, so settle those claims before export.

Whether China regulates these products as medical devices depends on the NMPA classification catalogue and classification determinations, and this article makes no statement on it. Product performance such as slip resistance and load capacity is covered in a separate Chinese-language article on this site. This article deals only with regulatory status and route to market.

FDA product codes for shower and commode chairs: ILS, INB, INM

The US classifies devices by product code. Three codes are common for these products. The table is based on the openFDA device classification database (read on 13 November 2026, data updated 2 November 2026) and the current regulation text; see 21 CFR 890.5050 on eCFR.

Product codeDevice name21 CFR sectionClass510(k)Part 820 exemption
ILSAdaptor, Hygiene890.5050 Daily activity assist deviceClass IExemptYes, if not labeled sterile; § 820.35 requirements retained in part
INBAttachment, Commode, Wheelchair890.3920 Wheelchair componentClass IExemptNo
INMChair, With Casters890.3100 Mechanical chairClass IExemptNo

Under 890.3920, a wheelchair component is a device generally sold as an integral part of a wheelchair that may also be sold separately as a replacement part. Under 890.3100, a mechanical chair is a manually operated device that helps a disabled person perform an activity the person would otherwise find difficult or be unable to do; the examples given are a chair with an elevating seat and a chair with casters used to move from place to place while seated. There are 15 product codes under 890.5050, all Class I and 510(k) exempt, including ILS and IKX "Aid, Transfer".

FDA registration and listing data show which codes companies actually use. On 13 November 2026 we counted openFDA listing records by proprietary-name keyword (data updated 2 November 2026):

Proprietary name keywordCode with most listing recordsSecond most common code
showerILS (112 records)INM (12 records)
commodeILS (82 records)INB (71 records)
toiletILS (91 records)INB (14 records)

These are listing record counts, not product or company counts, and they only show how the codes are used. Shower chairs and raised toilet seats cluster under ILS, commode chairs appear under both ILS and INB, and some wheeled shower commode chairs are listed under INM. The final choice must come from comparing your own product's intended use and design with the regulation text.

Note the limitations in 21 CFR 890.9. The exemption covers only devices with existing or reasonably foreseeable characteristics of commercially distributed devices in that generic type. A 510(k) is still required if the intended use is different (for example a different medical purpose, or lay use where the legally marketed device was for health care professionals only) or if the device uses a different fundamental scientific technology. A similar product name does not settle the question.

What still applies after the FDA 510(k) exemption

First, registration and listing. According to FDA's device registration and listing page, owners or operators of establishments involved in the production and distribution of devices intended for use in the US must register annually, and FDA collects an annual establishment registration fee. Establishments that must register are generally also required to list their devices. Under 21 CFR 807.40, a foreign establishment must give the name, address and phone number of its US agent when it registers, and may designate only one US agent. The fee amount and small business waiver criteria follow FDA's user fee notice for the relevant year.

Second, the quality system. Since 2 February 2026, 21 CFR part 820 has been the Quality Management System Regulation (QMSR), which incorporates ISO 13485:2016 by reference. For background see our article FDA replaced QSR with QMSR. The current text of 890.5050, as amended by technical amendments published on 4 December 2025, states that a daily activity assist device not labeled or otherwise represented as sterile is exempt from the current good manufacturing practice requirements of part 820, except for requirements concerning records and complaint files under § 820.35. Older sources that cite § 820.180 and § 820.198 reflect the pre-QMSR section numbers.

The regulations for INB and INM contain no such exemption, so manufacturers need a quality management system under the QMSR. However, § 820.10(c) requires design and development under ISO 13485 Clause 7.3 only for Class I devices automated with computer software and a short list of named devices. Ordinary shower and commode chairs are not on that list.

CE marking for shower and commode chairs under the EU MDR

Rule 1 of Annex VIII states that all non-invasive devices are Class I unless one of the later rules applies. An ordinary shower chair, commode chair or raised toilet seat is non-invasive and non-active, so it usually falls under Rule 1. A product with an active function such as powered height adjustment is an active device and must be checked against Rules 9 to 13 and the special rules; if no other rule applies, Rule 13 places it in Class I.

Under Article 52(7), a manufacturer of a Class I device draws up the technical documentation set out in Annexes II and III, issues the EU declaration of conformity under Article 19, and affixes the CE marking under Article 20. A notified body is involved only for Class I devices placed on the market sterile, with a measuring function, or that are reusable surgical instruments, and only for the sterility, metrology or reprocessing aspects. For an ordinary Class I shower chair, "CE certification" therefore means a manufacturer's self-declaration; there is no notified body certificate.

Self-declaration still involves work. Article 10 requires a quality management system proportionate to the risk class, a clinical evaluation under Article 61, and up-to-date technical documentation. Article 11 requires a non-EU manufacturer to designate a sole authorised representative. Articles 29 and 31 set out the registration of devices and manufacturers before placing on the market; for the EUDAMED registration dates see our article Two EUDAMED deadlines.

