What class is an oxygen concentrator under the EU MDR? An oxygen concentrator used to deliver oxygen enriched air directly to the patient is class IIb. The legal basis is Rule 12 in Annex VIII of the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745), and MDCG 2021-24 Rev.1 (April 2026) lists exactly this device as a Rule 12 class IIb example on page 48. Class IIb devices cannot be self-declared: a notified body must take part in the conformity assessment. This guide covers how Rule 12 works, the class IIb conformity assessment routes, common test standards and how to look them up on this site.
Key points
- Oxygen concentrator classification under the MDR follows Rule 12: active devices that administer or remove medicinal products, body liquids or other substances are class IIa, or class IIb where this is done in a potentially hazardous manner.
- On page 48, MDCG 2021-24 Rev.1 lists "oxygen concentrator used to deliver oxygen enriched air directly to the patient" among the Rule 12 class IIb examples. This is the most direct reference for a classification rationale.
- Classification is governed by the intended purpose (Annex VIII, implementing rule 3.1). Where several rules apply, the strictest rule giving the higher class applies (3.5). Accessories are classified in their own right (3.2).
- Under MDR Article 52(4), class IIb devices follow Annex IX (quality management system plus technical documentation assessment), or alternatively Annex X type examination coupled with Annex XI product conformity verification. Both routes involve a notified body.
- Common test standards are the particular standard ISO 80601-2-69 plus IEC 60601-1 and its EMC, alarm and home healthcare collateral standards. Where applicable, the file also needs evidence on gas pathway biocompatibility (ISO 18562 series), software, usability and batteries.
Background and scope
MDR Article 51 divides devices into classes I, IIa, IIb and III, taking into account their intended purpose and inherent risks, and requires classification in accordance with Annex VIII. Annex VIII contains 22 classification rules (Rule 1 to Rule 22). Section 6, "Active devices", covers Rules 9 to 13, and oxygen concentrators fall under Rule 12.
This guide covers the classification and conformity assessment of medical oxygen concentrators placed on the EU market. It does not cover the US FDA 510(k) route.
MDCG guidance is endorsed by the Medical Device Coordination Group, composed of representatives of all Member States. The cover page of MDCG 2021-24 Rev.1 states that its views are not legally binding and that only the Court of Justice of the European Union can give binding interpretations. A classification rationale should therefore cite both the MDR text and the MDCG example.
Oxygen concentrator MDR classification: how Rule 12 separates class IIa from class IIb
MDR Annex VIII, Section 6.4 (Rule 12) provides that all active devices intended to administer and/or remove medicinal products, body liquids or other substances to or from the body are class IIa, unless this is done in a manner that is potentially hazardous, taking account of the nature of the substances involved, of the part of the body concerned and of the mode of application, in which case they are class IIb.
MDCG 2021-24 Rev.1 explains that Rule 12 is intended primarily to cover drug delivery systems and anaesthesia equipment. Its page 48 table splits the examples into a class IIa row and a class IIb row; check the original in the MDCG 2021-24 Rev.1 PDF published by the European Commission.
| Class | Rule 12 condition | Examples on page 48 of MDCG 2021-24 Rev.1 |
|---|---|---|
| IIa | Active devices intended to administer or remove medicinal products, body liquids or other substances | Suction pump; feeding pumps; jet injectors for vaccination; elastomeric pumps or balloon pumps for infusion |
| IIb | Administration or removal in a potentially hazardous manner (nature of substances, body part, mode of application) | Infusion pumps; ventilators; anaesthesia machines; anaesthetic vaporisers; dialysis equipment; blood pumps for heart-lung machines; hyperbaric chambers; pressure regulators for medical gases with integrated flowmeter; medical gas mixers; moisture exchangers in breathing circuits if used on unconscious or non-spontaneously breathing patients; oxygen concentrator used to deliver oxygen enriched air directly to the patient |
The oxygen concentrator sits in the same row as ventilators, anaesthesia machines and medical gas mixers, all of which deliver gas to the patient. The qualifying words are "used to deliver oxygen enriched air directly to the patient". Compare the intended purpose and gas delivery described in the instructions for use and labelling against that wording. Where they match, the rationale can cite the page 48 example together with MDR Annex VIII, Section 6.4.
Three classification mistakes come up often:
- Applying Rule 13 just because the device is active. Rule 13 makes all other active devices class I, and page 49 of MDCG 2021-24 Rev.1 describes it as a fallback rule for active devices not covered by other rules. Rule 12 already applies to an oxygen concentrator.
- Stopping at one rule. Under implementing rule 3.5, where several rules or sub-rules apply to the same device, the strictest one resulting in the higher class applies. If the product has additional functions, check the other rules too.
- Treating accessories as part of the main unit. Under implementing rule 3.2, accessories such as nasal cannulas are classified on their own intended purpose rather than inheriting class IIb.
