For CE marking of a powered wheelchair charger, the first step is not picking standards but settling the charger's regulatory status in the EU: a component supplied with the wheelchair, an accessory under the Medical Device Regulation (MDR), a replacement part, or general electrical equipment with no medical intended purpose. The manufacturer's intended purpose decides which one applies. The first three are handled within the MDR framework; only a general-purpose charger with no medical intended purpose needs a separate assessment under the Low Voltage Directive (LVD) 2014/35/EU and other directives. In the US there is no separate product code for wheelchair chargers, and the charger is normally described in the 510(k) for the powered wheelchair (product code ITI).

Key points

  • Accessory status follows intended purpose: under Article 2(2) MDR, an accessory is intended by its manufacturer for use with one or several particular medical devices, specifically to enable their intended use. Only a charger designed for particular powered wheelchair models can meet that definition. Accessories are "devices" under the MDR (Article 1(4)) and are classified in their own right (Annex VIII, Section 3.2), not in the wheelchair's class by default.
  • Supplied with the wheelchair: when the charger is placed on the market together with the wheelchair, the wheelchair manufacturer normally covers it in the wheelchair's MDR technical documentation. Annex II, Section 1.1 MDR requires the technical documentation to describe the device's parts and the accessories intended to be used with it.
  • Where the LVD stops: the LVD applies to electrical equipment rated between 50 and 1,000 V AC or 75 and 1,500 V DC, but Annex II lists "electrical equipment for radiology and medical purposes" as outside its scope. Whether a given charger serves a medical purpose depends on the manufacturer's intended purpose and the legal text.
  • United States: 21 CFR 890.3860 covers only the powered wheelchair (Class II, product code ITI, 510(k) required). FDA has no separate code for wheelchair chargers. Of the 70 ITI 510(k) summaries archived on this site, 65 mention the charger in the body text (standard titles excluded).
  • Test basis: the FDA ITI page lists ISO 7176-25:2022 (lead-acid batteries and chargers), ISO 7176-31:2023 (lithium-ion battery systems and chargers) and ISO 7176-21 (EMC, including battery chargers) as recognized consensus standards. In the EU, the title of EN 12184 itself covers "their chargers".

The four possible EU statuses of a powered wheelchair charger

This article covers off-board and on-board chargers for powered wheelchairs and scooters. The same charger can have a different status depending on how it is supplied and what its manufacturer says it is for.

SituationLegal basisWhere the conformity evidence sitsWhat to watch
Supplied with the powered wheelchair as a componentMDR Annex II, Section 1.1 (documentation describes parts and accessories)The wheelchair's technical documentationCharger safety and EMC evidence is prepared with the wheelchair; explain the limits of any supplier component report
Charger supplied separately for particular powered wheelchairs (accessory)MDR Article 2(2), Article 1(4), Annex VIII Section 3.2The charger manufacturer's own MDR technical documentation, with separate classificationThe intended purpose names the particular devices; labels and instructions match
Replacement for the original chargerMDR Article 23The supplier keeps supporting evidence for the competent authoritiesA replacement that significantly changes performance, safety characteristics or intended purpose is a device and must meet the MDR
General-purpose charger with no medical intended purposeLVD 2014/35/EU, plus any other directives that may applyLVD Annex III technical documentation (Module A) and an EU declaration of conformityOnce it is claimed to be dedicated to a wheelchair model, reassess under Article 2(2) MDR

The table is a thinking aid, not a legal conclusion for any company. Which situation applies depends on the intended purpose stated in the technical documentation, labels and instructions; keep the reasoning in writing.

Is a wheelchair charger an MDR accessory?

Article 2(2) of the MDR (Regulation (EU) 2017/745) defines an "accessory for a medical device" as an article which, whilst not being itself a medical device, is intended by its manufacturer to be used together with one or several particular medical devices to specifically enable them to be used in accordance with their intended purpose, or to specifically and directly assist their medical functionality in terms of that intended purpose. See the official MDR text on EUR-Lex.

Three words decide the outcome: "particular", "specifically" and "directly". A universal charger does not point to a particular device and usually fails the "particular" element. A charger designed for the battery packs of specific powered wheelchairs, with compatible models named in its instructions, comes closer to the definition.

Once accessory status is confirmed, three further points follow:

  • Article 1(4) calls medical devices, accessories and Annex XVI products collectively "devices", so an accessory must meet MDR requirements too.
  • Annex VIII, Section 3.2 states that accessories are classified in their own right, separately from the device with which they are used. Electric wheelchairs themselves are listed as a Class I example under Rule 13 in MDCG 2021-24 Rev.1 (the April 2026 classification guidance), but the accessory's class has to be worked out under the rules separately rather than copied over.
  • Section 3.2 of MDCG 2021-24 Rev.1 advises manufacturers to decide, before applying the classification rules, whether the product is a medical device, an accessory, a replacement part or component under Article 23(2), or an Annex XVI product.

