If you supply wheelchairs or mobility scooters to Great Britain and show conformity under the Medical Devices Regulations 2002 (UK MDR 2002), you can rely on UK designated standards. As of 13 November 2026, the UK designated standards for wheelchairs are still EN 12183:2009 (manual wheelchairs) and EN 12184:2009 (electrically powered wheelchairs, scooters and their chargers); notice 0129/26 of 17 January 2026 did not replace them with the 2022 editions. A test report to the 2022 editions can still be used as evidence, but it does not by itself give the presumption of conformity, so your technical file needs an edition gap analysis.
Key points
- In the Department of Health and Social Care notice 0129/26 of 17 January 2026 and consolidated list version 2, wheelchairs are covered only by EN 12183:2009 and EN 12184:2009, with presumption of conformity from 1 January 2021 (notice 0034/21), no removal date, and no 2022 entry.
- Designated standards are voluntary. Under regulation 9(4) of the UK MDR 2002, a device that conforms to a relevant designated standard for an essential requirement is treated as complying with that requirement, unless there are reasonable grounds for suspecting otherwise.
- Reports to EN 12183:2022 or EN 12184:2022 can still serve as evidence, but the list cites the 2009 editions, so you need a gap analysis between the two editions and, where needed, supplementary testing or a written justification.
- The list supports the UK MDR 2002 route. As of 13 November 2026, GOV.UK guidance states that CE marked general medical devices compliant with the EU MDR can be placed on the Great Britain market until 30 June 2030; on that route, evidence is organised against the EU MDR.
- The CNAS scope of the SUNGO laboratory in Hefei covers EN 12183:2022 and EN 12184:2022. The IAS scope does not state an edition year; the edition that applies is the one stated in the report.
What UK designated standards are and where they apply
The core UK legislation for medical devices is the Medical Devices Regulations 2002 (S.I. 2002/618, often called the UK MDR 2002). It applies in Great Britain, meaning England, Scotland and Wales. Since 26 May 2021, Northern Ireland has applied the EU MDR (Regulation (EU) 2017/745), so its rules differ from those in Great Britain. In this guide, "UK designated standards" refers only to the Great Britain market.
Regulation 3A defines a designated standard as a technical specification adopted by a recognised standardisation body (CEN, CENELEC or BSI) or an international standardising body, for repeated or continuous application, with which compliance is not compulsory, and which the Secretary of State designates by publishing a reference to it and maintaining that publication. Regulation 9(4) sets out the effect: a device is treated as complying with an essential requirement if it conforms to a relevant designated standard for that requirement, unless there are reasonable grounds for suspecting that it does not comply (see regulation 9 on legislation.gov.uk). A designated standard is therefore a shortcut to a presumption of conformity, not the only route. You can show that the essential requirements in Annex I are met in other ways, but the work and the burden of proof then sit with the manufacturer.
Notices of designated standards are issued by the Department of Health and Social Care (DHSC) and collected on the GOV.UK page Designated standards: medical devices, which states that it applies to England, Scotland and Wales. The same page offers a consolidated list (an ODS spreadsheet) for business reference. The spreadsheet states that it is only a summary and does not as such generate legal effects; the notices themselves are the formal basis.
This guide is for manufacturers and export staff supplying manual wheelchairs, powered wheelchairs, mobility scooters and their chargers to Great Britain. It covers conformity evidence and the technical file, not registration procedures, and it is not legal advice.
Which edition the UK list names for wheelchairs
As of 13 November 2026, the GOV.UK page carries two notices: 0034/21 of 1 January 2021 and 0129/26 of 17 January 2026. Part 2 of Annex I to notice 0129/26 sets out the complete list of designated standards as at the date of that notice, and items 107 and 108 are still EN 12183:2009 and EN 12184:2009. The two entries in consolidated list version 2 (17 January 2026) read as follows:
| Standard | Title in the list | Presumption of conformity from | Notice | Removal date |
|---|---|---|---|---|
| EN 12183:2009 | Manual wheelchairs - Requirements and test methods | 1 January 2021 | 0034/21 | None |
| EN 12184:2009 | Electrically powered wheelchairs, scooters and their chargers - Requirements and test methods | 1 January 2021 | 0034/21 | None |
The same notice replaced a number of standards with newer editions, for example EN 556-1:2024, EN ISO 14971:2019 (with A11:2021) and EN ISO 15223-1:2021, each with presumption of conformity from 17 January 2026. In the consolidated list, the editions they replaced also show a removal date of 17 January 2026. The list is being maintained; the two wheelchair entries simply have not changed.
