An FDA ASCA-accredited testing laboratory is a lab that FDA has granted "ASCA Accreditation". When a manufacturer uses such a lab to test to an ASCA-included standard for a 510(k), De Novo, PMA or IDE, it can submit a declaration of conformity with an ASCA Summary Test Report, and FDA generally does not ask for more information on conformance with the standard or on test methods. Testing outside ASCA is submitted with a declaration of conformity and supporting material under FDA's 2018 standards guidance. This article explains what ASCA is, how the two routes differ, how to verify a lab on FDA's lists, and how to use this site's data to look up product codes and FDA recognition numbers.
The short answer
- ASCA (Accreditation Scheme for Conformity Assessment) is a voluntary FDA program based on section 514(d) of the FD&C Act. FDA implements it in accordance with the amendments made to that section by the Food and Drug Omnibus Reform Act of 2022 (FDORA) and with MDUFA V.
- ASCA applies only when several conditions are met: the submission is a 510(k), De Novo, PMA or IDE and is not submitted under the 510(k) Third Party Review Program; the lab is ASCA-accredited at the time of testing, with the standard and test method in its ASCA scope; and the standard is marked "Included in ASCA" in FDA's Recognized Consensus Standards database.
- On the ASCA route, the manufacturer submits a declaration of conformity for each standard plus the ASCA Summary Test Report exactly as the lab issued it. A complete test report is generally not needed unless the standards-specific guidance recommends one or the testing deviates from the standard.
- Off the ASCA route, the manufacturer submits a declaration of conformity under the 2018 guidance. Whether a results summary or a complete test report is also needed depends on the type of standard, and FDA may request the underlying data at any time.
- To verify a lab, check two lists: FDA's list of ASCA-accredited labs and its separate withdrawn list. FDA announced on September 18, 2025 that ASCA Summary Test Reports from labs with an "FDA Initiated Withdrawal" status will no longer be accepted.
What is FDA ASCA: legal basis and participants
ASCA began as a pilot. After the FDA Reauthorization Act of 2017 (FDARA) amended section 514, FDA set up the ASCA Pilot and issued guidance on it. On September 25, 2020, FDA issued three final guidances: the program guidance and two standards-specific guidances, one on basic safety and essential performance and one on biocompatibility. All three titles still include the words "Pilot Program".
There are four groups of participants: FDA, accreditation bodies, testing laboratories and device manufacturers. The process runs as follows:
- An accreditation body applies to FDA and is granted ASCA Recognition.
- A lab is first assessed by an ASCA-recognized accreditation body against ISO/IEC 17025:2017 plus the additional ASCA specifications, and then applies to FDA for ASCA Accreditation.
- Manufacturers do not apply. They simply choose to use an ASCA-accredited lab.
FDA does not charge accreditation bodies or labs a fee to take part. The program is funded through MDUFA V user fees, and manufacturers pay no additional fee to take part.
FDA chose the standards in ASCA from the biocompatibility series and the basic safety and essential performance series. As of November 13, 2026, filtering FDA's Recognized Consensus Standards database by "Included in ASCA" returned 112 entries. For standards outside that list, manufacturers can still submit a declaration of conformity under the 2018 guidance, but those standards are not eligible for ASCA's benefits.
