The EU Machinery Regulation (EU) 2023/1230 will apply from 20 January 2027, when Machinery Directive 2006/42/EC is repealed. Powered wheelchairs and mobility scooters stay under the Medical Device Regulation (MDR, Regulation (EU) 2017/745), but MDR Article 1(12) requires a device that is also machinery to meet, where a relevant hazard exists, the machinery requirements that are more specific than the MDR's own; from 20 January 2027 that reference points to Regulation (EU) 2023/1230. Exporters need to confirm whether the product is machinery and move the machinery part of the technical documentation from Annex I of the Directive to Annex III of the Regulation.
Key points: what the EU Machinery Regulation means for powered wheelchairs
- Dates: Regulation (EU) 2023/1230 was adopted on 14 June 2023, published in OJ L 165 on 29 June 2023 and entered into force on 19 July 2023. A corrigendum of 4 July 2023 (OJ L 169) replaced "14 January 2027" in the original text with 20 January 2027, which is both the application date and the repeal date of Directive 2006/42/EC.
- The market route does not change: a powered wheelchair remains an MDR device, generally class I under Rule 13 of MDR Annex VIII; MDCG 2021-24 Rev.1 lists electric wheelchairs as a Rule 13 class I example. The Machinery Regulation does not change the MDR class.
- Machinery requirements do not start in 2027: MDR Article 1(12) already applies them through Directive 2006/42/EC. From 20 January 2027, under Article 51(2) and the Annex XII correlation table of Regulation (EU) 2023/1230, the definition maps to Article 3 and the essential health and safety requirements to Annex III.
- The technical documentation is what changes: in the MDR demonstration of conformity with the general safety and performance requirements (GSPRs), add or update the applicable machinery requirements, standards and evidence. Replacing the directive number is not enough.
- Test evidence has limits: EN 12184:2022 results on moving parts, trapping of body parts, folding and adjustment mechanisms and emergency stop can be one piece of evidence for the related hazards; any presumption of conformity depends on the harmonised standard references published in the Official Journal.
How MDR Article 1(12) links to the Machinery Regulation
MDR Article 1(12) says that devices which are also machinery within the meaning of point (a) of the second paragraph of Article 2 of Directive 2006/42/EC shall, where a hazard relevant under that Directive exists, also meet the essential health and safety requirements of Annex I to that Directive to the extent they are more specific than the GSPRs in Chapter II of MDR Annex I. There is no class condition; only two questions matter: is the product machinery, and does a relevant hazard exist?
Regulation (EU) 2023/1230 has a matching provision. Under Article 9, where the risks addressed by the requirements in Annex III are wholly or partly covered for a product by more specific Union harmonisation legislation, the Regulation does not apply to that product to the extent that legislation covers those risks. The Machinery Regulation does not name the MDR, so MDR Article 1(12) and Article 9 have to be read together.
On classification, Rule 1 of MDR Annex VIII places non-invasive devices in class I by default, and Rule 13 places all other active devices in class I. An active device is one whose operation depends on a source of energy other than the human body or gravity. MDCG 2021-24 Rev.1 (April 2026) lists wheelchairs among the Rule 1 examples and electric wheelchairs among the Rule 13 examples. A mobility scooter, also an active device, generally falls under Rule 13 too; classification is decided case by case.
Is a powered wheelchair "machinery"? Article 3 of Regulation (EU) 2023/1230
Article 3, point (1)(a), of Regulation (EU) 2023/1230 defines machinery as an assembly fitted with, or intended to be fitted with, a drive system other than directly applied human or animal effort, consisting of linked parts or components, at least one of which moves, and joined together for a specific application. A first sorting looks like this:
- Powered wheelchairs and mobility scooters: a motor drive system and moving parts (drive wheels, transmission, often powered seat, backrest or leg-rest adjustment) mean they usually meet the wording of the definition and need a machinery hazard assessment under MDR Article 1(12).
- Manual wheelchairs with powered functions: a power-assist unit, powered seat lift or powered standing mechanism calls for the same assessment.
- Purely manual wheelchairs: propelled only by directly applied human effort, they generally do not meet point (a) of the definition.
Whatever the outcome, record the reasoning in the technical documentation rather than excluding the product by default.
