In Japan, an oxygen concentrator that falls under the certification standard for 酸素濃縮装置 (oxygen concentrators) is certified by a Registered Certification Body, and the standard cited for it is JIS T 7209. As of 13 November 2026, the PMDA standards database shows JIS T 7209:2018 as the latest edition. JIS T 7209:2018 is a modified adoption (MOD) of ISO 80601-2-69:2014, and ISO has since reached its third edition (2026), so an ISO test report is not automatically evidence of JIS conformity; the differences need a clause-by-clause assessment. This article covers classification, the certification standard, how the JIS and ISO editions line up, and how to work from an ISO report.

Key points

  • The Japanese generic name is 酸素濃縮装置. The PMDA standards database lists it as Class II and a specially controlled maintenance device (特定保守管理医療機器). Its certification standard is item 395 of Appended Table 3 of MHLW Notification No. 112 of 2005, and conforming products are certified by a Registered Certification Body.
  • Item 395 cites "T7209" with no publication year; the PMDA database shows JIS T 7209:2018 as the latest edition. MHLW Safety Information No. 424 states that oxygen concentrators manufactured and marketed from 1 February 2021 conform to JIS T 7209:2018.
  • JIS T 7209:2018 corresponds to ISO 80601-2-69:2014 (first edition), degree of correspondence MOD. The ISO second edition (2020) and third edition (published 10 April 2026) came later, and a new ISO edition does not by itself change the cited JIS.
  • PMDA training material states that the scope, specified values, shape and structure, test methods and test-specimen preparation in the cited standard must all be met. Its example: a concentrator with a 40% oxygen setting does not meet clause 201.12.4.102 of JIS T 7209:2018 and so does not fall under the certification standard.
  • The IAS scope of SUNGO's Hefei laboratory (TL-1119) lists ISO 80601-2-69 as a whole standard; JIS T 7209 is not in it. An ISO report is only a starting point, the JIS differences need separate assessment, and acceptance is for the Registered Certification Body to decide.

Oxygen concentrator classification in Japan: certification or approval?

Japan has three market routes (notification, certification and approval); this article covers only the certification route used for oxygen concentrators. Under Article 1(2) of MHLW Notification No. 112, the designated controlled medical devices are those listed in Appended Table 2 or 3 that meet the standard in that table and the Essential Principles standard; the notification does not apply where shape, structure and principle, method of use or performance clearly differ from existing controlled devices. An MHLW notice explains that each such device must be certified, item by item, by a certification body registered with the Minister.

The entry for oxygen concentrators is summarised below (as of 13 November 2026, based on MHLW Notification No. 112 and the PMDA standards database entry for 酸素濃縮装置):

ItemContent
Generic name酸素濃縮装置 (former name: 酸素濃縮式供給装置)
ClassClass II; specially controlled maintenance device: yes; QMS: applicable
Generic name code12873002
Certification standardNotification No. 112, Appended Table 3, item 395
Standard cited by the notificationJIS T 7209 (no year in the notification; latest edition shown in the database: JIS T 7209:2018)
Intended use or effectSeparating nitrogen or oxygen from ambient air to produce air with a high partial pressure of oxygen and supplying it to the patient
Definition, key pointsA device that separates nitrogen from room air using adsorption columns or a membrane with a large surface area; devices for medical gas pipeline systems (JIS T 7101) are excluded; where connected to a ventilator or similar device, only physical connection is covered, with no control or monitoring of either device

The entry's "cited standards" column also lists JIS T 0601-1, JIS T 0601-1-2 and JIS T 14971 (latest editions shown: 2023, 2023 and 2020), so general electrical safety, EMC and risk management should also be checked against these JIS editions.

JIS T 7209:2018 vs ISO 80601-2-69: how are they related?

The Japanese title of JIS T 7209:2018 is 医用電気機器-酸素濃縮装置の基礎安全及び基本性能に関する個別要求事項 (medical electrical equipment, particular requirements for the basic safety and essential performance of oxygen concentrator equipment). It is a particular standard in the JIS T 0601-1 family and is read together with the general standard JIS T 0601-1 and the relevant collateral standards.

The public-comment draft of this edition published on e-Gov states that the standard is based on the first edition of ISO 80601-2-69 issued in 2014, with technical content changed to take account of conditions in Japan. The changed passages are marked with dotted underlines in the text and listed with explanations in Annex JA. The corresponding international standard is ISO 80601-2-69:2014, the degree of correspondence is MOD (modified), and JIS T 7209:2007 is replaced. The official JIS text published by JISC prevails over the draft.

