De Novo requests must use eSTAR: FDA's requirement took effect on October 1, 2025, so unless exempt, originals, supplements and amendments must be prepared in eSTAR and, for CDRH, uploaded through the CDRH Portal, as 510(k)s have been since October 1, 2023. PreSTAR, the template for Pre-Submissions and other Q-Submissions, now covers more submission types, but as of November 11, 2026 its use remains voluntary. This article covers the mandatory scope, De Novo screening and hold rules, the draft Q-Sub guidance and 2026 template changes.
Key points
- De Novo requests must use eSTAR. Under the final guidance "Electronic Submission Template for Medical Device De Novo Requests" (August 23, 2024), originals, supplements and amendments must use eSTAR from October 1, 2025 unless exempted; FDA does not intend to grant waivers.
- 510(k)s have required eSTAR since October 1, 2023, covering Traditional, Special and Abbreviated 510(k)s. Both 510(k)s and De Novos to CDRH must go through the CDRH Portal.
- PreSTAR for Pre-Subs is still voluntary. The Q-Sub template guidance, issued as a draft in May 2025, is still labeled a draft as of November 11, 2026; it proposes a transition of at least one year after finalization before electronic Pre-Subs become mandatory.
- No Refuse to Accept (RTA) review; technical screening instead. FDA does not anticipate an RTA process for eSTARs and runs virus scanning and technical screening; a failed submission is held and considered withdrawn if no replacement arrives within 180 days of notification. For De Novo requests, screening starts only after the user fee is paid.
- Template content changed in 2026. The human factors content guidance was published on May 29, 2026; the eSTAR page gives August 1, 2026 as its effective date, and its content is built into the non-IVD and IVD eSTARs.
What is eSTAR, and which submissions must use it?
eSTAR (electronic Submission Template And Resource) is FDA's interactive PDF form. It shows only questions that apply, integrates FDA product codes and recognized consensus standards, and mirrors reviewers' internal review templates. Per FDA, a properly completed eSTAR should represent a complete submission.
The legal basis is section 745A(b) of the Federal Food, Drug, and Cosmetic Act (FD&C Act), as amended by section 207 of the FDA Reauthorization Act of 2017. It requires 510(k)s, De Novo requests, PMAs, IDEs, HDEs and pre-submissions, among others, to use the electronic format FDA specifies from a date set in final guidance, which also sets exemption and waiver criteria. Each type therefore depends on its own guidance; the Q-Sub one is still a draft.
Per FDA's eSTAR page as of November 11, 2026 (submissions to CDRH or CBER):
| Submission type | eSTAR requirement | Notes |
|---|---|---|
| 510(k) (Traditional, Special, Abbreviated, with supplements and amendments) | Mandatory unless exempted | In effect since October 1, 2023 |
| De Novo (original, supplements, amendments) | Mandatory unless exempted | In effect since October 1, 2025 |
| 510(k) and De Novo for combination products | Mandatory | Listed as mandatory on the eSTAR page |
| Dual 510(k)/CLIA Waiver (IVD, to CDRH) | Mandatory | Listed as mandatory on the eSTAR page |
| Original PMAs and listed PMA supplements | Voluntary | Panel Track, Real-Time, 180-Day and 30-Day Notice/135-Day supplements |
| Q-Submissions (Pre-Subs and five other types) | Voluntary, via PreSTAR | Q-Sub guidance still a draft |
| IDEs and 513(g) requests | Voluntary, via PreSTAR | Not for IDE reports |
De Novo: eSTAR mandatory from October 1, 2025
De Novo classifies novel devices with no legally marketed predicate into class I or II; such devices may be marketed and may later serve as 510(k) predicates. The procedures are in 21 CFR part 860, subpart D.
De Novo content was added to eSTAR in January 2022 for voluntary use. On August 23, 2024, FDA issued the final De Novo electronic submission guidance (Docket FDA-2023-D-3788), setting October 1, 2025 as the effective date with a transition period of at least one year. From that date, a De Novo request not provided electronically will not be received unless exempt. FDA's De Novo page adds that requests mailed on electronic media were only to be accepted if received before October 1, 2025.
What is exempt
The exemptions cover only ancillary documents:
- Interactive review responses: questions raised by a reviewer by phone or email are answered by email; formal Additional Information responses must still be submitted in eSTAR;
- Certain amendments: appeals or requests for supervisory review, substantive summary requests, change in correspondent, and add-to-files after the final decision;
- Withdrawal requests.
These remain subject to applicable eCopy requirements. FDA has identified no circumstances appropriate for a waiver and does not intend to grant waivers.
Technical screening, holds and withdrawal
With the acceptance review under 21 CFR 860.230 largely automated in eSTAR, technical screening checks that answers match the device description (e.g., no tissue-contacting components if so stated) and that each applicable attachment question has a relevant attachment. Per the guidance:
- FDA begins technical screening only once the De Novo user fee has been paid;
- Screening should occur within 15 calendar days of receipt; if it is not completed in that period, FDA accepts the request and notifies the requester;
- Screening checks for clear ineligibility, such as an existing predicate, classification regulation or approved PMA for the same device type; an ineligible request is placed on hold;
- A request proposing to classify more than one device type is also placed on hold until the device type is clarified with the review team;
- If the eSTAR is incomplete, FDA notifies the submitter by email; without a replacement within 180 days of that notification, the request is considered withdrawn and closed;
- For a submission that passes, the review clock starts on the day FDA received it; screening time does not affect the clock.