Under the MDR, electrical safety and EMC are covered by Annex I, Section 18: Sections 18.5 and 18.6 address electromagnetic interference and immunity, and Section 18.7 requires the risk of accidental electric shock to be avoided as far as possible in normal use and single fault conditions. For Class I devices, Article 52(7) has the manufacturer issue the EU declaration of conformity after drawing up the Annex II and III technical documentation; a notified body is involved only for the relevant aspects of sterile devices, devices with a measuring function or reusable surgical instruments.

Does the Low Voltage Directive apply to a powered wheelchair charger?

Article 1 of Directive 2014/35/EU says the directive applies to electrical equipment designed for use with a voltage rating of between 50 and 1,000 V AC and between 75 and 1,500 V DC, other than the equipment and phenomena listed in Annex II. Annex II includes "electrical equipment for radiology and medical purposes". See the official LVD text on EUR-Lex.

Most powered wheelchair chargers take mains input inside the LVD voltage range, so the real question is whether the charger serves a medical purpose. If it is regulated under the MDR as a component or accessory of the wheelchair, its electrical safety requirements already sit in MDR Annex I. If the manufacturer positions it as a general-purpose charger with no medical claim, the LVD route applies: under Article 6 the manufacturer draws up the Annex III technical documentation, carries out the Annex III procedure (Module A, internal production control), issues the EU declaration of conformity and affixes the CE marking. Whether the EMC Directive 2014/30/EU or other directives also apply should be checked against their texts.

FDA: does a powered wheelchair charger need its own 510(k)?

21 CFR 890.3860 identifies a powered wheelchair as a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position, and classifies it as Class II (performance standards). In the FDA classification database the matching product code is ITI: the pathway is 510(k), the review panel is Physical Medicine, and it is eligible for third-party review. See the FDA classification page for ITI.

As of 13 November 2026, a device-name search for "charger" in the FDA classification data (openFDA) returns only two product codes, both related to cardiac pacemakers or implantable pulse generators; there is no code for wheelchair chargers, and ITI is the only code under 21 CFR 890.3860. FDA's December 2017 guidance "Medical Device Accessories – Describing Accessories and Classification Pathways" explains that FDA has traditionally brought accessories into the parent device's classification through the parent's 510(k), and that accessories can also be classified separately on request under Section 513(f)(6) of the Federal Food, Drug, and Cosmetic Act. How a separately sold replacement or general-purpose charger is treated in the US is for FDA to decide.

Public 510(k) summaries show how chargers are handled in practice. The ITI page on this site archives 70 summaries (decision dates from September 2023 to August 2026). We searched the text of each one:

Search itemSummaries (out of 70)
Charger mentioned in the body text (standard titles excluded)65
Charger described as off-board60
ISO 7176-21 (EMC, including battery chargers)65
ISO 7176-25 (lead-acid batteries and chargers)54
ISO 7176-31 (lithium-ion battery systems and chargers)5
IEC 60601-1-2 (EMC of medical electrical equipment)53
IEC 62133 (battery safety)19

This is a text-search summary only: scanned pages or abbreviated references may be missed, and citing a standard does not mean every test in it was run on the charger. It does show that, in powered wheelchair 510(k)s cleared between September 2023 and August 2026, the charger is routinely written into the device description and test list of the whole wheelchair rather than filed separately. The FDA ITI page also lists ISO 7176-25:2022, ISO 7176-31:2023, ISO 7176-21 third edition (2025-04) and ANSI/RESNA WC-2:2019 Section 25 (batteries and chargers) as recognized consensus standards.

Common mistakes

  • Assuming the accessory inherits the wheelchair's class: Annex VIII, Section 3.2 MDR requires separate classification and separate reasoning.
  • Inconsistent status: calling the charger a wheelchair accessory in the technical documentation while issuing only an LVD declaration, or the reverse, is hard to defend in an audit. Decide the status first, then keep documents consistent.
  • Treating replacement chargers as unregulated: Article 23 MDR sets obligations for replacement parts, and a replacement that significantly changes performance or safety characteristics is regulated as a device.
  • Relying only on the supplier's component report: it covers the bought-in charger itself, not verification in the assembled wheelchair, and its limits should be stated in the technical documentation.

