For home-use medical devices sold in the US, FDA's design expectations are set out in the final guidance "Design Considerations for Devices Intended for Home Use". IEC 60601-1-11, the collateral safety standard for the home healthcare environment, has two entries in FDA's recognized consensus standards database: IEC 60601-1-11 Edition 2.1 (recognition number 19-38) is partially recognized, with Note 2 to entry in Sub-clause 3.1 excluded, while the US adoption ANSI/AAMI HA60601-1-11 (19-47) is recognized in full. This article explains what the guidance covers, how FDA recognizes these standards, and how both fit into a 510(k) for active home-use devices such as oxygen concentrators, powered breast pumps and compressible limb sleeves.

Key points

  • FDA's home use design guidance was issued on November 24, 2014 under docket FDA-2012-D-1161 and is listed as Final. It contains nonbinding recommendations, but it recommends that premarket submissions show how the hazards of the home environment and lay users were considered and addressed.
  • IEC 60601-1-11 Edition 2.1 (2020-07) carries FDA recognition number 19-38 and is partially recognized: Note 2 to entry in Sub-clause 3.1 (home healthcare environment) is not recognized because it conflicts with the guidance's definition of "professional healthcare facility."
  • ANSI/AAMI HA60601-1-11:2015 [including AMD1:2021], the modified US adoption of IEC 60601-1-11:2015, carries recognition number 19-47 and is recognized as a complete standard. The number cited in a 510(k) should match the version the test report was based on.
  • IEC 60601-1-2 Edition 4.1 (19-36), the EMC standard commonly used for home-use devices, is also partially recognized: the Figure 3 "nursing homes" example and the citation of Note k) in Table 8 of Sub-clause 8.9 are not recognized.
  • Under FDA's voluntary consensus standards guidance, a Declaration of Conformity (DOC) may refer only to standards with a recognition number and may not deviate from them. Standards answer only part of a 510(k); human factors, software, labeling and postmarket arrangements need their own documentation.

Which devices the FDA home use guidance covers

Section IV of the guidance defines a home use device as a device labeled for use in any environment outside a professional healthcare facility, including outdoor environments, offices, schools, vehicles, emergency shelters and independent living retirement homes. A device intended for use both inside and outside professional healthcare facilities also meets the definition.

Section III states that the guidance applies to prescription and over-the-counter devices of Class I, II and III. Its recommendations should also be considered for devices likely to be used in the home even if not intended solely for home use. The guidance is available on the FDA home use guidance page.

The guidance defines a "professional healthcare facility" as an environment where personnel with medical training are continually available to oversee or administer device use, naming hospitals, long-term care facilities, nursing homes, emergency medical services, clinics, physicians' offices and outpatient treatment facilities; a clinical laboratory also counts. FDA's decisions not to recognize Note 2 of IEC 60601-1-11 and the Figure 3 example in IEC 60601-1-2 both rest on this definition.

Two further points. The guidance dates from 2014, so the standard versions it cites are older, for example ANSI/AAMI HA60601-1-11:2011, ANSI/AAMI ES60601-1:2012, the fourth edition of IEC 60601-1-2 (2014) and IEC 60601-1-8 Edition 2.1 (2012); check the recognized versions in FDA's database at the time of submission. It was updated on August 8, 2016 to correct the transition date for IEC 60601-1-2. Also, Section VII is framed around the design controls of the former Quality System (QS) regulation in 21 CFR Part 820. The amended Part 820, titled the Quality Management System Regulation (QMSR), took effect on February 2, 2026, as FDA notes in the preface to the PDF of its voluntary consensus standards guidance.

What the FDA home use guidance covers

The guidance works as a design input checklist:

  • Environment (Section V): location, contaminants, water supply, temperature, humidity, atmospheric pressure, air flow, travel and international use, fluid exposure, storage. FDA recommends summarizing in the submission how these were accounted for where relevant.
  • Users (Section VI): physical, sensory and perceptual, cognitive and emotional differences. The submission should identify the intended users and summarize the actions taken to account for them.
  • Design (Section VII): lock-out mechanisms, maintenance, calibration, mechanical strength, and electrical issues: supply mains, internal power sources, permanent installation, outlets and adapters, power outages, EMC, wireless technology and alarm systems.
  • Human factors (Section VIII): FDA recommends a human factors validation study and refers to its human factors guidance issued February 3, 2016; there is also a subsection on user training.
  • Labeling (Section IX): must address applicable requirements in 21 CFR Part 801 and 809.10, and should cover emergencies, disposal and hygienic maintenance.
  • Postmarket (Section X): customer service and medical device reporting under 21 CFR Part 803.

Some Section VII recommendations bear directly on the 510(k). Portable devices should be tested for function after impact with the ground or other objects. Where an alarm is needed, FDA recommends signals in at least two of three modes: visual, auditory and tactile. For EMC, the submission should summarize the tests and how they were run, the functions and modes tested, pass/fail criteria, reference standards and any deviations or allowances, any modifications needed to pass, and the related labeling. Devices with RF wireless technology should fully describe the technology, frequency band, output power and functions.

