The Electric Wheelchair Registration Review Guideline (2025 Revision) is one of 39 revised guidelines issued by the Center for Medical Device Evaluation of China's NMPA (CMDE) under CMDE Notice No. 27 of 2025. It guides applicants preparing an electric wheelchair registration dossier in China and serves as a reference for technical reviewers. It is not enforced as regulation, but it sets out how to divide registration units, what goes into the product technical requirements, which standards apply, how test reports may be produced, which model to test, and what to submit on software, cybersecurity, biocompatibility and the clinical evaluation exemption. This article summarises the original Chinese text, without a clause-by-clause comparison with the earlier edition, and shows how to find the related standards in this site's testing database.

Key points

  • Publication: CMDE Notice No. 27 of 2025 is signed and dated 26 November 2025, and the electric wheelchair guideline is attachment 19 of 39. The reprint on the website of CMDE's Yangtze River Delta branch shows a publication date of 1 December 2025.
  • Registration units: the guideline suggests dividing units by intended use, technical principle, structure and performance specifications. Where outdoor and indoor types have markedly different performance requirements, it suggests separate registration units.
  • Technical requirements: performance specifications must not be lower than the applicable mandatory and industry standards. Beyond GB/T 12996, test methods must also follow the GB/T 18029 and GB/Z 18029 series. Electrical safety follows GB 9706.1, EMC follows YY 9706.102 and GB/T 18029.21, and products for home use must also meet YY 9706.111.
  • Test reports: applicants may submit a self-test report or a report from a qualified medical device testing institution. Within a registration unit, the model with the most complex structure, the most functions and the highest technical specifications is tested for performance and electrical safety; anything it cannot cover gets difference testing.
  • Other files: software in the product is classed at least at a moderate safety level, and electric wheelchairs are exempt from clinical evaluation, but the applicant must still submit the comparison documents required for exempt products.

China electric wheelchair guideline 2025: publication and scope

CMDE issued 39 guidelines together under Notice No. 27 of 2025, headed by the Electric Traction Device Registration Review Guideline (2025 Revision); the electric wheelchair guideline is attachment 19. The notice is dated 26 November 2025, and the reprint on the website of CMDE's Yangtze River Delta branch gives 1 December 2025 as the publication date. The original Chinese text can be downloaded from the guideline attachment on the CMDE website.

The scope section defines an electric wheelchair as a wheeled personal mobility device with seat support for people with disabilities, operated by the occupant or a carer, driven by one or more motors, with electrically controlled speed and manual or powered steering. The guideline's sample wording for intended use is transfer of patients with mobility impairment and compensation of walking function for people with disabilities or without full walking ability, indoors or outdoors, carrying one occupant only. For contraindications, the sample wording is that there is no absolute contraindication, but the product must not carry more than one person and must not be driven on motor vehicle lanes. The common product name is normally "electric wheelchair".

The guideline describes itself as a set of general requirements. Applicants decide which parts apply to their product, and where a part does not apply they must give the reasons and the scientific basis.

Registration units: should indoor and outdoor models be split?

Registration units are to be divided under the Guideline on Medical Device Registration Unit Division. Where a product has several models or configurations, the guideline suggests dividing them by key factors such as intended use, technical principle, structure and performance specifications. Where outdoor and indoor types have markedly different performance requirements, it suggests putting them in different registration units.

Where one unit contains several models, their differences in structure, performance and technical features must be shown in a comparison table or annotated diagrams. The guideline also suggests a configuration table in the product technical requirements covering key components, basic principle, main performance specifications, accessories and appearance. Models may be divided by intended use, use environment or construction, for example lateral-folding and longitudinal-folding types.

Device description: how much detail on motors, batteries, steering and brakes

The device description must show the product's functions and technical design. The guideline lists what to cover:

  • Power system: motor type, supplier, model, rated power and torque, heat dissipation and overload protection; controller model, manufacturer, core chip and operating logic, and the control strategy in each driving mode (such as normal and climbing).
  • Energy system: battery type, supplier, model, rated voltage, capacity (Ah) and charge-discharge cycles, the basis for range calculations, and the battery management system's overcharge protection, overdischarge protection and temperature monitoring.
  • Steering and braking: how steering is achieved (for example differential steering), steering sensors and minimum turning radius; brake type (electromagnetic, mechanical friction and so on), how braking is triggered, and braking distance.
  • Other items: tyre size and material; auxiliary features such as anti-tip wheels, seat belts and reversing alarms; overall and folded dimensions, weight including battery, and accessories such as the charger.

