To check whether a device is FDA 510(k) exempt, look up its product code, not its class. Confirm in FDA's classification database that the code's Submission Type is 510(K) Exempt, then read the 21 CFR regulation behind it, the .9 limitations of exemption and any exemption order in the Federal Register. Only a device that fits the exempt code's definition and stays within those limits can skip the 510(k). Below we walk through that chain, including a 2026 exemption order, the third-party review flag and our database lookup. Importers can use the same steps to test a supplier's exemption claim.

Key takeaways

  • Exemption is decided by product code, not by class. Each code in FDA's classification database has a Submission Type, and one 21 CFR regulation can hold both exempt and non-exempt codes.
  • Every exemption has limits. Each of the 16 parts in 21 CFR 862 to 892 has a .9 section: a new intended use, a different fundamental scientific technology or certain higher-risk in vitro diagnostic uses take a device outside the exemption; regulations may add partial limitations.
  • An exemption removes only the 510(k). Registration and listing, labeling and medical device reporting still apply; for the quality system, check the GMP Exempt? field and the regulation text.
  • The 6 October 2026 order (91 FR 63478) extends existing partial exemptions for 14 types of clinical toxicology drug test systems to Federal drug testing programs, but keeps the employment and insurance testing use and labeling conditions.
  • "Third-party review eligible" is only a 510(k) review route. FDA still decides, and the flag has nothing to do with exemption.

Background: FDA product codes, 21 CFR regulations and 510(k) exemptions

Each three-letter product code sits under a classification regulation (Regulation Number, e.g. 880.2910) that sets the device type's identification and class; the code then pins down the premarket pathway. 21 CFR Parts 862 to 892 are organized by specialty, e.g. Part 862 for clinical chemistry and clinical toxicology devices, Part 880 for general hospital and personal use devices.

The legal basis is Federal Food, Drug, and Cosmetic Act (FD&C Act) sections 510(l) and 510(m), for class I and class II devices respectively. FDA's "Class I and Class II Device Exemptions" page says most class I and some class II devices are exempt, subject to limitations; devices that may not need a 510(k) include class I devices not intended for a use of substantial importance in preventing impairment of health and not presenting a potential unreasonable risk, and devices FDA has specifically exempted. Section 3054 of the 21st Century Cures Act amended both provisions: FDA must publish in the Federal Register the class I and class II types it finds no longer need a 510(k) (for class II, at least once every 5 years). FDA may also exempt a class II type under section 510(m)(2), on its own initiative or on petition, so the list changes and must be tracked.

Step 1: how to look up an FDA product code and read its fields

The FDA Product Classification database can be searched by device name, product code or regulation number, and filtered by Submission Type to list 510(K) Exempt codes. The fields that matter (names as shown on the page):

FieldMeaningWatch for
Product CodeThree-letter codeConfirm the device fits the code's definition
Regulation Number21 CFR sectionExemption conditions are in paragraph (b) Classification
Device ClassClass 1, 2 or 3Class does not tell you whether it is exempt
Submission TypePathway, e.g. 510(k), 510(K) Exempt, Enforcement DiscretionCheck this first
DefinitionCode definitionPartially exempt codes cite the order and its limits
GMP Exempt?Whether quality system GMP requirements are exempted"Yes" may still carry retained requirements
Third Party ReviewWhether third-party 510(k) review is possibleEligible or Not Third Party Eligible
Recognized Consensus StandardsFDA-recognized standards linked to the codeSome exemption conditions cite them

Step 2: how to tell whether the 510(k) exemption applies

General limitations: 21 CFR 862.9 to 892.9 (880.9 as the example)

Sections 862.9 through 892.9 are all titled "Limitations of exemptions from section 510(k) of the Federal Food, Drug, and Cosmetic Act (the act)". Under 21 CFR 880.9, an exemption covers only devices with the existing or reasonably foreseeable characteristics of commercially distributed devices of that type; for in vitro diagnostics, only where misdiagnosis would not be associated with high morbidity or mortality. A 510(k) is still required, even within an exempt type, when:

  • (a) the intended use differs from a legally marketed device of that type, e.g. a different medical purpose, or lay use where earlier use was by health care professionals only;
  • (b) the device uses a different fundamental scientific technology, e.g. cutting tissue with a laser rather than a metal blade, or an in vitro diagnostic using DNA probe or nucleic acid hybridization instead of culture or immunoassay to identify infectious agents;
  • (c) an in vitro diagnostic is intended for any of nine uses: neoplastic disease diagnosis, monitoring or screening (except immunohistochemistry); screening or diagnosis of familial or acquired genetic disorders, including inborn errors of metabolism; measuring a surrogate marker for life-threatening diseases such as AIDS, chronic or active hepatitis, tuberculosis or myocardial infarction, or monitoring therapy; cardiovascular risk assessment; diabetes management; identifying microorganisms directly from clinical material; antibodies to microorganisms other than IgG, or IgG assays that are not qualitative, determine immunity or use matrices other than serum or plasma; noninvasive testing as defined in 21 CFR 812.3(k); and near-patient (point-of-care) testing.

