The EU MDR list of harmonised standards is the Annex to Implementing Decision (EU) 2021/1182; only editions whose references are published in the Official Journal of the European Union (OJ) give presumption of conformity. The list was amended four times in 2026; the fourth amendment, Decision 2026/2479, was published and entered into force on 10 November 2026, taking entry numbers up to 72. Whether a standard is harmonised, and how long an old edition still counts, is shown in the amending decisions' Annexes and Article 2; this article explains the rules, the 2026 changes and how to check, without interpreting any standard's technical content.
Key points
- Under Article 8(1) MDR, presumption of conformity applies only to harmonised standards whose references are published in the OJ. The MDR list is the Annex to Implementing Decision (EU) 2021/1182, and every update amends that Annex.
- Four amendments were adopted in 2026: 2026/193, 2026/760, 2026/1231 and 2026/2479. They added entries 37 to 72 (36 entries) and updated 7 existing entries through inserted entries such as 1a, 3a and 39a. Each entered into force on its OJ publication date.
- When a new edition replaces a listed one, the old reference is usually not deleted at once. Article 2 sets the date: 15 December 2027 and 17 June 2031 (as corrected) under 2026/1231, and 10 May 2028 under 2026/2479.
- Using standards is voluntary. A non-harmonised standard, or an edition not yet harmonised, can still be used, but gives no presumption of conformity; compliance must be justified in the technical documentation.
- The summary list on the Commission's single market page is dated 20 October 2025, so it is not the 2026 list. The UK (Great Britain) designated standards are a separate list.
What are EU MDR harmonised standards? Presumption of conformity and Decision 2021/1182
The legal basis
Article 8(1) of the MDR (Regulation (EU) 2017/745) provides that devices conforming to relevant harmonised standards, or parts of them, whose references are published in the OJ are presumed to conform to the requirements those standards cover. The second subparagraph extends this to system and process requirements for economic operators or sponsors, including quality management, risk management, post-market surveillance, clinical investigations, clinical evaluation and post-market clinical follow-up (PMCF), which is why the list also contains process standards.
Regulation (EU) No 1025/2012 defines a "harmonised standard" as a European standard adopted on a Commission request for the application of Union harmonisation legislation. For the MDR and IVDR, the Commission sent request M/575 to CEN and CENELEC on 14 April 2021 and amended it on 31 January 2023 and 27 May 2024.
Where the MDR list is
Implementing Decision (EU) 2021/1182 was adopted on 16 July 2021 and published on 19 July 2021. Its Annex first listed five standards, mainly on sterilization and irritation testing. Every update since has amended it: seven times between 2022 and 2025 (2022/6, 2022/757, 2023/1410, 2024/815, 2024/2631, 2025/681 and 2025/2078) and four times in 2026. As of 12 November 2026, entry numbers run up to 72.
IVDR harmonised standards are listed separately, in Implementing Decision (EU) 2021/1195 and its amendments (in 2026: 2026/197, 2026/1313 and 2026/2481). Check which Regulation a list belongs to before relying on it.
When presumption starts
Each amending decision's recitals state that presumption applies from the date the reference is published in the OJ, so each decision enters into force on publication; neither the adoption date nor CEN's publication date counts. Decision 2026/2479 was adopted on 9 November and published on 10 November 2026, so presumption runs from 10 November.
Annex Z: what the presumption covers
MDCG 2021-5 Rev. 1 (guidance on standardisation for medical devices, revised July 2024) explains that a harmonised standard's link to the legal requirements is set out in its foreword and an informative "Annex Z" (ZA, ZB and so on when several laws are covered), whose tables map clauses to requirements and flag those not covered or only partly covered. Presumption can be claimed only for requirements listed as covered. A standard without an adequate Annex Z cannot be cited in the OJ.