So an ISO 80601-2-69 report shows that a sample meets the ISO clauses; wherever the JIS deletes, adds or changes content, Japanese certification follows the JIS clause. A report issued against the 2020 or 2026 edition also carries the differences between ISO editions.

ISO 80601-2-69 2020 and 2026 editions: which one does Japan use?

StandardEditionPublication dateStatus as of 13 November 2026
ISO 80601-2-691st edition2014International standard corresponding to JIS T 7209:2018; 2nd and 3rd editions have followed
ISO 80601-2-692nd edition6 November 2020Withdrawn on 10 April 2026
ISO 80601-2-693rd edition10 April 2026Latest ISO edition
JIS T 72092018 edition2018Latest edition of the cited JIS shown in the PMDA standards database
JIS T 72092007 edition2007Replaced by the 2018 edition

The Japanese certification standard cites a JIS, not an ISO standard, so a new ISO edition does not by itself change it. Requirements change only after JIS T 7209 is revised and MHLW sets out how the certification standard applies. As of 13 November 2026, we did not find any announcement of a JIS T 7209 revision based on ISO 80601-2-69:2020 or 2026; check JISC and MHLW announcements for later developments.

How is conformity to JIS T 7209 judged under the certification standard?

PMDA's published training material on judging whether a product falls under a certification standard (certification-standard judgement points: conformity to the cited standard) sets out these points:

  • Designated controlled medical devices in Appended Table 3 must in principle show conformity to the JIS or IEC standard named in the notification; a product that does not conform does not fall under the certification standard.
  • The scope, specified values, shape and structure, test methods and test-specimen preparation in the cited standard must all be met.
  • Where the cited standard specifies a test method and the evaluation did not use it, the product does not fall under the certification standard.
  • Example: an oxygen concentrator with a 40% oxygen setting does not meet clause 201.12.4.102 of JIS T 7209:2018, cited by the oxygen concentrator certification standard, and therefore does not fall under it.

Clause 201.12.4.102 is the low oxygen concentration alarm clause. According to the public-comment draft on e-Gov, the oxygen concentrator must raise a low oxygen concentration alarm before the oxygen concentration of the product gas falls below 82% by volume; the official clause in JIS T 7209:2018 prevails. Specified values are therefore not free design choices; check design outputs against the JIS values first.

What happens when JIS T 7209 is revised?

When a JIS cited by the notification is revised, MHLW issues a notice on handling the revision that sets a transition period. One example is notice Yakushoku-ki Hatsu 0120 No. 1 of 20 January 2014, for the revision of JIS C 1010-1: the transition ended on 19 January 2019, after which products not conforming to the revised JIS could not be newly manufactured and marketed. During the transition, certification could follow the previous rules, but conformity to the revised JIS had to be confirmed before the end date, and products certified under the old edition and marketed before that date could still be sold afterwards.

For the change from the 2007 to the 2018 edition of JIS T 7209, MHLW's Pharmaceuticals and Medical Devices Safety Information No. 424 (November 2025) states that oxygen concentrators manufactured and marketed from 1 February 2021 conform to JIS T 7209:2018. It notes that the 2018 edition adds fire-prevention requirements, under which the oxygen outlet connector must have a means of stopping a flame from passing through it into the device. Check the original MHLW notice for the exact transition arrangements.

JIS T 7209 itself cites other JIS, ISO and IEC standards (second-level cited standards, 孫引き規格). The PMDA material quotes Q&A 1 of notice Yakushoku-ki Hatsu 0930 No. 1 of 30 September 2011: a cited JIS with a publication year applies in that edition; without a year, the latest edition applies in principle; if the latest edition creates inconsistent requirements, the earlier edition may be used with a stated justification. Conformity to standards other than these second-level cited standards is not accepted as meeting the cited JIS.

Using an ISO 80601-2-69 report for Japanese certification: how to do the gap assessment

  1. Confirm that the product falls within the definition and intended use of 酸素濃縮装置, for example whether it connects to a ventilator and, if so, whether control or monitoring is involved.
  2. Obtain the text of JIS T 7209:2018 (the PMDA database entry links to JISC for viewing and to the Japanese Standards Association) and use the comparison table in Annex JA as the checklist.
  3. Establish which ISO edition the report was issued against. JIS 2018 corresponds to the 2014 ISO edition; if the report was issued against the 2020 or 2026 edition, list the differences between ISO editions first, then compare with the JIS.
  4. Sort the clauses into three groups: those where the ISO report's test methods and results map directly to the JIS; those that need retesting by the JIS method or added explanation; and JIS-specific requirements the ISO report does not cover.
  5. Check specified values (such as the low oxygen concentration alarm and the oxygen setting), test methods, fire-prevention requirements, and marking and accompanying documents item by item; for the general parts, check the editions of JIS T 0601-1, JIS T 0601-1-2 and the IEC standards used.
  6. Before submitting the gap assessment, ask the Registered Certification Body you plan to use whether it will accept the material. Do not assume the outcome.