The key dates for this topic are:
| Date | Document | What it means for wheelchairs |
|---|---|---|
| 1 January 2021 | Notice 0034/21 | Designated standards published, including EN 12183:2009 and EN 12184:2009, with presumption of conformity from that date |
| 17 January 2026 | Notice 0129/26 and consolidated list version 2 | A number of standards updated; the two wheelchair entries remain the 2009 editions |
| 30 June 2030 | GOV.UK guidance (text as of 13 November 2026) | End date for placing CE marked general medical devices compliant with the EU MDR on the Great Britain market |
CE marking or UK MDR 2002: which route the list applies to
The GOV.UK guidance Regulating medical devices in the UK (last updated 20 February 2026) states that CE marked general medical devices compliant with the EU MDR can be placed on the Great Britain market until 30 June 2030. It also notes that the government will consult on indefinite recognition of CE marked medical devices. The designated standards list plays a different role on each route:
| Route | Legal basis | Role of the UK designated standards list |
|---|---|---|
| Show conformity under the UK MDR 2002 and affix the UKCA marking | S.I. 2002/618 | Conformity with the listed edition gives a presumption of conformity under regulation 9(4); for wheelchairs, that is the 2009 edition |
| Place on the Great Britain market with a CE marking under the EU MDR | EU MDR and the UK transitional arrangements | Evidence is organised against the EU MDR; not covered further in this guide |
The same guidance states that manufacturers of Class I medical devices, other than those that are sterile or have a measuring function, can self-declare conformity against the UK MDR 2002. The class of a specific product follows the classification rules in the regulations. With a self-declaration, no third party reviews the technical file beforehand, so the manufacturer alone is responsible for whether the explanation linking a 2022 report to the 2009 designated standard is adequate.
Using a 2022 edition test report in Great Britain
The published references carry the year, so the designated standards are EN 12183:2009 and EN 12184:2009. Notice 0034/21 also states that a reference removed from publication is no longer designated and no longer gives rise to a presumption of conformity from its removal date, which shows that the presumption follows the specific edition. A report to the 2022 editions cannot rely directly on regulation 9(4), so the chain of evidence needs to be completed. Common practice is:
- State the edition. Use the full number and year, for example EN 12184:2022, consistently in the declaration of conformity (where it lists the standards applied), the technical file and the test report. Do not write "EN 12184" without the year, and do not describe a 2022 report as showing conformity with EN 12184:2009.
- Prepare an edition gap analysis. Take the 2009 clauses as the baseline, find the corresponding 2022 clause for each, and mark each as "same requirement", "requirement or method changed", "new in 2022" or "in 2009 but not in 2022".
- Map it to the essential requirements. Link the gap analysis to your essential requirements checklist for Annex I of the UK MDR 2002, showing which report and clause covers each applicable requirement.
- Test or justify. Where the 2009 edition has a requirement that the 2022 report does not cover, or uses a different test method, add testing or give a written justification supported by your risk management file.
- Watch for new notices. For the standards replaced in notice 0129/26, the old edition was removed on the day the new one was listed. If the wheelchair entries are updated in future, recheck the technical file against the new notice.
Where a third-party conformity assessment body is involved, or an importer or distributor has its own document requirements, follow their written requirements.
Common misunderstandings
- "2022 is newer, so the UK accepts it automatically." The list does not name the 2022 editions, so a newer edition does not by itself carry the presumption of conformity.
- "Designated standards are mandatory." Regulation 3A says compliance is not compulsory; other means are allowed, with the burden of proof on the manufacturer.
- "The consolidated list is the legal text." It is a summary that does not as such generate legal effects; the notices are the formal basis.
- "Anything sold in Great Britain must be checked against the UK list." On the EU MDR CE marking route, evidence is organised against the EU MDR.
- "Northern Ireland works the same way." Northern Ireland applies the EU MDR; the UK designated standards notices relate to England, Scotland and Wales.
How to compare the 2009 and 2022 editions of EN 12184
The differences must be checked against the standard texts; this guide does not restate clause content or limits. A practical order is:
- Obtain the official texts of both editions from BSI or another national member of CEN. Standard texts are copyrighted; our database provides bibliographic records and official links only.
- Compare scope, normative references, requirements, test methods and information supplied by the manufacturer, clause by clause. The normative references clause lists the other standards, and their editions, on which the tests rely; check these item by item between the two editions.
