Does a contract manufacturer need FDA establishment registration when the brand owner holds the 510(k)? As a rule, yes. Under 21 CFR 807.40 and FDA's "Who Must Register, List and Pay the Fee" chart, a foreign contract manufacturer making finished devices for the US registers its establishment, lists its devices and pays the annual fee, even when the 510(k) is in the brand owner's name. A brand owner that initiates the specifications and sells the device under its own name registers and lists separately as a specification developer.

This guide is for Chinese factories making powered wheelchairs, mobility scooters, oxygen concentrators and similar devices for overseas brands, and for those brand owners. It covers each party's duties, listing under the brand's 510(k), deadlines and fees, names on test reports, and how to check them in databases.

Key points

  • The contract factory registers and lists. FDA's chart calls a factory that makes a finished device to another establishment's specifications a contract manufacturer; a foreign one registers, lists and pays under 807.40(a). Under 807.40(c), an unlisted device, or one not made at a registered foreign establishment, may not be imported.
  • The brand owner registers and lists too. Initiating specifications for a device made by a second party and sold under one's own name is manufacturing under 807.3(d)(3); the brand registers as a specifications developer under 807.20(a)(1). Neither registration covers the other.
  • The 510(k) submitter and establishment registration are separate. Section 807.87(b) asks only for the submitter's own establishment registration number, if applicable. The factory links its establishment to the 510(k) when it lists the device and provides the 510(k) number.
  • Deadlines are fixed; the fee is set each fiscal year. Initial registration and listing are due within 30 days of starting the activity; annual registration and listing review run October 1 to December 31. The FY 2027 fee is $13,785 per establishment.
  • Test reports follow the test contract. Client, manufacturer, trademark and model are supplied by the client, and the report as issued governs. Changes after issue are requested from the issuing laboratory under its procedure.

Scope: which contract manufacturing arrangements this covers

Contract manufacturing for the US usually takes one of three forms. In OEM, the brand supplies the design and specifications, a Chinese factory builds to them, and the brand sells the device and files the 510(k). In ODM, the factory supplies the design and the brand sells a chosen model under its own label. In split production, frames, electronics and final assembly happen at different plants.

Registration is done per establishment. Section 807.3(c) defines an establishment as a place of business under one management at one general physical location where a device is manufactured, assembled or otherwise processed. Each plant site, including each site in split production, is assessed on what it actually does, not on whose name ends up on the label.

This guide does not decide whether a device needs a 510(k). Confirm the product code, class and pathway first, as shown in the database section.

Does a contract manufacturer need FDA registration? What 807.20 and 807.40 say

Section 807.20(a) requires the owner or operator of an establishment engaged in the manufacture, preparation, compounding, assembly or processing of a device for human use to register and list devices in commercial distribution. Covered persons include anyone who initiates or develops specifications for a device made by a second party ((a)(1)); sterilizes or otherwise makes a device for or on behalf of a specifications developer or any other person ((a)(2)); repackages or relabels a device ((a)(3)); acts as an initial importer ((a)(5)); or makes components or accessories ready for a health-related use and packaged or labeled for commercial distribution for that use ((a)(6)). A factory building to a brand's specifications falls under (a)(2).

A Chinese factory is a foreign establishment, so 807.40 also applies. Section 807.40(a) requires foreign establishments making devices imported or offered for import into the US to register and list. Section 807.40(b) requires each to name one United States agent, who must reside or keep a place of business in the US. Section 807.40(c) bars import of any device that is not listed or not made at a registered foreign establishment.

FDA summarizes these rules in its "Who Must Register, List and Pay the Fee" chart, with separate tables for domestic and foreign establishments. It defines a contract manufacturer as one that "manufactures a finished device to another establishment's specifications", and a specification developer as one that develops specifications for a device distributed under its own name but performs no manufacturing. The full text is in 21 CFR Part 807 on eCFR.

