Rollator and walking frame CE marking under the EU Medical Device Regulation (MDR, Regulation (EU) 2017/745) is normally a Class I route under Annex VIII Rule 1. If the device is not placed on the market sterile and has no measuring function, the manufacturer draws up the technical documentation, issues the EU declaration of conformity and affixes the CE mark, with no notified body. ISO 11199 test reports are the main performance evidence, but the series is not an MDR harmonised standard, so the general safety and performance requirements (GSPRs) each report covers must be spelled out in the GSPR checklist. This guide sets out the steps and maps ISO 11199 clauses to MDR Annex I.
Key points
- MDR class: MDCG 2021-24 Rev.1 (April 2026), page 28, lists walking aids as Rule 1 Class I examples; the same page lists wheelchairs pushed by hand.
- Procedure: under MDR Article 52(7), Class I manufacturers draw up the technical documentation in Annexes II and III, then issue the declaration of conformity. A notified body is involved, for the relevant aspects only, if the device is sterile, has a measuring function or is a reusable surgical instrument.
- Registration: the first four EUDAMED modules, including actor registration and UDI/device registration, have been mandatory since 28 May 2026, and walking aid manufacturers must register.
- Report coverage: ISO 11199 reports support stability, strength, durability, moving parts, surfaces and edges, flammability and similar requirements. Biocompatibility, risk management, clinical evaluation and labelling are handled separately.
- Harmonised status: as of 13 November 2026, the European Commission's summary list of MDR harmonised standards does not include the EN ISO 11199 or EN ISO 11334 series, so testing to them gives no presumption of conformity.
What MDR class is a walking aid?
Here, walking aids means devices manipulated by both arms: walking frames (ISO 11199-1), rollators (ISO 11199-2) and walking tables (ISO 11199-3). The last section covers elbow crutches and multi-legged walking sticks (ISO 11334). Motorised products and products with a measuring or monitoring function are out of scope; MDCG 2021-24 Rev.1, for example, lists electric wheelchairs as Class I examples under Rule 13, not Rule 1. For walkers with seats, see Wheelchair or Walking Aid? Settling the Classification Boundary.
Rule 1 states that all non-invasive devices are Class I unless one of the later rules applies. The MDCG 2021-24 Rev.1 classification guidance says the manufacturer must consider all the rules, and the one giving the highest class decides. A walking aid gives external support and touches only intact skin, so with no other rule applying it stays in Class I. The guidance also says tyres of wheelchairs or walking frames made from rubber reinforced with carbon black nanomaterials, having no intended contact with users, should be exempt from classification under Rule 19 (nanomaterials). The class and the justification for the rule applied belong in the technical documentation (Annex II Section 1.1(f)).
Rollator CE marking under MDR Class I: self-declaration steps
MDR Article 52(7) says Class I manufacturers (custom-made and investigational devices aside) issue the EU declaration of conformity under Article 19 after drawing up the Annex II and III technical documentation. For devices placed on the market sterile, with a measuring function or that are reusable surgical instruments (Is, Im, Ir), Chapters I and III of Annex IX or Part A of Annex XI apply, and the notified body looks only at sterility, metrology or reprocessing. A standard walking aid is none of these, so the route is as follows.
| Step | What to do | MDR basis |
|---|---|---|
| 1. Classification | Class I under Rule 1; record why no other rule applies | Annex VIII; Annex II Section 1.1(f) |
| 2. Systems and people | Quality and risk management systems proportionate to the risk class and device type; a person responsible for regulatory compliance (PRRC). Micro and small enterprises need not have the PRRC in-house but must have one permanently and continuously available | Article 10(2) and (9); Article 15; Annex I Section 3 |
| 3. Clinical evaluation | Carry it out; if conformity based on clinical data is not deemed appropriate, justify this in the technical documentation from the risk management results | Article 61, including (10); Annex XIV |
| 4. Technical documentation | Description and specification, labels and IFU, design and manufacturing information, GSPR checklist, benefit-risk analysis and risk management, verification and validation (test reports go here), post-market surveillance documentation | Annexes II and III |
| 5. Declaration of conformity | Issue it with the content listed in Annex IV | Article 19; Annex IV |
| 6. CE marking | Visibly, legibly and indelibly on the device, or on the packaging if that is not possible | Article 20; Annex V |
| 7. UDI and EUDAMED | Register as an economic operator and obtain a single registration number (SRN); assign a Basic UDI-DI and submit it with the core data to the UDI database | Articles 27, 29 and 31 |
| 8. Post-market surveillance | Collect data under the surveillance plan; keep a post-market surveillance report, updated when necessary and available to the competent authority on request | Articles 83 to 85 |
Three points are easy to miss. First, a manufacturer not established in an EU Member State must designate a sole authorised representative under Article 11 before placing devices on the EU market. Second, under Article 10(8), the technical documentation and declaration of conformity must be kept available for at least 10 years after the last device covered by the declaration is placed on the market. Third, Commission Decision (EU) 2025/2371 was published in the Official Journal on 27 November 2025; under the transitional provisions of Regulation (EU) 2024/1860, the actor registration, UDI/device registration, notified bodies and certificates, and market surveillance modules have been mandatory since 28 May 2026. For devices placed on the market before that date, follow the arrangements on the Commission's EUDAMED page and its Q&A.