Voluntary, but documented in the technical documentation
MDCG 2021-5 Rev. 1 stresses that using standards is voluntary: manufacturers may apply any harmonised, European, international or national standard, or their own solutions, if they can show these meet the applicable requirements, for example through a deeper risk assessment or gap analysis in the technical documentation. Section 4 of Annex II MDR requires the technical documentation to list the harmonised standards, common specifications (CS) or other solutions applied; Section 4.5.1 of Annex VII requires notified bodies, where relevant, to consider available CS, guidance, best-practice documents and harmonised standards even if the manufacturer does not claim compliance.
One exception: Section 23.1(h) of Annex I MDR requires any symbol or identification colour used to conform to the harmonised standards or CS. MDCG 2021-5 Rev. 1 cites this as a case where standards can be regarded as mandatory, so changes to the labelling-symbols standard need close attention.
EU MDR harmonised standards amended in 2026: 2026/193 to 2026/2479
The table shows each decision's dates, entries changed and fields covered (from the decision titles), not current editions by product type; each decision's Annex is authoritative for exact numbers, ISO editions and amendments.
| Decision | Adopted | Published in OJ (in force) | Changes to the Annex | Fields covered |
|---|---|---|---|---|
| (EU) 2026/193 | 28 Jan 2026 | 30 Jan 2026 | New entries 37–48 (12) | Neurosurgical implants, biological evaluation, clinical investigation, non-active surgical implants, sterilization, biocompatibility of breathing gas pathways, small-bore connectors |
| (EU) 2026/760 | 1 Apr 2026 | 7 Apr 2026 | New entries 49–51 (3) | Small steam sterilizers, stainless steel steam boilers, hearing aid electroacoustics |
| (EU) 2026/1231 | 11 Jun 2026 | 17 Jun 2026 (corrigendum 22 Jun 2026) | New entries 52–65 (14); entries 1a, 7a, 12a, 14a and 20a inserted, with entries 1, 7, 12, 14 and 20 to be deleted | Biological evaluation, labelling symbols, medical electrical equipment, transfusion equipment, ophthalmic optics, non-active surgical implants, washer-disinfectors, prosthetics, sharps injury protection |
| (EU) 2026/2479 | 9 Nov 2026 | 10 Nov 2026 | New entries 66–72 (7); entries 3a and 39a inserted, with entries 3 and 39 to be deleted | Radiation sterilization, clinical investigation, natural rubber latex male condoms, biological evaluation, washer-disinfectors, medical electrical equipment, small-bore connectors |
2026/193 and 2026/760: additions only
These decisions only add entries, because the earlier editions they revise were not on the MDR list. The recitals of 2026/193 state that references to editions such as EN ISO 14630:2012 and EN ISO 17665-1:2006 were not published in the OJ; EN 13060:2014+A1:2018 and EN 60118-0:2015, revised under 2026/760, appear in no 2021–2025 Annex either. Easy to miss: 2026/193 added EN ISO 14155:2020 with A11:2024 as entry 39, which 2026/2479 has since replaced with the 2026 edition. EN 14222:2021+A1:2025 (stainless steel steam boilers) in 2026/760 is, per its recitals, newly drafted.
2026/1231: amendments and new entries
First, five listed standards received amendments (EN ISO 10993-23, EN ISO 10993-12, EN ISO 10993-17, EN IEC 60601-2-83 and EN ISO 15223-1); each old entry is replaced by a new entry including the amendment, with the deletion deferred. Second, 14 new entries were added, including EN ISO 10993-1:2025, the 2025 editions of EN ISO 15883 Parts 1, 2, 3 and 7, and the general medical electrical standard EN 60601-1:2006 with A13:2024 (entry 65). Per the recitals, EN 60601-1:2006 had not been cited under the MDR, so this is its first appearance on the MDR list.
EN ISO 15223-1:2021/A1:2025 adds a defined term for "authorized representative" and makes the EC REP symbol no longer specific to a country or region. The recitals give the old reference five years because labelling changes weigh heavily on economic operators in cost and time. Labelling teams can also read the MDCG 2021-5 Rev. 1 appendix on the transition to the "EU REP" symbol.