How to search this in the database

This site can help you map US registration data and EU guidance before you plan tests. Figures below were read from the pages on 13 November 2026:

  1. Open the ITI product code page to see the device name, device class, 21 CFR reference, medical specialty, marketing pathway, third-party review flag and the FDA classification definition. As of 13 November 2026, it shows 98 510(k) clearances in its three-year data window, 70 summaries archived for direct download, 586 clearances in all years according to FDA, and a most recent decision dated 21 August 2026. Download a few summaries and search for "charger" or the ISO 7176 part numbers to see how similar products describe the charger.
  2. Use the FDA product code index to check neighbouring device types. As of 13 November 2026, it lists 7,091 product codes grouped by FDA medical specialty, 220 of them under Physical Medicine. Motorized three-wheeled scooters are not under ITI but under product code INI (21 CFR 890.3800), in the same group.
  3. Open the EU medical device regulation page. As of 13 November 2026, it lists 172 regulations and guidance documents, including MDCG 2021-24 Rev.1 and the borderline and classification manual, a starting point for accessory and classification questions.

This site provides records, data summaries and links to official sources; current legal and standard texts are those published by the issuing bodies.

Test evidence and SUNGO

Whatever the charger's status, the charger clauses in wheelchair standards are a common source of evidence. The table shows how the three certificates of Hefei SUNGO Medical Technology Co., Ltd. cover them; the certificates of Shanghai SUNGO Medical Technology Co., Ltd. do not include these standards.

StandardHefei CNAS (L24279)Hefei IAS (TL-1119)Hefei CMA (260015342945)
EN 12184:2022 clause 12.3 (battery chargers) and 12.4 (charging connectors)Covered, including clauses 12.3 and 12.4Covered; the certificate excludes clause 6.6 (biocompatibility and toxicity) and clause 12.3.1Not covered
ISO 7176-25:2022 (lead-acid batteries and chargers)7 items: 5.2.3, 5.2.4, 5.2.5, 5.2.6, 5.3, 5.5, 6.3Whole standardNot covered
ISO 7176-31:2023 (lithium-ion battery systems and chargers)8 items: 4.2.2, 4.2.4, 5.2.1.3, 5.2.1.5, 5.2.1.6, 5.2.1.7, 6, 7Whole standardCovered, the same 8 items
GB/T 18029.25-2022 Chapter 5 (battery chargers)Covered; the certificate notes "EMC not tested"Whole standardNot covered
ISO 7176-21 (EMC, including battery chargers)Not coveredWhole standardNot covered

Three notes. First, charger EMC testing in Hefei is accredited only by IAS (TL-1119); the CNAS scope notes that GB/T 18029.25 Chapter 5 excludes EMC. Second, the IAS scope excludes EN 12184 clause 12.3.1, while the CNAS scope lists clause 12.3. Third, the IAS certificate states no edition years, and the FDA ITI page lists both the 2009 and 2025 editions of ISO 7176-21, so confirm the edition before sending samples. Accredited reports cover only clauses within the certificate scope.

IEC 60335-2-29, IEC 62368-1, IEC 61010-1 and IEC 62133 are outside SUNGO's accreditation scope, so SUNGO cannot provide all the testing a standalone charger needs for the LVD or CE marking. The certificates are the reference for scope; see the SUNGO laboratory page.

Frequently asked questions

Does a powered wheelchair charger need separate CE marking?

A charger supplied with the wheelchair is normally covered by the wheelchair's MDR technical documentation and CE marked with it. If sold separately, first decide whether it is an MDR accessory, a replacement part or general electrical equipment, then prepare the file under the MDR or the LVD.

Is a wheelchair charger an MDR accessory?

It depends on the manufacturer's intended purpose. Only a charger intended for particular medical devices, specifically to enable their intended use, meets the Article 2(2) definition; a universal charger usually fails the "particular" element.

Does the Low Voltage Directive apply to a wheelchair charger?

LVD Annex II excludes "electrical equipment for radiology and medical purposes". For a charger regulated under the MDR, electrical safety sits in MDR Annex I, Section 18; a general-purpose charger with no medical purpose is assessed under the LVD.

Does a wheelchair charger need its own FDA 510(k)?

FDA has no separate product code for wheelchair chargers. Of the 70 ITI 510(k) summaries we searched, 65 mention the charger in the wheelchair submission; how a separately sold charger is treated is for FDA to decide.

Which standard covers EMC for a wheelchair charger?

ISO 7176-21 names battery chargers in its title and is an FDA-recognized standard on the ITI page; 53 of the summaries we searched also cite IEC 60601-1-2. SUNGO Hefei holds ISO 7176-21 accreditation under IAS (TL-1119) only.

Settle the charger's status first; standards, report scope and documentation follow from it. Further reading: Powered Wheelchair Chargers: Electrical Safety and Test Prep, Do You Need a Notified Body for Wheelchair CE Marking? and EU MDR technical documentation for wheelchair exporters. This article is based on official public information as of 13 November 2026 and is subject to the latest documents from regulators.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.