Common misreadings

  • Misreading 1: an ISO 80601-2-69 report equals conformity to JIS T 7209. The JIS is a modified adoption with the changes listed in Annex JA, and all JIS requirements, including test methods, must be met.
  • Misreading 2: the JIS T 7209:9999 (draft) file on e-Gov is a new revision. It states that it replaces JIS T 7209:2007, uses JIS T 0601-1:2017 as the general standard and corresponds to ISO 80601-2-69:2014. It is the public-comment draft of the 2018 edition, not a later revision.
  • Misreading 3: Japanese certification switches as soon as ISO publishes the 2026 edition. Japan follows the cited JIS and MHLW's transition arrangements.
  • Misreading 4: conformity to another standard can be used instead. The PMDA material states that conformity to standards other than the second-level cited standards does not meet the cited JIS requirements.
  • Misreading 5: the oxygen setting is the manufacturer's own choice. The PMDA 40% example shows that a product whose specified values miss the JIS does not fall under the certification standard.

How to check this in the database

This site gives starting points for Japanese regulations and related standards; the JIS text and the original notification come from Japanese official sources. Page details below are as of 13 November 2026:

  • The Japan medical device regulations page names PMDA / 厚生労働省 as the authorities, shows an update date of 2026-09-18 and lists 11 regulation and guidance entries, including the PMD Act (Act No. 145 of 1960), its enforcement regulation and the QMS ministerial ordinance (MHLW Ordinance No. 169 of 2004). Certification-standard Notification No. 112 is not listed separately; read the original in the MHLW legal database.
  • The ISO 80601-2-69 standard record page shows the title, edition year 2020 and the FDA recognised consensus standard number, without the standard text. ISO published the third edition on 10 April 2026, so check the edition when citing. The standards index groups 4,792 standard pages by publishing body, including 530 ISO and 141 IEC entries; JIS is not a separate group, so obtain JIS T 7209 from JISC or the Japanese Standards Association.
  • The site's registration data sources (see the data sources page) do not include PMDA certification or approval listings, so Japanese certifications cannot be looked up here; the only Japan-related sources are company identity records.
  • For the US route, search product code CAW in the regulation search: Generator, Oxygen, Portable, Class II, 21 CFR 868.5440, 510(k) required, with 22 clearances in the three-year data window and 288 in FDA's all-time total.

Test evidence and SUNGO

  • Hefei SUNGO Medical Technology Co., Ltd. holds IAS accreditation (certificate TL-1119, effective 4 June 2026). The scope lists ISO 80601-2-69, EN ISO 80601-2-69 and YY 9706.269 as whole standards, together with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and related whole standards. None of these entries carries an edition year.
  • Medical electrical safety and EMC are within the Hefei IAS scope only; they are not covered by CNAS or CMA accreditation. The Shanghai laboratory does not perform this testing.
  • JIS T 7209, the JIS T 0601-1 series and the Japanese modifications are not in SUNGO's accreditation scope. SUNGO can issue an IAS-accredited report for ISO 80601-2-69; the JIS differences need a separate assessment, and whether and how far the report can be used for Japanese certification is for the Registered Certification Body to decide.

For an introduction to the laboratory, see the SUNGO Lab page.

Frequently asked questions

What certification does an oxygen concentrator need for Japan?

One that meets the 酸素濃縮装置 certification standard (Notification No. 112, Appended Table 3, item 395) is a designated controlled medical device, certified item by item by a registered certification body. A product outside that standard cannot use this route, and its market route must be confirmed separately.

Can an ISO 80601-2-69 report be used directly for Japanese certification?

Not directly as evidence of conformity to JIS T 7209, which is a modified adoption of ISO 80601-2-69:2014; all JIS requirements must be met. The ISO report can be the basis of a gap assessment agreed in advance with the certification body.

What is the latest edition of JIS T 7209?

As of 13 November 2026, the PMDA standards database shows the 2018 edition, which replaced the 2007 edition.

After ISO 80601-2-69:2026, does Japanese certification move to the new edition?

The certification standard cites JIS T 7209, so an ISO revision does not by itself change requirements. Any change depends on a JIS revision and MHLW's transition arrangements.

Can an oxygen concentrator with a 40% oxygen setting be certified?

According to the PMDA example, no: it does not meet clause 201.12.4.102 of JIS T 7209:2018 and does not fall under the certification standard.

This article is based on official public information as of 13 November 2026. The latest documents from the regulators prevail.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.