2026/2479: superseding and new standards
The recitals sort the standards into two groups. "Superseding standards" are new editions of listed standards: EN ISO 11137-1:2026 replaces EN ISO 11137-1:2015+A2:2019 and EN ISO 14155:2026 replaces EN ISO 14155:2020+A11:2024, inserted as entries 3a and 39a. "New standards", whose earlier editions were not on the MDR list, are entries 66–72: natural rubber latex male condoms, biological evaluation Parts 6 and 7, washer-disinfectors Part 6, particular standards for laser and non-laser light source equipment, and small-bore connectors Part 1.
When are old harmonised standards withdrawn? Read Article 2
When a new edition supersedes a listed one, the decision deletes the old entry, inserts a new one and sets a separate application date for the deletion in Article 2. Until that date both references are listed and either supports presumption; afterwards the old one does not. The 2026 deletions, all falling after this article's publication:
| Reference to be deleted (entry) | Replacement entry | Deletion applies from | Source |
|---|---|---|---|
| EN ISO 10993-23:2021 (entry 1) | Entry 1a, with A1:2025 | 15 Dec 2027 | 2026/1231 |
| EN ISO 10993-12:2021 (entry 7) | Entry 7a, with A1:2025 | 15 Dec 2027 | 2026/1231 |
| EN IEC 60601-2-83:2020 + A11:2021 (entry 14) | Entry 14a, with A1:2025 | 15 Dec 2027 | 2026/1231 |
| EN ISO 10993-17:2023 (entry 20) | Entry 20a, with A1:2025 | 15 Dec 2027 | 2026/1231 |
| EN ISO 15223-1:2021 (entry 12) | Entry 12a, with A1:2025 | 17 Jun 2031 (as corrected) | 2026/1231 and corrigendum |
| EN ISO 11137-1:2015 + A2:2019 (entry 3) | Entry 3a, EN ISO 11137-1:2026 | 10 May 2028 | 2026/2479 |
| EN ISO 14155:2020 + A11:2024 (entry 39) | Entry 39a, EN ISO 14155:2026 | 10 May 2028 | 2026/2479 |
Check corrigenda too. Article 2 of 2026/1231 originally set the deletion of EN ISO 15223-1:2021 for 15 June 2031; a corrigendum published on 22 June 2026 changed it to 17 June 2031, and the harmonised standards page of the Commission's Directorate-General for Health and Food Safety (DG SANTE) links it next to 2026/1231. The OJ text prevails.
Common misconceptions about MDR harmonised standards
- "A new ISO edition is automatically harmonised." No. Harmonisation depends on OJ publication. For EN ISO 14155:2026, MDR presumption runs from 10 November 2026.
- "The EN year matches the ISO year." Not always. The Annex gives the ISO edition in brackets: EN ISO 21535:2024 corresponds to ISO 21535:2023, EN ISO 15883-1:2025 to ISO 15883-1:2024, and EN ISO 11137-1:2026 to ISO 11137-1:2025.
- "Only the base standard matters." Some entries combine a base standard with an amendment; entry 55 lists EN ISO 10993-5:2009 together with EN ISO 10993-5:2009/A11:2025. Check that every listed amendment has been applied.
- "A non-harmonised standard cannot be used." Incorrect. Use is voluntary; the manufacturer must demonstrate compliance without relying on presumption of conformity.
- "The summary list is the current list." The Commission states it is for information only, with no legal effect. As of 12 November 2026, the file on the single market page is still dated 20 October 2025 and omits the 2026 amendments.
UK designated standards vs EU harmonised standards
The GOV.UK page "Designated standards: medical devices", from the Office for Product Safety and Standards (OPSS) and the Department of Health and Social Care (DHSC), applies to England, Scotland and Wales. As of 12 November 2026, it carries two notices of publication (0034/21 of 1 January 2021 and 0129/26 of 17 January 2026) and a consolidated list (version 2, 17 January 2026). Notice 0129/26 designates standards under regulation 3A of the Medical Devices Regulations 2002 (S.I. 2002/618).