Which performance specifications go into the product technical requirements

Section 3.2 asks applicants to set functional and safety specifications in line with the Guideline on Writing Product Technical Requirements for Medical Devices. Operating conditions are not listed as performance specifications; test conditions are stated in the test methods instead. The general performance items in section 3.2.2 can be grouped as follows (specific requirements stated in the original are in brackets):

CategorySpecifications listed in the guideline
Basic parametersOverall length, width and height, wheelbase, seat height, width and depth; total weight including battery, weight of removable parts; maximum load
Drive performanceMotor rated power, speed range and torque; stall protection (when the motor stalls, current no more than 1.5 times rated value, triggering a protective shutdown)
Driving performanceMaximum speed, acceleration and deceleration, climbing ability at full load, range (theoretical distance on a flat surface at normal temperature)
Safety performanceBraking distance on level ground and on slopes, parking slope angle, static and dynamic stability; battery cycle life, charging inhibit (no driving while charging); insulation resistance, dielectric strength, overvoltage protection (automatic cut-off at 1.33 times rated input voltage, response time no more than 0.2 s)
OperationControl modes and response, joystick operating force, minimum turning radius, obstacle climbing height, seat adjustment, shock absorption, flammability rating of seat and back cushions
Special functions (where applicable)Stair climbing (step height, gradient, anti-tip protection device); intelligent assistance functions (such as automatic obstacle avoidance and remote control)

For products with software, the technical requirements must state the software name, model, release version and version naming rules, and set out software requirements such as functions and use limitations under the Guideline on Medical Device Software Registration Review (2022 Revision), consistent with the instructions for use. Electrical safety must meet GB 9706.1 and YY 9706.111 (where applicable). EMC must meet YY 9706.102, GB/T 18029.21 and YY 9706.111 (where applicable).

Electric wheelchair registration testing in China: standards, self-testing and which model to test

Table 2 of the guideline lists the related standards and says the current valid edition of each should be applied, with applicants tracking updates. The table below groups them by purpose; edition years come from the guideline's reference list:

PurposeStandards
Product standardGB/T 12996-2024 Electrically powered wheelchairs (issued 26 October 2024, in force since 1 February 2025)
Test methodsGB/T 18029.1, .2, .6, .7, .8, .13, .14, .15, .22, .25; GB/Z 18029.3, .4, .5, .9; Parts 10, 11 and 16 (numbering explained below)
Electrical safetyGB 9706.1-2020; YY 9706.111-2021 (for use in the home healthcare environment)
EMCYY 9706.102-2021; GB/T 18029.21-2012
OtherGB/T 14710-2009 (environmental requirements), GB/T 29634-2013 (motor drive units), GB/T 18029.19-2014 (wheeled mobility devices used as seats in motor vehicles), GB/T 18029.26-2014 (vocabulary)

One point that is easy to get wrong: the guideline writes Parts 10, 11 and 16 as GB/T 18029.10, GB/T 18029.11 and GB/T 18029.16 (2021 editions in its reference list). On China's national standards information platform, the current 2021 editions are numbered GB/Z 18029.10-2021, GB/Z 18029.11-2021 and GB/Z 18029.16-2021. In product technical requirements and test reports, use the number and year shown on the standards platform.

For test reports, applicants may submit a self-test report or a report from a qualified medical device testing institution. A self-test report must be accompanied by the documents required under the Provisions on Self-Testing for Medical Device Registration (NMPA Announcement No. 126 of 2021).

The tested model must represent the safety and effectiveness of the other products in its registration unit, so the model with the most complex structure, the most functions and the highest technical specifications is tested for performance and electrical safety. If one model cannot cover all the performance and functions of the others, the parts it does not cover are tested separately, with at least an analysis of differences in circuit design, function and intended use. For how to write a coverage rationale for multiple configurations, see Writing a Wheelchair Configuration Coverage Rationale That Holds Up; for narrowing the test items in the technical requirements, see Defining the NMPA Registration Test List for Wheelchairs.

Software, cybersecurity, biocompatibility and clinical evaluation files

  • Software: if the product contains software, its safety level is at least moderate. Research data follow the Guideline on Medical Device Software Registration Review (2022 Revision), including off-the-shelf software and interoperability data and a GB/T 25000.51-2016 self-test report.
  • Cybersecurity: products with electronic data exchange, remote control or user access functions provide research data under the Guideline on Medical Device Cybersecurity Registration Review (2022 Revision).
  • Biocompatibility: the parts expected to contact the patient are mainly the seat, armrests and controller. Short-term skin contact is evaluated under GB/T 16886.1.
  • Cleaning and disinfection: normally no sterilisation; routine wiping or low- to intermediate-level disinfection. If high-level disinfection or sterilisation applies, the recommended process and validation data are required.
  • Home use: products used in the home healthcare environment must meet YY 9706.111, and the instructions for use must state conditions such as the maximum gradient, the maximum obstacle height and the maximum ditch width.
  • Service life and transport: service life studies follow the guideline on service life of active medical devices; environmental and transport simulation may refer to GB/T 14710 and GB/Z 18029.9.
  • Usability: a use error assessment report is submitted as for a moderate use-risk device.
  • Clinical evaluation: electric wheelchairs are devices exempt from clinical evaluation, and the applicant submits the comparison documents required for products on the exemption list. A product outside the catalogue description, or one that cannot show basic equivalence with a device already registered in China, needs clinical evaluation by another route.