EU Machinery Regulation 2023/1230 timeline
All dates in the table follow the corrected text.
| Date | Event | Basis |
|---|---|---|
| 14 June 2023 | Regulation (EU) 2023/1230 adopted by the European Parliament and the Council | Title of the Regulation |
| 29 June 2023 | Published in the Official Journal, OJ L 165 | Official Journal |
| 4 July 2023 | Corrigendum published in OJ L 169, correcting dates in Articles 51, 52, 54 and others | Corrigendum, points 5, 6, 10 and others |
| 19 July 2023 | Regulation enters into force (20th day after publication) | Article 54, first paragraph |
| 20 January 2027 | Regulation applies; Directive 2006/42/EC repealed | Article 54, second paragraph, and Article 51(2), as corrected |
| Before 20 January 2027 | Member States may not impede the making available of products placed on the market in conformity with Directive 2006/42/EC | Article 52(1), as corrected |
Article 52(1) concerns products placed on the market under Directive 2006/42/EC. A powered wheelchair is placed on the market under the MDR, and Regulation (EU) 2023/1230 has no transitional provision written for medical devices, so how the switch applies to devices placed on the market after 20 January 2027 is for the competent authorities to confirm. The safe course is to finish the update before that date.
What to change in the technical documentation before and after 2027
The role of each text:
| Text | Role | What it means for a powered wheelchair |
|---|---|---|
| MDR (EU) 2017/745 | Device classification, conformity assessment and market access | Remains the basis for market access; powered wheelchairs are generally class I |
| MDR Article 1(12) | Adds machinery requirements more specific than the GSPRs for devices that are also machinery | Applies to powered wheelchairs and scooters with a drive system and moving parts |
| Directive 2006/42/EC | Former Machinery Directive, repealed on 20 January 2027 | The text MDR Article 1(12) refers to until that date |
| Regulation (EU) 2023/1230 | New Machinery Regulation, applies from 20 January 2027 | References converted through Annex XII; requirements in Annex III |
| EN 12184:2022 | Requirements and test methods for electrically powered wheelchairs, scooters and their chargers | Can provide test evidence for moving parts, trapping, folding and adjustment, and emergency stop hazards |
Following the structure of the MDR technical documentation, the machinery requirements can be handled in six steps:
- Record the machinery decision. Check the drive system, moving parts and specific application against Article 3, point (1)(a), and write the conclusion and reasons into the technical documentation.
- Identify machinery hazards. Follow Part B (General principles) of Annex III: set the limits of the product (intended use and reasonably foreseeable misuse), identify hazards and hazardous situations, estimate and evaluate the risks, then eliminate hazards or reduce risks. Feed the results into the risk management file under Section 3 of MDR Annex I.
- Find the more specific machinery requirements. Compare Annex III with Chapter II of MDR Annex I, list the requirements that are more specific than the GSPRs and apply to the product, and give reasons for those that do not. Relevant examples include sections 1.3.7 (risks related to moving parts), 1.2.4.3 (emergency stop) and 1.2.1 (safety and reliability of control systems). For software-based controllers, also compare section 1.1.9 (protection against corruption) with Section 17 of MDR Annex I (electronic programmable systems) to see which is more specific.
- Write them into the GSPR demonstration. Section 4 of MDR Annex II requires the applicable requirements, the methods used to demonstrate conformity, the harmonised standards or other solutions applied, and the location of the evidence. Treat the machinery requirements the same way, with test reports, calculations and inspection records as evidence.
- Update the references. From 20 January 2027, a reference to Article 2 of Directive 2006/42/EC corresponds to Article 3 of Regulation (EU) 2023/1230, and a reference to Annex I corresponds to Annex III. Check each clause against the Annex XII correlation table.
- Check the declaration of conformity. The EU declaration of conformity follows MDR Article 19 and Annex IV. Under Article 19(2), where a device is also subject, for aspects not covered by the MDR, to other Union legislation requiring a manufacturer's EU declaration of conformity, a single declaration is drawn up for all applicable Union acts. Whether to list Regulation (EU) 2023/1230 depends on how far it applies to the product (see Article 9 above); record the reasoning and confirm with the competent authority if needed.
Part A of Annex IV to Regulation (EU) 2023/1230 lists the elements of a machinery technical file, such as a complete description and intended use, risk assessment documentation, drawings and circuit schemes, the harmonised standards or common specifications applied, and calculations and test reports. It works as a checklist for the machinery part of the MDR file. For the overall MDR file structure, see EU MDR technical documentation for wheelchair exporters.
On responsibility, MDR Article 10 requires manufacturers to ensure that devices they place on the market are designed and manufactured in accordance with the MDR, including the machinery requirements brought in by Article 1(12). Under MDR Article 11, a manufacturer not established in a Member State must designate a sole authorised representative before placing a device on the Union market. The manufacturer owns this argument; exporters should not assume the buyer covers it.
Common mistakes with the Machinery Regulation for powered wheelchairs
- "Class I devices are not affected": MDR Article 1(12) has no class condition. It only asks whether the product is also machinery and whether a relevant hazard exists.
- "Machinery law reaches wheelchairs only in 2027": the machinery requirements already apply through Directive 2006/42/EC. What changes in 2027 is the text referred to and the detailed requirements.
