China's draft Guideline for Registration Review of Artificial Intelligence Medical Devices (2026 Revision) was published for comment on 14 September 2026 by the Center for Medical Device Evaluation of the National Medical Products Administration (CMDE); the comment period closed on 12 October 2026, and as of 11 November 2026 CMDE had not issued a final text, so the 2022 guideline remains the version in force. Compared with 2022, the eight-chapter structure is unchanged; the main changes are four new sections in Chapter 5 (large models, multimodal products, multi-disease products and cross-modality image synthesis) and more detailed rules on algorithm updates and algorithm documentation. Below we compare the two texts chapter by chapter ("the draft" always means Attachment 1 as published by CMDE) and explain how the draft treats a device that calls a general-purpose large model, a medical AI agent built on such a model, and a large model developed in-house.
Key points
- It is still a draft. As of 11 November 2026, CMDE's notices page showed no final version of the revision. The version CMDE has issued is still the 2022 guideline.
- The eight-chapter structure is unchanged; the changes sit mainly in Chapter 5, "Technical considerations". Four new sections were added: cross-modality image synthesis, multimodal products, multi-disease products and large models.
- Large models and AI agents appear for the first time. Neither term occurs in the 2022 text. Under the draft, a call to a general-purpose large model is treated either as a non-medical-device function or as off-the-shelf software, depending on whether it is essential to the intended use. A medical AI agent can be regarded as medical application software. A self-developed foundation model can be regarded as medical middleware and a self-developed fine-tuned model as medical application software.
- The continuous learning section is word-for-word the same as in 2022. The algorithm update section adds a rule for performance optimisation where the core algorithm stays the same.
- Algorithm documentation is more detailed. A cross-subgroup analysis is now required, and the "if applicable" qualifier on the algorithm performance factor analysis report has been removed.
Background and scope
The guideline in force is the Guideline for Registration Review of Artificial Intelligence Medical Devices issued with CMDE Notice No. 8 of 2022, signed on 7 March 2022 and posted on 9 March 2022. The 2026 notice states that the revision follows NMPA's plan for drafting and revising registration review guidelines, and that comments were to be returned by email.
The scope is the same in both versions: registration of Class II and Class III standalone AI software and of medical devices containing AI software components, including in vitro diagnostic devices. It applies to self-developed software; off-the-shelf software components are handled by reference, and the external software environment is out of scope. An "AI medical device" is still defined as a device that uses AI technology on medical device data to achieve a medical purpose. Both versions state that the guideline is guidance for applicants, reviewers and inspectors, does not involve administrative approval matters and is not enforced as a regulation.
What changed from the 2022 guideline
Both versions have eight chapters: scope, main concepts, basic principles, life-cycle processes, technical considerations, algorithm documentation, supplementary notes on registration dossiers, and references. The table lists only the places that changed.
| Location | 2022 version | 2026 draft |
|---|---|---|
| Ch. 2, intended-use categories | Automatic measurement, automatic segmentation and 3D reconstruction listed as examples of non-decision-support functions | Automatic segmentation and measurement of lesions or target volumes (including organs at risk), image fusion and EEG signal decoding listed as decision-support examples; segmentation and measurement of organs and tissues that are not lesions or targets stay non-decision-support; "assisted assessment" and "assisted screening" added as decision-support types |
| Ch. 2, algorithm updates | Algorithm-driven and data-driven updates | Same rules kept, plus a new rule for performance optimisation with an unchanged core algorithm |
| Ch. 4, data collection | Raw, base, annotated and augmented "databases" | Renamed "datasets"; traceability written into data collection and dataset construction |
| Ch. 5, technical considerations | 16 sections, with separate sections for stress testing, adversarial testing and each of four learning methods, plus "human factors and usability" | Still 16 sections: the two test types merged; ensemble, transfer, reinforcement and federated learning merged; "human factors and usability" renamed "usability engineering"; four new sections |
| Ch. 5, third-party databases; Ch. 7, product technical requirements | Evaluation database assessment materials submitted under the device master file registration requirements (Ch. 5); the evaluation database's master file registration number given in the technical requirements appendix (Ch. 7) | Both references to master file registration no longer appear |
| Ch. 6, algorithm study report | Performance factor analysis report provided "if applicable" | Report required outright; cross-subgroup analysis added; the basis for training targets to draw on literature review and clinical expert discussion |
| Ch. 7, dossier notes | Initial registration, change registration and renewal | Only initial registration and change registration listed |
| Ch. 8, references | 54 entries | 90 entries |
How large-model medical software is registered: three scenarios
Section 13 of Chapter 5 describes large models as deep-learning models that usually use attention mechanisms and follow a "pre-training plus fine-tuning" approach, with general and emergent capabilities coexisting with risks such as hallucination and bias. The draft's view is that it is difficult to deploy a large model directly in a medical device, and it describes three workable approaches. All three must take account of relevant national laws and regulations and of data security requirements.
