Whether a mobility scooter is a medical device or a vehicle in the EU comes down to two separate tests: whether Article 2(2) of Regulation (EU) No 168/2013 excludes it from the L-category vehicle rules (maximum design speed not exceeding 6 km/h, or exclusively intended for use by the physically handicapped), and whether the intended purpose stated by the manufacturer meets a medical purpose under Article 2(1) of the Medical Device Regulation (EU) 2017/745 (MDR). Falling outside Regulation 168/2013 does not make a scooter a medical device automatically. Below are the legal text, the use of MDCG 2021-24 Rev.1, and how to look things up in this database.

Key points

  • Article 2(2)(a) of Regulation 168/2013 excludes vehicles with a maximum design speed not exceeding 6 km/h. Article 2(2)(b) excludes vehicles exclusively intended for use by the physically handicapped. The word is "exclusively"; it cannot be read as "mainly".
  • Being outside Regulation 168/2013 does not put a scooter into the MDR. Whether it is a medical device depends on MDR Article 2(1): is it intended for a medical purpose such as alleviation of, or compensation for, an injury or disability?
  • Under MDR Article 2(12), the intended purpose is read from the label, the instructions for use, and promotional or sales materials and statements. The same scooter can reach a different answer depending on how it is described.
  • MDCG 2021-24 Rev.1 gives walking aids and wheelchairs as Class I examples under Rule 1 and electric wheelchairs as a Class I example under Rule 13. It does not name mobility scooters, so a scooter's class has to be worked out under Annex VIII rather than quoted from the guidance.
  • For a Class I scooter that is not sterile and has no measuring function, the manufacturer draws up the technical documentation and issues the EU declaration of conformity (MDR Article 52(7)). EN 12184:2022 test data are evidence in that documentation, not a regulatory conclusion in themselves.

Which EU rules can apply to a mobility scooter

EU law divides products by their nature and intended use. Regulation (EU) No 168/2013 covers the approval and market surveillance of two- or three-wheel vehicles and quadricycles. Under Article 2(1) it applies to such vehicles ("L-category vehicles") intended to travel on public roads, and to the systems, components, separate technical units, parts and equipment designed for them. The MDR starts instead from the definition of a medical device: a product intended by the manufacturer for human beings for specific medical purposes.

So the same product raises two questions: does vehicle law apply, and is it a medical device? Looking at vehicle law first is only a way to organise the analysis, not a legal procedure. Article 1(3) of Regulation 168/2013 also states that it is without prejudice to road safety legislation. Where a scooter may be ridden and at what speed is a matter for each Member State's road traffic rules and is not covered here.

Article 2(2) of Regulation 168/2013: which scooters are not L-category vehicles

Article 2(2) lists exclusions (a) to (k) (Regulation text on EUR-Lex). Points (a) and (b) are the most directly relevant to scooters. As of 13 November 2026, Article 2 is unchanged in the latest consolidated version on EUR-Lex (dated 2 August 2026).

Product situationArticle 2(2) of Regulation 168/2013Next step
Maximum design speed not exceeding 6 km/hExcluded under (a)Regulation 168/2013 does not apply; then assess whether it is a medical device
Design speed above 6 km/h, but exclusively intended for use by the physically handicappedExcluded under (b)Regulation 168/2013 does not apply; then assess whether it is a medical device
Design speed above 6 km/h, marketed for general mobility, not "exclusively" for handicapped users, and no other exclusion appliesNot excludedIf intended to travel on public roads, it may fall within L-category vehicle law and must be assessed under that law (not covered here)

Either exclusion is enough, and speed and user group are separate questions. A product with a design speed of 6 km/h or less is outside Regulation 168/2013 even if sold to ordinary consumers, but its medical device status is a separate question. Conversely, a product that mentions disabled users only in its marketing while being sold to everyone is unlikely to meet the "exclusively" wording of point (b).

Other exclusions, such as point (k) on seat R-point height, may matter for particular products; check the text.

Is a mobility scooter a medical device? MDR Article 2(1) and intended purpose

MDR Article 2(1) defines a medical device as any instrument, apparatus, appliance or other article intended by the manufacturer to be used, alone or in combination, for human beings for one or more specific medical purposes, including the diagnosis, monitoring, treatment, alleviation of, or compensation for, an injury or disability. A scooter that the manufacturer positions as compensating for reduced walking ability in people with mobility impairment comes close to that purpose.

