Device registration in swissdamed has been mandatory in Switzerland since 1 July 2026, and the transitional period ends on 31 December 2026. Before placing devices on the Swiss market, a manufacturer based outside Switzerland must also have designated a Swiss authorised representative (CH-REP), and the CH-REP and the Swiss importer must each register with Swissmedic for a Swiss Single Registration Number (CHRN); a CE certificate and an EU authorised representative do not replace these steps. Written for Swiss importers and their Chinese suppliers, this guide draws on the Medical Devices Ordinance (MedDO, SR 812.213), the IVD Ordinance (IvDO, SR 812.219) and Swissmedic information sheet MU600_00_016 (version 8.0, valid from 26 October 2026).

Key points

  • A manufacturer not domiciled in Switzerland may only place devices on the market after designating a CH-REP domiciled in Switzerland by written mandate (MedDO Art. 51 para. 1). Per Swissmedic this also applies to EU/EEA manufacturers, so an EU authorised representative does not replace a CH-REP; Liechtenstein manufacturers are exempt.
  • The CH-REP and the importer are separate roles, and each registers for a CHRN (MedDO Art. 55, IvDO Art. 48). A company acting in both roles needs two CHRNs. Distributors do not register.
  • Devices are registered in swissdamed by Swiss manufacturers, persons who assemble systems and procedure packs, or the CH-REP of a foreign manufacturer. Importers do not register devices; they link their CHRN to UDI-DIs already registered.
  • Devices with sales units placed on the Swiss market after 1 July 2026 must be registered by 31 December 2026. Devices for which a serious incident, field safety corrective action (FSCA) or trend must be reported to Swissmedic have no transitional period. Importers have until 1 April 2027 to complete their links.
  • swissdamed and EUDAMED are not connected, and an EU SRN does not replace a CHRN; EUDAMED device data can be downloaded as XML and uploaded to swissdamed.

Background and scope: after the Swiss-EU MRA was not updated

Per the Swissmedic information sheet, under the former EU directives the Swiss-EU Mutual Recognition Agreement (MRA) gave Switzerland equal-partner access to the EU medical device market. When the revised MedDO entered into force on 26 May 2021, the MRA was not updated. To limit the consequences, notably the lack of access to EUDAMED and of cooperation in market surveillance, Switzerland introduced the CH-REP, registration of economic operators and devices in swissdamed, reporting of serious incidents to Swissmedic, and recognition of EU certificates of conformity.

On EU certificates, MedDO Art. 25 para. 4 deems certificates from bodies designated under EU law in the EU/EEA, but not recognised by an international agreement, equivalent to Swiss certificates if it can be credibly demonstrated that the procedures meet Swiss requirements and the body has an equivalent qualification; the ordinance text governs the details. The remaining work lies in the Swiss obligations on top of the certificate.

  • As of 11 November 2026, the consolidated MedDO and IvDO on Fedlex were last amended on 1 July 2026; the registration obligation comes from the amendment of 20 November 2024 (AS 2024 742), in force since that date.
  • For devices placed on the market under MedDO/IvDO, "in Switzerland" means the common market of Switzerland and Liechtenstein (customs union).
  • MedDO cross-refers to the EU MDR: Article 11 for the CH-REP's rights and obligations, Article 12 for a change of CH-REP and Article 13 for the importer's further obligations. IVD equivalents: IvDO Articles 44, 45, 48 and 90.

Swiss authorised representative (CH-REP): who needs one and what it does

  • Mandate and responsibility. A CH-REP is a person domiciled in Switzerland that has accepted a written mandate from a manufacturer located abroad (Art. 4 para. 1). Per the information sheet, the designation must cover at least all devices of the same generic device group; non-Swiss persons who assemble systems and procedure packs must also designate one (Art. 51 para. 5). The CH-REP is responsible for the formal and safety-related aspects of placing the device on the market (Art. 51 para. 2), and its rights and obligations follow MDR Article 11 (para. 3).
  • Technical documentation. The CH-REP keeps a copy, or the parties agree that the manufacturer submits it directly to Swissmedic on request, in which case the CH-REP must ensure it arrives within seven days (Art. 51 para. 3bis).
  • PRRC. The CH-REP must permanently and continuously have at least one person responsible for regulatory compliance (Art. 52 para. 1); Art. 49 paras. 2 to 4 apply mutatis mutandis, including a deputy.
  • Vigilance. The CH-REP is responsible for reporting serious incidents and FSCAs in Switzerland to Swissmedic, and submits trend reports unprompted. The transfer of these obligations must be agreed in writing in the mandate (Art. 66 para. 2bis).
  • Verification and retention. Per the information sheet, the CH-REP checks the declaration of conformity, technical documentation and conformity assessment, and from 1 January 2027 ensures devices are registered in swissdamed. Declarations and certificates are kept for 10 years after the last device is placed on the market (15 for implantables).
  • Mandate in swissdamed. The mandate is entered in the Actors module with a validity period; mandates reported earlier were not imported and must be entered again. Without a valid mandate, no devices may be placed on the market (Art. 51); Swissmedic may inactivate an expired mandate, marking linked UDI-DIs "No longer placed on the market".