| Scenario | Characterisation | Dossier points |
|---|---|---|
| Calling a general-purpose large model through an interface, where the call is not essential to the intended use | Usually a non-medical-device function | Technical requirements briefly describe the interface function; the instructions for use give basic model information such as name and version and state that it is a non-medical-device function |
| Calling a general-purpose large model, where the call is essential | Can be regarded as an off-the-shelf software component | State the model's name, model/specification, version, operating environment and interface specification, and provide study documentation; technical requirements and instructions for use give model information, interface specification and performance requirements |
| Developing a medical AI agent on a general-purpose large model | The agent can in principle be regarded as medical application software; the large model can be regarded as an off-the-shelf software component | A standalone software product consists of the agent; the large model follows the off-the-shelf component requirements above |
| Developing a model in-house using the large-model approach | The self-developed foundation (pre-trained) model can be regarded as medical middleware; the self-developed scenario (fine-tuned) model can be regarded as medical application software | Where the two are version-controlled separately, technical requirements state the release version and version naming rules of each |
Common requirements for all three: risk management documentation for hallucination, bias and similar risks, with warnings in the instructions for use; clinical evaluation according to the AI clinical evaluation guidelines; and, where applicable, study documentation on emergent capabilities and model distillation (emergent capabilities need not be addressed where the call is a non-medical-device function). In the application form, agents and self-developed models may be noted in the product structure by way of remarks.
Whether a medical large model or medical AI agent is regulated as a medical device at all is, according to the draft, decided under the relevant classification guidelines, with a formal classification request where necessary. NMPA issued the Guideline for Classification of AI-based Medical Software Products with NMPA Notice No. 47 of 2021, and the draft lists it among its references.
Algorithm updates and continuous learning: when change registration is needed
The update framework is the same as in 2022. An update that affects safety or effectiveness is a significant software update and requires change registration; otherwise it is a minor update controlled through the quality management system. Algorithm-driven updates are usually significant; for data-driven updates, the test is whether algorithm performance differs statistically from the previous registration. The draft adds the following rule:
| Performance optimisation with an unchanged core algorithm | Update type |
|---|---|
| More training data | Assessed under the data-driven update rules |
| Affects user decisions (decision capability, results, workflow or user actions) | Significant software update |
| Purely faster computation or lower resource use | Minor software update |
| A mix of the above | Highest risk applies: treated as significant |
For a data-driven update, the draft requires the update report to describe the change at least under algorithm risk management, data quality control, and algorithm verification and validation. The continuous/adaptive learning section is identical in both versions: the self-learning function should be switched off, or switched on but not put into use; an update produced by self-learning must be verified and validated, with change registration where necessary, and may be used only after approval.
Datasets and clinical evaluation
The data chapter changes mainly in terminology and traceability. In both the data collection and the annotation personnel requirements, the 2022 text asked applicants to specify qualification requirements for any overseas staff; the draft deletes this in both places. The requirement that training, tuning and test sets do not overlap, verified by duplicate checking, is unchanged. Algorithm validation adds a cross-subgroup analysis combining the disease dimension with dimensions such as target population and acquisition equipment; where sample sizes are too small or performance varies too much, use limitations must be stated and warned of in the instructions for use.