How to look up electric wheelchair standards in the database

As of 13 November 2026, the China (NMPA) section of this site's regulatory database is still marked "in progress", so the regulatory search does not return this guideline; use the CMDE attachment linked above. Standard records are in the testing database:

  1. Open Testing search and enter "12996" or "18029". The results list the record pages for GB/T 12996 and each part of GB/T 18029, including Part 21 (EMC) and Part 14 (power and control systems).
  2. A standard record page shows the number, title, edition years and the date it took effect in China. For example, the GB/T 12996 page shows the 2024 edition and its effective date in China, 1 February 2025. This site provides records and official links only, not standard texts.
  3. Edition information is not yet complete on some part pages; for example, the GB/T 18029 Part 10 page lists only the 2009 edition. Confirm the current edition on China's national standards information platform.
  4. The "Laboratories covering this standard" section on a record page is only a lead. Whether a laboratory holds CNAS accreditation or CMA approval for a standard should be checked in the official directories and the certificate scope.

As of 13 November 2026, the standards index of the testing database lists 4,792 medical device standards with record pages.

Test evidence and SUNGO laboratory scope

SUNGO's electric wheelchair testing is carried out at its Hefei laboratory (Hefei SUNGO Medical Technology Co., Ltd.), under CNAS L24279, CMA 260015342945 and IAS TL-1119. The Shanghai laboratory's certificate scope does not include these items. CNAS and IAS are both signatories to the ILAC MRA. The table below lists the standards relevant to this guideline, taken from the certificate scopes:

StandardCNAS accreditationCMA approvalIAS accreditation
GB/T 12996-2024 clauses 6.3 to 6.10 (19 items)YesYesNo
GB/T 18029.1-2024, .2-2022, .6-2024, .7-2009, .8-2024, .13-2008, .14-2012, .22-2024YesYesYes
GB/Z 18029.3, .4, .5, .9, .10, .11, .16 (all 2021 editions); GB/T 18029.15-2008YesNoYes
GB/T 18029.25-2022Chapter 5 battery chargers only, with EMC excludedNoYes
GB 9706.1, YY 9706.111 (electrical safety); YY 9706.102, GB/T 18029.21 (EMC)NoNoYes

Limits to note: electrical safety and EMC are covered by IAS accreditation only, with no CMA approval, so SUNGO's reports for these items cannot serve as the CMA reports needed for entrusted testing in a China registration. The same applies to the other items marked "No" for CMA. GB/T 18029.19 and ISO 7176-19 (wheelchairs used as seats in motor vehicles), GB/T 14710 and GB/T 29634 are outside SUNGO's accredited scope. SUNGO does not perform the biological tests behind the biocompatibility evaluation in-house. They can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the subcontracted items and subcontractor are stated in the report. For the full accredited scope, see the SUNGO laboratory page.

Frequently asked questions

Is China's electric wheelchair registration review guideline mandatory?

No. It is not enforced as regulation, and other methods that meet regulatory requirements may be used with detailed research and verification data. The mandatory standards it cites still apply.

Do electric wheelchairs need a clinical trial for NMPA registration?

Normally not. The guideline states that electric wheelchairs are exempt from clinical evaluation; the applicant submits the comparison documents for exempt products. Products outside the catalogue description, or without demonstrated equivalence, need another clinical evaluation route.

Must indoor and outdoor electric wheelchairs be separate registration units?

The guideline uses the word "suggest": where their performance requirements differ markedly, separate registration units are suggested. The decision also follows the Guideline on Medical Device Registration Unit Division.

Which electric wheelchair model should be sent for registration testing?

The model with the most complex structure, the most functions and the highest technical specifications. Parts it cannot cover are tested separately on the other models.

Do existing registration certificates need retesting under the 2025 revision?

The guideline sets out no transition arrangements for existing certificates at change or renewal. Whether more data or testing is needed is for the registration authority to decide; confirm with it before filing.

Standard numbers and editions should be confirmed on China's national standards information platform. This article is based on official public information available as of 13 November 2026; refer to the latest documents from the regulator.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.