The two lists do not move in step. The UK designated EN ISO 12870:2025 and EN ISO 22675:2025 on 17 January 2026; the EU harmonised both only through 2026/1231 (published 17 June 2026). Conversely, the complete list in notice 0129/26 still shows EN ISO 10993-1:2009 and EN ISO 14155:2011, whereas the EU list has since added EN ISO 10993-1:2025 and EN ISO 14155:2026. For devices sold in both the EU and Great Britain, check both lists.
How to check in the database
This database helps you locate a standard and the related official documents; the final check is always the OJ text. A suggested order:
- Find the standard. Enter the number (for example 15883-1) in the site's Testing search, or open the EN standards list. As of 12 November 2026, the standards index shows 4,792 standard pages, 293 of them under the EN publishing body. The EN list has a "Recognised by" column showing EU and/or UK recognition.
- Read two fields on the standard page. "EU harmonised (OJ)" shows a decision number and OJ date; "UK designated" shows the UK notice number (such as 0034/21 or 0129/26). The EN 15883-1 page lists editions 2009 and 2025, and its EU field shows the base Decision 2021/1182 with an OJ date of 2026-06-17, which matches 2026/1231 on the Commission's list of amending decisions. The EN 13795-2 page shows 2025/2078 (OJ 2025-10-20); the EN 60118-0 page shows 2026/760 (OJ 2026-04-07).
- Separate MDR from IVDR. A standard can be on both lists, and the EU field shows a single record, not necessarily the MDR one; the EN 11137-1 page, for example, shows the IVDR base Decision 2021/1195. If you see a 2021/1195-series number, confirm against the MDR decisions.
- Confirm against the official text. Database updates lag behind the OJ: as of 12 November 2026, the EU field on the EN 14155 page still shows 2026/193 (OJ 2026-01-30), not 2026/2479. The reference is the decisions listed on the DG SANTE harmonised standards page and their EUR-Lex texts, such as Decision 2026/2479. For deletions, check the Article 2 date and any corrigendum.
- Check guidance and the UK side. The EU regulations section lists Commission and MDCG regulations and guidance; as of 12 November 2026 it shows 172 entries, updated 2026-09-18. MDCG 2021-5 Rev. 1 has its own page linking to the official text, and its EU REP symbol appendix is in the same list. The UK regulations section includes the Medical Devices Regulations 2002 (S.I. 2002/618); for the designated standards themselves, use the GOV.UK designated standards page.
FAQ
From what date can presumption be claimed for a newly harmonised standard?
From the day the amending decision is published in the OJ; for the standards in 2026/2479, 10 November 2026.
Which standards did Decision 2026/2479 add?
Two superseding standards: EN ISO 11137-1:2026 (entry 3a) and EN ISO 14155:2026 (entry 39a). Seven new entries (66–72): EN ISO 4074:2026, EN ISO 10993-6:2026, EN ISO 10993-7:2026, EN ISO 15883-6:2026, EN IEC 60601-2-22:2020 with A11:2026, EN IEC 60601-2-57:2026 with A11:2026, and EN ISO 80369-1:2026.
Once an old edition is deleted, can tests performed to it still be used?
After the deletion date the old edition gives no presumption. Earlier results can still be relied on if the technical documentation shows, usually through a gap analysis, that the differences between editions do not affect conformity with the relevant general safety and performance requirements; where a notified body is involved, discuss it with them.
How do I find which MDR requirements a harmonised standard covers?
Look at the standard's Annex Z (ZA and so on). Presumption applies only to requirements listed as covered; others need separate evidence.
Which list applies if the device is also sold in the UK?
For Great Britain, the designated standards published on GOV.UK under regulation 3A of S.I. 2002/618. That list is maintained separately, and the same standard may appear in different editions on each side.
The MDR list has been amended every year since 2022, four times in 2026, and deletion dates sit in each decision's Article 2 and corrigenda. In your standards register, record each standard's decision number, OJ date and any deletion date, and update it with each new decision. This article is based on official public information available as of 12 November 2026; the latest documents from the regulators prevail.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.