The basic clinical evaluation principles are the same in both versions: decision-support functions are compared with equivalent devices on the core algorithm, the clinical evidence for the equivalent device should in principle come from clinical trials (including retrospective studies), and new functions, algorithms or uses in principle require clinical trials. All four new sections refer clinical evaluation to the AI clinical evaluation guidelines. CMDE consulted on the following related drafts in 2026:
| Notice date | Draft for comment | Comments due |
|---|---|---|
| 17 June 2026 | Guideline for Registration Review of Clinical Evaluation of AI-Assisted Diagnosis Medical Devices | 10 July 2026 |
| 17 June 2026 | Guideline for Clinical Trials of Diabetic Retinopathy Assisted Diagnosis Software | 10 July 2026 |
| 16 September 2026 | Guideline for Registration Review of Clinical Evaluation of AI Multi-Disease Decision-Support Medical Devices | 14 October 2026 |
The English titles above are our translations; the official titles are in Chinese. The other three new sections in brief: a cross-modality synthetic image (for example a "CT-like image" generated from MR images) may not be used alone for diagnosis, and the user interface and instructions for use must say so. A multimodal product is one algorithm that takes two or more data modalities and outputs decision support for a single disease. A multi-disease product is one algorithm that supports decisions on two or more diseases. Simply bundling two or more single-modality, single-disease products does not count as either: same-disease bundles are not multimodal products, and different-disease bundles are not multi-disease products.
Other dossier changes
- Intended-use terms in product names extended to six: assisted triage, assessment, detection, diagnosis, screening and treatment, each with its typical setting.
- The same AI software function running on different computing platforms may use one algorithm study dossier if equivalence of performance across platforms can be shown.
- The user training plan now includes assessment of trainees.
Notes for manufacturers outside China
Both versions require imported AI medical devices to consider risks arising from differences between China and other countries, such as ethnicity, epidemiological characteristics and clinical practice norms. The draft's reference list includes the IMDRF Good Machine Learning Practice guiding principles (N88) and essential principles for predetermined change control plans (N90), several FDA documents on AI-enabled devices, and the EU's MDCG 2025-6 on the interplay between MDR/IVDR and the AI Act. The reference list shows only what the drafters consulted. For each market, prepare the dossier against that market's own regulations and guidance.
How to look this up in the database
First, what the database covers. As of 11 November 2026, the regulations database overview shows 282 regulations and guidance entries and 589 update records. China (NMPA) is marked "in progress", and neither version of the guideline discussed here is in the database yet, so read the original texts on the CMDE website. What the database does cover is overseas counterparts, FDA product codes and standard records:
- In regulation search, as of 11 November 2026, "artificial intelligence" returns 7 results (1 regulation, 2 FDA product codes, 4 510(k) records) and "machine learning" returns 11 FDA product codes. Each product code page shows the device class, the 21 CFR section, the marketing pathway and 510(k) clearances from the last three years.
- For EU software qualification, see MDCG 2019-11 rev.1, whose page shows the number, issuer, status and a link to the official text.
- In testing search, Chinese standards are listed under their Chinese titles. Entering 人工智能医疗器械 ("AI medical device") returned 11 standard records as of 11 November 2026: the five parts of YY/T 1833, GB/T 47941-1, and YY/T 1858, 1907, 1949, 1990 and 1991. The site gives records and official links only. For whether a standard is current, check the announcements of the standardisation authorities.
- Under browse by product category, the "Digital Medicine, Data & Artificial Intelligence" category listed 121 companies as of 11 November 2026. Categories come from exhibition classifications and count companies, not registration certificates. See Digital Medicine, Data & Artificial Intelligence: registration footprint of 121 companies.
FAQ
Is the 2026 revision of China's AI medical device guideline in force?
No. As of 11 November 2026 it is still a draft for comment, and the version CMDE has issued is still the one published with Notice No. 8 of 2022. The content and implementation arrangements of the revision are whatever CMDE formally issues.
Our product calls a general-purpose large model. Does the model go into the dossier?
First decide whether the call is essential to the intended use. If not, it is usually a non-medical-device function, identified as such in the instructions for use. If it is essential, the model can be regarded as an off-the-shelf software component, and model information, interface specification and study documentation are needed.
Is a medical AI agent a medical device in China?
Not necessarily. Whether it is regulated as a device is decided under the classification guidelines, with a classification request where necessary. If it is a device, the agent can in principle be regarded as medical application software.
Does the 2026 draft give large models their own chapter?
No. The draft still has eight chapters. Large models are Section 13 of Chapter 5, "Technical considerations".
For AI products already on the market, the parts to read first are the update rules, the more detailed algorithm documentation and the shifted decision-support boundary. For developers of large-model or agent products, the draft sets out a framework for characterising the three scenarios. Whether these provisions survive into the final text depends on what CMDE issues. The originals are CMDE's notice on the 2026 draft and Notice No. 8 of 2022. This article is based on official public information as of 11 November 2026; the latest documents from the regulator prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.