Oxygen concentrator CE marking as class IIb: conformity assessment routes
MDR Article 52 sets out the conformity assessment procedures by class. The table compares class IIb, class IIa and ordinary class I (custom-made and investigational devices excluded).
| Class | MDR provision | Procedure | Scope of technical documentation assessment |
|---|---|---|---|
| IIb (non-implantable) | Article 52(4) | Annex IX, Chapters I and III, including technical documentation assessment under Section 4; or Annex X type examination coupled with Annex XI product conformity verification | Under Annex IX, at least one representative device per generic device group |
| IIa | Article 52(6) | Annex IX, Chapters I and III, including technical documentation assessment under Section 4; or technical documentation under Annexes II and III coupled with Annex XI, Section 10 or Section 18 | At least one representative device for each category of devices |
| I (not sterile, no measuring function, not a reusable surgical instrument) | Article 52(7) | EU declaration of conformity after drawing up the technical documentation in Annexes II and III | No notified body involvement |
Annex IX is based on a quality management system and assessment of technical documentation, Annex X on type-examination, and Annex XI on product conformity verification, with Part A (production quality assurance) and Part B (product verification). The right route depends on the manufacturer's quality system and product range and should be agreed with the notified body.
One provision is easy to miss. MDR Article 54 requires notified bodies to follow the clinical evaluation consultation procedure, in addition to Article 52, for class IIb active devices intended to administer and/or remove a medicinal product as referred to in Rule 12; Article 54(2) lists exceptions. This guide does not decide whether this applies to oxygen concentrators; confirm it with the notified body at the start of the project.
Choosing a notified body: check the designation codes
Commission Implementing Regulation (EU) 2017/2185 sets out the list of codes used to specify the scope of designation of notified bodies. The active device codes start with MDA, and MDA 0307 is "active non-implantable respiratory devices". Horizontal codes include MDS (devices with specific characteristics) and MDT (devices for which specific technologies or processes are used). Judging by the code titles, MDA 0307 may be among the codes relevant to an oxygen concentrator, but the notified body assesses which codes finally apply.
When comparing notified bodies, confirm that the MDR designation scope covers those codes, then ask what they expect from test reports, such as standard editions, accreditation marks and models covered. SUNGO is a testing laboratory, not a notified body, and does not provide CE certification or registration services; technical documentation review and certification rest with the notified body.
Oxygen concentrators with an MDD certificate: transition milestones
Under MDR Article 120 as amended by Regulation (EU) 2023/607, class IIb devices with a valid MDD certificate (other than the implantable devices covered by point (a) of Article 120(3a)) may continue to be placed on the market if the conditions are met. The key milestones are:
| Milestone | Date | Provision |
|---|---|---|
| Quality management system in accordance with MDR Article 10(9) in place | No later than 26 May 2024 | Article 120(3c)(d) |
| Formal MDR application lodged with a notified body | No later than 26 May 2024 | Article 120(3c)(e) |
| Written agreement signed with the notified body | No later than 26 September 2024 | Article 120(3c)(e) |
| End date for placing these class IIb legacy devices on the market | 31 December 2028 | Article 120(3a)(b) |
The devices must also continue to comply with the Directive, have no significant changes in design and intended purpose, and present no unacceptable risk. Products that do not meet these conditions cannot rely on the extension and need an MDR certificate first. The current consolidated MDR text on EUR-Lex prevails.
Which test reports an oxygen concentrator CE technical file needs
Under MDR Article 8, devices that conform to harmonised standards whose references are published in the Official Journal of the European Union are presumed to conform to the MDR requirements those standards cover. Oxygen concentrator safety and performance evidence is usually organised around the standards below; whether each is listed as an MDR harmonised standard, and in which edition, follows the Official Journal list.
| Standard | Subject | Role in the technical file |
|---|---|---|
| ISO 80601-2-69 / EN ISO 80601-2-69 | Particular requirements for the basic safety and essential performance of oxygen concentrator equipment | Particular standard for oxygen concentrators |
| IEC 60601-1 / EN 60601-1 | General requirements for basic safety and essential performance of medical electrical equipment | General electrical safety |
| IEC 60601-1-2 / EN 60601-1-2 | Electromagnetic disturbances: requirements and tests | Electromagnetic compatibility (EMC) |
| IEC 60601-1-8 / EN 60601-1-8 | General requirements for alarm systems | Alarm functions |
| IEC 60601-1-11 / EN 60601-1-11 | Requirements for equipment used in the home healthcare environment | Home-use models |
| ISO 18562 series | Biocompatibility evaluation of breathing gas pathways in healthcare applications | Gas pathway biocompatibility |
| IEC 62304, IEC 62366-1, IEC 62133 and others | Software life cycle, usability engineering, battery safety | Models with software, usability needs or batteries |
Two mistakes are common. One is preparing only general standard reports: as a particular standard, ISO 80601-2-69 supplements or modifies clauses of the general and collateral standards, so reports need to reflect its requirements (the standard text governs). The other is a mismatch between the edition used in a report and the edition cited in the technical file. As of 13 November 2026, the ISO 80601-2-69 page on this site shows the edition year 2020. If a new edition is published, confirm with the notified body which edition a new application should use and whether older reports need a gap assessment, relying on ISO's official publication information.