Once the user fee is paid and a valid eSTAR provided, FDA emails an acknowledgment letter with the De Novo number: "DEN" or "BR" followed by six digits, the first two digits being the year of receipt.
What is PreSTAR, and is it required for Pre-Subs?
PreSTAR, the Early Submission Requests eSTAR, is the eSTAR template for Q-Subs, IDEs and 513(g) requests; as of November 11, 2026 all of these uses are voluntary. The Q-Submission (Q-Sub) program is how sponsors request written feedback or meetings with FDA. On May 29, 2025, FDA issued an updated final Q-Sub guidance (Docket FDA-2018-D-1774), superseding the June 2, 2023 version. It details Pre-Submissions, Submission Issue Requests, Study Risk Determinations, Informational Meetings and PMA Day 100 Meetings, states that Q-Subs are subject to eCopy requirements under section 745A(b), notes the voluntary PreSTAR for Pre-Subs, and recommends the CDRH Portal for CDRH packages.
The same day, FDA announced the draft guidance "Electronic Submission Template for Medical Device Q-Submissions" (Docket FDA-2025-D-1082; comments requested by July 28, 2025). Under the draft:
- Once finalized, all Pre-Subs (originals, supplements, amendments) must be submitted electronically unless exempted;
- FDA intends a transition period of at least one year from the final guidance; until then, eSTAR may be used voluntarily;
- Proposed exemptions include interactive review responses, appeals, change in correspondent or legal entity, amendments after decision, meeting minutes and disagreements with them, presentation slides, and withdrawals;
- PreSTAR elements would replace the Pre-Sub Acceptance Checklist in Appendix 1 of the Q-Sub guidance;
- Screening of a Pre-Sub is anticipated within 15 calendar days; a failed Pre-Sub is held, is considered withdrawn without a response within 180 days, and the review clock restarts at day 0 when new information arrives.
When the draft was issued, PreSTAR held only Pre-Sub and 513(g) content. The June 1, 2026 eSTAR page update shows PreSTAR available for voluntary use with five more Q-Sub types: Submission Issue Requests, Informational Meetings, Study Risk Determinations, PMA Day 100 Meetings and Accessory Classification Requests. As of November 11, 2026, FDA's guidance page still labels the document a draft not for implementation; finalization and any mandatory start date depend on later FDA notices.
| Item | As of November 11, 2026 | After finalization and transition (per the draft) |
|---|---|---|
| Format | eCopy, or PreSTAR voluntarily | eSTAR required unless exempted |
| Acceptance | Checklist in Appendix 1 of the Q-Sub guidance; technical screening if PreSTAR is used | Technical screening; PreSTAR elements replace the checklist |
| Meeting minutes and slides | eCopy via the CDRH Portal | On the proposed exemption list |
| Transition | Not applicable | At least one year from the final guidance |
Two changes to eSTAR content in 2026
Human factors content
On May 29, 2026, FDA published the final guidance "Content of Human Factors Information in Medical Device Marketing Submissions" (Docket FDA-2015-D-4599), a risk-based framework for the human factors information to include in a marketing submission to CDRH, complementing "Applying Human Factors and Usability Engineering to Medical Devices". The Federal Register notice says FDA generally does not expect submissions received before August 1, 2026 to include the newly recommended information. The eSTAR page update of June 1, 2026 says the non-IVD and IVD eSTARs include the guidance's content and gives August 1, 2026 as its effective date. In the updated template, human factors content for 510(k)s, De Novos and PMAs follows this guidance; check what to provide against the template and the guidance.
eSTAR version numbers and version rules
As of November 11, 2026, FDA's eSTAR page lists the nIVD eSTAR Version 7.1 and IVD eSTAR Version 7.1 (510(k), De Novo and PMA), and PreSTAR Version 3.1. Major version updates reflect policy or regulatory changes from final guidance or final rules (e.g., 1.2 to 2.0); other changes are minor. Once FDA acknowledges an original submission, it is grandfathered to that version, and responses to Additional Information or Technical Screening requests should update the original eSTAR rather than switch versions.