Swiss importer obligations and CHRN registration

An importer is not designated: any person domiciled in Switzerland who places a device from abroad on the Swiss market is the importer (Art. 4 para. 1). Per the information sheet, pharmacies, supermarkets and online shops buying directly from abroad are importers, and two companies importing identical devices in parallel are both importers. Main obligations:

  • Before placing on the market, verify the conformity marking, the declaration of conformity, that the manufacturer is identified and has designated a CH-REP under Art. 51, compliant labelling and instructions for use, and the UDI where applicable (Art. 53 para. 1); non-conforming devices must not be placed on the market (para. 3).
  • Indicate its name, place of business and contact address on the device, its packaging or an accompanying document (Art. 53 para. 2).
  • Register for a CHRN before first placing a device on the market, and report changes within one week (Art. 55 paras. 1 and 2).
  • Per chapter 5 of the information sheet, from 1 January 2027 verify that imported devices are registered in swissdamed.
  • Link its CHRN in swissdamed to registered UDI-DIs with the status "On the market" (possible since 1 October 2026). Until 1 April 2027, Swissmedic will not take enforcement measures solely for a missing link; after that, no delay between placing on the market and linking is accepted.
ItemCH-REPImporterDistributor
How the role arisesWritten mandate from a foreign manufacturerPlacing a device from abroad on the Swiss marketAny other operator making the device available
CHRNRequiredRequiredNot required and not possible
PRRCRequiredNo obligationNo obligation
swissdamedRegisters mandated devicesLinks its CHRN onlyNot applicable
Serious incident and FSCA reportsEnsures reports reach SwissmedicNot requiredNot required
Document retentionTechnical documentation copy or seven-day agreement; declarations and certificates 10 years (15 for implants)Declarations and certificates 10 years (15 for implants)No requirements under therapeutic products legislation

The CHRN is applied for in the swissdamed Actors module with evidence of domicile in Switzerland (commercial register extract or certificate of residence). It is issued per role; one CHRN cannot cover several roles. Swissmedic has in principle a maximum of 30 days to process an application; after an objection, the 30 days restart once the shortcomings are rectified.

Swiss labelling: where to indicate the CH-REP and importer

The manufacturer must always be on the label. For the CH-REP, chapter 6 of the information sheet applies:

Device typeWhere the CH-REP is indicated
MDR devicesOn the label
IVDR devices not for self-testingOn the label or in an accompanying document
IVDR devices for self-testingOn the label
Devices CE-marked under the former directivesDepends on whether there is an EU/EEA manufacturer or EU authorised representative; see the information sheet

The importer may be indicated on the device, the packaging or an accompanying document. In the Swiss interpretation this can be a delivery note, guarantee certificate, customs documents, invoice, sticker or the instructions for use, but it must travel with the device through the supply chain so distributors can carry out their checks; Swissmedic notes this differs from the EU view in MDCG 2021-27, question 8. Product information must in principle be in all three official languages of Switzerland (Art. 16 para. 2); para. 3 sets out when fewer languages or English may be used.

swissdamed device registration: deadlines and requirements

swissdamed, operated by Swissmedic, has an Actors module and a UDI Devices module, resembles EUDAMED, and makes registered device data publicly viewable.

DateMilestoneBasis
August 2025Voluntary device registration opensswissdamed Q&A
1 July 2026Registration mandatory; devices subject to serious incident, FSCA or trend reporting registered immediatelyMedDO Art. 17 para. 5, Art. 108; IvDO Art. 16 para. 5, Art. 90
1 October 2026Importer linking availableSwissmedic importers page
9 November 2026Contact lenses requiring a Master UDI-DI start registering; complete by 31 December 2026swissdamed Q&A
31 December 2026Registration transitional period endsMedDO Art. 108 para. 2; IvDO Art. 90 para. 2
January 2027First registration fee invoices (amounts per Swissmedic)Swissmedic announcement
1 April 2027Importer linking transitional period endsImporters page; information sheet
1 November 2028Spectacle frames, spectacle lenses and ready-to-wear reading spectacles start registering; complete by 31 December 2028swissdamed Q&A

Who registers, and which devices

  • Swiss manufacturers (including of systems and procedure packs) and CH-REPs of foreign manufacturers register devices, once registered as actors. Under a specific mandate, the CH-REP can give a foreign manufacturer's user the "UDI Editor" role to upload devices and set market status.
  • The obligation covers devices under current legislation (including Annex 1 MedDO products without an intended medical purpose) and legacy devices placed on the market under MedDO Art. 101 or IvDO Art. 82, as long as further sales units are placed on the Swiss market after 1 July 2026. If the same device is already registered as an MDR/IVDR device, the legacy device generally needs no separate registration unless it has a vigilance case.
  • Placing on the market refers to each individual unit, so a device on sale before 1 July 2026 and still supplied afterwards must be registered too; once a UDI-DI is registered, further units need no new registration.