.9 works as an exclusion: if any item applies, a code marked 510(K) Exempt cannot be relied on. Example: class I regulation 862.1510 (nitrite, nonquantitative, test system) is exempt subject to 862.9, yet in FDA's database its code JMT is 510(K) Exempt while NGJ, named "Over The Counter", is 510(k). The difference can be read against 862.9(a) on lay use.

Partial exemptions: one regulation, several codes (880.2910)

FDA can exempt only part of a device type and give exempt and non-exempt devices separate codes, as with 21 CFR 880.2910 (clinical electronic thermometer). The final order of 18 June 2025 (90 FR 25889), effective on publication, exempts the type only if all three conditions are met: no telethermographic functions; no continuous temperature measurement functions; and appropriate analysis and testing validates specifications and performance, for example under the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, ASTM E1965, ASTM E1112 or ASTM E1104. As of 12 November 2026, FDA's database lists these codes under 880.2910:

CodeNameSubmission TypeNotes
SDVClinical Electronic Thermometer510(K) ExemptNew in 2025; intermittent thermometers meeting the three conditions
SDWStationary Infrared Thermometer510(k)New in 2025; telethermographic thermometers, e.g. kiosks with an infrared thermometer, wall- or ceiling-mounted units
FLLContinuous Measurement Thermometer510(k)Renamed from "Thermometer, Electronic, Clinical"; continuous products such as surgical temperature probes and skin patches
OZKThermometer, Exhaled Breath510(k)Exhaled breath thermometer
PWW, PXHFever Monitoring Kit, Thermometer KitEnforcement DiscretionConvenience kits under FDA's kit guidance

Six codes, three pathways. The SDV definition says no 510(k) is needed only within 880.9 and the order's specific limitations, and the third condition shows that exemption does not mean no validation testing.

Step 3: obligations that remain after a 510(k) exemption

FDA's exemptions page states that exempt devices must still meet other regulatory requirements unless their classification regulation expressly exempts them. The October 2026 order lists requirements that may apply to manufacturers and importers, including registration and listing (21 CFR Part 807), labeling (Part 801, plus Part 809 for in vitro diagnostics), quality management system (Part 820, the QMSR) and medical device reporting (Part 803).

Read GMP exemptions in the regulation text. For the class I stand-on patient scale (21 CFR 880.2700, code FRI), the database shows GMP Exempt? Yes; the current eCFR text exempts it from the QMSR's good manufacturing practice requirements except records and complaint files under § 820.35. As of 12 November 2026, the FRI page note still cites the old QSR sections 820.180 and 820.198; where they differ, follow the eCFR. For QMSR background, see FDA replaced QSR with QMSR.

Reading the 6 October 2026 clinical toxicology exemption order (91 FR 63478)

On 6 October 2026 FDA published the final order "Medical Devices; Exemptions From Premarket Notification: Class II Devices; Certain Clinical Toxicology Test Systems" (91 FR 63478, FR Doc. 2026-20448, Docket No. FDA-2026-N-4268). Issued under FD&C Act section 510(m)(2), it took effect on publication and widens existing partial exemptions rather than creating new ones.

DateDocumentWhat it did
11 July 201782 FR 31976Class II exemptions under the Cures Act procedure, including these drug test systems
30 December 201984 FR 71794CFR text amended to reflect the 2017 exemptions
1 May 202691 FR 23427Notice of intent to expand; comments due 30 June 2026
6 October 202691 FR 63478Final order as proposed, amending paragraph (b) of 14 regulations

Before the order, these exemptions covered only test systems for employment and insurance testing, labeled as intended solely for that use, and excluded Federal drug testing programs (the order cites SAMHSA, Department of Transportation and US military programs). The order removes only that exclusion; the use and labeling conditions and 862.9 still apply. FDA said the labeling statement needs no specific wording, but a device without it still needs a 510(k); no other standards or requirements replace the 510(k), and other applicable requirements continue.

21 CFR sectionAnalyteExempt codeCorresponding non-exempt codes
862.3100AmphetaminePUX10
862.3150BarbituratesPUY8
862.3170BenzodiazepinesPUZ5
862.3250Cocaine and metabolitePVA11
862.3270CodeinePVB3
862.3580LSDPVC2
862.3610MethamphetaminePVD4
862.3620MethadonePVE7
862.3630MethaqualonePVF1
862.3640MorphinePVG10
862.3650OpiatesPVH7
862.3700PropoxyphenePVI5
862.3870CannabinoidsPVJ4
862.3910Tricyclic antidepressantsPVK5

Each non-exempt code has a corresponding exempt code in the same row, and a device within the limitations falls under the exempt code. Fentanyl tests generally fall under 862.3650 (opiate test system); devices formerly under DJG move to PVH if they stay within the limitations.