US and EU routes compared

ItemUS FDAEU MDR
Classification basis21 CFR section linked to the product code (890.5050, 890.3920, 890.3100)Annex VIII classification rules (Rule 1 for most non-active products)
Typical classClass IClass I
Premarket route510(k) exempt, subject to 890.9Technical documentation, manufacturer's EU declaration of conformity, CE marking
Quality systemILS not labeled sterile: only § 820.35 records and complaint requirements; INB and INM: full QMSRArticle 10 quality management system proportionate to risk class
RegistrationEstablishment registration and device listing, annual feeRegistration in EUDAMED under Articles 29 and 31
Local representativeUS agent (21 CFR 807.40)EU authorised representative (Article 11)

Four mistakes come up often. First, treating a similar product name as the same generic type and ignoring the 890.9 limits on intended use and technology. Second, reading a 510(k) exemption as no registration and no fee. Third, keeping pre-QMSR section numbers and requirements instead of updating quality documents to the current part 820. Fourth, assuming that because the US route is exempt, the EU technical documentation and clinical evaluation can be skipped. The two markets need separate preparation.

Where ISO 17966 test reports fit for FDA and CE

Buyers often ask for ISO 17966 or GB/T 43386 test reports. GB/T 43386 is titled, in English, "Assistive products for personal hygiene that support users - Requirements and test methods", matching the title of ISO 17966. The Chinese standards MZ/T 119 and GB/T 24434 cover shower chairs and commode chairs respectively. The requirements are those of the standard texts.

In the US, a 510(k)-exempt product has no premarket submission, so no test report goes to FDA before marketing. Such reports serve mainly as the manufacturer's own verification and quality records and as evidence for buyer acceptance. In the EU, a report can be one piece of evidence in the technical documentation that the product meets the general safety and performance requirements of Annex I. In neither market should anyone state that "FDA or the MDR requires ISO 17966". Annex II requires the technical documentation to state the methods and standards used to demonstrate conformity, so which standard to use, and why, is for the manufacturer to explain.

How to look up FDA codes for shower and commode chairs in this database

The pages below were opened and checked on 13 November 2026. The data come from openFDA; classification and pathways follow FDA's current publications, and this site does not infer a classification for any specific product.

  1. Open the FDA Physical Medicine product code list. As of 13 November 2026 the database holds 7,091 FDA product codes, 220 of them in the Physical Medicine panel. The list shows each code's device name, class and 21 CFR section, so you can find ILS and IKX under 890.5050, and INB and INM under 890.3920 and 890.3100.
  2. Open the ILS product code page. It shows the device class, 21 CFR 890.5050, the marketing pathway "510(K) Exempt", a "GMP exempt" flag, and a link to FDA's official classification page.
  3. Compare the INB product code page: 21 CFR 890.3920, also 510(k) exempt, but with no "GMP exempt" flag. The INM page is the same. That is the quality system difference between the codes.
  4. For Chinese standard records, enter GB/T 43386 or GB/T 24434 on the testing search page to see the standard title, edition year and effective date. Standard texts are copyrighted and are not provided on this site.

Test evidence and Sungo

The IAS accreditation scope of Hefei Sungo Medical Technology Co., Ltd. (certificate TL-1119) includes ISO 17966, GB/T 43386, MZ/T 119 (shower chairs), GB/T 24434 (commode chairs), and ISO 21856 / EN ISO 21856 (assistive products, general requirements and test methods). The scope has exclusions. For ISO 17966 and GB/T 43386, clause 4.3 (clinical evaluation and investigation) and clause 5.3 (biocompatibility and toxicity) are excluded. For MZ/T 119, Chapter 7 (materials) is excluded. For ISO 21856 / EN ISO 21856, clauses 4.3, 5.2, 5.3, 8.3, 11 and 17.4 are excluded.

Where biocompatibility testing is needed, it is subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the report identifies the subcontracted items and the subcontractor. These standards are within the IAS scope only, not within CNAS or CMA scope, and the Shanghai laboratory's accreditation scope does not include them. Whether a regulator or buyer accepts a particular report depends on that party's requirements.

Frequently asked questions

Does a shower chair need an FDA 510(k)?

If the product falls within ILS or another 890.5050 code and its intended use and technology match legally marketed devices of that type, it is 510(k) exempt. Establishment registration, device listing and the annual fee still apply.

Is the FDA product code for a commode chair ILS or INB?

Both are used. INB is "Attachment, Commode, Wheelchair" under 890.3920 and has no part 820 exemption; ILS is "Adaptor, Hygiene" under 890.5050. Choose by comparing the product's design and intended use with the regulations, and confirm in FDA's classification database.

Does a shower chair need a notified body for CE marking?

A Class I shower chair that is not sterile and has no measuring function does not. The manufacturer draws up the technical documentation and issues its own EU declaration of conformity. A non-EU manufacturer still needs an EU authorised representative.

Do 510(k)-exempt products still need a quality system?

It depends on the regulation for the code. ILS devices not labeled sterile keep only the § 820.35 records and complaint file requirements; INB and INM have no such exemption and follow the QMSR.

Is the ISO 17966 report a buyer asks for a legal requirement?

No. Neither FDA nor the MDR names ISO 17966 as mandatory. It is common test evidence for technical files and buyer acceptance.

For shower and commode chairs, fix the intended use first, then work out the code, class and procedure separately for each market; test reports support the technical file and buyer acceptance. This article is based on official public information as of 13 November 2026; the regulators' latest documents prevail. For an introduction to Sungo's laboratory, see Sungo Lab.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.