RoleTypical situationRegisterListAnnual feeBasis
Contract manufacturer (foreign factory)Makes the finished device to the brand's specificationsYesYesYes807.20(a)(2), 807.40(a)
Specification developer (brand owner)Initiates specifications, has a factory make the device, sells under its own nameYesYesYes807.3(d)(3), 807.20(a)(1)
Initial importer (US importer)Imports from a foreign manufacturer and furthers its marketing without changing packaging or labelingYesGives the manufacturer's name and address onlyYes807.20(a)(5)
Repackager or relabelerChanges packaging or labeling before further distributionYesYesYes807.20(a)(3)
Component maker supplying only finished-device makersComponents not classified as finished devices, sold only to a finished-device manufacturerNoNoNo807.65(a)

ODM cases are assessed on actual activities. If the factory develops the specifications and makes the device, it is the manufacturer. A brand owner that did not initiate the specifications cannot simply be treated as a specification developer; check whether it imports, relabels or performs another listed activity. For individual cases, the eCFR text and FDA's chart govern.

How the factory lists when the brand owner holds the 510(k)

A 510(k) is a premarket notification filed by a submitter. Section 807.87(b) requires "the establishment registration number, if applicable, of the owner or operator submitting the premarket notification submission". It covers the submitter only, not every manufacturing site. So the applicant field in FDA's 510(k) database shows who filed the notification; it does not mean the device is made at the applicant's site.

The factory is linked to the 510(k) at listing. FDA's registration and listing page says that if a device requires marketing authorization, the owner or operator should also provide the premarket submission number, such as the 510(k) number. Under 807.28(f), FDA assigns one listing number per premarket submission number for devices not exempt from premarket notification. The factory lists the devices made there and its activities, with the device's premarket submission number, so the brand owner has to give the factory the 510(k) number.

Under 807.28(a), a listing must be updated when an additional establishment begins manufacturing, developing specifications, repackaging, relabeling or otherwise processing a listed device, or when an establishment starts or stops an activity on it. Whether a change of manufacturing site also needs a new 510(k) is a separate assessment under FDA's change criteria and is outside this guide.

Deadlines and the annual registration fee

Time pointActionBasis
Within 30 days of starting manufacturing or another covered activityInitial registration, with listing information807.22(a)
Before first import or offer for importForeign establishment registered and device listed807.40(c), 807.28(c)
October 1 to December 31 each yearAnnual registration; review and update all listing information, confirming it even if nothing changed807.22(b)(1), (b)(3)
Within 30 days of a change in registration informationUpdate registration information807.22(b)(2)
Within 10 business days of a change in the US agent's name, address or phone numberReport the change to FDA807.40(b)(3)
When a site is added or dropped, or its activities changeUpdate the listing807.28(a)

Under 807.22(c), missing the deadlines puts an establishment in "failed to register" or "failed to list" status, and FDA does not consider it active. Annual registration for FY 2027 (October 1, 2026 to September 30, 2027) has to be completed between October 1 and December 31, 2026.

FDA published the FY 2027 medical device user fee rates in the Federal Register on July 30, 2026 (document 2026-15335). Each establishment that is registered, or required to register, pays the annual establishment registration fee: $13,785 for FY 2027. FDA may, but need not, waive the annual fee for a qualifying small business facing financial hardship; the waiver excludes initial registration. Where the brand owner and the factory each register, each pays.

Quality system: under QMSR both parties count as manufacturers

The Quality Management System Regulation (QMSR, 21 CFR Part 820) has applied since February 2, 2026. Section 820.1(a) states that manufacturers subject to the part include those performing specification development, and that a manufacturer engaged in only some operations need only comply with the requirements for those operations. The part applies to finished devices imported or offered for import into the US. The quality agreement should state which party handles production controls and which handles design controls. For background, see FDA replaced QSR with QMSR.

Common misreadings

  • "The 510(k) is in the brand's name, so the factory does not register." Registration depends on what an establishment does; FDA's chart requires foreign contract manufacturers to register, list and pay, whoever submitted the 510(k).
  • "An FDA registration number means the product is FDA-approved." FDA's registration and listing database states that registration, a registration number or a listing does not in any way denote FDA approval of the establishment or its products. Section 807.39 adds that implying official approval because of registration is misbranding.
  • "The 510(k) applicant in the database is the factory." The applicant filed the notification. Which establishments make the device is shown in FDA's registration and listing database, searchable by product code or company name.

Test reports: whose name goes on the report

The client on a test report is the party that places the test order and supplies the samples: the brand owner or the factory, as they agree. The manufacturer name and address, trademark, model and similar details are supplied by the client when the order is placed, and the report as issued by the laboratory governs.