ASCA vs non-ASCA testing: the differences
| Item | ASCA route | Non-ASCA route (2018 standards guidance) |
|---|---|---|
| Eligible submissions | 510(k), De Novo, PMA, IDE; not submissions under the 510(k) Third Party Review Program | Not subject to the ASCA conditions |
| Testing laboratory | ASCA-accredited at the time of testing, with the standard and method in its ASCA scope | Chosen by the manufacturer; if a third-party lab is used, give its name and address and reference its accreditations |
| Standards that qualify | FDA-recognized standards marked "Included in ASCA" | Standards with an FDA recognition number can support a declaration of conformity; standards without one, or used with deviations, fall under "general use" |
| What to submit | Cover letter stating that the submission contains ASCA testing, the lab's name and location, its ASCA ID and the standards used; a declaration of conformity for each standard; the unaltered ASCA Summary Test Report | Declaration of conformity; depending on the type of standard (Table 1 of the guidance), acceptance criteria and a results summary, or a complete test report |
| FDA review | FDA generally relies on the results without asking for more information on conformance with the standard, and does not intend to question test methods or results within the lab's ASCA scope except in circumstances set out in the guidance | No data is required with the declaration, but FDA may request at any time the data or information the declaration relies on |
| What stays the same | The manufacturer must still document how the testing supports marketing authorization | Same |
ASCA Summary Test Report vs a complete test report
Under ASCA, the lab completes the testing and gives the manufacturer both the complete test results and the ASCA Summary Test Report. The manufacturer includes the summary report, unaltered, as supporting documentation for the declaration of conformity. FDA's Manufacturers: How to Participate in ASCA page says such submissions generally do not need the complete test reports for ASCA standards, unless the standards-specific guidance recommends them or deviations from the standard are noted. The appendices of the two standards-specific guidances include example Summary Test Reports.
The manufacturer remains responsible for the test plan and should develop it with the lab, considering relevant FDA guidance (for example, on EMC and on use of ISO 10993-1) and whether any change to a recognized standard's methods or acceptance criteria is permitted by that standard.
Outside ASCA, the governing document is FDA's September 2018 guidance, Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices (docket FDA-2014-D-0456). It describes two ways to use standards: a declaration of conformity, and "general use". A declaration of conformity may cite only standards that have an FDA recognition number, and it may not deviate from the standard. Table 1 of the guidance sets out when supporting material is needed:
- If the standard includes both the test method and the acceptance criteria, a declaration of conformity is generally enough.
- If the standard gives the method but not all acceptance criteria, supporting documentation should follow ISO/IEC 17050-2, meaning the acceptance criteria and a results summary. The guidance cites IEC 60601-1-2 as an example.
- If the standard includes neither the method nor the acceptance criteria, a complete test report is expected.
The guidance text itself governs.
IEC 60601 and IEC 61326-2-6: which editions are in ASCA
Most of the 112 "Included in ASCA" entries are IEC 60601-2-x and ISO 80601-2-x particular standards. The entries for general medical electrical requirements and IVD equipment EMC are:
| Standard and edition | FDA recognition no. | Extent of recognition |
|---|---|---|
| IEC 60601-1 Edition 3.2 (2020-08) | 19-49 | Complete, with US national differences applied |
| IEC 60601-1-2 Edition 4.1 (2020-09) | 19-36 | Partial |
| IEC 60601-1-8 Edition 2.2 (2020-07) | 5-131 | Complete |
| IEC 60601-1-11 Edition 2.1 (2020-07) | 19-38 | Partial |
| ISO 80601-2-69 Second edition (2020-11) | 1-148 | Complete |
| IEC 61326-2-6 Edition 3.0 (2020-10) | 19-43 | Partial |
| IEC 61326-2-6 Edition 4.0 (2025-06) | 19-57 | Complete; date of entry July 28, 2025 |
(Read from FDA's Recognized Consensus Standards database on November 13, 2026.)
"Included in ASCA" is an attribute of the standard. It shows that the standard is within ASCA's scope; it does not show that any particular lab holds ASCA status. For partially recognized entries, some clauses are not recognized, so check the standard's Supplementary Information Sheet (SIS) clause by clause.
On July 28, 2025, FDA announced full recognition of IEC 61326-2-6 Edition 4, stating that it is part of the IEC 61010 family of standards and within the scope of the ASCA basic safety and essential performance standards-specific guidance. FDA said the edition resulted from collaboration between CDRH and industry and greatly expands IVD manufacturers' ability to use ASCA.