MDR Article 2(12) defines intended purpose by the data the manufacturer supplies on the label, in the instructions for use or in promotional or sales materials or statements, and in the clinical evaluation. The decisive question is therefore not whether the product looks like a wheelchair, but whom it is supplied to and for what use:

  • Claims aimed at people with mobility impairment that stress compensation for, or alleviation of, a functional limitation are close to an MDR medical purpose;
  • Claims limited to general travel, shopping or leisure do not usually point to a medical purpose, although all the material still has to be considered;
  • If the instructions, packaging and online listings mix both kinds of claim, the position becomes hard to defend; the manufacturer needs one consistent story.

When in doubt, manufacturers can consult the competent authority of the Member State concerned. MDR Article 4 also allows the Commission, on a duly substantiated request from a Member State and after consulting the MDCG, to decide by implementing acts whether a product or group of products is a medical device. The Manual on borderline and classification under the MDR and IVDR (Version 5, April 2026) contains no entry found by searching for "scooter" or "wheelchair". A scooter placed on the market neither as a medical device nor as an L-category vehicle is a consumer product and must also take into account rules such as the General Product Safety Regulation (EU) 2023/988 (GPSR, applicable since 13 December 2024), which is not covered here.

MDR classification of mobility scooters: how to use MDCG 2021-24 Rev.1

MDR Article 51 divides devices into classes I, IIa, IIb and III under Annex VIII. Revision 1 of the classification guidance MDCG 2021-24 was published in April 2026 (PDF on the European Commission website). Its examples section opens by stating that the examples do not imply the products are a priori qualified as devices, and that the classification rules apply only after qualification has been established. For scooters: first decide whether it is a device, then decide the class.

Place in the guidanceRuleClass I examples relevant here
Page 28Rule 1: all non-invasive devices are Class I unless a later rule appliesDevices intended in general for external patient support, such as hospital beds, patient hoists, walking aids and wheelchairs
Page 49Rule 13: all other active devices are Class I (the fallback rule for active devices)Electric wheelchairs, electric hospital beds, patient hoists and others

The guidance does not name electric mobility scooters anywhere, so "the guidance makes scooters Class I" is not correct. Working through Annex VIII, a battery-powered scooter is an active device within the meaning of Article 2(4). Rules 9 to 12 cover active therapeutic devices that administer or exchange energy, devices for diagnosis and monitoring, software, and devices that administer or remove medicinal products, body liquids or other substances. A product with none of those functions will usually end up under the Rule 13 fallback. Section 3.5 of Annex VIII also requires the strictest rule, giving the higher class, to apply where several rules apply. The manufacturer has to check each rule and record the reasoning; the electric wheelchair example can inform it but cannot replace it.

After Class I under the MDR: self-declaration and technical documentation

Under MDR Article 52(7), manufacturers of Class I devices (other than custom-made or investigational devices) issue the EU declaration of conformity referred to in Article 19 after drawing up the technical documentation set out in Annexes II and III. A notified body is involved, within a limited scope, only for Class I devices placed on the market sterile, with a measuring function, or as reusable surgical instruments. For more on this, see Do You Need a Notified Body for Wheelchair CE Marking?

Where the manufacturer is not established in a Member State, MDR Article 11 requires it to designate a sole authorised representative before the device can be placed on the Union market. The technical documentation has to show that the product meets the general safety and performance requirements in Annex I (Article 5(2)). EN 12184:2022 (electrically powered wheelchairs, scooters and their chargers, requirements and test methods) can serve as one testing basis for this kind of product, but it does not replace MDR conformity assessment and does not decide whether the product is a medical device.

Maximum design speed and ISO 7176-6 testing: why speed matters for both routes

Maximum design speed is the threshold in Article 2(2)(a) of Regulation 168/2013, so speed information directly affects the vehicle-law exclusion; it is also part of the medical device technical documentation. EN 12184:2022 contains a maximum speed clause (8.1.8), and ISO 7176-6:2018 is the method standard for determining the maximum speed of electrically powered wheelchairs.