FDA recognition of IEC 60601-1-11: 19-38 and 19-47

On the 19-38 page of FDA's recognized consensus standards database, IEC 60601-1-11 Edition 2.1 (2020-07 consolidated version) is partially recognized, entered on December 21, 2020. The standard covers basic safety and essential performance of medical electrical equipment and systems in the home healthcare environment, whether operated by lay persons or trained healthcare personnel. Note 2 to entry in Sub-clause 3.1 is not recognized; FDA states it conflicts with existing final guidance, pointing to the "professional healthcare facility" definition in the home use guidance.

The database also lists ANSI/AAMI HA60601-1-11:2015 [including AMD1:2021], titled as IEC 60601-1-11:2015 MOD, under 19-47, entered on May 30, 2022, as a complete standard. The table summarizes the collateral standards most often involved; the FDA detail pages hold the exact wording.

Standard and versionFDA recognition numberDate of entryExtentPart not recognized, per the FDA page
IEC 60601-1-11 Edition 2.1 (2020-07)19-382020-12-21PartialNote 2 to entry in Sub-clause 3.1 (home healthcare environment)
ANSI/AAMI HA60601-1-11:2015 [including AMD1:2021]19-472022-05-30CompleteNone
IEC 60601-1-2 Edition 4.1 (2020-09)19-362020-12-21PartialFigure 3, "nursing homes" as a home healthcare environment example; citation of Note k) in Table 8 of Sub-clause 8.9, row "Conducted disturbances induced by RF fields"
IEC 60601-1-8 Edition 2.2 (2020-07, alarm systems)5-1312020-12-21CompleteNone

For 19-36, FDA says Figure 3 conflicts with the guidance's "professional healthcare facility" definition, and that Note k), which excludes SIP/SOP cables shorter than 1 m from the IEC 61000-4-6 test, is inconsistent with published literature. EMC files for home-use devices should therefore not treat nursing homes as a home healthcare environment, and excluding short cables under Note k) falls outside FDA's recognition.

Home use guidance items mapped to standards and evidence

Clause numbers are not listed; the standard text is the reference.

Guidance item (section)Standards or documents named in the guidanceTypical 510(k) evidence
Mechanical strength (VII.D)HA60601-1-11 content on mechanical strength for transit-operable and non-transit-operable equipment; FDA also recommends reviewing ASTM D4169Test report plus design description
Supply mains and voltage range (VII.E.1)Applicable specifications of ES60601-1 and HA60601-1-11, the latter specifying voltage ranges; supply interruptions point to IEC 60601-1-2Test report
Internal power and power outages (VII.E.2, VII.E.5)Mainly labeling and design measuresOperating and post-outage use time in the IFU; backup power description
EMC (VII.E.6)IEC 60601-1-2 (recognition number 19-36)Test report and test summary
Wireless technology (VII.E.7)FDA RF wireless technology guidance (August 13, 2013)Wireless description, coexistence and risk management documentation
Alarm systems (VII.E.8)IEC 60601-1-8 (recognition number 5-131)Test report plus alarm mode description
Software (Section VII)IEC 62304 and FDA software guidancesSoftware documentation
Human factors (VIII)FDA human factors guidance, IEC 62366 and othersHuman factors file and validation study
Labeling (IX)21 CFR Part 801 and 809.10Labels and instructions for use

Test reports answer only part of this list. Design descriptions, software documentation, human factors files and instructions for use are prepared by the manufacturer.

How to cite standards and home-use suitability in a 510(k)

FDA's guidance "Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices" (September 2018) describes two uses. A Declaration of Conformity (DOC) may refer only to standards FDA has recognized or decided to recognize, with a recognition number, and may not deviate from them. Its elements include the applicant's name and address, device identification with product codes and model numbers, a statement of conformity, the standards list with each FDA recognition number, the date and place of issuance, and a signature. General use means using a standard without a DOC, for example by choice, because the method was changed (a deviation), or because the standard has no recognition number; FDA recommends stating the basis together with supporting information or data.

The guidance also cites IEC 60601-1-2 as an example where supporting documentation should accompany a DOC, asks that third-party laboratories be identified by name and address with their accreditations referenced, and notes that FDA may request the data behind a DOC at any time. See the FDA voluntary consensus standards guidance page.

Taking both guidances together, a 510(k) for an active home-use device can be organized as follows:

  1. Describe the intended use environment and users in terms consistent with the home use guidance; for example, long-term care facilities and nursing homes are professional healthcare facilities there.
  2. Give the version and recognition number of each standard. Citing IEC 60601-1-11 under 19-38 or 19-47 depends on the version tested. Cite 19-36 for EMC and state whether any unrecognized part was involved.
  3. Attach an EMC test summary covering the items in Section VII.E.6.
  4. Summarize design measures for environment, users, power outages, alarms, wireless, human factors and labeling, using the table above as an index.
  5. Explain deviations or uncertain points as general use with a stated basis; FDA's Q-Submission (Pre-Submission) program can be used for early confirmation.