- Record the result as a gap table, keep it as an annex to the technical file, and keep the edition consistent across the test report, the declaration of conformity and the instructions for use.
How to check this in our database
Our database helps you find UK regulatory entry points, legislation numbers and standard records. It does not replace the official notices on GOV.UK.
- UK regulations page (/en/regulatory/uk/): as of 13 November 2026, the page shows 36 regulations and guidance entries and 271 update records, names the authority as MHRA, and is dated 2026-09-18. The list includes S.I. 2002/618 and entries for its parts, such as Part I and Part II, as well as S.I. 2023/627, The Medical Devices (Amendment) (Great Britain) Regulations 2023, under which GOV.UK sets out the timelines for CE marked devices.
- Regulation search (/en/regulatory/search/): the page states that it covers regulations and guidance, FDA product codes and recent 510(k) clearances. Searching "S.I. 2002/618" lists the regulations themselves and entries for their parts and schedules.
- Testing search (/en/testing/search/): searching "EN 12184" or "EN 12183" leads to the standard record, where the "UK designated" field shows 0034/21, the notice under which the standard is designated in the UK. Standard texts are not provided.
- UK registration data: for an overview of registered devices in the UK, see our article United Kingdom device registration in numbers.
Our data is updated with a time lag. For the current status, check the notice reference, presumption start date and any removal date on the GOV.UK page "Designated standards: medical devices".
Test evidence and the SUNGO laboratory
SUNGO's wheelchair testing is carried out by the laboratory of Hefei SUNGO Medical Technology Co., Ltd.; the Shanghai laboratory does not test to these two standards. The accredited scope relevant to this topic is shown below; the certificates prevail.
| Accreditation | Certificate | EN 12183 | EN 12184 |
|---|---|---|---|
| CNAS | L24279 | EN 12183:2022, manual wheelchairs, 32 items | EN 12184:2022, electrically powered wheelchairs, scooters and their chargers, 47 items |
| IAS | TL-1119 | No edition year stated; except clause 5.6 (biocompatibility and toxicity) | No edition year stated; except clause 6.6 (biocompatibility and toxicity) and clause 12.3.1 |
The CNAS accreditation covers the 2022 editions, not the 2009 editions. The IAS scope does not state an edition year, but that cannot be read as covering the 2009 editions; the edition that applies is the one stated in the report. Neither laboratory's CMA scope includes these two standards. CNAS and IAS are both signatories to the ILAC MRA. Biocompatibility clauses are outside the accredited scope above. Where a project needs them, they are subcontracted to a laboratory with the appropriate qualifications, with the client's prior agreement, and the report identifies the subcontracted items and the subcontractor. SUNGO does not promise that any UK regulator or other party will accept a particular report. For the wheelchair testing services, see the manual wheelchair testing and electric wheelchair testing pages.
Frequently asked questions
Which edition of EN 12184 is a UK designated standard?
As of 13 November 2026, it is EN 12184:2009, and EN 12183 is also the 2009 edition, with presumption of conformity from 1 January 2021 (notice 0034/21). Notice 0129/26 of 17 January 2026 did not replace them with the 2022 editions.
Can I use an EN 12184:2022 test report in Great Britain?
It can be used as conformity evidence, but it does not by itself give the presumption of conformity. Add a 2009 versus 2022 gap analysis to the technical file, show how the essential requirements are covered, and test further where needed.
Are UK designated standards mandatory?
No. Regulation 3A of the UK MDR 2002 defines a designated standard as a technical specification with which compliance is not compulsory. You may show compliance with the essential requirements in other ways, but the burden of proof is on the manufacturer.
If my wheelchair is CE marked, do I need to check the UK designated standards list?
As of 13 November 2026, GOV.UK states that CE marked general medical devices compliant with the EU MDR can be placed on the Great Britain market until 30 June 2030; on that route, evidence is organised against the EU MDR. If you plan to move to the UK MDR 2002 route, check the editions against the UK list at that point.
Where can I find the UK designated standards list for medical devices?
On the GOV.UK page "Designated standards: medical devices", where you can download the notices and the consolidated list. The consolidated list does not as such generate legal effects; the notices prevail.
For wheelchair exporters to Great Britain, the essentials are a consistent edition across the test report, the declaration of conformity and the technical file, and a clear record of the differences between the 2022 and 2009 editions. This guide is based on official public information as of 13 November 2026; the latest documents from the regulator prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.