How to verify an ASCA-accredited testing laboratory on FDA's list
- Open FDA's ASCA-Accredited Testing Laboratories database. As of November 13, 2026 it held 98 records, all with the status "Accreditation Granted": 93 in the basic safety and essential performance category and 5 in biocompatibility. The table can be sorted by column and exported to Excel.
- Click a lab's name to open its detail page. It shows the ASCA status and start date, the accreditation body, and the "Scope of ASCA Accreditation", which lists each standard's status, date, FDA recognition number, designation and edition, and any exclusions.
- Match against your test dates. FDA requires the lab to be ASCA-accredited at the time of testing, with the standard and method in its ASCA scope at that time. A lab's name alone is not enough.
- Check the withdrawn list. It is a separate page with two statuses. "Voluntarily Withdrawn" means the lab asked to leave, often for business reasons such as closing or moving the lab or limiting its services; "FDA Initiated Withdrawal" means FDA withdrew the accreditation. As of November 13, 2026, the withdrawn list held 23 records, 5 of them FDA-initiated. FDA states on the withdrawn list that these labs had their ASCA Accreditation withdrawn for failing to meet the requirements of section 514(d) and the relevant ASCA guidance, and that for some of them it identified data integrity concerns with testing submitted in premarket submissions.
- Do not treat other accreditations or numbers as ASCA. FDA states that ASCA Accreditation exists only within ASCA and is separate from any accreditation an accreditation body grants a lab for other purposes. ASCA IDs on FDA's list take the form "TL-" plus a number, which looks similar to some accreditation bodies' lab certificate numbers, so rely on FDA's list.
Where to check in the database: product codes, recognition numbers and the ASCA mark
This site does not replace FDA's ASCA lists. It can help you pin down the product code, the marketing pathway and a standard's FDA recognition number before you confirm the ASCA mark and the lab status on FDA's pages.
- Start with the product code and third-party review eligibility: for portable oxygen concentrators, the FDA product code CAW page shows 21 CFR 868.5440, Class 2, 510(k) as the pathway, and the flag "Third-party review eligible". As of November 13, 2026, the page listed 22 510(k) clearances in its three-year window, with 19 summaries archived for download, and 288 clearances in FDA's records across all years. If you plan to use the 510(k) Third Party Review Program, ASCA does not apply, so settle the route before choosing a lab.
- Then find the standard's FDA recognition number: in the testing database, the ISO 80601-2-69 page shows recognition number 1-148 and the 2020 edition, and the IEC 61326-2-6 page lists recognition numbers 19-43 and 19-57, for the 2020 and 2025 editions.
- Finally, confirm the ASCA mark at FDA: the standard pages on this site do not show the ASCA mark. Look up the recognition number in FDA's list of standards included in ASCA, or open the standard's Supplementary Information Sheet. Only standards marked "Included in ASCA" next to the designation are within ASCA.
To see which standards earlier submissions in the same product code cited, download the 510(k) summaries from the CAW page. Whether an existing test report can be used directly in a submission is covered in Can a Wheelchair Test Report Be Used Directly for FDA and CE Filings?
Biocompatibility testing under ASCA: what to watch
As of November 13, 2026, there were 12 ASCA entries in the biocompatibility group. They include ISO 10993-4, -5, -11 and -12, ISO 10993-10 and ISO 10993-23 (both partially recognized), ASTM F756, ASTM F720 and the USP pyrogen test (rabbit test, partially recognized). FDA's database governs the editions and extent of recognition.
On the same date, FDA's ASCA-accredited lab list held only 5 records in the biocompatibility category, so the choice is narrower than for electrical safety. Certain devices are also excluded from the ASCA biocompatibility scope; see Section II (Scope) of the ASCA biocompatibility standards-specific guidance. Which biological tests are needed depends on the device's contact type, contact duration and biological evaluation, and ASCA does not change that assessment.
Test evidence and Sungo
Sungo is not an ASCA-accredited laboratory. As of November 13, 2026, Sungo did not appear on FDA's list of ASCA-accredited labs or on the withdrawn list.