Before filing, check whether the device name, model and specifications in the report correspond to the device in the 510(k), and whether the manufacturer details match the actual production site. Where the report client differs from the 510(k) submitter, explain each party's role in the submission. This is a self-check, not a Part 807 requirement; FDA's expectations come through its review. Changes to client or manufacturer details after issue go to the issuing laboratory under its procedure; never alter a report yourself.

How to look it up in the database

All pages and figures below were read online on 13 November 2026.

  1. Find the product code. The site's FDA device product code index lists 7,091 codes as of 13 November 2026, browsable by FDA medical specialty. Each code page gives the 21 CFR section, device class and marketing pathway.
  2. See who files 510(k)s under a code. The ITI code page (Wheelchair, Powered) shows 21 CFR 890.3860, Class 2 and the 510(k) pathway. As of 13 November 2026 it lists 98 clearances in its three-year data window, 70 with a Summary for direct download, and 586 across all years at FDA. For mobility scooters use INI (21 CFR 890.3800: 40 in the window, 30 Summaries, 351 across all years); for portable oxygen concentrators use CAW (21 CFR 868.5440: 22, 19 and 288), also as of 13 November 2026. Rows show 510(k) number, decision date, applicant, device name and Summary status, with no manufacturing-site column.
  3. Check a company's US listings. Use the site search to find a company by name and open its page. The "United States FDA openFDA" block shows the number of listed device types, registration records and 510(k)s, with some listed device names. The search page notes that device rows are a sample of up to 8 per company per market, so a missing record does not mean none exists. For an overview, see Chinese device manufacturers in United States.
  4. Confirm production sites in FDA's own database. FDA's Establishment Registration and Device Listing database can be searched by product code, establishment type (for example Contract Manufacturer or Specification Developer) and establishment country together. For example, enter ITI, choose Contract Manufacturer and select China to look for Chinese establishments registered as contract manufacturers for that code. The page notes that the public database is generally updated weekly.

This site and FDA's databases may update at different times; FDA's databases govern.

Test evidence and SUNGO

Among standards relevant to these three device types, the accreditation scope of Hefei SUNGO Medical Technology Co., Ltd. (the Hefei laboratory) is as follows; the certificate schedules govern the exact parts, editions and clauses. ISO 7176 series: 17 parts under CNAS (L24279), 14 under CMA (260015342945) and 20 under IAS (TL-1119). ANSI/RESNA WC-1 and WC-2 are within the IAS scope only. Medical electrical safety standards IEC 60601-1 and IEC 60601-1-11, the oxygen concentrator standard ISO 80601-2-69, and the medical EMC standard IEC 60601-1-2 are within the Hefei IAS (TL-1119) scope only. CNAS and IAS are both signatories to the ILAC MRA. The Shanghai laboratory's scope does not include these items. Whether a report can be used in a given 510(k) is for FDA to decide in review. See the lab page.

Frequently asked questions

Does a contract manufacturer need FDA registration if the brand owner holds the 510(k)?

Yes. FDA's chart lists foreign contract manufacturers as required to register, list and pay the annual fee under 807.40(a), and 807.40(c) requires imported devices to be made at a registered foreign establishment.

Which 510(k) number does the contract manufacturer use when listing?

The 510(k) under which the device is marketed; if the brand owner holds it, the brand owner's number. FDA asks for the premarket submission number when listing devices that need marketing authorization, and 807.28(f) assigns listing numbers by premarket submission number.

Does a foreign contract manufacturer need a US agent?

Yes. Section 807.40(b) requires each foreign establishment that must register to name one United States agent. Changes to the agent's name, address or phone number must be reported to FDA within 10 business days.

Does a component-only supplier have to register?

A maker of components not classified as finished devices and sold only to a finished-device manufacturer is exempt under 807.65(a). A maker of components or accessories packaged or labeled for a health-related use and distributed for that use must register and list under 807.20(a)(6).

Must the test report client match the 510(k) applicant?

The client is set by the test contract. Check that the device name, model and manufacturer details correspond to the 510(k), explain roles where they differ, and request changes from the issuing laboratory.

Settle each party's role under Part 807 first, then align the 510(k) number, the listings and the device details in the test reports. This article is based on official public information as of 13 November 2026. Refer to the latest documents issued by the regulatory authorities.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.