How to upload

  • Three methods: online editor, XML upload (EUDAMED "GET DEVICE" or "POST DEVICE" format) and machine-to-machine (REST API). Economic operators must upload the data themselves.
  • Manual XML uploads require a 1:1 relationship between Basic UDI-DI and UDI-DI, up to 300 pairs per file. Only EMDN nomenclature is used; devices with electronic instructions for use need the eIFU URL; UDI-DIs left in "Draft" for 30 days are deleted.
  • For certain high-risk devices whose data must be confirmed by a notified body, Swissmedic has a separate enforcement arrangement; see its device registration page.

Common misconceptions

  • "An EU authorised representative is enough": it does not replace a CH-REP.
  • "The importer will register the devices": importers only link to registered devices.
  • "Devices on sale before 1 July do not count": units placed afterwards still require registration.

swissdamed registration checklist before 31 December 2026

  1. List every UDI-DI still supplied to Switzerland after 1 July 2026 and note whether it falls under MDR/IVDR or the former directives.
  2. Review the CH-REP mandate: written, covering the generic device group, stating how technical documentation is kept and how vigilance reporting is transferred.
  3. Check that the CH-REP appears in swissdamed "Search for actors" and confirm with it that the mandate is entered and valid; allow CHRN review time for any new CH-REP or importer.
  4. Decide who uploads: the CH-REP, or a manufacturer user with the UDI Editor role.
  5. Prepare data: download XML from EUDAMED; check EMDN codes, eIFU URLs and the XML rules above.
  6. Register immediately any device with a serious incident, FSCA or trend report.
  7. Check labelling: CH-REP on the label of MDR devices; importer details travel with the goods; languages meet Art. 16.
  8. Importers: confirm your CHRN and plan to complete all links before 1 April 2027.

How to look it up in our database

First, the limits: as of 11 November 2026, this site has no Swiss registration data. The data sources and licences page lists registration data from official registers in 14 markets, without swissdamed, and our regulatory database has no Swiss entries (read MedDO and IvDO on Fedlex). The site helps in three ways:

  1. Swiss-based companies. As of 11 November 2026, the Switzerland country page shows 538 companies, 479 with registrations found. These are companies in our directory whose country is Switzerland, with record counts for China, the EU, the US, the UK, Brazil, Saudi Arabia and Canada and no Swiss column; it is neither Swiss registration data nor a list of CH-REPs or importers. To check a CH-REP or importer, use "Search for actors" on the swissdamed website; Swissmedic states that an actor visible there is officially registered.
  2. A manufacturer's EUDAMED footprint. Use site search to open the manufacturer's company page. The EU entry shows its UDI-DI records in EUDAMED (only a lower bound such as "≥300" for large result sets), and the Regulatory identity section lists the EU SRN and EU authorised representative where public. This roughly indicates how many UDI-DIs could be exported as XML for swissdamed. For EU-side dates, see our article on the two EUDAMED deadlines.
  3. EU provisions MedDO refers to. The EU section of our regulatory database includes the MDR (2017/745) and the authorised representative guidance MDCG 2022-16, with reference number, issuing body, status and a link to the official text; as of 11 November 2026 it holds 172 regulations and guidance documents. Swissmedic documents govern Swiss practice.

Frequently asked questions

Do manufacturers with a CE certificate and an EU authorised representative still need a Swiss CH-REP?

Yes. The CH-REP must be domiciled in Switzerland, and Swissmedic states the requirement also applies to EU/EEA manufacturers; only those in Liechtenstein are exempt.

Can a Chinese manufacturer register its devices in swissdamed itself?

A foreign manufacturer is not required to register as an actor in Switzerland; its CH-REP registers the devices and can give a manufacturer user the UDI Editor role under the mandate to upload the data.

Our Swiss company is both CH-REP and importer. How many CHRNs do we need?

Two. CHRNs are issued per role, each with its own application, and the company carries both sets of obligations.

Do Swiss distributors need to register?

No, and they cannot obtain a CHRN. But before making a device available, distributors must check the conformity marking, declaration of conformity, product information, importer details and UDI (MedDO Art. 54 para. 1).

Is there a fee for swissdamed registration?

Yes. Fees are charged per UDI-DI, first invoices will be sent from January 2027, and devices registered before then are counted retrospectively. See Swissmedic's fee information for amounts.

In short, Switzerland expects a registered CH-REP, a registered and linked importer, and publicly searchable device data in swissdamed. Primary texts: MedDO on Fedlex, the Swissmedic information sheet Obligations Economic Operators CH and the swissdamed questions and answers. This article is based on official public information available as of 11 November 2026; the latest documents from the regulators prevail. The accreditation scope of the Sungo laboratories is available on the testing laboratory page.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.