Since 17 January 2026, the EN 60601-1 entry on the UK designated standards list for medical devices has been the combination EN 60601-1:2006, AC:2010, A1:2013 plus A13:2024; the older combination without A13 was removed on the same day. This combination is not written the same way as the current IEC edition, IEC 60601-1:2005+AMD1:2012+AMD2:2020 (edition 3.2), and the collateral standards EN 60601-1-2, -1-8 and -1-11 on the UK list are also older than their current IEC editions. Below, the four standards are compared in one table, followed by how to use IEC-based reports for Great Britain and how to check the data on ssy.

Key points

  • In the UK consolidated list v2 (17 January 2026, published with notice 0129/26), the current EN 60601-1 combination is EN 60601-1:2006, AC:2010, A1:2013 and A13:2024. The removal date of the older combination and the start date of the new one are both 17 January 2026; the list gives no parallel transition period.
  • EN 60601-1-2:2015, EN 60601-1-8:2007+AC:2010+A11:2017 and EN 60601-1-11:2010 remain on the UK list and were not updated this time. The last two carry an informative note that they may not cover the requirements introduced by S.I. 2008/2936.
  • All four UK list entries are older than the current IEC consolidated versions. The widest gap is EN 60601-1-11: the list cites IEC 60601-1-11:2010, while the current IEC version is the 2015 edition with AMD1:2020.
  • In the EU, EN 60601-1:2006 with A13:2024 has been on the MDR harmonised standards list since 17 June 2026. As of 13 November 2026, EN 60601-1-2, -1-8 and -1-11 are not on the MDR harmonised list. The UK and EU lists need to be checked separately.
  • Designated standards are voluntary. A report written to the current IEC editions can be used as evidence, but it does not by itself give a presumption of conformity; an edition gap analysis is needed. EN ISO 80601-2-69 is not on UK list v2.

What UK designated standards are and which products they cover

The base legislation for medical devices in the UK is The Medical Devices Regulations 2002 (S.I. 2002/618, "UK MDR 2002"). Under regulation 3A, a designated standard is a technical specification adopted by CEN, CENELEC, BSI or an international standardising body, with which compliance is not compulsory, and which the Secretary of State designates by publishing a reference to it. When the reference is removed from publication, the standard is no longer a designated standard. Regulation 9(4) provides that a general medical device is treated as complying with an essential requirement if it conforms, as respects that requirement, to a relevant designated standard, unless there are reasonable grounds for suspecting that it does not comply.

The list is published by notice of the Department of Health and Social Care (DHSC). The UK government's designated standards page carries both the notice PDFs and a consolidated list in ODS format. Notice 0129/26 of 17 January 2026 states that the listed standards are designated in relation to England, Wales and Scotland, that is, Great Britain. Northern Ireland follows different rules and is outside the scope of this article. As of 13 November 2026, the page had last been updated on 17 January 2026.

Designated standards support the UKCA route under UK MDR 2002. The MHRA guidance Regulating medical devices in the UK (as of 13 November 2026, last updated 20 February 2026) states that where third-party conformity assessment is required for UKCA marking, a UK approved body is needed; that general medical devices compliant with the EU MDR and CE marked can be placed on the Great Britain market up until 30 June 2030; and that the government has announced it will consult on indefinite recognition of CE marked medical devices. The CE route refers to EU harmonised standards; only the UKCA route refers to the UK designated standards discussed here. Many manufacturers prepare both routes, so both lists need checking.

This article is for exporters of active (powered) medical devices to Great Britain. It covers only the general standard EN 60601-1 and three collateral standards: -1-2 (electromagnetic disturbances), -1-8 (alarm systems) and -1-11 (home healthcare environment). Consumables, non-active devices and 60601-2-x particular standards are out of scope.

Which EN 60601-1 editions are on the UK designated standards list

The table follows UK consolidated list v2, checked on 13 November 2026. The cited IEC edition is the reference in brackets after each title on the list; current IEC versions come from the IEC webstore; EU status comes from Implementing Decision (EU) 2021/1182 and its amendments.

StandardCombination and dates on the UK listIEC edition cited in the listCurrent IEC consolidated versionEU MDR harmonisation status
EN 60601-1 (older combination)EN 60601-1:2006, AC:2010, A1:2013; from 1 January 2021 (notice 0034/21), removed 17 January 2026 (notice 0129/26)IEC 60601-1:2005IEC 60601-1:2005+AMD1:2012+AMD2:2020 (edition 3.2)Not listed
EN 60601-1 (current combination)EN 60601-1:2006, AC:2010, A1:2013, A13:2024; from 17 January 2026 (notice 0129/26)Not stated in the listAs aboveEN 60601-1:2006 with A13:2024 listed from 17 June 2026
EN 60601-1-2EN 60601-1-2:2015; from 1 January 2021 (notice 0034/21), no noteIEC 60601-1-2:2014IEC 60601-1-2:2014+AMD1:2020 (edition 4.1)Not listed as of 13 November 2026
EN 60601-1-8EN 60601-1-8:2007, AC:2010, A11:2017; from 1 January 2021 (notice 0034/21), with noteIEC 60601-1-8:2006IEC 60601-1-8:2006+AMD1:2012+AMD2:2020 (edition 2.2)Not listed as of 13 November 2026
EN 60601-1-11EN 60601-1-11:2010; from 1 January 2021 (notice 0034/21), with noteIEC 60601-1-11:2010IEC 60601-1-11:2015+AMD1:2020 (edition 2.1)Not listed as of 13 November 2026