Testing to IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11, ISO 80601-2-69 and IEC 61326-2-6 at Hefei Sungo Medical Technology Co., Ltd. is within the scope of its IAS accreditation (certificate TL-1119). CNAS and CMA do not cover these items. Such reports follow the non-ASCA route: Sungo issues test reports within its IAS accreditation scope, and the manufacturer submits them under FDA's 2018 standards guidance. Whether FDA accepts them depends on the complete submission and the review.
Biological tests are outside Sungo's accreditation scope. They can only be subcontracted to a laboratory with the appropriate qualifications, with the client's prior consent, and the subcontracted items and subcontractor are stated in the report. For an introduction to the laboratories, see the Sungo laboratory page.
Frequently asked questions
Does an FDA 510(k) require an ASCA-accredited lab?
No. ASCA is voluntary. Manufacturers can use data from non-ASCA labs and submit a declaration of conformity under the 2018 standards guidance, or rely on a standard through general use, without ASCA's benefits.
Is an ASCA-accredited lab the same as an IAS- or CNAS-accredited lab?
No. As of November 13, 2026, IAS was on FDA's list of ASCA-recognized accreditation bodies, but holding ISO/IEC 17025 accreditation from IAS or any other body does not give a lab ASCA Accreditation. Only FDA's ASCA lab list shows ASCA status.
Can I use an ASCA Summary Test Report in a 510(k) Third Party Review?
FDA's manufacturers page states that submissions under the 510(k) Third Party Review Program are not eligible for ASCA. Choose between the two routes before you file.
What happens to earlier reports after FDA withdraws a lab's ASCA accreditation?
FDA announced on September 18, 2025 that ASCA Summary Test Reports from labs with an "FDA Initiated Withdrawal" status will no longer be accepted. The FDA ASCA pages reviewed for this article give no general answer on whether the existing data can be used on the non-ASCA route, so raise it with FDA for the specific submission.
If a standard is "Included in ASCA", can any lab issue an ASCA report for it?
No. The mark only shows that the standard is within ASCA's scope. The lab must also hold ASCA Accreditation, with that standard and edition in its ASCA scope at the time of testing.
ASCA reduces follow-up questions on conformance with standards but does not replace the manufacturer's responsibility for the test plan and a complete submission. Check the submission type, any plan for third-party review and the ASCA mark on your standards first, then verify the lab on FDA's list. This article is based on official public information as of November 13, 2026. Regulatory agencies' latest documents govern.
About SUNGO Lab: What We Test
SUNGO Lab is made up of two independent companies, Shanghai SUNGO Medical Technology Co., Ltd. and Hefei SUNGO Medical Technology Co., Ltd., and focuses on medical device testing. Our core service lines are:
- Wheelchair and mobility aid testing (Hefei laboratory): manual and powered wheelchairs, scooters, walking aids and crutches to the ISO 7176 series (excluding Part 19 crash testing and Part 20 standing wheelchairs), ANSI/RESNA WC-1/WC-2, EN 12183, EN 12184, the GB/T 18029 series and GB/T 12996-2024. Which accreditation (CNAS, CMA or IAS) covers each standard is as listed in our certificate scopes. See Wheelchair testing for FDA 510(k) and RESNA.
- Medical electrical safety and EMC (Hefei laboratory, IAS accredited, certificate TL-1119): IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-1-11 and ISO 80601-2-69, with the corresponding Chinese standards GB 9706.1 and YY 9706.102.
- Testing for registration in China and overseas: we plan testing to the standards your target market requires (NMPA, FDA 510(k), EU CE marking and others) and issue the test reports for your submission; whether a report is accepted is decided by the regulator.
Our Shanghai laboratory also tests medical masks, gloves, surgical gowns, sterile packaging and sterilization-related items. CNAS and IAS are both signatories to the ILAC MRA. Send your product, target markets and standards through our contact page and we will confirm which laboratory and which accredited scope covers each test.