- "Just change the number to 2023/1230": the European Commission lists changes such as provisions for AI-powered safety functions, cyber-safety for compliance-relevant software and safety control systems, and conditions for digital instructions and declarations of conformity. Map Annex III clauses through Annex XII and review each one.
- "The application date is 14 January 2027": that date appears in the original Official Journal text and was corrected to 20 January 2027 by the corrigendum.
- "An EN 12184 report means the machinery requirements are met": a test report is one piece of evidence; it still needs the risk assessment and a clause-by-clause argument.
How to check this in the database
Our regulatory database shows the numbers, status and official source links for the MDR and related guidance; it does not replace the official texts. As read on 13 November 2026:
- European Union medical device regulations: authority shown as the European Commission / MDCG; 172 regulations and guidance entries, updated 18 September 2026, each with type, number, title and status. The list includes the MDR (32017R0745), the classification guidance MDCG 2021-24 rev.1 and the class I manufacturer guidance MDCG 2019-15 rev.1.
- MDR (32017R0745) entry page: shows type, issuer, status and a short summary, with a link to the official text on EUR-Lex, where Article 1(12), Article 19 and Annexes II and VIII cited here can be read.
- Regulation search: the page states that it covers regulations and guidance, FDA product codes and recent 510(k) clearances. Searching by number gives the most precise results, for example "MDCG 2021-24" or "32017R0745".
As of the same date, Regulation (EU) 2023/1230 itself is not in our EU list. To read it, start from the European Commission's machinery page, which links to the Regulation, its latest consolidated version and the Machinery Directive application guide. For the classification examples, see MDCG 2021-24 Rev.1; the Rule 13 examples are on page 49. Whether a notified body is involved for a wheelchair is covered in Do You Need a Notified Body for Wheelchair CE Marking?
Test evidence and SUNGO
SUNGO does not carry out conformity assessment under the Machinery Regulation or the MDR and does not issue declarations of conformity on behalf of manufacturers; EN ISO 12100 is outside SUNGO's testing scope. What is relevant here is EN 12184 testing:
- The CNAS accreditation scope of Hefei SUNGO Medical Technology Co., Ltd. (certificate L24279) lists EN 12184:2022 under "electrically powered wheelchairs, scooters and their chargers", including clauses 6.10 (safety of moving parts), 6.11 (prevention of trapping of body parts), 6.12 (folding and adjustment mechanisms) and 12.6 (emergency stop).
- The Hefei laboratory's IAS accreditation (certificate TL-1119) covers EN 12184 except clause 6.6 (biocompatibility and toxicity) and clause 12.3.1.
- EN 12184 is not within the CMA scope, and the Shanghai laboratory does not test wheelchairs.
For the biological tests under EN 12184 clause 6.6, SUNGO subcontracts to a suitably qualified laboratory, with the client's prior consent, and states the subcontracted items and the subcontractor in the report. A test report is one piece of evidence in the manufacturer's argument; acceptance depends on that argument and on the regulator, and SUNGO does not promise that any authority will accept a given report.
Frequently asked questions
Is a powered wheelchair machinery under the EU Machinery Regulation?
Under Article 3 of Regulation (EU) 2023/1230, a powered wheelchair or scooter with a motor drive and moving parts joined for a specific application usually qualifies. It stays an MDR device, the machinery requirements apply through MDR Article 1(12), and the decision and reasons belong in the technical documentation.
When does the EU Machinery Regulation 2023/1230 apply?
It entered into force on 19 July 2023 and applies from 20 January 2027, when Directive 2006/42/EC is repealed. The original Official Journal text said 14 January 2027; the corrigendum of 4 July 2023 corrected it.
How is the 2006/42/EC reference in MDR Article 1(12) read after the repeal?
Under Article 51(2) of Regulation (EU) 2023/1230, references to Directive 2006/42/EC are construed as references to the Regulation and read in accordance with the Annex XII correlation table: the definition maps to Article 3 and the essential health and safety requirements map to Annex III.
Does a powered wheelchair need a separate Machinery Regulation declaration of conformity?
The declaration follows MDR Article 19 and Annex IV. If other Union legislation requiring a declaration applies to aspects not covered by the MDR, Article 19(2) requires one combined declaration; whether to list Regulation (EU) 2023/1230 depends on its Article 9 and the competent authority's view.
Does an EN 12184 test report prove the machinery requirements are met?
Not on its own. It is one piece of evidence for hazards such as moving parts, trapping of body parts, folding and adjustment, and emergency stop, alongside the risk assessment and a clause-by-clause argument. Any presumption of conformity depends on harmonised standard references published in the Official Journal.
The key to the switch is updating the machinery part of the MDR technical documentation to Regulation (EU) 2023/1230, using the corrected date of 20 January 2027. This article is based on official public information as of 13 November 2026; always follow the latest documents issued by the regulators.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.