"With note" refers to the informative note on the list: the standard does not necessarily cover the requirements introduced into UK MDR 2002 by S.I. 2008/2936, and manufacturers are advised to check whether all relevant essential requirements are appropriately covered. The two EN 60601-1 combinations and EN 60601-1-2:2015 carry no such note.

On the IEC side, as of 13 November 2026 the consolidated general standard on sale in the IEC webstore is still edition 3.2 (published 20 August 2020), and the IEC 60601-1 series pack published on 10 July 2026 contains the same four consolidated versions shown in the table. The UK list names the amendments AC:2010, A1:2013 and A13:2024 and does not list a separate amendment corresponding to IEC AMD2:2020. How the two map clause by clause has to be checked against the text of the EN amendments.

EN 60601-1 with A13:2024: from when, and is the EU aligned

Notice 0129/26 replaced the whole EN 60601-1 row of the 2021 list with the combination that includes A13:2024. In the consolidated list, the removal column for the older combination is headed "end of presumption of conformity" and shows 17 January 2026; the start of presumption of conformity for the new combination is also 17 January 2026. In other words, from 17 January 2026, a presumption of conformity on the UKCA route has to be claimed against the combination that includes A13:2024.

On the EU side, Commission Implementing Decision (EU) 2026/1231 was adopted on 11 June 2026, published in the Official Journal on 17 June 2026 and entered into force on that day. It added EN 60601-1:2006 with EN 60601-1:2006/A13:2024 as entry 65 of the MDR harmonised standards list. The recitals of the decision state that A13:2024 resulted from the revision of EN 60601-1:2006 by CEN and CENELEC under the standardisation request for the MDR. The UK list also writes AC:2010 and A1:2013 into the combination, while the EU entry lists only EN 60601-1:2006 and A13:2024. The two lists use different wording, so cite each exactly as published.

DateEventSource
1 January 2021Older EN 60601-1 combination and -1-2, -1-8, -1-11 start giving presumption of conformityNotice 0034/21
17 January 2026EN 60601-1 changes to the combination with A13:2024; older combination removed the same dayNotice 0129/26, consolidated list v2
17 June 2026EN 60601-1:2006 with A13:2024 added to the EU MDR harmonised standards listImplementing Decision (EU) 2026/1231
30 June 2030End date for placing EU MDR-compliant, CE marked general medical devices on the Great Britain market (as stated in MHRA guidance as of 13 November 2026)MHRA guidance

UK technical documentation prepared against the older combination should have an assessment of A13:2024 added at its next update. The clauses changed by A13:2024 are set out in its own text; this article does not review them clause by clause.

Using a report written to the current IEC editions in Great Britain

A common situation: a manufacturer tested to IEC 60601-1 edition 3.2, IEC 60601-1-2 edition 4.1 and IEC 60601-1-11 edition 2.1 for other markets and reuses the reports for its UK technical documentation. Under regulation 9(4) of UK MDR 2002, presumption of conformity comes from a "relevant designated standard". Where the report's edition does not match the list, it cannot simply be described as conforming to the UK designated standard; a gap analysis is needed. In practice:

  1. Build a comparison table: list the standard numbers, years, amendments and clauses in the report and compare them with the combinations on UK list v2, marking matches and differences.
  2. Explain differences in both directions. For amendments in the current IEC version that the UK combination does not list separately (for example AMD2:2020 to IEC 60601-1, or the 2015 changes to IEC 60601-1-11), explain how they support the argument for the essential requirements. For amendments in the UK combination that the IEC version lacks (for example A13:2024), check whether further assessment or testing is needed.
  3. Handle entries with a note: for EN 60601-1-8 and EN 60601-1-11, check whether the essential requirements introduced by S.I. 2008/2936 are covered, and fill gaps with other evidence.
  4. Record it in the technical documentation and declaration of conformity: state the full edition of each standard applied, including year and amendments, not just "IEC 60601-1" or "EN 60601-1".
  5. Agree the edition in writing when commissioning tests: if you need the EN combination, write the full reference, such as EN 60601-1:2006+AC:2010+A1:2013+A13:2024, on the test request so the report matches it.

As for responsibilities, the manufacturer prepares the gap analysis in its technical documentation; for devices that need third-party conformity assessment, the UK approved body decides during its review whether the evidence is sufficient. This article does not guarantee that a report to any edition will be accepted in Great Britain.