eSTAR mandatory dates: timeline
| Date | Event |
|---|---|
| February 2020 | CDRH eSTAR pilot for 510(k)s |
| September 2020 | 510(k) eSTAR open to all submitters voluntarily |
| January 2022 | De Novo content added to eSTAR |
| October 1, 2023 | eSTAR mandatory for 510(k)s |
| August 23, 2024 | Final De Novo electronic submission guidance issued |
| May 29, 2025 | Updated final Q-Sub guidance issued; draft Q-Sub template guidance announced |
| July 28, 2025 | Comment date for the draft |
| October 1, 2025 | eSTAR mandatory for De Novo requests |
| May 29, 2026 | Final human factors content guidance published |
| June 1, 2026 | eSTAR page update: human factors content; PreSTAR extended to five more Q-Sub types |
| August 1, 2026 | Human factors content guidance effective (date given on the eSTAR page) |
| As of November 11, 2026 | Q-Sub template guidance still a draft; PreSTAR voluntary |
How to submit an eSTAR: CDRH Portal, software and attachments
Accounts and submission
Anyone can register for a CDRH Portal account; identity verification uses Okta, so corporate mail filters should let emails from @okta.com addresses through. The Portal accepts CDRH submissions only; CBER submissions go through FDA's Electronic Submissions Gateway (ESG). Submissions received before 4 PM ET on a business day are processed that day, later ones the next business day. Only the official correspondent (the eSTAR "Contact" or "Primary Correspondent/Consultant") and designated delegates can track progress, so settle this before submitting.
Software and attachments
- eSTAR cannot be opened in a browser. Save it, then open it with Adobe Acrobat Pro (recommended), PDF-XChange Editor 10.2.0+ or Foxit PDF Reader 12.1+; Adobe Acrobat Reader cannot save changes or add attachments.
- Through the CDRH Portal, attachments are limited to 1 GB each and the eSTAR to 4 GB; larger submissions are mailed in electronic form as the Portal page describes. Attachment names are limited to 124 characters, and file types such as zip are rejected by the template.
- Only one person can edit the file at a time; keep attachments on a local drive before attaching them.
- Form FDA 3881 (Indications for Use), Form FDA 3514 (cover sheet) and, if applicable, the Declaration of Conformity are built into eSTAR.
- For testing, provide complete test reports as FDA requires, under the relevant attachment questions. The Standards section dropdowns list FDA-recognized consensus standards; non-recognized standards can be typed in, leaving the recognition number blank.
Common mistakes
- "Finished means ready." The CDRH Portal accepts an eSTAR only when its status reads "eSTAR COMPLETE". That is completeness of form; inaccurate answers (e.g., a wrong dropdown choice that skips later questions) can still lead to an early hold.
- "A screening hold can wait." The 180 days run from the hold notification; after that the submission is considered withdrawn. A De Novo Additional Information letter in substantive review likewise needs a complete response within 180 calendar days, with no extensions, per FDA's De Novo page.
- "Rebuild the response in the newest template." Modify the original eSTAR instead, select the Additional Information sub-type, answer each question, and leave unaffected sections unchanged.
How to look it up in the database
Before preparing an eSTAR, settle three things: the product code and classification, the applicable regulations, and the standards you will cite.
- Use the product code to choose the pathway. FDA device product codes lists 7091 codes as of November 11, 2026, grouped by medical specialty. Each code page shows the device name, class, 21 CFR section, marketing pathway, third-party review eligibility and 510(k) clearances in the last three years. Where a code and classification regulation already exist for the same device type, assess 510(k) or another existing pathway first; under the De Novo guidance, a request with an existing predicate, classification regulation or approved PMA for the same device type may be found ineligible and held at technical screening. A code is only a lead; judge device type against the regulation text.
- Search by number. Regulation search covers regulations and guidance, FDA product codes and recent 510(k) clearances. The 510(k) records come from the openFDA 510(k) dataset, which includes a few DEN-prefixed records; under FDA's numbering these are De Novo numbers, so count them separately from K-numbered 510(k)s.
- Find the procedural regulation. The 21 CFR 860 entry links to the official eCFR text of the classification procedures, including the format, content and review of De Novo requests. The United States regulations page shows 44 regulations and guidance entries and 318 update records as of November 11, 2026, including 21 CFR 807 (510(k)), 21 CFR 814 (PMA) and 21 CFR 812 (IDE).
- Check editions and FDA recognition numbers. The standards index holds 4792 medical device standards with pages as of November 11, 2026. Standard pages show editions; entries sourced from FDA's recognized consensus standards database also show the FDA recognition number to compare with the eSTAR Standards dropdown. Standard texts are not provided.
FAQ
Can a De Novo request still be mailed on a disc or USB drive?
Not as a routine route. FDA only intended to accept mailed media received before October 1, 2025; non-exempt requests to CDRH go through the CDRH Portal as eSTARs, except files over the Portal's size limits, which are mailed as the Portal page describes.
When will PreSTAR become mandatory for Pre-Subs?
Only after the Q-Sub template guidance is finalized; the start date will be set in the final guidance, and the draft says FDA intends at least a one-year transition. As of November 11, 2026 the guidance is still a draft.
What if I use an older eSTAR version?
Per FDA, if you do not use the major version FDA has implemented and its changes apply to your device, you may get additional information requests about them; minor version gaps typically do not.
Do Pre-Sub meeting minutes go into PreSTAR?
No. Per FDA's eSTAR page, Pre-Sub meeting minutes and presentation slides are submitted as eCopy via the CDRH Portal.
510(k)s and De Novo requests are fully on eSTAR, while Pre-Subs await the final guidance. Official sources: FDA eSTAR Program, De Novo electronic submission guidance and draft Q-Sub electronic submission template guidance. This article is based on official public information available as of November 11, 2026; the latest documents from the regulators prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.