How to look up oxygen concentrator EU regulations and standards in the database
- Open the EU medical device regulations section. As of 13 November 2026, it shows 172 regulation and guidance entries (marked as updated on 18 September 2026), listed by type, number, title and status. The MDR (32017R0745), Regulation (EU) 2023/607 and Implementing Regulation (EU) 2017/2185 are all listed, and their entries link to the official text on EUR-Lex.
- Open the MDCG 2021-24 rev.1 entry. It shows the number, type (guidance), issuer (MDCG), status and last verification date (8 September 2026), with a link to the official European Commission PDF, so you can go straight to page 48 to check the Rule 12 examples.
- Open the ISO 80601-2-69 standard page. As of 13 November 2026, it shows the issuing body, number, edition year, FDA recognition number and the official channel for the standard text. Other standards, such as ISO 18562-1 and IEC 60601-1-11, can be found by standard number in the testing database search box. On the same date, the testing database home page shows 4,792 standards with a published page and 11 testing bodies.
The database provides bibliographic records and official links only, not standard texts or classification decisions. The MDR text, MDCG guidance and the notified body's view govern classification.
Test evidence and SUNGO
SUNGO's testing capability relevant to this topic sits with the laboratory of Hefei SUNGO Medical Technology Co., Ltd. Its IAS accreditation (certificate TL-1119) covers the following whole standards: ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269; IEC 60601-1, EN 60601-1 and GB 9706.1; IEC 60601-1-2, EN 60601-1-2 and YY 9706.102; IEC 60601-1-8, EN 60601-1-8 and YY 9706.108; IEC 60601-1-11, EN 60601-1-11 and YY 9706.111. These items are within the IAS accreditation only, so the reports carry the IAS mark and are not CNAS or CMA reports. CNAS and IAS are both signatories to the ILAC MRA. Whether a notified body accepts a given report is the notified body's decision.
The ISO 18562 series, and biological tests in the ISO 10993 series such as cytotoxicity, sensitisation and irritation, are outside the certificates of both the Shanghai and Hefei laboratories. Where these tests are needed, they can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior agreement, and with the subcontracted items and the subcontractor stated in the report. Software (IEC 62304), usability (IEC 62366-1, IEC 60601-1-6) and battery (IEC 62133) evidence is also outside SUNGO's accreditation. An introduction to the SUNGO laboratories is on the testing laboratory page.
Frequently asked questions
What class is an oxygen concentrator under the EU MDR?
An oxygen concentrator used to deliver oxygen enriched air directly to the patient is class IIb under Rule 12 of MDR Annex VIII, and MDCG 2021-24 Rev.1 lists it as a class IIb example on page 48. If the intended purpose differs from the example, assess the product against the Annex VIII rules one by one.
Does oxygen concentrator CE marking need a notified body?
Yes. Under MDR Article 52(4), class IIb devices follow Annex IX, or Annex X coupled with Annex XI, and both involve a notified body. The self-declaration route, issuing only an EU declaration of conformity, applies to ordinary class I devices under Article 52(7).
Is the nasal cannula supplied with an oxygen concentrator also class IIb?
Not necessarily. Implementing rule 3.2 of Annex VIII classifies accessories in their own right. For reference, page 56 of MDCG 2021-24 Rev.1 lists "oxygen delivery system with a nasal cannula unless treating life-threatening conditions" as a Rule 20 class IIa example.
Is MDCG 2021-24 legally binding?
No. Its cover page states that only the Court of Justice of the European Union can give binding interpretations. It is classification guidance endorsed by the MDCG; cite the document number, revision and page together with the MDR text.
What if the manufacturer and the notified body disagree on classification?
Under MDR Article 51(2), a dispute between the manufacturer and the notified body arising from the application of Annex VIII is referred for decision to the competent authority of the Member State where the manufacturer has its registered place of business. The same paragraph sets out what happens when the manufacturer has no registered place of business in the EU.
For an oxygen concentrator entering the EU, classification is the first step: write the rationale on Rule 12 and the MDCG example, choose a notified body for the class IIb route, then line up the test evidence standard by standard. This guide is based on official public information as of 13 November 2026. Always check the latest documents from the regulatory authorities.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.