This is an approach drawn from FDA's public guidance, not a fixed FDA template. Whether FDA clears a submission depends on the device, the data and the review.

Three mistakes are common: assuming an IEC 60601-1-11 report alone satisfies the home use guidance, which also covers human factors, labeling and postmarket issues; copying the 2011 to 2014 standard versions cited in the guidance; and treating nursing homes as a home healthcare environment.

How to look it up in the database

The pages and figures below were read on this site as of November 13, 2026:

  1. The testing standards library has an IEC 60601-1-11 standard record showing edition year 2020 and FDA recognition number 19-38, and a record for ANSI/AAMI HA60601-1-11 showing 19-47. Standard texts are copyrighted; the site provides bibliographic records and official links only.
  2. For oxygen concentrators, see the CAW (Generator, Oxygen, Portable) product code page: Class 2, 21 CFR 868.5440, 510(k) pathway, third-party review eligible. It shows 22 510(k) clearances in its "last 3 years" window, 19 Summaries archived on the site and 288 clearances across all years per FDA. Most of those 22 device names are "Oxygen Concentrator" or "Portable Oxygen Concentrator."
  3. For powered breast pumps, see the HGX (Pump, Breast, Powered) page: Class 2, 21 CFR 884.5160, 79 clearances in the window, 64 Summaries archived, 250 across all years. Compressible limb sleeves use code JOW (Sleeve, Limb, Compressible): Class 2, 21 CFR 870.5800, 21 clearances, 14 Summaries, 352 across all years. All three codes are third-party review eligible.
  4. Code pages list the 510(k) number, decision date, applicant, device name and Summary link of each clearance. Summaries of similar devices show which standards and versions they cited for non-clinical testing. The site's regulatory search also accepts a product code or 510(k) number.

These pages are for quick orientation; recognition status is determined by FDA's recognized consensus standards database.

Test evidence and SUNGO

Electrical safety and EMC test reports are one part of a 510(k) for an active home-use device. SUNGO's only accreditation in this area is IAS accreditation TL-1119, held by Hefei SUNGO Medical Technology Co., Ltd. The Shanghai laboratory is not involved, and the CNAS and CMA certificates of both sites do not cover medical electrical safety or EMC.

  • Within scope (whole standards; no edition year specified on the certificate): IEC/EN 60601-1 and GB 9706.1; IEC/EN 60601-1-11 and YY 9706.111; IEC/EN 60601-1-2 and YY 9706.102; IEC/EN 60601-1-8 and YY 9706.108.
  • The only particular standard in scope is ISO/EN ISO 80601-2-69 and YY 9706.269 (oxygen concentrator equipment).
  • Outside scope: US adoptions such as ANSI/AAMI HA60601-1-11, IEC 62366-1 human factors, IEC 62304 software, IEC 62133 batteries, cybersecurity, and any 60601-2-x or 80601-2-x particular standard other than ISO 80601-2-69.

IAS accreditation is not FDA recognition and does not mean FDA will accept any particular report. An introduction to SUNGO's laboratories is on the SUNGO laboratory page.

Frequently asked questions

Is the FDA home use device guidance mandatory?

No. It contains nonbinding recommendations, and FDA states that using consensus standards is generally not mandatory unless a standard is incorporated by reference into a regulation. The guidance does recommend describing how home-use risks were addressed, which FDA says helps it determine whether safety and effectiveness requirements are met.

Is IEC 60601-1-11 fully recognized by FDA?

IEC 60601-1-11 Edition 2.1 is partially recognized under 19-38, with Note 2 to entry in Sub-clause 3.1 excluded. ANSI/AAMI HA60601-1-11:2015 [including AMD1:2021] is fully recognized under 19-47. Cite the number that matches the version tested.

Do nursing homes count as a home use environment for FDA?

No. Under the home use guidance, long-term care facilities and nursing homes are professional healthcare facilities, while independent living retirement homes are listed among home use environments.

What should home-use devices watch for with IEC 60601-1-2?

Edition 4.1 is partially recognized under 19-36: the Figure 3 nursing homes example is not recognized, nor is the citation of Note k), which excludes SIP/SOP cables under 1 m from IEC 61000-4-6 testing. The EMC test summary should state whether either point was involved.

Is the 2014 FDA home use guidance still in effect?

As of November 13, 2026, the FDA guidance page lists it as Final. Its cited standard versions are older, so use the versions recognized in FDA's database at the time of submission, and follow the QMSR, in effect since February 2, 2026, for quality system matters.

For active home-use devices entering the US, the FDA home use guidance serves as a design input checklist, test reports to recognized standards provide part of the evidence, and product code pages and 510(k) Summaries show how similar devices were handled. This article is based on official public information available as of November 13, 2026. Please refer to the latest documents published by the regulatory authorities.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.