Watch for database lag: the order says exempt code definitions will cite it instead of 82 FR 31976, yet as of 12 November 2026 FDA's PUX definition still cites 82 FR 31976, while the eCFR text of 862.3100 already lists 91 FR 63486 in its history. Rely on the Federal Register and the current eCFR.

How to track new 510(k) exemptions

  • FDA's "Class I and Class II Device Exemptions" page tables exemption notices issued under the Cures Act procedure; as of 12 November 2026, the latest row is 91 FR 63478, affecting 14 product codes. The table excludes exemptions granted on petition.
  • On federalregister.gov, search FDA documents for "Exemptions From Premarket Notification" and separate requests for comments from final orders. For the class II request for comments of 6 February 2026 (91 FR 5485, Docket No. FDA-2026-N-0232), no final order could be found as of 12 November 2026.
  • Each eCFR section ends with a bracketed amendment history.

What "third-party review" means on a product code page

The Third Party Review field in FDA's database and the "Third-party review eligible" flag on our code pages refer to the 510(k) Third Party Review Program, formally the Accredited Persons Program. FDA describes it as voluntary: a 510(k) for an eligible device can go first to an FDA-recognized review organization, which applies FDA's criteria and sends FDA its review with a substantially equivalent or not substantially equivalent recommendation; FDA makes the final decision. FDA publishes the eligible device list (with exceptions in its guidance) and the list of recognized organizations.

The flag says nothing about exemption, and since the criteria match FDA's, testing data and other submission content are not reduced. A review organization reviews 510(k) files; it is not a laboratory issuing test reports. Compare two class II 510(k) codes: FMI (single lumen hypodermic needle, 21 CFR 880.5570), eligible in FDA's database and flagged on our page, and NBW (over-the-counter blood glucose test system, 21 CFR 862.1345), Not Third Party Eligible and unflagged.

Common 510(k) exemption misreadings

MisreadingRealityCheck
Class I means no 510(k)Most class I devices are exempt, not all; e.g. NGJ under 862.1510The code's Submission Type
The regulation is exempt, so my device isOnly some codes may be exempt, as with 880.2910Settle the code, then read its definition
The code says Exempt, so I can rely on itThe device must stay within .9 and any partial limitationsEach .9 item and the FR order cited in the definition
The database is always currentDefinitions can lag behind FR orders, as with PUXThe FR text and current eCFR text

How to look it up in our database

  1. Browse by specialty. As of 12 November 2026, our FDA device product codes page holds 7091 codes grouped by FDA medical specialty, e.g. 322 under General Hospital and 268 under Clinical Toxicology. A specialty page such as Clinical Toxicology lists code, device name, class, 21 CFR section and 510(k)s over three years.
  2. List all codes under one regulation. In Regulation search, enter "880.2910" to list its six codes, or "862.3100" for exempt code PUX and its 10 non-exempt codes, the quickest way to spot a split regulation; codes and device names also work.
  3. Read the code page. A page such as SDV shows class, 21 CFR section, specialty, marketing pathway and FDA's definition. Flags such as "Third-party review eligible" and "GMP exempt" appear where they apply (see FMI). Codes with 510(k) records show FDA's all-years total and list the last three years' clearances.
  4. Check back with FDA. Each code page links to FDA's classification page; our data comes from openFDA and defers to current FDA publications, so confirm definitions and recognized standards with FDA and the eCFR.
  5. Look at listing data. Exempt devices still need registration and listing; for US listing data, see United States device registration in numbers.

FAQ

Where can I look up an FDA product code?

In FDA's Product Classification database, by device name, code or regulation number. In our database, browse the product code pages by specialty or enter a regulation number in Regulation search.

Do all class I devices skip the 510(k)?

No. Most are exempt, but not all, and every exemption is subject to .9; go by the code's Submission Type and the regulation text.

Is my device exempt if its regulation has an exempt code?

First confirm it fits the exempt code's definition, then check every .9 item and every partial limitation. If any applies, a 510(k) is needed.

Not sure whether a device is 510(k) exempt: can I ask FDA?

The October 2026 order encourages manufacturers with questions to discuss them with FDA and to consider a request for information under FD&C Act section 513(g). FDA's exemptions page adds that CDRH's Office of Industry and Consumer Education (OICE) can help identify applicable requirements.

A supplier says its device is 510(k) exempt. What should I ask for?

The product code, the text of the 21 CFR regulation, an explanation of why the device stays within .9 and any partial limitations (intended use, technology, labeling statement), and its registration and listing information, all checkable in FDA's database and the eCFR.

In short: settle the code, read the regulation, test the device against .9 and partial limitations, and check the Federal Register for new orders. An exemption removes only the 510(k). This article is based on official public information available as of 12 November 2026; the latest regulatory documents take precedence.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.