A maximum speed measured in a test is not the same thing as the "maximum design speed" in the regulation. How design speed is established, and what evidence is needed, should be checked against the legal text and the competent authority's position. For products with a speed dial or several driving profiles, state the test sample's speed setting and make sure it matches production; otherwise the data will not match the marketed model. For how scooter and powered wheelchair test plans differ, see Mobility Scooter vs Powered Wheelchair Testing: What Differs.

How to check this in the database

As of 13 November 2026, the EU medical device regulations page on this site shows the authority as the European Commission / MDCG, an update date of 18 September 2026, and 172 regulations and guidance entries. The list includes the MDR (32017R0745), MDCG 2021-24 rev.1 (Guidance on classification of medical devices) and Version 5 of the borderline and classification manual, all marked in_force.

  1. Open the regulation search. "MDCG 2021-24" goes straight to the classification guidance; "2017/745" finds the MDR and related MDCG documents; "borderline" finds the borderline and classification manual and related documents.
  2. Each detail page shows the number, type, issuer, status, a last-verified date (8 September 2026 for MDCG 2021-24 rev.1) and a link to the official text.
  3. Regulation 168/2013 is vehicle legislation. As of 13 November 2026 it is not in this database and a search by its number returns no results, so read it on EUR-Lex.

The database is compiled from official public sources for reference only; the official current text prevails.

Test evidence and SUNGO

Testing relevant to this topic is carried out by Hefei SUNGO Medical Technology Co., Ltd. (the Hefei laboratory), within the scope listed on its accreditation schedules:

  • CNAS L24279: 47 clause items of EN 12184:2022 (products: electrically powered wheelchairs, mobility scooters and their chargers; clause 6.6 on biocompatibility is not included), and ISO 7176-6:2018 "determination of maximum speed on a level surface" (products: wheelchairs and mobility scooters), which is also within the scope of CMA 260015342945.
  • IAS TL-1119: EN 12184 (except 6.6 biocompatibility and toxicity, and 12.3.1); ISO 7176, 20 parts, including ISO 7176-6.

CNAS and IAS are both signatories to the ILAC MRA. The Shanghai laboratory's scope does not include these standards. If a project needs clause 6.6 of EN 12184 (biocompatibility), that work can only be subcontracted to a suitably qualified laboratory, with the client's prior consent, and the report names the subcontracted item and the subcontractor. L-category vehicle type-approval is outside SUNGO's scope of work. Whether a product is a medical device or a vehicle, and which route it takes to market, is for the manufacturer to decide on the basis of the legal text and the competent authority's position; a test report provides test data.

Frequently asked questions

Is every mobility scooter a medical device in the EU?

No. It depends on whether the manufacturer's intended purpose meets a medical purpose under MDR Article 2(1), such as alleviation of, or compensation for, a disability. A product presented only as general transport is not usually a medical device, but all the material counts.

Does a mobility scooter with a design speed of 6 km/h or less need L-category type-approval?

Under Article 2(2)(a) of Regulation 168/2013, vehicles with a maximum design speed not exceeding 6 km/h are outside that regulation. That settles only the vehicle-law question; whether the product is a medical device, or subject to other rules, is decided separately.

Does labelling a scooter "for disabled people" bring it under Article 2(2)(b)?

Not automatically. The wording is "exclusively intended". If the product is in fact sold to ordinary consumers and disabled users are mentioned only in marketing, that condition is unlikely to be met. Intended users, instructions and sales channels are read together.

What MDR class is a mobility scooter? Can the electric wheelchair Class I example be applied directly?

MDCG 2021-24 Rev.1 gives electric wheelchairs as a Rule 13 Class I example and does not name scooters. The manufacturer has to work through Annex VIII rule by rule and record the reasoning; the guidance example cannot be quoted as the conclusion.

Does a Class I mobility scooter need a notified body?

Under MDR Article 52(7), a Class I device that is not sterile and has no measuring function is placed on the market on the manufacturer's own EU declaration of conformity, without notified body involvement. A higher class follows the procedure for that class.

For a scooter heading to the EU, settle "vehicle or medical device" first, then choose the route and the evidence. This article is based on official public information as of 13 November 2026; refer to the latest documents issued by the regulators.

About SUNGO Lab: What We Test

SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:

  • Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
  • Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
  • Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.

Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.