Mapping ISO 11199 test reports to the Annex I GSPRs
Annex II Section 4 requires the technical documentation to list the applicable GSPRs and explain why the others do not apply. For each one it states the method used, the harmonised standards, common specifications or other solutions applied, and the precise identity and location of the evidence documents. Annex II Section 6.1 adds the results of engineering and laboratory tests and detailed information on, among other things, biocompatibility, including identification of all materials in direct or indirect contact with the patient or user. Test reports hang off this checklist.
The table shows a common way to map walking aid evidence. Clause numbers follow ISO 11199-1:2021 and ISO 11199-2:2021 (ISO 11199-3:2005 for walking tables is numbered differently). The final mapping is for the manufacturer's risk analysis and the standard text to decide.
| Annex I section | What it requires | Related ISO 11199-1/-2 clauses | Covered by the report? |
|---|---|---|---|
| Section 6 | Performance must not degrade to the point of compromising safety under normal-use stresses during the stated lifetime | Static strength (16), durability (17) | Test data; the lifetime is stated and justified by the manufacturer |
| Section 20.1 | Protection against mechanical risks from resistance to movement, instability and moving parts | Static stability (15), moving parts (10), trapping of body parts (11), folding, adjusting and locking mechanisms (12) | The test part |
| Section 5; Section 14.2(a) | Reduce injury and use-error risks from ergonomic and physical features such as dimensions | Handgrips (6.7), surfaces, corners and edges (14), ergonomic principles (18) | Partly; usability analysis for intended users is the manufacturer's |
| Section 10.1(a) | Material toxicity and, where relevant, flammability | Flammability (7.2) | Flammability only; toxicity evaluated separately |
| Section 10.1(b) | Compatibility of materials with tissues, cells and body fluids | Biocompatibility and toxicity (7.3) | No; separate biological evaluation needed |
| Sections 11.1 and 11.2 | Reduce infection risk; where necessary, allow safe cleaning and disinfection | Infection and microbial contamination (7.4) | The clause check; cleaning and disinfection methods go in the IFU |
| Section 6; Section 14.2(b) | Risks from foreseeable environmental conditions such as humidity and temperature | Corrosion resistance (7.5), ingress of liquids (8) | The test part |
| Section 20.6 | Accessible parts must not reach potentially dangerous temperatures in normal use | Temperature of parts in contact with the user (9) | Yes |
| Section 7 | Transport and storage must not adversely affect characteristics and performance | Packaging (19) | Partly; transport and storage conditions verified by the manufacturer |
| Section 23 (Chapter III) | Label and instructions for use | Information supplied by the manufacturer (20) | Partly; languages, symbols and IFU content checked item by item against Section 23 |
| Section 3 | Risk management | None | No; done separately under a risk management standard |
| Sections 1 and 8; Article 61 | Acceptable benefit-risk; clinical evaluation | None | No |
Rollators also need attention to the edition. The ISO website shows ISO 11199-2:2021/Amd 1:2024, published in November 2024, titled "Removal of brake requirements" and one page long; check the amendment itself for exactly what it changes. Reports and the technical documentation should state whether they rely on ISO 11199-2:2021 or the amended edition, and the clause numbers in the GSPR checklist should match the report. Either way, brake-related risks still need control measures and verification evidence in the risk analysis. For sample preparation for each ISO 11199-2 test, see Rollator Testing to ISO 11199-2: Scope, Items and Sample Prep; for organising the technical documentation, see EU MDR technical documentation for wheelchair exporters, which applies equally to walking aids.
Is ISO 11199 an MDR harmonised standard?
MDR Article 8 says devices conforming to harmonised standards, or relevant parts of them, whose references are published in the Official Journal are presumed to conform to the requirements those standards cover. As of 13 November 2026, the summary list on the European Commission's MDR harmonised standards page includes EN ISO 14971:2019 (with A11:2021), EN ISO 15223-1:2021 and some parts of EN ISO 10993, among others, but not the EN ISO 11199 or EN ISO 11334 series, nor EN ISO 20417.
Two consequences follow. First, the technical documentation cannot say "conforms to EN ISO 11199 and is therefore presumed to conform to the MDR". Second, ISO 11199 reports remain usable: cite the standard under "other solutions" in the GSPR checklist, explain which requirements its tests cover and which they do not, and list the report as evidence. Risk management to EN ISO 14971:2019/A11:2021, which is on the list, can benefit from the presumption within that standard's scope. The list changes with each implementing decision; check the official page before citing it.