Common mistakes

  • Treating A13:2024 as IEC AMD2:2020. The IEC consolidated version contains only AMD1:2012 and AMD2:2020 and has no A13; A13:2024 is a European amendment to EN 60601-1:2006.
  • Assuming the UK list and the EU harmonised list are the same. EN 60601-1-2, -1-8 and -1-11 are on the UK list but, as of 13 November 2026, not on the MDR harmonised list; EN 60601-1 includes A13:2024 on both, but the combinations are written differently.
  • Assuming the older combination still has a transition period. On the list, the removal date of the older combination and the start date of the new one are both 17 January 2026.
  • Assuming designated standards are mandatory. Regulation 3A itself says compliance is not compulsory; other evidence can show conformity with the essential requirements.
  • Checking only the standard number, not the year and amendments. A combination on the list must be compared as a whole, and a report without its edition cannot later be traced.

How to check in the database

ssy (the medical device TIC database) lets you run a first pass over regulation entries and standard records before confirming against the official texts:

  • UK regulations: the UK medical device regulations page lists regulations and guidance by type, number, title and status. As of 13 November 2026, it showed MHRA as the authority, data updated on 18 September 2026, 36 regulation and guidance entries and 271 update records. The entries include S.I. 2002/618 and its parts, and S.I. 2008/2936, the regulation named in the list's note; each entry opens to a short description and a link to the official text on legislation.gov.uk.
  • Standard search: on the testing search page, type "60601-1" to list the EN, IEC and ANSI/AAMI records together.
  • EN standards list: the EN standards list has a "Recognised by" column. As of 13 November 2026, the standards library held 4,792 standards, including 293 EN and 141 IEC records. On page 2 of the EN list, EN 60601-1 is marked "EU, UK", while EN 60601-1-2, -1-8 and -1-11 are marked "UK" only, matching the findings above.

The site provides bibliographic records and official links only; standard texts are copyright-protected and not available for download. The amendments in each combination should still be confirmed against the UK list and the EU decision. For UK registration volumes, see Chinese device manufacturers in the United Kingdom.

Test evidence and SUNGO

At SUNGO, medical electrical safety and EMC testing is carried out only by Hefei SUNGO Medical Technology Co., Ltd., under its IAS accreditation (certificate TL-1119). The accredited scope lists whole standards without year designation: IEC 60601-1 and EN 60601-1, IEC/EN 60601-1-2, IEC/EN 60601-1-8 and IEC/EN 60601-1-11, together with the corresponding Chinese standards GB 9706.1, YY 9706.102, YY 9706.108 and YY 9706.111. The only particular standards in scope are ISO 80601-2-69 and EN ISO 80601-2-69 (oxygen concentrators); other 60601-2-x and 80601-2-x particular standards are outside SUNGO's scope. None of these items is within the CNAS accreditation or CMA designation of either site, and the Shanghai laboratory does not perform them.

The edition to be tested and reported, for example the EN combination on the UK list or the current IEC version, is agreed in writing when the test is commissioned, and the report states the full edition. Whether a report supports conformity in Great Britain is for the manufacturer to argue and the reviewer to decide. For an introduction to the SUNGO laboratories, see the testing laboratory page.

Frequently asked questions

Which edition of EN 60601-1 is on the UK designated standards list now?

From 17 January 2026, it is the combination EN 60601-1:2006, AC:2010, A1:2013 plus A13:2024 (notice 0129/26). The older combination without A13 was removed the same day and no longer gives a presumption of conformity.

Can a report to IEC 60601-1 edition 3.2 be used for UKCA marking?

It can be used as evidence, but it is not the designated edition on the list, so it does not by itself give a presumption of conformity. The edition differences need to be documented in the technical file and A13:2024 assessed separately; for devices that need third-party assessment, the UK approved body decides whether the evidence is sufficient.

Are UK designated standards mandatory?

No. Regulation 3A of UK MDR 2002 defines a designated standard as a technical specification with which compliance is not compulsory. A manufacturer that does not use one can show conformity with the essential requirements by other evidence.

Are the UK and EU 60601 standard lists the same?

No. As of 13 November 2026, EN 60601-1-2:2015, EN 60601-1-8 and EN 60601-1-11:2010 are on the UK list but not on the EU MDR harmonised standards list. EN 60601-1 includes A13:2024 on both lists, but the combinations are written differently.

Why is EN ISO 80601-2-69 not on the UK list?

Neither UK consolidated list v2 nor notice 0129/26 includes this standard. For oxygen concentrators tested to ISO 80601-2-69, the technical documentation needs to explain how the test results support the argument for the essential requirements.

The UK list is updated from time to time, so check its version and date on the UK government page before preparing technical documentation or shipping. This article is based on official public information available as of 13 November 2026. Always refer to the latest documents from the regulator.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.