Do crutches and walking sticks follow the same Class I route?
Elbow crutches and multi-legged walking sticks are also walking aids for external support. Classification follows Rule 1 and the self-declaration steps are the same. The evidence side is ISO 11334: ISO 11334-1 for elbow crutches and ISO 11334-4 for walking sticks with three or more legs. ISO 11334-4:2024 also has clauses on static stability, static strength, durability, corrosion resistance and information supplied by the manufacturer, which map to Annex I as in the table above. ISO 11334-2 and ISO 11334-3 are outside SUNGO's accredited scope.
Common mistakes
- Treating a test report as a certificate. Class I self-declaration has no notified body certificate; conformity rests on the manufacturer's declaration of conformity.
- Filing reports without a GSPR checklist. Annex II Section 4 requires the method and evidence location for each requirement; a "pass" in a report does not replace it.
- Mismatched editions. The edition tested, the edition on the report and the edition cited in the file must match; for ISO 11199-2, watch Amd 1:2024.
- Forgetting EUDAMED. The first four modules are mandatory; for devices newly placed on the market, actor and device registration come first.
How to search the database
For EU regulations and guidance, open our European Union medical device regulations page. As of 13 November 2026 it lists 172 regulations and guidance documents, with an update date of 18 September 2026. It includes Regulation (EU) 2017/745 (32017R0745), the MDCG 2021-24 rev.1 classification guidance and the EUDAMED gradual roll-out Q&A, with a status column showing in_force or repealed. Check anything newer on the Commission's website.
To search by number, use the regulation search page, which covers regulations and guidance, FDA product codes and recent 510(k) clearances; numbers give the most precise results. Entering "MDCG 2021-24" opens the guidance's detail page with number, type, issuing body, status and a link to the official text; "32017R0745" finds the MDR itself.
For standards, enter the standard number on the testing search page, which covers standards records and testing bodies. Standard texts are copyrighted; the site gives records and official links only, and full texts come from the publisher's official channels. As of 13 November 2026, the standards index shows 4,792 medical device standards with pages, 530 of them from ISO.
Site information is for reference only; official regulatory texts and accreditation bodies' directories govern.
Test evidence and SUNGO
Walking aid and crutch testing is done by the laboratory of Hefei SUNGO Medical Technology Co., Ltd.; the Shanghai laboratory does not do it. The Hefei laboratory holds CNAS accreditation L24279, whose scope includes ISO 11199-1:2021 walking frames (23 items), ISO 11199-2:2021/Amd 1:2024 rollators (22 items; neither includes 7.3 biocompatibility and toxicity), ISO 11199-3:2005 walking tables (10 items; biocompatibility under 4.9 materials and finish not tested), ISO 11334-1:2007 elbow crutches (6 items) and ISO 11334-4:2024 walking sticks with three or more legs (15 items). It also holds IAS accreditation TL-1119 covering ISO 11199-1 and ISO 11199-2 (both except 7.3 biocompatibility and toxicity), ISO 11199-3 (except 4.9 materials and finish), ISO 11334-1, ISO 11334-4 (except 4.4 materials) and ISO 9227 salt spray testing. CNAS and IAS are both signatories to the ILAC MRA. ISO 11199 and ISO 11334 are not within either laboratory's CMA scope. Service overview: walking aid and crutch testing.
Biocompatibility clauses are outside the accredited scope above. Where needed, they can only be subcontracted to a laboratory with appropriate qualifications, with the client's prior agreement, and the report states the subcontracted items and the subcontractor. Whether a regulator or importer accepts a given report depends on their own requirements.
FAQ
Does rollator CE marking need a notified body?
A walking aid not placed on the market sterile and with no measuring function is Class I and self-declared, with no notified body. If it is sterile or measures, a notified body assesses those aspects.
What MDR class is a walker or rollator?
Normally Class I under Annex VIII Rule 1; MDCG 2021-24 Rev.1 lists walking aids as Rule 1 Class I examples. Motorised or measuring products must be checked against the other rules.
Is ISO 11199 harmonised under the MDR?
As of 13 November 2026, the EN ISO 11199 series is not on the Commission's summary list, so it gives no presumption of conformity. Reports can still serve as evidence, with their coverage explained in the GSPR checklist.
Do Class I walking aids need EUDAMED registration?
Yes. MDR Articles 29 and 31 require device and economic operator registration, and the relevant EUDAMED modules have been mandatory since 28 May 2026. Authorised representatives of non-EU manufacturers and importers register too.
For walking aids, the hard part of MDR Class I self-declaration is linking every applicable GSPR to its evidence. This article is based on official public information as of 13 November 2026. Always refer to the latest documents issued by the regulators.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.